K013575 · Organ Recovery Systems, Inc. · KDN · Jan 25, 2002 · Gastroenterology, Urology
Device Facts
Record ID
K013575
Device Name
KPS-I, THE KIDNEY PERFUSION SOLUTION
Applicant
Organ Recovery Systems, Inc.
Product Code
KDN · Gastroenterology, Urology
Decision Date
Jan 25, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5880
Device Class
Class 2
Indications for Use
KPS-I™, The Kidney Perfusion Solution is intended to be used for flushing and continuous hypothermic machine perfusion of kidneys at the time of their removal from the donor, in preparation for storage, transportation and eventual transplantation into a recipient.
Device Story
KPS-I is a sterile, non-pyrogenic, non-toxic chemical solution used for the preservation of explanted kidneys. It functions as an extracellular solution with an osmolarity of ~300mOsM, sodium concentration of 100mEq/L, potassium concentration of 25mEq/L, and pH of ~7.4. The solution is cooled to 4-8°C and utilized within a compatible perfusion machine to maintain hypothermic conditions. It supports organ preservation for a mean duration of 29 hours (± 8 hours). The solution is intended for use by clinical professionals during the organ recovery and transport process to reduce metabolism and maintain physiological conditions, ultimately facilitating successful transplantation.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on identical chemical formulation and intended use compared to the predicate device.
Indicated for in-vitro flushing and temporary continuous hypothermic machine perfusion of explanted donor kidneys in preparation for transplantation.
Regulatory Classification
Identification
An isolated kidney perfusion and transport system and accessories is a device that is used to support a donated or a cadaver kidney and to maintain the organ in a near-normal physiologic state until it is transplanted into a recipient patient. This generic type of device may include tubing, catheters, connectors, an ice storage or freezing container with or without bag or preservatives, pulsatile or nonpulsatile hypothermic isolated organ perfusion apparatus with or without oxygenator, and disposable perfusion set.
{0}------------------------------------------------
KPS-ITM, The Kidney Perfusion Solution 510(k) Application: K013575 1/15/2002 Page 6 of 10
K013575
Page 1 of 3
# JAN 2 5 2002
#### 510(k) SUMMARY
## Organ Recovery Systems, Inc. KPS-I™ The Kidney Perfusion Solution
# 1. SUBMITTER INFORMATION
- A. Organ Recovery Systems, Inc., 701 E. Bay St., Suite 433 MSC 1119 Port City Center Charleston, SC 29403
Federal Identification Number: 36-4256620
### B. Official Correspondent:
Stanley J. Harris, Director, Regulatory and Clinical Affairs
Phone: (843) 853-6756 ex. 29 Fax: (843) 722-6657
# 2. DEVICE IDENTIFICATION
| A. Classification Name: | Isolated kidney perfusion and transport system and<br>accessories (21 CFR 876.5880) |
|-------------------------|-------------------------------------------------------------------------------------|
| B. Classification: | Class II, Gastroenterology/Urology Panel |
| C. Common/Usual Name: | Cold Storage Solution |
| D. Proprietary Name: | KPS-ITM, The Kidney Perfusion Solution |
# 3. PREDICATE DEVICE
Belzer MPS™ (UW Kidney Preservation Solution), Trans-Med Corporation (Elk River, MN), 510(k) Notification Number K972066
Indications for Use: Belzer MPSTM is the UW-gluconate perfusion solution indicated for flushing and continuous hypothermic machine perfusion of explanted donor kidneys.
{1}------------------------------------------------
KPS-17M, The Kidney Perfusion Solution 510(k) Application: K013575 1/15/2002 Page 7 of 10
K013575
Page 2 of 3
Device Description: Belzer MPS™ is a clear, sterile, non-pyrogenic, non-toxic solution for the in-vitro flushing and temporary continuous perfusion preservation of explanted kidneys. This solution has an approximate calculated osmolarity of 300mOsM, a sodium concentration of 100mEq/L, a potassium concentration of 25mEq/L, and a pH of approximately 7.4 at room temperature. Based upon the sodium/potassium ratio, the composition of Belzer MPSTM is thus consistent with that of an extracellular solution.
