TRIDENT POROUS TITANIUM ACETABULAR COMPONENT WITH PERI-APATITE COATING
Applicant
Howmedica Osteonics Corp.
Product Code
LPH · Orthopedic
Decision Date
Dec 20, 2001
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3358
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The indications for use of the Trident® Porous Titanium Acetabular Component with Peri-Apatite™ Coating, in keeping with those of other legally marketed Howmedica Osteonics acetabular component systems, are as follows: - Painful, disabling joint disease of the hip resulting from: degenerative arthritis, post-traumatic arthritis or late stage avascular necrosis. - Clinical management problems where arthrodesis or alternative reconstructive techniques are less likely to achieve satisfactory results. - Revision of previously unsuccessful femoral head replacement, cup arthroplasty or other procedure. - Where bone stock is of poor quality or inadequate for other reconstructive techniques as indicated by deficiencies of the acetabulum.
Device Story
Trident® Porous Titanium Acetabular Component is a single-use, cementless artificial hip replacement shell; designed for implantation into the prepared acetabulum. Device consists of a Ti6Al-4V ELI alloy shell with a CP Titanium porous coating and a 20-micron Peri-Apatite™ hydroxyapatite-based coating. Used in conjunction with Trident® UHMWPE acetabular inserts and compatible Howmedica Osteonics hip stems/heads. Implanted by orthopedic surgeons during total hip arthroplasty to resurface the acetabulum; provides an articulating surface for the femoral head. Optional fixation via cancellous bone screws. Benefits include improved prosthetic fixation and joint function for patients with degenerative hip disease or revision needs.
Clinical Evidence
Bench testing only. Characterization of the Peri-Apatite™ coating was performed according to the FDA document 'Calcium Phosphate [Ca-P] Coating Draft Guidance for Preparation of FDA submissions for Orthopedic and Dental Endosseous Implants.' No clinical data provided.
Indicated for patients with painful, disabling hip joint disease (degenerative/post-traumatic arthritis, late-stage avascular necrosis), clinical management problems requiring reconstruction, or revision of failed prior hip procedures, including cases with poor bone stock. Contraindicated in patients with active/latent hip infection, mental/neuromuscular disorders risking instability/fixation failure, compromised bone stock, skeletal immaturity, or obesity.
Regulatory Classification
Identification
A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.
{0}------------------------------------------------
DEC 2 0 2001
# (013475
#### 510(k) Summary
## Trident® Porous Titanium Acetabular Component with Peri-Apatite™ Coating
The Trident® Porous Titanium Acetabular Shells with Peri-Apatite™ Coating described in this 510(k) submission consist of single use devices which are intended for cementless fixation within the prepared acetabulum. The Trident® Porous Titanium Acetabular Shells with Peri-Apatite™ coating are intended for mating with the commercially available Trident® UHMWPE Acetabular Inserts (N2Vac packaged and Crossfire™ styles), also single use devices.
If supplemental bone screw fixation is deemed necessary, Osteonics® 5.5mm and 6.5mm Cancellous Bone Screws (K894124, K873251) can be placed through the shells' dome screw holes without interfering with the seating of the insert.
The Trident® Porous Titanium Acetabular Shells with Peri-Apatite™ Coating are compatible with any appropriately selected, legally marketed Howmedica Osteonics hip stem/head combination.
#### Indications:
The indications for use of the Trident® Porous Titanium Acetabular Shell with Peri-Apatite™ Coating, in keeping with those of other legally marketed Howmedica Osteonics acetabular component systems, are as follows:
- Painful, disabling joint disease of the hip resulting from: degenerative arthritis, . post-traumatic arthritis or late stage avascular necrosis.
- Clinical management problems where arthrodesis or alternative reconstructive ● techniques are less likely to achieve satisfactory results.
- Revision of previously unsuccessful femoral head replacement, cup arthroplasty . or other procedure.
- Where bone stock is of poor quality or inadequate for other reconstructive . techniques as indicated by deficiencies of the acetabulum.
{1}------------------------------------------------
NOLLS
#### Contraindications:
As with other Howmedica Osteonics hip replacement acetabular component systems, the contraindications for the Trident® Porous Titanium Acetabular Component with Peri-ApatiteTM Coating include:
- Any active or suspected latent infection in or about the hip joint. .
- Any mental or neuromuscular disorder which would create an unacceptable risk ● of prosthesis instability, prosthesis fixation failure, or complication in postoperative care.
- Bone stock compromised by disease, infection, or prior implantation which . cannot provide adequate support and fixation of the prosthesis.
- Skeletal immaturity. .
- Obesity. An overweight or obese patient can produce loads on the prosthesis . which can lead to failure of the fixation of the device or to failure of the device itself.
#### Device Description
The Trident® Porous Titanium Acetabular Component with Peri-Apatite™ Coating, when used with the commercially available Trident® UHMWPE Acetabular Insert, is an artificial total hip replacement device which consists of an acetabular shell and a mating insert. Each insert/shell assembly is intended to resurface the acetabulum thereby providing a suitable articulating surface for a mating artificial stem/head combination.
