← Product Code [DER](/productcode/DER) · K013249

# TINA-QUANT APOLIPOPROTEIN VER.2 (K013249)

_Roche Diagnostics Corp. · DER · Nov 16, 2001 · Immunology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K013249

## Device Facts

- **Applicant:** Roche Diagnostics Corp.
- **Product Code:** [DER](/productcode/DER.md)
- **Decision Date:** Nov 16, 2001
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.5580
- **Device Class:** Class 2
- **Review Panel:** Immunology

## Indications for Use

Immunoturbidmetric assay for the in vitro quantitative determination of apolipoprotein A-1 in human serum and plasma on automated clinical chemistry analyzers. A lipoprotein test system is a device intended to measure lipoprotein in serum and plasma. Lipoprotein measurements are used in the diagnosis and treatment of lipid disorders and atherosclerosis.

## Device Story

Tina-quant Apolipoprotein A-1 ver.2 is an in vitro diagnostic immunoturbidimetric assay. It uses anti-apolipoprotein A-1 antibodies to react with antigens in human serum or plasma samples, forming antigen/antibody complexes. These complexes are measured turbidimetrically on automated clinical chemistry analyzers. The device is intended for use by laboratory professionals in clinical settings. Results assist healthcare providers in diagnosing and managing lipid disorders and atherosclerosis. The assay provides quantitative measurements of apolipoprotein A-1, supporting clinical decision-making regarding patient lipid profiles.

## Clinical Evidence

Bench testing only. Performance characteristics include precision (within-run CV 0.6-1.0%, between-day CV 1.2-2.4%), analytical sensitivity (0.6 mg/dL), and measuring range (20-400 mg/dL). Method comparison against nephelometric method yielded y = 2.45 + 1.073 mg/dL (r = 0.781). Interference testing showed no significant impact from icterus (up to 60 mg/dL), hemolysis (up to 1000 H index), or lipemia (up to 1000 L index). No cross-reactivity with apolipoprotein B or A-II.

## Technological Characteristics

Immunoturbidimetric assay; utilizes anti-apolipoprotein A-1 antibodies (sheep origin). Designed for use on automated clinical chemistry analyzers. Reagent storage at 2-8°C. Standardized to IFCC reference preparation SP1-01. Quantitative measurement via turbidimetry.

## Regulatory Identification

An alpha-1-lipoprotein immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the alpha-1-lipoprotein (high-density lipoprotein) in serum and plasma. Measurement of alpha-1-lipoprotein may aid in the diagnosis of Tangier disease (a hereditary disorder of fat metabolism).

## Predicate Devices

- Dade Behring N Antisera to Human Apolipoprotein A-1 and Apolipoprotein B assay ([K860894](/device/K860894.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# NOV 1 6 2001

| 510(k) Summary                   | K.013249                                                                                                                                                                                                                                                |
|----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Introduction                     | According to the requirements of 21 CFR 807.92, the following information provides sufficient detail to understand the basis for a determination of substantial equivalence.                                                                            |
| Submitter name, address, contact | Roche Diagnostics Corporation<br>9115 Hague Road<br>Indianapolis, IN 46250<br>(317) 521 - 3544<br><br>Contact Person: Helen T. Torney<br><br>Date Prepared: September 25, 2001                                                                          |
| Device Name                      | Proprietary name: Tina-quant Apolipoprotein ver.2<br><br>Common name: Apolipoprotein A-1<br><br>Classification name: Alpha-1- lipoprotein immunological test system                                                                                     |
| Device Description               | A device for the measurement of human apolipoprotein A-1 in serum or plasma. Anti-apolipoprotein A-1 antibodies react with the antigen in the sample to form antigen/antibody complexes which, following agglutination, are measured turbidimetrically. |
| Intended use                     | Immunoturbidmetric assay for the in vitro quantitative determination of apolipoprotein A-1 in human serum and plasma on automated clinical chemistry analyzers.                                                                                         |
| Indications for Use              | A lipoprotein test system is a device intended to measure lipoprotein in serum and plasma. Lipoprotein measurements are used in the diagnosis and treatment of lipid disorders and atherosclerosis.                                                     |

{1}------------------------------------------------

#### Substantial The Tina-quant Apolipoprotein A-1 ver.2 is substantially equivalent to other devices legally marketed in the United States. We claim equivalence to the Equivalence Dade Behring N Antisera to Human Apolipoprotein A-1 and Apolipoprotein B assay (K860894).

