UNI-T SPINAL SYSTEM
K013301 · Spine Vision, Inc. · MNI · Feb 13, 2002 · Orthopedic
Device Facts
| Record ID | K013301 |
| Device Name | UNI-T SPINAL SYSTEM |
| Applicant | Spine Vision, Inc. |
| Product Code | MNI · Orthopedic |
| Decision Date | Feb 13, 2002 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 888.3070 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
The UNI-T Uni-Thread Spinal System is a pedicle screw system indicated for skeletally mature patients who: have severe spondylolisthesis (Grades 3 and 4) at the L5-S1 vertebra; receive fusions using autogenous bone graft only; have the device fixed or attached to the lumbar and sacral spine (L3 to sacrum); and have the device removed after the development of a solid fusion. In addition, the UNI-T Uni-Thread Spinal System is a pedicle screw system intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: Degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; failed previous fusion (pseudoarthrosis).
Device Story
UNI-T Uni-Thread Spinal System consists of titanium alloy screws, rods, and connectors; assembled into constructs for posterior spinal fixation. Used by surgeons to provide immobilization and stabilization of spinal segments during fusion procedures. Components are implanted in the thoracic, lumbar, and sacral spine to treat various pathologies including deformities and instabilities. System provides mechanical support to facilitate bone fusion; device is intended for removal following successful fusion. Benefit includes stabilization of spinal segments in patients with severe spinal disorders.
Clinical Evidence
Bench testing only; no clinical data provided.
Technological Characteristics
Titanium alloy components (ASTM F136). Pedicle screw system consisting of rods, screws, and connectors for posterior spinal fixation. Mechanical assembly.
Indications for Use
Indicated for skeletally mature patients requiring spinal immobilization/stabilization as an adjunct to fusion for severe L5-S1 spondylolisthesis (Grades 3-4) or thoracic/lumbar/sacral instabilities/deformities including degenerative spondylolisthesis with neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, or pseudoarthrosis. Requires autogenous bone graft; device removal required after solid fusion.
Regulatory Classification
Identification
(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
Special Controls
*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.
Predicate Devices
- SCS Claris Spinal System
- CD Horizon Sofamor Danek
Related Devices
- K021510 — UNI-T UNI-THREAD SPINAL SYSTEM · Spine Vision, Inc. · Aug 1, 2002
- K052406 — UNI-THREAD SPINAL SYSTEM · Spine Vision, Inc. · Nov 7, 2005
- K070367 — ENDIUS TITLE 2 POLYAXIAL SPINAL SYSTEM · Endius, Inc. · Apr 12, 2007
- K182478 — Whistler Modular Pedicle Screw System · Evolution Spine, LLC · Jan 15, 2019
- K022374 — CADENCE SPINAL FIXATION SYSTEM · Sulzer Spine-Tech · Sep 24, 2002
Submission Summary (Full Text)
{0}------------------------------------------------
FEB 1 3 2002
K013301
Image /page/0/Picture/2 description: The image shows the SpineVision logo. The logo consists of a stylized "S" shape above the word "SpineVision". The "S" shape is gray and black, and the word "SpineVision" is black. There is also a small "TM" symbol to the right of the word "SpineVision".
Spine systems for specialists
## 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
| NAME OF FIRM: | SpineVision, Inc.<br>3003 Summit Blvd.<br>Suite 1500<br>Atlanta, GA 30319<br>Phone: 404-460-5077 |
|--------------------------------------|--------------------------------------------------------------------------------------------------|
| 510(K) CONTACT: | Lynnette Whitaker<br>Vice President, Regulatory Affairs<br>219-269-3701 |
| TRADE NAME: | UNI-T™ Uni-Thread Spinal System |
| COMMON NAME: | Rod and Screw Spinal Instrumentation |
| CLASSIFICATION: | 888.3070 Pedicle Screw Spinal System |
| DEVICE PRODUCT CODE: | Product code: 87 MNH, MNI |
| SUBSTANTIALLY<br>EQUIVALENT DEVICES: | SCS Claris Spinal System, Orthotec, Inc.<br>CD Horizon Sofamor Danek |
## DEVICE DESCRIPTION:
The UNI-T Uni-Thread Spinal System instrumentation consists of screws, rods and connectors that can be assembled in a variety of constructs. The instrumentation is combined and placed to treat a variety of spinal disorders, with the constructs varying according to the nature of the spinal pathology. The UNI-T System components are available in titanium alloy, manufactured according to ASTM F136.