## 4. DESCRIPTION OF DEVICE
KPS-I™, The Kidney Perfusion Solution (University of Wisconsin Machine Perfusion Solution Formulation)
Indications for Use: KPS-I™, The Kidney Perfusion Solution is intended to be used for in-vitro flushing and temporary continuous hypothermic machine perfusion of kidneys at the time of their removal from the donor, in preparation for storage, transportation and eventual transplantation into a recipient.
Device Description: KPS-I™, The Kidney Perfusion Solution (UW Machine Perfusion Solution) is a clear, sterile, non-pyrogenic, non-toxic solution for the invitro flushing and temporary continuous perfusion preservation of explanted kidneys. This solution has an approximate calculated osmolarity of 300mOsM, a sodium concentration of 100mEq/L, a votassium concentration of 25mEg/1, and a pH of approximately 7.4 at room temperature. Based upon the sodium/potassium ratio, the composition is thus consistent with that of an extracellular solution.
KPS-I™ should be cooled to about 5 ℃ (4℃ to 8 ℃) prior to use and should be used in a perfusion machine that is capable of maintaining temperature within the above specified range.
It is recommended that the KPS-I™ be stored between 2 ℃ and 8 ℃. The solution should not be frozen or exposed to excessive heat.
KPS-I™, The Kidney Perfusion Solution is suitable for a mean perfusion time of 29 hours +/- 8 hours.
## 5. SUBSTANTIAL EQUIVALENCE
- A. Indications for Use: KPS-I™, The Kidney Perfusion Solution are identical in intended use to Belzer MPSTM
- B. Technological Characteristics: KPS-I™, The Kidney Perfusion Solution and Belzer MPS™ are both clear, sterile, non-pyrogenic solutions containing various components in the categories of ions, pH buffers, impermeants, colloids and pharmacologics intended to reduce metabolism and preserve
{2}------------------------------------------------
KPS-I'M, The Kidney Perfusion Solution 510(k) Application: K013575 1/15/2002 Page 8 of 10
K013575
Page 3 of 3
physiological conditions of explanted organs during cold storage. They have the same pH value, 7.4 +/-. 15 and an osmolarity of 300 +/-15mOsm/L. . The solutions are the same in that both the KPS-I™, The Kidney Perfusion Solution and Belzer MPS™ contains a sodium/potassium ratio (high/low) characteristic of an extracellular solution enables it to function well for hypothermic machine perfusion.
#### 6. Conclusion
KPS-I™, The Kidney Perfusion Solution has the same intended use as Belzer MPS. The solution formulation is identical and packaging system that is employed is identical to the predicate device. Data contained in the 510(k) demonstrates that KPS-IM, The Kidney Perfusion Solution is substantially equivalent to Belzer MPS.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, placed inside a circle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the circumference of the circle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN 2 5 2002
Mr. Stan Harris Director, Regulatory and Clinical Affairs Organ Recovery Systems Charleston Research Center 701 East Bay Street, Suite 433 MSC 1119, Port City Center CHARLESTON SC 29403
Re: K013575 Trade/Device Name: KPS-ITM, The Kidney Perfusion Solution Regulation Number: 21 CFR §876.5880 Regulation Name: Isolated kidney perfusion and transport system and accessories Regulatory Class: II Product Code: 78 KDN Dated: October 22, 2001 Received: October 29, 2001
Dear Mr. Harris:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C Brogdon
Nancv C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
510(k) Number (if known): K013575
Device Name: KPS-I™ The Kidney Perfusion Solution
Indications For Use:
KPS-I™ The Kidney Perfusion Solution is intended to be used for flushing and continuous hypothermic machine perfusion of kidneys at the time of their removal from the donor, in preparation for storage, transportation and eventual transplantation into a recipient.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
David A. Seppan
(Optional Format 3-10-98)
**Prescription Use**
**(Per 21 CFR 801.109)**
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.