Substantial equivalence similarities

The following table compares the Tina-quant Apolipoprotein A-1 ver.2 Assay with the predicate device.

| Feature                           | Tina-quant<br>Apolipoprotein A-1 ver.2                                                                                                                                                                                                                                                                             | Apolipoprotein A-1<br>(predicate)                                                                                                                                                                                                                                                                                  |
|-----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use                      | Immunoturbidmetric assay<br>for the in vitro quantitative<br>determination of<br>apolipoprotein A-1 in<br>human serum and plasma<br>on automated clinical<br>chemistry analyzers.                                                                                                                                  | In vitro diagnostic reagent for<br>the quantitative determination<br>of apolipoprotein A-1 in<br>human serum with the<br>Behring nephelometers.                                                                                                                                                                    |
| Indication for<br>Use             | For the quantitative<br>determination of<br>apolipoprotein A-1 in serum<br>and plasma. A lipoprotein<br>test system is a device<br>intended to measure<br>lipoprotein in serum and<br>plasma. Lipoprotein<br>measurements are used in<br>the diagnosis and treatment<br>of lipid disorders and<br>atherosclerosis. | For the quantitative<br>determination of<br>apolipoprotein A-1 in serum<br>and plasma. A lipoprotein<br>test system is a device<br>intended to measure<br>lipoprotein in serum and<br>plasma. Lipoprotein<br>measurements are used in<br>the diagnosis and treatment<br>of lipid disorders and<br>atherosclerosis. |
| Assay Protocol                    | Immunoturbidometric                                                                                                                                                                                                                                                                                                | Immunoturbidometric                                                                                                                                                                                                                                                                                                |
| Traceability /<br>Standardization | Standardized with regard to<br>the IFCC reference<br>preparation SP1-01.                                                                                                                                                                                                                                           | Not provided in insert                                                                                                                                                                                                                                                                                             |
| Calibration<br>Interval           | • After each lot<br>• as required by QC<br>procedures                                                                                                                                                                                                                                                              | • After each lot<br>• as required by QC<br>procedures                                                                                                                                                                                                                                                              |

{2}------------------------------------------------

### Substantial equivalence differences

The following table compares the Tina-quant Apolipoprotein A-1 ver.2 Assay with the predicate device.

| Feature              | Tina-quant Apolipoprotein<br>A-1 ver.2                                              | Apolipoprotein A-1<br>(predicate)                                                                                                                                                                                                                                                                                  |
|----------------------|-------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sample Type          | Serum and plasma (heparin,<br>EDTA)                                                 | Serum                                                                                                                                                                                                                                                                                                              |
| Reagent<br>Stability | • Store at 2-8°C,<br>unopened.<br>• 42 days opened and<br>refrigerated on analyzer. | • Store at 2-8°C,<br>unopened.<br>• Use within 4 weeks, if<br>directly after use if vials<br>are stopped, capped and<br>stored at 2-8°C.<br>• Do not use remaining<br>antiserum if left open on<br>nephelometer for longer<br>than 5 days at 8 hours<br>daily or comparable<br>period of time.<br>• Do not freeze. |
| Calibrator           | C.f.a.s. Lipids                                                                     | N Apoliporprotein Standard<br>Serum (human)                                                                                                                                                                                                                                                                        |
| Controls             | Precinorm L, Precipath L                                                            | Apolipoprotein Control<br>Serum CHD (human)                                                                                                                                                                                                                                                                        |
| Expected<br>Values   | Females: 108 - 225 mg/dL<br>Males: 104 - 202 mg/dL                                  | Females: 1.25 - 2.15 g/L<br>Males: 1.10-2.05 g/L                                                                                                                                                                                                                                                                   |
| Instrument           | Roche/Hitachi Clinical<br>Chemistry Analyzers                                       | Dade Behring<br>Nephelometers                                                                                                                                                                                                                                                                                      |
| Measuring<br>Range   | 20 - 400 mg/dL                                                                      | Not provided in insert                                                                                                                                                                                                                                                                                             |

{3}------------------------------------------------

Substantial equivalence performance characteristics The performance characteristics of the Tina-quant Apolipoprotein A-1 ver.2 Assay and the predicate device are compared in the table below.