{1}------------------------------------------------
K013301 2/2
## INDICATIONS FOR USE:
The UNI-T Uni-Thread Spinal System is a pedicle screw system indicated for skeletally mature patients who:
- have severe spondylolisthesis (Grades 3 and 4) at the L5-S1 vertebra; .
- receive fusions using autogenous bone graft only; .
- have the device fixed or attached to the lumbar and sacral spine (L3 to sacrum); and ●
- have the device removed after the development of a solid fusion. ●
In addition, the UNI-T Uni-Thread Spinal System is a pedicle screw system intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine:
- Degenerative spondylolisthesis with objective evidence of neurologic impairment .
- fracture .
- dislocation ●
- scoliosis
- . kyphosis
- spinal tumor .
- failed previous fusion (pseudoarthrosis) .
#### BASIS OF SUBSTANTIAL EQUIVALENCE:
The components of the UNI-T Uni-Thread Spinal System are similar in design, material, and indications to other spinal instrumentation systems that have been cleared by FDA for posterior spinal use. Mechanical testing was performed to demonstrate the equivalence of the construct design to currently marketed spinal systems.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized design featuring three parallel, curved lines that resemble a human figure. The lines are thicker at the top and taper towards the bottom, giving the impression of movement or flow.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB 1 3 2002
Ms. Lynnette Whitaker Vice President, Reģulatory Affairs SpineVision, Inc. 3003 Summit Blvd., Suite 1500 Atlanta, Georgia 30319
Re: K013301
> UNI-TTM Uni-Thread Spinal System Regulation Number: 888.3070 Regulation Names: Pedicle Screw Spinal System and Spondylolisthesis Spinal Fixation Device System Regulatory Class: II Product Codes: MNI and MNH Dated: January 14, 2002 Received: January 15, 2002
Dear Ms. Whitaker:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Ms. Lynnette Whitaker
This letter will allow you to begin marketing your device as described in your Section 510(k) I mis lotter will and w yourse of substantial equivalence of your device to a legally premated noticated. " results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of additionally 21 (301) 594-4659. Additionally, for questions on the promotion and advertising of Compliance at (301) 597 - the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Miriam C. Provost
for Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# 510(k) Number (if known): __ K013301
#### UNI-T™ Uni-Thread Spinal System Device Name:
#### Indications for Use:
The UNI-T Uni-Thread Spinal System is a pedicle screw system indicated for skeletally mature patients who:
- have severe spondylolisthesis (Grades 3 and 4) at the L5-S1 vertebra; .
- receive fusions using autogenous bone graft only; .
- · have the device fixed or attached to the lumbar and sacral spine (L3 to sacrum); and
- have the device removed after the development of a solid fusion. .
In addition, the UNI-T Uni-Thread Spinal System is a pedicle screw system intended to provide In addition, the UNF-1 Unl-Tiread Spinal System is a pkelers on on adjunct to fusion in fusion in fusion in immobilization and Sabilization of Spiralis in Skolouny Industry Industry of the thoracic, lumbar, and sacral spine:
- Degenerative spondylolisthesis with objective evidence of neurologic impairment .
- fracture ◆
- . dislocation
- scoliosis .
- kyphosis .
- spinal tumor .
- failed previous fusion (pseudoarthrosis) .
Concurrence of CDRH, Office of Device Evaluation
1 Prescription Use
OR (Per 21 CFR 801.109) Over-The Counter Use _
Miriam C. Provost
(Division Sign-Off) Division of General, Restorative and Neurological Devices
510(k) Number K013301