← Product Code [DFC](/productcode/DFC) · K013128

# N LATEX LP(A) (K013128)

_Dade Behring, Inc. · DFC · Jan 18, 2002 · Immunology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K013128

## Device Facts

- **Applicant:** Dade Behring, Inc.
- **Product Code:** [DFC](/productcode/DFC.md)
- **Decision Date:** Jan 18, 2002
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.5600
- **Device Class:** Class 2
- **Review Panel:** Immunology

## Indications for Use

N Latex Lp(a) is an in vitro diagnostic reagent for the quantitative determination of lipoprotein(a) [Lp(a)] in human serum and heparinized plasma by means of particle enhanced immunonephelometry using BN™ Systems. Measurement of Lp(a) aids in the identification of individuals at risk from cardiovascular disease in specific populations when used in conjunction with clinical evaluation.

## Device Story

In vitro diagnostic reagent; uses polystyrene latex particles coated with anti-human Lp(a) antibodies. Operates via particle-enhanced immunonephelometry on BN™ Systems. Input: human serum or heparinized plasma samples. Mechanism: agglutination of latex particles upon binding to Lp(a) in sample; intensity of scattered light measured by BN™ Systems is proportional to Lp(a) concentration. Output: quantitative Lp(a) concentration determined by comparison against known standards. Used in clinical laboratory settings by trained personnel. Results assist healthcare providers in assessing cardiovascular disease risk and lipid metabolism disorders.

## Clinical Evidence

Bench testing only. Correlation study compared N Latex Lp(a) to a commercially available immunonephelometric method using 86 serum samples (range 0.02–1.31 g/L), yielding a correlation coefficient of 0.96 (Passing-Bablok slope 0.92, intercept -0.005). Intra-assay precision (n=20) showed CVs of 1.8–4.1%. Inter-assay reproducibility (n=10) showed CVs of 2.8–5.3%.

## Technological Characteristics

Polystyrene latex particles coated with specific antibodies to human Lp(a). Principle: particle-enhanced immunonephelometry. Energy source: light scattering measurement via BN™ Systems. Form factor: liquid reagent. Connectivity: integrated with BN™ Systems instrumentation.

## Regulatory Identification

A low-density lipoprotein immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the low-density lipoprotein in serum and other body fluids. Measurement of low-density lipoprotein in serum may aid in the diagnosis of disorders of lipid (fat) metabolism and help to identify young persons at risk from cardiovascular diseases.

## Predicate Devices

- Beckman Coulter Array® System LPA assay ([K000121](/device/K000121.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# JAN 1 8 2002

## 510(k) Summary for N Latex Lp(a) Reagent

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) number is: _______________________________________________________________________________________________________________________________________________

#### Manufacturer's Name, Address, Telephone, and Contact Person, Date of 1. Preparation:

Manufacturer:

Dade Behring Marburg GmbH Emil-von-Behring Str. 76 Marburg/Germany

Contact Information:

Dade Behring Inc. Glasgow Site P.O. Box 6101 Newark, Delaware 19714 Attn: Rebecca S. Ayash Tel: 302-631-6276

Preparation date:

January 4, 2002

#### Device Name/ Classification: 2.

| N Latex Lp(a) Reagent: | Low-density lipoprotein immunological test system, |
|------------------------|----------------------------------------------------|
|                        | Class II (866.5600)                                |
| Product Code:          | 81 DFC                                             |

#### Identification of the Legally Marketed Device: 3.

Beckman Coulter Array® System LPA assay (K000121)

#### 4. Device Description:

Polystyrene latex particles coated with specific antibodies to human Lp(a) are agglutinated when mixed with samples containing Lp(a). The intensity of scattered light in the BN™ Systems depends on the Lp(a) concentration in the sample. The concentration can therefore be determined by comparison with dilutions of a standard of known concentration.

{1}------------------------------------------------

#### 5. Device Intended Use:

N Latex Lp(a) is an in vitro diagnostic reagent for the quantitative determination of lipoprotein(a) [Lp(a)] in human serum and heparinized plasma by means of particle enhanced immunonephelometry using BN™ Systems. Measurement of Lp(a) aids in the identification of individuals at risk from cardiovascular disease in specific populations when used in conjunction with clinical evaluation.

#### Medical device to which equivalence is claimed and comparison information: 6.

There are a number of in vitro diagnostic products in commercial distribution, which employ immunoassay techniques for the quantitative determination Lp(a) in human serum or plasma. One such product is the Beckman Coulter Array® System LPA assay (K000121). The N Latex Lp(a) Reagent is substantially equivalent in intended use and results obtained to the Array System LPA assay. The N Latex Lp(a) Reagent, like the Array® System LPA assay is intended to be used for the quantitative determination of Lp(a) activity in human serum or plasma.

#### 7. Device Performance Characteristics:

### Correlation:

The N Latex Lp(a) assay was compared to a commercially available Lp(a) immunonephelometric method by evaluating 86 laboratory serum samples ranging from 0.02 to 1.31 g/L. A correlation coefficient of 0.96 was obtained with a y-intercept value of -0.005 and a slope of 0.92 (Passing-Bablok).

### Precision:

The N Latex Lp(a) Reagent was used to measure the N Lp(a) Control SY and 6 patient samples or pools with Lp(a) concentrations between 0.29 and 1.77 g/L and yielded coefficients of variation between 1.8 and 4.1% for the intra-assay precision (n=20). N Lp(a) Control SY and 5 patient samples or pools with Lp(a) concentrations between 0.17 and 1.75 g/L were used to determine the inter-assay reproducibility (n=10) and the coefficients of variation were between 2.8 and 5.3%.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure, represented by three curved lines that suggest feathers or wings.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

JAN 1 8 2002

Ms. Kathleen A. Dray-Lyons Manager, Regulatory Affairs Dade Behring Inc. Glasgow Site, P.O. Box 6101 Newark, DE 19714

> k013128 Trade/Device Name: N Latex Lp (a) Reagent Regulation Number: 21 CFR 866.5600 Regulation Name: Low-density lipoprotein immunological test system Regulatory Class: Class II Product Code: DFC Dated: January 4, 2002 Received: January 7, 2002

Dear Ms. Dray-Lyons:

Re:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Steven Gutman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# 4013128 Indications Statement

N Latex Lp(a) Reagent Device Name:

## Indications for Use:

N Latex Lp(a) Reagent is an in vitro diagnostic test for the quantitative determination of Lipoprotein(a) [Lp(a)] in human serum or plasma. N Latex Lp(a) aids in the diagnosis of disorders of lipid (fat) metabolism and helps to identify persons at risk from cardiovascular disease in specific populations when used in conjunction with clinical evaluation.

(Division Sign-(
Division of Clir. oratory Devices
510(k) Number. K013 128

(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF . NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use
(Per 21 CFR 801.109)

Over-The-Counter-Use (Optional Format 1-2-96)

000056

---

**Source:** [https://fda.innolitics.com/device/K013128](https://fda.innolitics.com/device/K013128)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
