← Product Code [DJG](/productcode/DJG) · K013482

# ABUSCREEN ONLINE OPIATES 300 / 2000 (K013482)

_Roche Diagnostics Corp. · DJG · Dec 21, 2001 · Clinical Toxicology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K013482

## Device Facts

- **Applicant:** Roche Diagnostics Corp.
- **Product Code:** [DJG](/productcode/DJG.md)
- **Decision Date:** Dec 21, 2001
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.3650
- **Device Class:** Class 2
- **Review Panel:** Clinical Toxicology

## Indications for Use

An in vitro diagnostic test for the qualitative and semi-quantitative detection of morphine and its metabolites in human urine on automated clinical chemistry analyzers at cutoff concentrations of 300 and 2000 ng/mL. Semi- quantitative test results may be obtained that permit laboratories to assess assay performance as part of a quality control program.

## Device Story

Abuscreen OnLine Opiates 300/2000 is an in vitro diagnostic assay for detecting morphine and metabolites in human urine. It utilizes Kinetic Interaction of Microparticles in Solution (KIMS) technology. In the absence of drug, drug-polymer conjugates bind to antibody-coated microparticles, causing aggregation and increased light absorbance. Presence of sample drug inhibits this aggregation, decreasing absorbance proportionally to drug concentration. Used in clinical laboratories on automated chemistry analyzers (Hitachi 900/Modular series). Results are interpreted by laboratory personnel to determine presence of opiates relative to 300 or 2000 ng/mL cutoffs. Semi-quantitative outputs assist in quality control monitoring.

## Clinical Evidence

No clinical data provided; device performance is established through analytical bench testing of the assay's reagent formulation and its application on automated clinical chemistry analyzers.

## Technological Characteristics

In vitro diagnostic assay using Kinetic Interaction of Microparticles in Solution (KIMS). Measures changes in light transmission (absorbance) to detect drug-antibody-microparticle aggregation. Reagent-based system for use on automated clinical chemistry analyzers (Hitachi 900/Modular series).

## Regulatory Identification

An opiate test system is a device intended to measure any of the addictive narcotic pain-relieving opiate drugs in blood, serum, urine, gastric contents, and saliva. An opiate is any natural or synthetic drug that has morphine-like pharmocological actions. The opiates include drugs such as morphine, morphine glucoronide, heroin, codeine, nalorphine, and meperedine. Measurements obtained by this device are used in the diagnosis and treatment of opiate use or overdose and in monitoring the levels of opiate administration to ensure appropriate therapy.

## Special Controls

*Classification.* Class II (special controls). An opiate test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).

## Predicate Devices

- Abuscreen OnLine II Opiates 300/2000 ([K974840](/device/K974840.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## 510(k) Summary

K013482

Introduction

According to the requirements of 21 CFR 807.92, the following information provides sufficient detail to understand the basis for a determination of substantial equivalence.

Roche Diagnostics Corporation 1) Submitter name, address, 9115 Hague Rd. contact Indianapolis, IN 46250 (317) 521-7637

Contact Person: Kerwin Kaufman

Date Prepared: October 18, 2001

Proprietary name: Abuscreen OnLine Opiates 300/2000 2) Device name

Common name: Opiates Test System

Classification name: Enzyme immunoassay, opiates

3) Predicate device

We claim substantial equivalence to the currently marketed Abuscreen OnLine II Opiates 300/2000 assay (K974840).

Continued on next page

{1}------------------------------------------------

4) Device Description The Abuscreen OnLine Opiates 300/2000 assay is an in vitro diagnostic test for the qualitative and semi-quantitative detection of morphine and its metabolites in human urine on automated clinical chemistry analyzers at cutoff concentrations of 300 and 2000 ng/mL

#### Principal of procedure

The Abuscreen OnLine Opiates 300/2000 assay is based on the kinetic interaction of microparticles in a solution (KIMS) as measured by changes in light transmission. In the absence of sample drug, soluble drug-polymer conjugates bind to antibody-bound microparticles, causing the formation of particle aggregates.

