MODIFIED INS-1 INTRAOPERATIVE NERVE SURVEILLANCE SYSTEM
K013215 · Nuvasive, Inc. · BXM · Oct 16, 2001 · Anesthesiology
Device Facts
Record ID
K013215
Device Name
MODIFIED INS-1 INTRAOPERATIVE NERVE SURVEILLANCE SYSTEM
Applicant
Nuvasive, Inc.
Product Code
BXM · Anesthesiology
Decision Date
Oct 16, 2001
Decision
SESE
Submission Type
Special
Regulation
21 CFR 868.2775
Device Class
Class 2
Indications for Use
The NeuroVision JJB System is intended to provide intraoperative electromyographic (EMG) surveillance to assist in the location of nerves during percutaneous surgery of the spine, by administration of brief electrical stimulus pulses to tissues and nerves at the operative site, and EMG monitoring of muscle groups associated with those nerves. The NeuroVision JJB System is designed for use in conjunction with the NuVasive guided spinal arthroscopy system to assist in gaining controlled percutaneous access to the spinal nerve root, foramina, intervertebral disc, and surrounding tissues of the spine via uniportal or biportal posterior or posterolateral approach, where anatomical restrictions safely permit.
Device Story
NeuroVision JJB System is an intraoperative EMG monitor/stimulator for spinal surgery. Inputs: electrical signals from EMG electrodes placed on muscle groups; stimulation pulses delivered via cannula or hand-held probe. Operation: device delivers electrical pulses to tissues; monitors EMG response to detect nerve proximity. 'Detection' routine alerts surgeon when cannula tip nears nerve; 'Direction' routine uses four orthogonal electrodes to determine nerve orientation, displayed on touch screen. 'Screw test' algorithm uses hand-held probe to check for pedicle wall breach by measuring depolarization threshold current. Output: visual alerts and threshold data on touch screen. Used in OR by surgeons to guide percutaneous access and verify screw placement. Benefits: assists in avoiding nerve damage during spinal procedures.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
System includes reusable patient module, control unit with embedded computer, touch screen, conductive probe cables, electrodes, and leads. Employs electrical nerve stimulation and EMG monitoring. Connectivity includes touch screen interface and hard drive for data recording. Sterilization method not specified.
Indications for Use
Indicated for intraoperative EMG surveillance to assist in nerve location during percutaneous spinal surgery. Used in conjunction with NuVasive guided spinal arthroscopy for controlled access to spinal nerve roots, foramina, intervertebral discs, and surrounding tissues via uniportal or biportal posterior or posterolateral approaches.
Regulatory Classification
Identification
An electrical peripheral nerve stimulator (neuromuscular blockade monitor) is a device used to apply an electrical current to a patient to test the level of pharmacological effect of anesthetic drugs and gases.
Predicate Devices
NuVasive INS-1 Intraoperative Nerve Surveillance System (K002677)
Submission Summary (Full Text)
{0}------------------------------------------------
OCT 1 6 2001
Image /page/0/Picture/1 description: The image shows the logo for Nuvasive, a medical device company specializing in spine surgery. The logo consists of a stylized geometric shape to the left of the company name, "NUVASIVE," in all caps. Below the company name is the tagline "Creative Spine Technology" in a smaller, sans-serif font. The logo is simple, modern, and professional, reflecting the company's focus on innovation in the medical field.
K0132/5
### 510(k) Summary
| Submitter: | NuVasive, Inc.<br>10065 Old Grove Road, Suite A<br>San Diego CA 92131 |
|----------------------|------------------------------------------------------------------------------|
| Contact person: | Sean M. Curry<br>16787 Bernardo Center Drive, Suite A<br>San Diego, CA 92128 |
| Phone: | (858) 675-8200 |
| FAX: | (858) 675-8201 |
| Proprietary name: | NeuroVision JJB System |
| Common name: | Electromyography (EMG) monitor/stimulator |
| Classification: | 868.2775 |
| Product Code: | BXM |
| Classification name: | Nerve Stimulator, AC-Powered |
Substantial equivalence claimed to:
K002677: NuVasive INS-1 Intraoperative Nerve Surveillance System
#### Description:
The NeuroVision JJB System (NV JJB) is an electronic device which uses arthroscopic surgical instruments, electrodes, and probes to stimulate nerves with electrical energy, and through the use of electromyography (EMG) electrodes, monitors the sensitivity, and assists in determining the location of nerves during percutaneous surgery of the spine.
The device employs a "Detection" routine which directs brief electrical stimulus pulses through the distal tip of the NuVasive Dilating Cannula as it penetrates tissue to the operative target. Through a series of EMG electrodes placed on the muscle groups associated with the nerves at the operative level of the spine, the NeuroVision JJB System monitors EMG activity in these muscle groups and, when the Cannula tip draws sufficiently near to one such nerve that the electrical energy it emits is strong enough to depolarize the nerve and evoke an EMG response, alerts the surgeon that the Cannula tip is in proximity to that nerve.
