K013164 · Terumo Medical Corp. · FPA · Oct 15, 2001 · General Hospital
Device Facts
Record ID
K013164
Device Name
SURSHIELD SAFETY WINGED BLOOD COLLECTION SET
Applicant
Terumo Medical Corp.
Product Code
FPA · General Hospital
Decision Date
Oct 15, 2001
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.5440
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The TERUMO® SURSHIELD™ SAFETY WINGED BLOOD COLLECTION SET is a winged blood collection needle intended for venipuncture to collect blood specimens from patients. The TERUMO® SURSHIELD™ SAFETY WINGED BLOOD COLLECTION SET is also indicated for intravenous administration of fluids after removing the attached luer adapter from the blood collection set connector and attaching a syringe, or other compatible/appropriate device. This device may be used for any patient population with consideration given to patient size, appropriateness for the solution being infused, and duration of therapy. Additionally, after withdraw of the needle from the patient's vein, the attached needle safety shield can be manually activated to cover the needle immediately after use to minimize risk of accidental needlestick.
Device Story
Sterile, single-use winged blood collection set; consists of needle, winged hub, tubing, and female connector with multi-sample luer adapter. Manual operation; user performs venipuncture for blood collection or fluid infusion. Features hinged, transparent safety shield attached to wing; user manually rotates shield 180 degrees and pushes to latch over needle post-withdrawal to prevent needlestick injuries. Used in clinical settings; operated by healthcare professionals. Benefits include minimized risk of accidental needlestick via single-handed safety mechanism.
Clinical Evidence
Bench testing only. Evaluated leakage at connector/luer junction, blockage, connection strength, and valve protector fit. Biocompatibility tested per ISO-10993. Pyrogen-free status confirmed via LAL gel clot test. Sterilization validated per ISO11135/EN550.
Technological Characteristics
Manual winged needle set; materials identical to predicates K010103 and K983490. Includes hinged safety shield mechanism. Sterilized via Ethylene Oxide (EtO) to SAL 10^-6. Tubing lengths 180mm-300mm. No electronic components or software.
Indications for Use
Indicated for venipuncture to collect blood specimens and for intravenous fluid administration in any patient population, considering patient size, therapy duration, and solution appropriateness.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
Predicate Devices
TERUMO® SURSHIELD™ SAFETY WINGED INFUSION SET (K010103)
VACUTAINER® SAFETY-LOK™ BLOOD COLLECTION SET (K980414)
Submission Summary (Full Text)
{1}------------------------------------------------
# OCT 1 5 2001
K013164
# SURSHIELD™ SAFETY WINGED BLOOD COLLECTION SET Section II - Summary & Certification
# SUMMARY OF SAFETY & EFFECTIVENESS INFORMATION PERTAINING TO SUBSTANTIAL EQUIVALENCE
# A. Device Name
Proprietary Device Name: TERUMO® SURSHIELD™ Safety Winged Blood Collection Set
Classification Name: Tubes, vials, systems, serum separators, blood collection and Intravascular Administration Sets
Product Code(s): 75 JKA and 80 FPA
## B. Reason for Submission:
This premarket notification (510(k)) is being submitted for the SURSHIELD™ SAFEY WINGED BLOOD COLLECTION SET, which is blood collection device comprised of two cleared devices both manufactured by Terumo Corporation in Tokyo, Japan. The base predicate device is the SURSHIELD™ SAFETY WINGED INFUSION SET cleared under 510(k) number K010103 and the attached device is the TERUMO® VENOJECT® LUER ADAPTER cleared under 510(k) number K983490. The focus of this submission is to address any potential issues of safety and effectiveness of the device being formed after joining the two cleared devices.
# C. Intended Use:
The TERUMO® SURSHIELD™ SAFETY WINGED BLOOD COLLECTION SET is a winged blood collection needle intended for venipuncture to collect blood specimens from patients.
The TERUMO® SURSHIELD™ SAFETY WINGED BLOOD COLLECTION SET is also indicated for intravenous administration of fluids after removing the attached luer adapter from the blood collection set connector and attaching a syringe, or other compatible/appropriate device. This device may be used for any patient population with consideration given to patient size, appropriateness for the solution being infused, and duration of therapy.
