K013518 · Irvine Scientific Sales Co., Inc. · MQL · Jan 18, 2002 · Obstetrics/Gynecology
Device Facts
Record ID
K013518
Device Name
EMBRYO FREEZE MEDIUM, EMBRYO THAW MEDIUM
Applicant
Irvine Scientific Sales Co., Inc.
Product Code
MQL · Obstetrics/Gynecology
Decision Date
Jan 18, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.6180
Device Class
Class 2
Indications for Use
The Embryo Freeze Media Kit is intended for use in the preparation for, and cryopreservation of, human Cleavage Stage (C-S) embryos. The Embryo Thaw Media Kit is intended for use in the thawing and recovery of cryopreserved human embryos.
Device Story
Embryo Freeze and Thaw Media Kits facilitate cryopreservation and recovery of human cleavage-stage embryos. Freeze kit contains two sequential media (F1, F2) with 1.5M propanediol (PROH), sucrose, human serum albumin (HSA), and gentamicin sulfate. Thaw kit contains three sequential media (T1, T2, T3) with varying concentrations of PROH and sucrose, plus HSA and gentamicin sulfate. Media protect embryos during rapid freezing, storage, and warming. Used in clinical embryology laboratories by trained technicians/embryologists. Output is preserved or recovered embryo suitable for implantation. Benefits include enabling future implantation for patients with excess embryos or history of unsuccessful procedures.
Clinical Evidence
Bench testing only. Five independent field laboratories evaluated the kits using mouse embryo models. Protocol involved freezing, thawing, and recovery procedures as per product inserts. Results demonstrated performance at least as well as control freeze/thaw media currently in use.
Technological Characteristics
Media based on modified human tubal fluid. Components: propanediol (PROH), sucrose, human serum albumin (HSA), and gentamicin sulfate. Sequential use design. Non-patient contacting. Sterility, endotoxin, and mouse embryo assay performance testing performed as release criteria.
Indications for Use
Indicated for use in assisted reproductive procedures for the cryopreservation, thawing, and recovery of human cleavage-stage embryos.
Regulatory Classification
Identification
Reproductive media and supplement are products that are used for assisted reproduction procedures. Media include liquid and powder versions of various substances that come in direct physical contact with human gametes or embryos (including water, acid solutions used to treat gametes or embryos, rinsing solutions, sperm separation media, supplements, or oil used to cover the media) for the purposes of preparation, maintenance, transfer or storage. Supplements are specific reagents added to media to enhance specific properties of the media (e.g., proteins, sera, antibiotics, etc.).
Special Controls
*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, biocompatibility testing, and clinical testing). The device, when it is phosphate-buffered saline used for washing, and short-term handling and manipulation of gametes and embryos; culture oil used as an overlay for culture media containing gametes and embryos; and water for assisted reproduction applications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.
Predicate Devices
Notice of Final Rule, 63 FR 48428, Docket number 97N-0335
Submission Summary (Full Text)
{1}------------------------------------------------
Irvine Scientific
# JAN 1 8 2002
K013518
October 18, 2001
Page 1 of 3
## 510(k) SUMMARY AS REQUIRED BY SECTION 807.92(c)
Submitted by:
Irvine Scientific Sales Co., Inc. 2511 Daimler Street Santa Ana, CA 92705-5588
Telephone: (800) 437-5706 Facsimile: (949) 261-6522
Contact: Wendell Lee, PhD.
Date Submitted: January 27, 2000
Device Identification:
| Trade Name: | Embryo Freeze Media Kit<br>Embryo Thaw Media Kit |
|----------------------|--------------------------------------------------------------------|
| Common Name: | Embryo cryopreservation media<br>Embryo thawing and recovery media |
| Classification Name: | Reproductive Media (21 CFR, 888.6180) |
#### Predicate Device:
Notice of Final Rule, 63 FR 48428, Docket number 97N-0335
#### Description:
The five media that comprise the two kits, Embryo Freeze Media Kit and Embryo Thaw Media Kit are all based upon the formulation of modified human tubal fluid (K983586). The two media in the Embryo Freeze Media kit
{2}------------------------------------------------
October 18, 2001 Y
are intended to be used sequentially, for the preparation for, and cryopreservation of, human blastocysts. The first medium to be used, F1, which is used in preparation for freezing, contains 1.5M PROH. The second medium, F2, to be used during cryostorage, containing 1.5M PROH and 0.1M sucrose. Both F1 and F2 are also supplemented with human serum albumin (HSA).