| Feature                            | Tina-quant Apolipoprotein<br>A-1 ver.2                                                                                                                                                                                                                              | Apolipoprotein A-1<br>(predicate)                                                                                                                                            |
|------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Precision                          | Within run CV<br>1.0% @ 40 mg/dL (serum)<br>0.6% @ 176 mg/dL (serum)<br>1.0% @ 157 mg/dL (control)<br>1.2% @ 83 mg/dL (control)<br>Between Day CV<br>2.4% @ 47 mg/dL (serum)<br>1.6% @ 179 mg/dL (serum)<br>1.2% @ 171 mg/dL (control)<br>2.4% @ 84 mg/dL (control) | Inter-assay Precision<br>2.2% CV @ 1.58 g/L<br><br>Intra-assay Precision<br>5.7% CV @ 1.45 g/L                                                                               |
| Method<br>Comparison               | Bablok/Passing:<br>Tina-quant Apolipoprotein A-1<br>ver.2 (Y) / Nephelometric<br>method (X).<br>y = 2.45+1.073 mg/dL<br>r = 0.781                                                                                                                                   | Dade Behring N Antisera<br>Apo A-1 (Y) /<br>radioimmunodiffusion<br>commerical method (X):<br>y(BN)= 1.0 (RID) – 0.04<br>g/L<br>r= 0.98                                      |
| Hook Effect                        | No effect up to 600 mg/dL                                                                                                                                                                                                                                           | NA                                                                                                                                                                           |
| Analytical<br>sensitivity<br>(LDL) | 0.6 mg/dL                                                                                                                                                                                                                                                           | Established by the lower<br>limit of the reference curve<br>and depends therefore<br>upon the concentration of<br>the proteins in the N<br>Apolipoprotein Standard<br>Serum. |

{4}------------------------------------------------

Substantial equivalence performance characteristics, cont.

The performance characteristics of the Tina-quant Apolipoprotein A-1 ver.2 Assay and the predicate device are compared in the table below.

| Feature     | Tina-quant Apolipoprotein<br>A-1 ver.2                                                                                                                                                                                                                                                                                                      | Apolipoprotein A-1                                                                                                                                                                                                                                                                                                                                                                                                                         |
|-------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Limitations | Icterus: No significant interference up to an I index of 60 mg/dL (conjugated and unconjugated) Hemolysis: No significant interference up to an H index of 1000. Lipemia: No significant interference up to an L index of 1000. Anti-human apolipoprotein A-1 antibodies from sheep show no cross-reactivity with apolipoprotein B or A-II. | Turbidity and particles in the sample can interfere with the test. Therefore particulates resulting from incompleted coagulation or denaturation of proteins should be removed prior to assay by centrifugation. In isolated cases excessive concentrations of triglycerides or hyperlipemic samples may disturb the Apo B assay. In such cases the effect of the disturbance can be reduced by retesting the sample in a higher dilution. |

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the circle is a stylized image of an eagle.

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Ms. Helen T. Torney Regulatory Submissions, Centralized Diagnostics Roche Diagnostics Corporation 9115 Hague Road P.O. Box 50457 Indianapolis, IN 46250-0457

NOV 1 6 2001

Re: k013249

> Trade/Device Name: Tina-quant Apolipoprotein A-1 ver.2 Regulation Number: 21 CFR 866.5580 Regulation Name: Alpha-1-lipoprotein immunological test system Regulatory Class: Class II Product Code: DER Dated: September 25, 2001 Received: September 28, 2001

Dear Ms. Torney:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{6}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and 1 additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, (1) = 1 = 1 = 1 = 1 = 1 = 1 = 1 = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrb/dsma/dsmamain.html".

Sincerely yours,

Steven Gutman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory-Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

### Indications for Use Statement

Ko13249 510(k) Number (if known): N/A

Device Name: Tina-quant Apolipoprotein A-1 ver.2

Indications For Use:

Immunoturbidmetric assay for the in vitro quantitative determination of apolipoprotein A-1 in human serum and plasma on automated clinical chemistry analyzers.

A lipoprotein test system is a device intended to measure lipoprotein in serum and plasma. Lipoprotein measurements are used in the diagnosis and treatment of lipid disorders and atherosclerosis.

Prescription Use
(Per 21 CFR 801.109)

OR

Over-The-Counter Use

(Optional Format 1-2-96)

(Division Sign-Off)
Division of Clinical Laboratory Levices
510(k) Number.

vision of Clinical Laboratory Devices
510(k) Number K013249

---

**Source:** [https://fda.innolitics.com/device/K013249](https://fda.innolitics.com/device/K013249)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