When a urine sample containing the drug in question is present, this drug competes with the drug conjugate for microparticle-bound antibody. Antibody bound to sample drug is no longer available to promote particle aggregation, and subsequent particle lattice formation is inhibited.

As the aggregation reaction proceeds in the absence of sample drug, the absorbance increases. Conversely, the presence of sample drug diminishes the increasing absorbance in proportion to the concentration of drug in the sample. Sample drug content is determined relative to the value obtained for a known cutoff concentration of drug.

#### Negative Sample

drug-polymer conjugate + antibody-bound microparticle = particle aggregates (↑ absorbance)

#### Positive Sample

sample drug + antibody-bound microparticle = particle aggregation inhibited drug-polymer conjugate + antibody bound microparticle = particle aggregates

Continued on next page

{2}------------------------------------------------

### 510(k) Summary, Continued

The Abuscreen OnLine Opiates 300/2000 assay is an in vitro diagnostic test 5.) Intended for the qualitative and semi-quantitative detection of morphine and its Use metabolites in buman urine on automated clinical chemistry analyzers at cutoff concentrations of 300 and 2000 ng/mL. Semi-quantitative test results may be obtained that permit laboratories to assess assay performance as part of a quality control program.

6.) Comparison to the Predicate Device

The Roche Abuscreen OnLine Opiates 300/2000 assay is substantially equivalent to other products in commercial distribution intended for similar use. Most notably, it is substantially equivalent to the currently marketed Roche Abuscreen OnLine II Opiates 300/2000 (K974840).

The Abuscreen OnLine Opiates 300/2000 assay utilizes the same reagent formulation as the currently marketed Abuscreen OnLine II Opiates 300/2000 assay (K974840). Differences between this application and the cleared assay include:

- addition of a 300 ng/mL semi-quantitative cutoff claim, and ●
- application to the Hitachi 900 series and Modular series of . analyzers.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird with three curved lines representing its body and wings. The bird is facing to the right. Encircling the bird is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

DEC 2 1 2001

Mr. Kerwin Kaufman Regulatory Affairs Consultant Roche Diagnostics Corporation 9115 Hague Road P.O. Box 50457 Indianapolis, IN 46250-0457

Re: k013482

> Trade/Device Name: Roche Diagnostics Abuscreen OnLine Opiates 300/2000 Regulation Number: 21 CFR 862.3650 Regulation Name: Lead test system Regulatory Class: Class II Product Code: DJG Dated: October 18, 2001 Received: October 19, 2001

Dear Mr. Kaufman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and 1 additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrb/dsma/dsmamain.html".

Sincerely yours,

Steven Butman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory-Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

# Indications for Use Statement

| 510(k) Number (if known): | K013482                                                                                                                                                                                                                                                                                                                                                             |
|---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name:              | Roche Diagnostics Abuscreen OnLine Opiates 300/2000                                                                                                                                                                                                                                                                                                                 |
| Indications for Use:      | An in vitro diagnostic test for the qualitative and semi-quantitative detection of morphine and its metabolites in human urine on automated clinical chemistry analyzers at cutoff concentrations of 300 and 2000 ng/mL. Semi- quantitative test results may be obtained that permit laboratories to assess assay performance as part of a quality control program. |

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER
PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

| Prescription Use (Per 21 CFR 801.109) | <span style="text-decoration: overline;">✓</span> OR Over-the-Counter Use ______ |
|---------------------------------------|----------------------------------------------------------------------------------|
| (Optional format 1-2-96)              |                                                                                  |

(Division Sign-Off)

Division Clinical Laboratory Devices

510(k) Number: K013482

______________________________________________________________________________________________________________________________________________________________________________

---

**Source:** [https://fda.innolitics.com/device/K013482](https://fda.innolitics.com/device/K013482)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