The three media in the Embryo Thaw Media kit are also intended for sequential use in the thawing and recovery of cryopreserved human embryo. The first medium used in the thawing process, T1, contains 1.0M PROH and 0.2M sucrose. The second medium, T2 contains 0.5M PROH and 0.2M sucrose. T3 contains 0.2M sucrose. Each T1, T2 and T3 contain HSA. All five media contain qentamicin sulfate.
#### Intended Use:
The Embryo Freeze Media Kit is intended for use in the preparation for, and cryopreservation of, human Cleavage Stage (C-S) embryos. The Embryo Thaw Media Kit is intended for use in the thawing and recovery of cryopreserved human embryos.
#### Technological Characteristics:
Embryos are routinely stored for use in future assisted reproductive procedures. In some instances, excess eggs will be retrieved from the patient, and fertilized. If development of these fertilized eggs indicates a potential for viability during implantation, they may be frozen for future use. In the event that the current transfer is unsuccessful, and does not result in a clinical pregnancy, the patient has embryos in reserve that may be used for implantation in future procedures. Embryos are also routinely frozen when patients have a history of unsuccessful implantation procedures, and also for those patients who desire multiple children. Media to protect the embryos during the preparation for cryopreservation, during
{3}------------------------------------------------
K013518
October 18, 2001 Page 3 of 3
The media in the Embryo Freeze Media Kit, F1 and F2 are designed to be used sequentially for the preparation of embryos for cryopreservation, and as the protective media during cryostorage. The media in the Embryo Thaw Kit, T1, T2 and T3 are also designed for sequential use, in the thawing and recovery of cryopreserved human embryos. None of the media are intended to contact the patient.
### Performance Data:
The Embryo Freeze Media Kit and the Embryo Thaw Media kit have been studied by five independent field laboratories, using mouse embryos and the protocol for embryo freezing, thawing and recovery presented in the product inserts for these kits. The conclusion from all laboratories that evaluated the kits was that the Embryo Freeze Media kit and Embryo Thaw Media kit performed at least as well as the control freeze/thaw media in use in the laboratories.
### Additional Information:
Endotoxin, mouse embryo freezing and recovery assay performance and sterility tests will be performed as a condition of release for these products. Results of all release assays performed will be reported on a lot-specific certificate of analysis, and will be indicated on the labeling.
#### Conclusion:
The results from the field testing of these products demonstrates that Embryo Freeze Media Kit and Embryo Thaw Media kit are suitable for their intended use, and meet the criteria outlined in the Notice of Final Rule, 63 FR 48428. Docket number 97N-0335.
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# JAN 1 8 2002
Wendell Lee, Pharm.D. Vice President Ouality Systems and Regulatory Affairs Irvine Scientific Sales Co., Inc. 2511 Daimler Street SANTA ANA CA 92705-5588
Re: K013518
Trade/Device Name: Embryo Freeze Media Kit and Embryo Thaw Media Kit Regulation Number: 21 CFR §884.6180 Regulation Name: Reproductive media and supplements Regulatory Class: II Product Code: 85 MOL Dated: October 18, 2001 Received: October 23, 2001
Dear Mr. Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
{6}------------------------------------------------
#### INDICATIONS FOR USE STATEMENT (page 1 of 2)
510(K) Number: KO13518
Device Name: Embryo Freeze Media Kit; Embryo Thaw Media Kit
Indications for Use:
Embryo Freeze Media Kit is intended for use in the assisted reproductive procedure, of embryo cryopreservation. The two media kit is designed to protect human cleavage-stage embryos during rapid freezing procedures and during frozen storage.
Embryo Thaw Media Kit is intended for use in the assisted reproductive procedure of thawing frozen cleavage-stage embryos. The three media kit is designed to protect human cleavage-stage embryos during warming and thawing after cryopreservation.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Nancy C. Brogdon
Diwalon
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number.
K013518
**Prescription Use**
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.