MICROLIFE UPPER ARM AUTOMATIC BLOOD PRESSURE MONITOR, MODEL BP-3BTO-1, WITH OPTIONAL THERMAL PRINTER, MODEL PR 1KA1
Applicant
Microlife Corp.
Product Code
DXN · Cardiovascular
Decision Date
Jan 17, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1130
Device Class
Class 2
Indications for Use
The Microlife Upper Arm Automatic Blood Pressure Monitor, Model BP-3BTO-1, with Optional Thermal Printer, Model PR 1KA1, is a device intended to measure the systolic and diastolic blood pressure and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. This device can be used in connection with the Microlife Thermal Printer.
Device Story
Device measures systolic/diastolic blood pressure and pulse rate using oscillometric method. Input: pressure variations in inflatable upper arm cuff detected by electronic capacitive pressure sensor. Operation: microprocessor analyzes electrical signals from sensor to determine blood pressure/pulse; utilizes fuzzy logic for inflation control; employs split slope resolution (three sections: 0-100, 100-200, 200-300 mmHg) to reduce nonlinearity error to 0.33%. Output: blood pressure and pulse rate values displayed on device; optional thermal printer for hard copy. Used in home or clinical settings by adults. Healthcare providers use output for clinical assessment of patient cardiovascular status. Benefits: non-invasive, automated monitoring of blood pressure.
Clinical Evidence
Clinical testing performed per ANSI/AAMI SP10-1992 'National Standard for Electronic or Automated Sphygmomanometers'. Device met all relevant requirements of the standard.
Technological Characteristics
Oscillometric measurement; electronic capacitive pressure sensor; microprocessor-based; fuzzy logic inflation control; split slope resolution software (three linear segments); solenoid valve for pressure release; optional thermal printer accessory; field calibration access via battery/key sequence.
Indications for Use
Indicated for adult individuals to measure systolic and diastolic blood pressure and pulse rate via non-invasive upper arm cuff.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
Predicate Devices
Microlife Automatic Blood Pressure Monitor, Model BP-2BHO (K970211)
Submission Summary (Full Text)
{0}------------------------------------------------
# EXHIBIT # 1
# 510(K) SUMMARY
This summary of 5l0(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
The assigned 5l0(k) number is:
#### 1. Submitter's Identification:
Microlife Corporation 9F, 431 Rui Guang Road Nei Hu. Taipei 114 Taiwan, Republic of China
## Contact:
Mr. Laurence Yanq
Date Summary Prepared: October 17, 2001
#### 2. Name of the Device:
Microlife Upper Arm Automatic Blood Pressure Monitor, Model BP-3BTO-1, with Optional Thermal Printer, Model PR 1KA1.
#### 3. Predicate Device Information:
The Microlife Upper Arm Automatic Blood Pressure Monitor, Model BP-3BTO-1 is substantially equivalent to the Microlife Automatic Blood Pressure Monitor, Model BP-2BHO, K# 970211.
#### 4. Device Description:
The Microlife Upper Arm Automatic Blood Pressure Monitor, Model BP-3BTO-1 is designed to measure the systolic and diastolic blood pressure and pulse rate of an individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. Our method to define systolic and diastolic pressure is similar to the auscultatory method but uses an electronic capacitive pressure sensor rather than a stethoscope and mercury manometer. The sensor converts tiny alterations in cuff pressure to electrical signals, by analyzing those signals to define the systolic and diastolic blood pressure and calculating pulse
{1}------------------------------------------------
rate, which is a well - known technique in the market called the "oscillometric method".
An optional printer is included as an accessory which can be sold separately in a separate box with the printer user manual.
#### Intended Use: 5.
The Microlife Upper Arm Automatic Blood Pressure Monitor, Model BP-3BTO-1, with Optional Thermal Printer, Model PR 1KA1, is a device intended to measure the systolic and diastolic blood pressure and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. This device can be used in connection with the Microlife Thermal Printer.
#### Comparison to Predicate Devices: 6.
Both devices use the well known oscillometric method within the software algorithm to determine the systolic and diastolic blood pressure and pulse rate. An upper arm cuff is inflated automatically; deflate rate is controlled by a factory set bleed valve and the deflation pressures are transferred via tubing to a sensor in both units. Each device uses a similar capacitance-type pressure sensor to translate the pressure variations to electrical signals that can be interpreted by an integrating circuit. Once the reading is determined each unit operates a solenoid valve to release the pressure to zero. Our Blood Pressure Monitor, Model BP-3BTO-1, differs from the predicate device in the fuzzy logic inflation.
