K013678 · Victorch Meditek, Inc. · DHA · Dec 27, 2001 · Clinical Chemistry
Device Facts
Record ID
K013678
Device Name
VICTORCH MEDITEK'S HCG TEST CARD
Applicant
Victorch Meditek, Inc.
Product Code
DHA · Clinical Chemistry
Decision Date
Dec 27, 2001
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1155
Device Class
Class 2
Attributes
3rd-Party Reviewed
Intended Use
The VicTorch Meditek Human Chorionic Gonadotropin (hCG) Urinary Pregnancy Test Card device (also The VicTorch hCG Test Card) is designed as a rapid in vitro diagnostic (IVD) qualitative lateral flow color-conjugated immuno-chromatographic (CCI) assay for the early detection of pregnancy, now bolor only gated "immal test for the placental hormone, hCG, at the required cut-off level of 20 mlU hCG/ml of human urine. VicTorch hCG Test Card devices are intended to meet all requirements for yielding rapid, easily read, qualitative results for the purpose of early pregnancy detection via CCI assay of hCG, a placental hormone that may be present in human plasma or urine, according to the requirements set forth in Title 21, section 862.1155 of the Code of Federal Regulation or CFR, and in subsequent federal Guidance Documents. VicTorch hCG Test Card devices are not intended for quantitative results, nor for over the counter (OTC) sales. VicTorch hCG Test Card devices are designed for professional use only, and provide only preliminary analytical data. For a final diagnosis of pregnancy, a more specific alternative clinical method must be used to obtain a confirmed analytical result.
Device Story
Lateral flow qualitative color-conjugated immunochromatographic (CCI) assay; detects hCG in human urine; cutoff 20 mIU/mL. Professional use only; clinic setting. Input: human urine sample. Transformation: immunochromatographic reaction produces visual color change indicating presence/absence of hCG. Output: rapid, visual, qualitative result. Healthcare provider interprets visual result; provides preliminary data; requires confirmation by alternative clinical method for final diagnosis. Benefits: rapid early pregnancy detection.
Clinical Evidence
Clinical accuracy study conducted at an OBGYN center with 172 female patients (91 hCG positive, 81 hCG negative). Results showed 100% agreement with the predicate device (ABI's SureStep hCG Test). 95% confidence interval: 99.1-99.8%. Bench testing included sensitivity (20 mIU/mL cutoff), specificity (no cross-reactivity with LH, TSH, FSH), interference (no interference from common drugs/pH), and stability (1-year shelf life).
Technological Characteristics
Lateral flow immunochromatographic assay; qualitative visual readout; 20 mIU/mL sensitivity; professional use format. No electronic components, software, or energy source.
Indications for Use
Indicated for professional use in the early detection of pregnancy by identifying human Chorionic Gonadotropin (hCG) in human urine at a cutoff concentration of 20 mIU/mL. Not for quantitative use or over-the-counter sales.
Regulatory Classification
Identification
A human chorionic gonadotropin (HCG) test system is a device intended for the early detection of pregnancy is intended to measure HCG, a placental hormone, in plasma or urine. A human chorionic goadotropin (HCG) test system is a device intended for any uses other than early detection of pregnancy (such as an aid in the diagnosis, prognosis, and management of treatment of persons with certain tumors or carcinomas) is intended to measure HCG, a placental hormone, in plasma or urine.
Predicate Devices
ABI's SureStep hCG Test Card
Related Devices
K013702 — VICTORCH HCG TEST STRIP · Victorch Meditek, Inc. · Dec 27, 2001
K032661 — ACON QUICK-CHECK II HOME PREGNANCY TEST DEVICE (CASSETTE) · ACON Laboratories, Inc. · Oct 14, 2003
K965227 — OSOM CLASSIC HCG-URINE TEST (115) · Wyntek Diagnostics, Inc. · Feb 4, 1997
{0}------------------------------------------------
#### 510k Submission for VicTorch Meditek Human Chorionic Gonadotropin (hCG) Urinary Pregnancy Test Card Device
VicTorch Meditek, Inc.
# roprietary Information DEC 2 7 2001 Revision B, September 19, 2001
## 9. SUMMARY STATEMENT OF SAFETY AND EFFECTIVENESS
The VicTorch hCG Test Card device is a lateral flow qualitative color-conjugated immunochromatographic (CCI) in vitro diagnostic (IVD) assay for human Chorionic Gonadotropin (hCG) on the cutoff level of 20 mlUml of human urine, for detection of early pregnancy as commonly used by professionals. Measurements obtained by the qualitative VicTorch hCG Test Cards are used to aid in the diagnosis of pregnancy. VicTorch hCG Test Card devices are classified as about II Devices, as described in Section 21 CFR 862.1155 Human Chorionic Gonadotropin (hCG) Test Systems, having product code: 91 DHA.
These VicTorch hCG Test Card devices are IVD tests designed to give rapid, visual, qualitative results and are intended for professional use only. VicTorch hCG Test Card devices are not intended for quantitative results, nor for over-the-counter sales. The VicTorch hCG Test Cards provide only preliminary analytical data. A more specific alternative clinical method must be used to obtain a confirmed analytical result.