The Trident® Porous Titanium Acetabular Component with Peri-Apatite™ Coating is manufactured from ASTM F-136 Ti6Al-4V ELI alloy, and employs a porous coating fabricated from ASTM F-67 Commercially Pure (CP) Titanium. Overlaying the porous coating is a 20 micron thick layer of Peri-Apatite™ coating.
The Trident® Porous Titanium Acetabular Shells with Peri-Apatite™ Coating are characterized by the following features:
- The outer shell has a single radius (hemispherical) geometry .
- A shell substrate composed of titanium alloy (Ti6Al-4V ELI) comparable to that . used in other Howmedica Osteonics acetabular components
{2}------------------------------------------------
- A titanium porous coating (CP Titanium) which meets the definition of porous ● coating outlined in 21 CFR 888.3358, and is comparable to that used in other Howmedica Osteonics hip and knee components
- Peri-Apatite™ coating that is used on the Vitalock® Solid Back Shell with Peri-. Apatite™ coating.
- An interior geometry which allows a mating with the Trident® UHMWPE inserts . through maximum conformity and a wireless locking mechanism.
- Availability with screw hole options and cluster screw holes which accept . Osteonics® 5.5mm and 6.5mm Cancellous bone Screws (K873251, K894124). A dome hole which is compatible with the optional, currently marketed Osteonics® Acetabular Dome Hole Plugs (K942809). The Osteonics® 5.0mm Cancellous Screws are available for use with peripheral screw holes.
- A range of outer diameters from 40 through 82mm in 2mm increments. .
- The following shell configurations will be available: .
- dome hole only, no screw holes ﺖ
- 2-3 cluster screw holes, and dome hole ・
- 3-5 cluster screw holes and dome hole -
- 5 cluster screw holes with 4 inferior screw holes, and dome hole -
- 8-12 multiple dome screw holes and dome hole -
- 5-7 peripheral screw holes and dome hole ـــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــ
Characterization of the Peri-Apatite™ coating was presented to the extent possible using the testing outlined in the FDA document "Calcium Phosphate [Ca-P] Coating Draft Guidance for Preparation of FDA submissions for Orthopedic and Dental Endosseous Implants."
The Trident® Porous Titanium Acetabular Component with Peri-Apatite™ Coating is substantially equivalent to other legally marketed devices. These products are listed below:
{3}------------------------------------------------
#### 510(k) Summary
Trident® Porous Titanium Acetabular Comp with Peri-Apatite™ Coating
- Trident® Porous Titanium Acetabular Component K010170 1) Howmedica Osteonics Corp.
- Vitalock® Solid Back Shell with Peri-Apatite™ Coating K971206 2)
A discussion of the equivalent features was presented.
For further information please contact:
Margaret F. Crowe Regulatory Affairs Consultant Howmedica Osteonics Corp. 59 Route 17 Allendale, New Jersey 07401 (201) 934-4359
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an abstract eagle or bird-like figure, composed of three curved lines that suggest feathers or wings.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC 2 0 2001
Ms. Margaret F. Crowe Regulatory Affairs Consultant Howmedica Osteonics Corporation 59 Route 17 Allendale, New Jersey 07401-1677
Re: K013475 Trade Name: Trident™ Porous Titanium Acetabular Component w/Peri-Apatite Regulation Number: 888.3358 Regulation Name: Prothesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented; Prosthesis, Hip, Semi-Constrained, Metal/Polymer, non-porous, Calcium Regulatory Class: II; unclassified Product Code: LPH; MEH Dated: October 18, 2001 Received: October 19, 2001
Dear Ms.Crowe:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
Page 2 -- Ms. Margaret Crowe
This letter will allow you to begin marketing your device as described in your Section 510(k) I mo lotel with a motification. The FDA finding of substantial equivalence of your device to a legally premarket notication. "The a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of additionally 21 011 594-4659. Additionally, for questions on the promotion and advertising of Compinates at (301) 594 - the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Celia M. Witten, Ph.D., M.D.
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
K0013475
page 1 of 1
### INDICATIONS FOR USE
510(k) Number (if known):
Device Name: Trident® Porous Titanium Acetabular Component with Peri-Apatite™ Coating
Indications for Use:
The indications for use of the Trident® Porous Titanium Acetabular Component with
Peri-Apatite™ Coating, in keeping with those of other legally marketed Howmedica
Osteonics acetabular component systems, are as follows:
- Painful, disabling joint disease of the hip resulting from: degenerative arthritis, ● post-traumatic arthritis or late stage avascular necrosis.
- . Clinical management problems where arthrodesis or alternative reconstructive techniques are less likely to achieve satisfactory results.
- Revision of previously unsuccessful femoral head replacement, cup arthroplasty . or other procedure.
- Where bone stock is of poor quality or inadequate for other reconstructive . techniques as indicated by deficiencies of the acetabulum.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use X OR (per 21 CFR 801.109)
Over-the-Counter Use
(Optional Format 1-2-96)
Signature
(Division Sign-Off) (Division Sign Cral, Restorative Division ological Devices
510(k) Number k013475
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.