Rage 2 1 3 1 NU V A S I V E ", I n c . . 10065 Old Roo ad , S a n Diego , Grove toll free 800.4555.1476 t : 8 5 8 . 2 7 1 . 7 0 7 0 w w w . n u v a s i v e . c o m
{1}------------------------------------------------
By the same method, a "Direction" routine may be used to sequentially activate four orthogonal stimulation electrodes surrounding the Cannula tip, and thereby determine the orthogolial similation closest nerve lies. This direction is displayed on a graphical touch screen.
An additional feature of the NeuroVision JJB System is a hand-held electrical probe All auditional teature of the from transpedicular screw has violated the pedicle wall. which is used to abonain whomer a the screw hole prior to screw insertion or placed on The ban tip of the prooe is placed in as been breached by the screw or tap, the the mstalled screw nead. If the pearlers was the adjacent nerve roots and they will depolarize at a lower stimulation current. The screw test algorithm determines the depolarization (threshold) current for all 8 EMG channels. The surgeon may also set a baseline threshold) current by stimulating a nerve root directly with the probe. The surgeon may choose to display the screw test threshold current relative to this baseline.
The NeuroVision JJB System consists of a reusable Patient Module, a Control Unit comprised of an embedded computer with touch screen controls and an interface card, comprised of an emboduced compater ressable conductive probe cables, electrodes, and electrode leads.
#### Intended use:
The NeuroVision JJB System is intended to provide intraoperative electromyographic (EMG) surveillance to assist in the location of nerves during percutaneous surgery of the (EMO) survermance to access in the lectrical stimulus pulses to tissues and nerves at the spire, by administration of birer elser cross associated with those nerves. The operative site, and EM20 is designed for use in conjunction with the NuVasive guided Nour Islen 30D by bystem to assist in gaining controlled percutaneous access to the Spinal antirosoopy bystein to assist our ounding tissues of the spine via uniportal or biportal ronamina, r posterolateral approach, where anatomical restrictions safely permit.
# Summary of technological characteristics:
The NeuroVision JJB System has Indications for Use which are identical to the predicate I no route roomposed of the same or equivalent materials, has equivalant design features, and has functional characteristics which are equivalent. The user interface is via touch screen display replacing the LED display and push button interface of the previous device, and the use of a hard drive allows recording system parameters during use.
Due to the equivalency of indications for use, materials of composition, design features, method of use, and functional characteristics, the device raises no new safety or effectiveness issues.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is a stylized image of an eagle with its wings spread, with three horizontal lines above the eagle's head. The seal is black and white.
OCT 1 6 2001
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Sean Curry Chief Operating Officer Certified Software Solutions, Inc. 16787 Bernardo Center Drive Suite A San Diego, California 92128
Re: K013215
Trade Name: Neuro Vision JJB System Regulation Number: 868.2775 Regulation Name: Electrical peripheral nerve stimulator Regulatory Class: Class II Product Code: BXM Dated: September 25, 2001 Received: September 26, 2001
Dear Mr. Curry:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
## Page 2 - Mr. Sean Curry
This letter will allow you to begin marketing your device as described in your Section 510(k) This letter will anow you to begin manoung your manteing your device of your device to a legally prematication. The PDF intellig sification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and If you desire specific arrive for your avitro diagnostic devices), please contact the Office of additionally 21 CFR Part 007.10 Idditionally, for questions on the promotion and advertising of Compliance at (301) 597-1697 First of Compliance at (301) 594-4639. Also, please note the your device, produce of movending by reference to premarket notification" (21CFR Part 807.97). It guiation entitied, "Misorananing of soonsibilities under the Act may be obtained from the Oiler general information on your seems and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Susan
Celia M. W.
Sincerely yours,
![]()
Celia M. Witten, Ph.D., MD Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Number (if known):
Device Name: NuVasive NeuroVision JJB System
Indications for Use:
The NeuroVision IJB System is intended to provide intraoperative electromyographic (EMG) surveillance to assist in the location of nerves during percutaneous surgery of the spine, by administration of brief electrical stimulus pulses to tissues and nerves at the operative site, and EMG monitoring of muscle groups associated with those nerves. The NeuroVision JJB System is designed for use in conjunction with the NuVasive guided spinal arthroscopy system to assist in gaining controlled percutaneous access to the spinal nerve root, foramina, intervertebral disc, and surrounding tissues of the spine via uniportal or biportal posterior or posterolateral approach, where anatomical restrictions safely permit.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of General, Restorative
and Neurological Devices
Prescription Use
(Per 21 CFR 801.109)
OR
510(k) Number
K013215
Over-the-Counter Use
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.