The interface between the sensor and the microprocessor determines the system's accuracy. For our Pressure Monitor, Model BP-3BTO-1, the software is capable of a split slope resolution to improve accuracy over the entire range. Since the range is "split" into the three sections (0 to 100mmHg) (100 to 200mmHg) (200 to 300mmHg) error due to nonlinearity is reduced by the ability to adjunct the slope to best fit the output curve. A nonlinearity of 1% is reduced to 0.33% by splitting the span into three separated linear relations. This way the sensor is matched to the software by using a series of jumpers that profile the slopes to the output of the sensor.
The Microlife Automatic Blood Pressure Monitor, Model BP-3BTO-1, has field calibration access; this model is initiated by the following steps: (a) Remove one battery from compartment first, (b) When inserting back the battery, press and hold "I/O" key, and don't release it until the battery is inserted well.
### Discussion of Non-Clinical Tests Performed for Determination of 7. Substantial Equivalence are as follows:
{2}------------------------------------------------
Testing information demonstrating safety and effectiveness of the Microlife Automatic Blood Pressure Monitor, Model BP-3BTO-1 in the intended Automatio Drosa is supported by testing that was conducted in accordance with the FDA November 1993 Draft "Reviewer Guidance for Premarket Notification Submissions", DCRND, which outlines Electrical, Mechanical and Environmental Performance Requirements.
The following testing was conducted:
- General Functions Test a.
- Reliability Test Operation Conditions b.
- Reliability Test Drop Testing C.
- Reliability Test Storage ರ.
- Reliability Test Vibrating Testing e.
- f. EMC Test
- IEC 60601-1 Safety Test g.
- Blood Pressure Monitor Printer Port Function Test Report h.
- Thermal Printer Software Validation Report ....
- EMC Test Report Blood Pressure Monitor Connected with j. Thermal Printer
None of the testing demonstrated any design characteristics that violated the requirements of the Reviewer Guidance or resulted in any safety hazards. It was our conclusion that the Microlife Upper Arm Automatic Blood Pressure Monitor, Model BP-3BTO-1 tested met all relevant requirements of the aforementioned tests.
#### Discussion of Clinical Tests Performed: 8.
ANSI/AAMI SP10-1992 "National Standard for Electronic or Automated Sphygmomanometers" testing was performed. All relevant sections were addressed and testing conducted. The BP-3BTO-1 met all relevant requirements of this standard.
#### 9. Conclusions:
We have demonstrated that the Microlife Automatic Blood Pressure Monitor, Model BP-3BTO-1, is as safe and effective as the predicate, the Microlife Automatic Blood Pressure Monitor, Model BP-2BHO based on electrical, mechanical and environmental testing results as well as the FDA DCRND November 1993 Draft "Reviewer Guidance for Premarket Notification Submissions", and, the ANSI/AAMI Voluntary Standard, SP10-1992 testing results.
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States of America. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized eagle-like symbol, which is the official emblem of the HHS. The seal is black and white.
JAN 1 7 2002
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Microlife Corporation c/o Ms. Susan D. Goldstein-Falk MDI Consultants, Inc. 55 Northern Blvd., Suite 200 Great Neck, New York 11021
Re: K013485
Trade Name: Microlife Upper Arm Automatic Blood Pressure Monitor, Model BP-3BTO-1, with Optional Thermal Printer, Model PR 1KA1 Regulation Number: 21 CFR 870.1130 Regulation Name: Non-Invasive Blood Pressure Measurement System Regulatory Class: Class II (two) Product Code: DXN Dated: October 17, 2001 Received: October 19, 2001
Dear Ms. Goldstein-Falk:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Ms. Susan D. Goldstein-Falk
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Kala Teler
Bram D. Zuckerman, M.D. Acting Director Division of Cardiovascular and Respiratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Exhibit B
Page __ 1_____________________________________________________________________________________________________________________________________________________________________
510(k) Number (if known): __ KO13485
Device Name: Microlife Upper Arm Automatic Blood Pressure Monitor, Model BP-3BTO-1, with Optional Thermal Printer, Model PR 1KA1
### Indications For Use:
The Microlife Upper Arm Automatic Blood Pressure Monitor, Model BP-3BTO-1 with Optional Thermal Printer, Model PR 1KA1 is a device intended to measure the systolic and diastolic blood pressure and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. This device can be used in connection with the Microlife Thermal Printer.
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
**Division of Cardiovascular & Respiratory Devices**
5'10(k) Number K01348
Prescription Use (Per 21 CFR 801.109)
OR
**Over-The-Counter Use** X
(Optional Format 1-2-96)