← Product Code [MQP](/productcode/MQP) · K013382

# MODIFICATION TO: STACKABLE CAGE SYSTEM (K013382)

_Depuyacromed · MQP · Nov 7, 2001 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K013382

## Device Facts

- **Applicant:** Depuyacromed
- **Product Code:** [MQP](/productcode/MQP.md)
- **Decision Date:** Nov 7, 2001
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3060
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Stackable Cage System is indicated for use in the thoracolumbar spine (i.e., T1 to L5) to replace a diseased vertebral body resected or excised for the treatment of tumors, to achieve anterior decompression of the spinal cord and neural tissues, and to restore the height of a collapsed vertebral body. The Stackable Cage System is also indicated for treating fractures of the thoracic and lumbar spine. The Stackable Cage System is designed to restore the biomechanical integrity of the anterior, middle and posterior spinal column even in the absence of fusion for a prolonged period. The Stackable Cage System is intended for use with supplemental internal fixation. The supplemental internal fixation systems that may be used with the Stackable Cage System include DePuy AcroMed titanium plate or rod systems (i.e., Kaneda SR, University Plate, M-2, ISOLA, VSP, Moss Miami, TiMX, Monarch, Profile).

## Device Story

Stackable Cage System; spinal intervertebral body fixation orthosis; used for vertebral body replacement in thoracolumbar spine (T1-L5) following tumor resection or fracture treatment. Device restores height of collapsed vertebral bodies and biomechanical integrity of spinal column. Implanted by surgeons; requires supplemental internal fixation using DePuy AcroMed titanium plate or rod systems (e.g., Kaneda SR, ISOLA, Moss Miami).

## Clinical Evidence

No clinical data provided; substantial equivalence based on design and material characteristics.

## Technological Characteristics

Materials: Carbon-fiber reinforced polymer and titanium alloy. Form factor: Stackable cage system for vertebral body replacement. Intended for use with supplemental internal fixation systems.

## Regulatory Identification

A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.

## Predicate Devices

- Stackable Cage System ([K001340](/device/K001340.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

KD13382

#### 510 (k) Summary X.

SUBMITTER:

DePuy AcroMed™, Inc. 325 Paramount Drive Ravnham, MA 02780

Karen F. Jurczak CONTACT PERSON:

October 10, 2001 DATE PREPARED:

Implant, Fixation Device CLASSIFICATION NAME: Spinal Intervertebral Body Fixation Orthosis Device

Stackable Cage System PROPRIETARY NAME:

Stackable Cage System (K001340) PREDICATE DEVICES:

The Stackable Cage System is indicated for use in the INTENDED USE: thoracolumbar spine (i.e., T1 to L5) to replace a diseased vertebral body resected or excised for the treatment of tumors, to achieve anterior decompression of the spinal cord and neural tissues, and to restore the height of a collapsed vertebral body.

> The Stackable Cage System is also indicated for treating fractures of the thoracic and lumbar spine.

> The Stackable Cage System is designed to restore the biomechanical integrity of the anterior, middle and posterior spinal column even in the absence of fusion for a prolonged period.

> The Stackable Cage System is intended for use with supplemental internal fixation. The supplemental internal fixation systems that may be used with the Stackable Cage System include DePuy AcroMed titanium plate or rod systems (i.e., Kaneda SR, University Plate, M-2, ISOLA, VSP, Moss Miami, TiMX, Monarch, Profile).

MATERIALS:

Carbon-fiber reinforced polymer and titanium alloy

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, symbolizing the department's mission to protect and promote the health and well-being of the nation.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

NOV 0 7 2001

Mr. Frank Maas Director, Regulatory Affairs DePuy Acromed 325 Paramount Drive Raynham, Massachusetts 02767-0350

Re: K013382

> Trade Name: Stackable Cage System Regulation Number: 21 CFR 888.3060 Regulation Name: Vertebral Body Replacement Device Regulatory Class: II Product Code: MQP Dated: October 10, 2001 Received: October 12, 2001

Dear Mr. Maas:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{2}------------------------------------------------

Page 2 - Mr. Frank Maas

This letter will allow you to begin marketing your device as described in your 510(k) premarket I his letter will anow you to begin marketing your device to a legally marketed notification. The FDA infollig of substainler of July 2005, 1975. This your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFF Part 801 and If you desire specific at rice for your actively splease contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please 594-4037. Additionally, 101 questions of (301) 594-4639. Also, please note the regulation entitled, Colliact the Office of Complance at (301) of 1001.97). Other general information wilsolanding of reicled to prematics not nearlied from the Division of Small Manufacturers on your responsibilities ander the 110 ha3-2041 or at (301) 443-6597, or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

SP

Signature

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# NOV 0 7 2001

# 510(k) Number (if known): K013382

## Device Name: Stackable Cage System

### Indications For Use:

The Stackable Cage System is indicated for use in the thoracolumbar spine (i.e., T1 -L5) to replace a diseased vertebral body resected or excised for the treatment of tumors, to achieve anterior decompression of the spinal cord and neural tissues, and to restore the height of a collapsed vertebral body.

The Stackable Cage System is also indicated for treating fractures of the thoracic and lumbar spine.

The Stackable Cage is designed to restore the biomechanical integrity of the anterior, middle and posterior spinal column even in the absence of fusion for a prolonged period.

The Stackable Cage System is intended for use with supplemental internal fixation. The supplemental internal fixation that may be used with the Stackable Cage System include DePuy AcroMed titanium plate or rod systems (i.e., Kaneda SR, University Plate, M-2, ISOLA, VSP, Moss Miami, TiMX, Monarch, Profile).

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Optional Format 3-10-98)

(Division Sign-Off) Division of General, Restorative and Neurological Devices

KO13382 510(k) Number -

---

**Source:** [https://fda.innolitics.com/device/K013382](https://fda.innolitics.com/device/K013382)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