In order to determine if the VicTorch hCG Test Card devices can function as effectively as other FDA approved and currently marketed similar hCG test devices (such as ABI's SureStep hCG Test Card devices), a number of evaluation studies were done with the VicTorch hCG Test Cards as listed herein. These studies included: (1) Sensitivity studies that includes 20% above and below 20 mlU hCG/ml cutoff concentration and other 0 to 100 mlU hCG/ml levels that indicated the VicTorch hCG Test Card can consistently detect 20 mlU hCG /ml or greater hCG in human urine as it claimed. (2) Specificity studies against other structurally related hormones such as 500 mlU/ml of LH, 1000 µlU/ml of TSH, and 1000 mlU/ml of FSH that shown that these hormones do not cross-react with the VicTorch hCG Test Card at the indicated test levels. (3) Interference studies against analytes commonly found in OTC, prescription and/or abuse drugs proved that those drugs, chemical analytes, and pH (varied within normal range for human urine) did not interfere VicTorch hCG Test Card performance, yielding no false positive or false negative results. (4) The Specimen Collection and Storage Studies on both hCG negative and hCG positive urine specimens validated the VicTorch hCG Test Card devices when tested for urine specimens refrigerated for up to three days. (5) Accuracy studies at an external OBGYN clinical center on 172 female patients (with 91 hCG positive and 81 hCG negative) gave complete agreement (accuracy = 100%) with the ABI's SureStep hCG Test, the predicate approved hCG assay device. (6) In house stability study for both accelerated and real time evaluations showed that the performance of VicTorch hCG Test Card was stable for at least one year from the time of being manufactured. Support data collected from the above evaluation and accuracy studies (i.e., Relative Accuracy = Total # of VicTorch hCG Test Card agreeing with SureStep / vs. Total # of ABI SureStep tested x 100% = 172/172 x 100% = 100%, with at 95% Confidence = 99.1-99.8%) indicate substantial equivalence of this newly designed VicTorch hCG Test Card with the predicate approved hCG assay device.
We can conclude that VicTorch hCG Test Card devices are substantial equivalent to currently marketed ABI's SureStep hCG pregnancy rapid tests devices for professional use qualitative urinary assay of hCG at the required cut-off of 20 mlU/mL. Additional information on this submission may be obtained by contacting: Dr. Pierce Liu, President, at VicTorch Meditek, Inc, via Tel.: (858)-530-9191. Fax: (858)-530-9192. Email: Vtorch99@aol.com.
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal is circular and contains the department's name around the perimeter. In the center of the seal is an emblem featuring a stylized caduceus, a symbol often associated with medicine and healthcare. The caduceus consists of a staff with two snakes coiled around it, topped with wings.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
DEC 2 7 2001
VicTorch Meditek, Inc. c/o Raymond Wilson, Pharm. D. California Department of Health 714/744 P Street P.O. Box 942732 (MS-357) Sacramento, CA 94234-7320
Re: k013678
> Trade/Device Name: VicTorch Meditek Human Chorionic Gonadotropin (hCG) Urinary Pregnancy Test Card Assay (VicTorch hCG Test Card) Regulation Number: 21 CFR 862.1155 Regulation Name: Human Chorionic Gonadotropin (hCG) test system Regulatory Class: Class II Product Code: JHI Dated: December 12, 2001 Received: December 13, 2001
Dear Dr. Wilson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
Page 2 -
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/odrh/dsma/dsmamain.html".
Sincerely yours,
Steven Putman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory-Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### 510k Submission for VicTorch Meditek Human Chorionic Gonadotropin (hCG) Urinary Preqnancy Test Card Device
VicTorch Meditek, Inc.
Proprietary Information
Revision B, September 19, 2001
Identity of The Classification: 21 CFR 862.1155;
K013678
**Product Code:** DHA
Device Name: VicTorch Meditek Human Chorionic Gonadotropin (hCG) Urinary Pregnancy Test Card Assay (VicTorch hCG Test Card)
### INDICATIONS FOR USE STATEMENT:
The VicTorch Meditek Human Chorionic Gonadotropin (hCG) Urinary Pregnancy Test Card device (also The VicTorch hCG Test Card) is designed as a rapid in vitro diagnostic (IVD) qualitative lateral flow color-conjugated immuno-chromatographic (CCI) assay for the early detection of pregnancy, now bolor only gated "immal test for the placental hormone, hCG, at the required cut-off level of 20 mlU hCG/ml of human urine. VicTorch hCG Test Card devices are intended to meet all requirements for yielding rapid, easily read, qualitative results for the purpose of early pregnancy detection via CCI assay of hCG, a placental hormone that may be present in human plasma or urine, according to the requirements set forth in Title 21, section 862.1155 of the Code of Federal Regulation or CFR, and in subsequent federal Guidance Documents. VicTorch hCG Test Card devices are not intended for quantitative results, nor for over the counter (OTC) sales. VicTorch hCG Test Card devices are designed for professional use only, and provide only preliminary analytical data. For a final diagnosis of pregnancy, a more specific alternative clinical method must be used to obtain a confirmed analytical result.
#### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Jlan Cocer
Division Sign-Off)
vision of Clinical Laboratory
K013678
Prescription Use: or (Per 21 CFR 801.109)
Over-the-Counter Use: (Optional Format 1-2-96)
ર (દ
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.