Additionally, after withdraw of the needle from the patient's vein, the attached needle safety shield can be manually activated to cover the needle immediately after use to minimize risk of accidental needlestick.
Note: This is the same intended use as the predicate device, Surshield Winged Infusion Set -K010103.
{2}------------------------------------------------
# D. Description
The Surshield Safety Winged Blood Collection Set is a sterile, single-use device consisting of a needle attached to a winged type hub, tubing, a female connector with a multi-sample luer adapter attached for blood collection. The Surshield Safety Winged Blood Collection Set is the same as the Surshield Safety Winged Infusion Set (K010104) except for the attached luer adapter (K983490).
A hinged safety shield, previously cleared under K010103, is attached to the wing just below the needle-to-wing junction. The safety shield can be turned 180 degrees on the hinge. As the needle is removed from the patient's vessel, the user's finger actively pushes the shield cover until it latches onto needle using a single-handed technique. The shield cover is designed to allow the user's finger to remain behind the needle point so that the risk of needlestick injury is minimized. The shield cover is transparent for easy confirmation of the needle held in it.
## E. Substantial Equivalence
The SURSHIELD™ SAFETY WINGED BLOOD COLLECTION SET is substantially equivalent in intended use, design, technology/principals of operation, materials, and performance to the following cleared devices:
- TERUMO® SURSHIELD SAFETY WINGED INFUSION SET K010103 1.
- TERUMO® VENOJECT® LUER ADAPTER K983490 2.
- VACUTAINER® SAFETY-LOK™ BLOOD COLLECTION SET K980414 3.
Differences between the devices do not raise any significant issues of safety and effectiveness.
#### F. Principals of Operation/Technology
The Surshield Safety Winged Blood Collection Set, Surshield Safety Winged Infusion Set, Venoject Luer Adapter, and Vacutainer Safety-Lok Blood collection set are all operated manually.
#### G. Materials
The Surshield Safety Winged Blood Collection Set is comprised of the same materials as the cleared Surshield Safety Winged Infusion Set (K010103) and Venoject Luer Adapter (K983490).
{3}------------------------------------------------
# H. Specifications
| Cannula<br>gauge | Color code | Product code | Cannula<br>length | Tubing<br>length |
|------------------|------------|-----------------------|-------------------|------------------|
| 19G | Cream | MN*SVS19B303/4"(19mm) | | 300mm±10% |
| | | MN*SVS19B183/4"(19mm) | | 180mm±10% |
| 21G | Green | MN*SVS21B303/4"(19mm) | | 300mm±10% |
| | | MN*SVS21B183/4"(19mm) | | 180mm±10% |
| 23G | Light blue | MN*SVS23B303/4"(19mm) | | 300mm±10% |
| | | MN*SVS23B183/4"(19mm) | | 180mm±10% |
| 25G | Orange | MN*SVS25B303/4"(19mm) | | 300mm±10% |
| | | MN*SVS25B183/4"(19mm) | | 180mm±10% |
Note: The 180mm length tubing is new for this blood collection device but is bracketed by the 90 and 300mm lengths cleared under K010103.
# I. Performance
The following performance tests were performed on the Surshield Safety Winged Blood Collection Set:
- Leakage at connector to luer adapter junction
- . Blockage
- t Connection strength of luer adapter to connector
- . Valve Protector Fit
Additionally, a risk analysis was conducted and any potential issues were addressed through design modification, and/or labeling as appropriate. None of the data raises any new issues of safety and effectiveness.
# J. Additional Safety Information
- a. Sterilization
The EtO sterilization conditions are validated according to ISO11135/EN550 to provide a Sterility Assurance Level (SAL) of 10-6.
Ethylene Oxide residual levels resulting from EtO sterilization will not exceed the maximum residue levels proposed for Part 821 of Title 21 in the Federal Register Notice issued June 23, 1978, and indicated as follows:
| Ethylene Oxide | 25 ppm |
|-----------------------|---------|
| Ethylene Chlorohydrin | 25 ppm |
| Ethylene Glycol | 250 ppm |
{4}------------------------------------------------
# b. Biocompatibility
The device's blood contacting materials were tested in accordance with the tests recommended in the FDA General Program Memorandum #G95-1 (5/1/95): Use of International Standard ISO-10993. "Biological Evaluation of Medical Devices Part-1: Evaluation and Testing. Results of the testing demonstrate that the blood contacting materials are biocompatible. Therefore, no further testing was deemed necessary.
c. Expiration Dating
Expiration Dating for the Surshield Safety Winged Blood Collection Set will be 36 months (3 years).
d. Pyrogen Testing
The Surshield Safety Winged Blood Collection Set is pyrogen free. Testing is performed on each lot of the product using the Limulus Amebocyte Lysate (LAL) gel clot test according to the US Pharmacopoeia XXIV.
# K. Conclusion
In summary, the Surshield Safety Winged Blood Collection Set is substantially equivalent in intended use, design, technology/principals of operation, materials, and performance to the following cleared devices:
- 1. TERUMO® SURSHIELD SAFETY WINGED INFUSION SET K010103
- TERUMO® VENOJECT® LUER ADAPTER K983490 2.
- 3. VACUTAINER® SAFETY-LOKTM BLOOD COLLECTION SET-K980414
Differences between the devices do not raise any significant issues of safety and effectiveness.
Terumo's statement that this device is substantially equivalent to any other device is done solely to comply with the requirements of the Federal Food, Drug and Cosmetic Act and is not intended whatsoever to be the basis for a patent infringement action.
Date Prepared: September 20, 2001
Prepared by: Barbara Smith Regulatory Affairs Specialist Terumo Medical Corporation 125 Blue Ball Road Elkton, Maryland 21921 Phone#: 410-392-7241 Fax#: 410-398-6079
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized eagle-like symbol with three curved lines representing its wings or feathers. The logo is in black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT 1 5 2001
Ms. Barbara Smith Regulatory Affairs Specialist Terumo Medical Corporation 125 Blue Ball Road Elkton, Maryland 21921
Re: K013164
Trade/Device Name: Surshield™ Safety Winged Blood Collection Set Regulation Number: 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: II Product Code: FPA Dated: September 20, 2001 Received: September 21, 2001
Dear Ms. Smith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{6}------------------------------------------------
Page 2 - Ms. Smith
You must comply with all the Act's requirements, including, but not limited to: registration 1 ou interest 011); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours.
\$\substack{f \ y}\$ \$\substack{S \ Ti}\$
noves Timothy A. Ulatowski
Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
510(k) Number (if known):
# K013/64
#### SURSHIELD™ SAFETY WINGED BLOOD COLLECTION SET Device Name:
### Indications For Use:
The TERUMO® SURSHIELD™ SAFETY WINGED BLOOD COLLECTION SET is a winged blood collection needle intended for venipuncture to collect blood specimens from patients.
The TERUMO® SURSHIELD™ SAFETY WINGED BLOOD COLLECTION SET is also indicated for intravenous administration of fluids after removing the attached luer adapter from the blood collection set connector and attaching a syringe, or other compatible/appropriate device. This device may be used for any patient population with consideration given to patient size, appropriateness for the solution being infused, and duration of therapy.
Additionally, after withdraw of the needle from the patient's vein, the attached needle safety shield can be manually activated to cover the needle immediately after use to minimize risk of accidental needlestick.
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) |
|--------------------------------------------------------|
|--------------------------------------------------------|
| Prescription Use<br>(Per 21 CFR 801.109) | OR | Over-The-Counter Use |
|------------------------------------------|----|----------------------|
|------------------------------------------|----|----------------------|
| (Division Sign-Off) | B. Brede |
|---------------------------------------------------------------------|----------|
| Division of Dental, Infection Control, and General Hospital Devices | |
| 510(k) Number | K013164 |
| (Optional Format 1-2-96) |
|--------------------------|
|--------------------------|
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.