K013112 · Computerized Medical Systems, Inc. · MUJ · Dec 5, 2001 · Radiology
Device Facts
Record ID
K013112
Device Name
FOCAL SIM
Applicant
Computerized Medical Systems, Inc.
Product Code
MUJ · Radiology
Decision Date
Dec 5, 2001
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The FOCAL workstation software is a computer software package intended to be used as an accessory to a radiation treatment planning system. FOCAL Sim is intended to permit CT simulation to be performed on the FOCAL workstation. The CT scan is read into the radiation treatment planning system and then sent to the FOCAL workstation. On FOCAL, the user is able to identify patient isocenters, place treatment beams and identify beam modifiers (blocks, wedges, etc.). This information is then passed back to the radiation treatment planning system for storage and documentation of the resultant treatment plan and calculation of patient dose based on this information. The resultant plan is to be evaluated, modified as necessary, approved and delivered by qualified medical personnel.
Device Story
FOCAL Workstation is a remote PC-based software accessory for radiation treatment planning (RTP) systems. It offloads contouring and simulation tasks from high-powered UNIX-based RTP workstations. Inputs include CT and MR images transferred from the RTP system. Users (radiation oncologists/technicians) perform manual contouring, autosegmentation, and CT simulation (identifying isocenters, placing beams, and defining modifiers like blocks/wedges). Outputs are processed data sent back to the RTP system for final dose calculation and documentation. The device enables remote viewing of treatment plans, dose volume histograms (DVH), and digitally reconstructed radiographs (DRR). It facilitates clinical decision-making by allowing clinicians to compare alternate treatment plans and select the optimal prescription. Benefits include improved workflow efficiency by freeing up primary RTP system resources.
Clinical Evidence
No clinical data. Bench testing only. The device is a software module that migrates existing RTP system tasks to a remote workstation; clinical testing was deemed unnecessary to demonstrate safety and effectiveness.
Technological Characteristics
Software module for PC (Windows-based). Supports standard CT scanners and Varian, Siemens, and Philips/Elekta linear accelerators. Features include manual contouring, autosegmentation, and CT simulation tools. Connectivity via data transfer to/from primary RTP system.
Indications for Use
Indicated for use as an accessory to radiation treatment planning systems to perform CT simulation, including identification of patient isocenters, placement of treatment beams, and identification of beam modifiers for radiation therapy planning.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
Predicate Devices
Picker International Acqsim Simulator/Localizer (K923851)
{0}------------------------------------------------
DEC 0 5 2001
K013112
# 510(k) Summary of Safety and Effectiveness Computerized Medical Systems Premarket Notification FOCAL CT Simulation (FOCAL SIM)
| Submitter Name: | Computerized Medical Systems, Inc. |
|--------------------------|----------------------------------------------------------------------------------------------------------------------------|
| Submitter Address: | 1145 Corporate Lake Drive<br>St. Louis, MO 63132-1716 |
| Submitter Phone: | 314 993 0003 |
| Submitter Fax: | 314 993-0075 |
| Contact Person: | Michael A. Parsons - Director - Quality Assurance<br>and Regulatory Affairs |
| Date Summary Prepared: | September, 2001 |
| Device Trade Name: | FOCAL Workstation |
| Device Common Name: | Accessory to Radiation Treatment Planning System |
| Device Classification: | System, Simulation, Radiation Therapy per<br>21CFR892.5840 |
| Substantial Equivalence: | Picker International Acqsim Simulator/Localizer -<br>K923851<br>GE Advantage Sim - K951830<br>ADAC Advantage Sim - K993923 |
Device Description: (See Section 7 of this submittal for a more detailed Device Description). The FOCAL Workstation was initially cleared for marketing under K981535. (At that time, the device was referred to as the FOCUS Pilot Contouring Workstation. However, by the time the Premarket Notification was cleared, the CMS Marketing Department had selected another trade name.)
The initial release of the product had, as its intended use, the remote contouring of patient outlines, structures and tumors as part of radiation therapy planning. The FOCAL
{1}------------------------------------------------
### 510(k) Summary of Safety and Effectiveness Computerized Medical Systems Premarket Notification FOCAL CT Simulation (FOCAL SIM) Page 2 of 3
Workstation was designed to work with our FOCUS Radiation Treatment Planning (RTP) W orkstation was designed to work while of the same of the most recently under System willer was intraily ordains repically performed by the Radiation Oncologist, is K002147. The task of contouring, typediation therapy planning and requires a minimum of the most thing consuming task in rad to free up the higher powered UNIX-based RTP Computer Capaonity. It was our goal goal calculation-intensive activities of treatment Workstation for performing the ing to a remote device. This remote device was a prailing by moving the contourning to a common Microsoft products (Windows 95, Fersonal Computer fouded with the FOCAL contouring software running on that While was 20, While was complete, the information was returned to the RTP System to continue the treatment planning process.
The first release of FOCAL contained only manual contouring capability and was given The first ferease of FOCAL Ease". The second release enhanced the users ability to the trade hame of 'I OCAL Lase ' - I vision of CT and MR images as well as providing an autosegmentation capability. This added functionality was given the trade name "FOCAL Fusion".
A later release of FOCAL software provided the capability to view the results of treatment planning performed earlier on the FOCUS RTP System. This included the treatment planning perfermed vans as well as Dose Volume Histograms (DVH's) and abinty to view isodose assations (DRR's). This was given the name "FOCAL Vue" Dighany Roconsalation at a remote capability to view and compare alternate treatment plans and select the one which best satisfied her/his prescription.
All of these development efforts were performed per the CMS software development An of these de relopment of the Premarket Notification clearance. This included inhouse validation testing to assure clinical efficacy and safety.
The subject of this Premarket Notification is the addition of the ability to perform CTV The subject of this Promation, a feature we call "FOCAL Sim". This addition simulation off the contouring of patient targets or viewing of treatment planning results into a more active role in the treatment planning process. For a description of the CT simulation process, refer to Section 7 of this submittal.
{2}------------------------------------------------
## 510(k) Summary of Safety and Effectiveness Computerized Medical Systems Premarket Notification FOCAL CT Simulation (FOCAL SIM) Page 3 of 3
Device Intended Use: The FOCAL Workstation will continue to be used for patient contouring and viewing of treatment plan outputs. With this new software it will also be econtring and for the mations and forward this information to the RTP System for dose calculation.
Summary of Technological Characteristics Compared to Predicate Devices: The CT Simulation capability described in this 510(k) will be a software module "added-on" to the existing FOCAL Workstation previously cleared under K981535 in April, 1999. The FOCAL System with CT Simulation capability incorporates no technological characteristics not currently in the predicate CT Simulation devices.
The FOCAL System supports the most popular CT Scanners as well as the three most common linear accelerators (Varian, Siemens and Philips/Elekta).
Summary of Clinical Testing: Clinical testing was not performed as part of the development of this feature. Clinical testing is not required to demonstrate substantial equivalence or safety and effectiveness. This software merely takes some of the tasks previously performed on the RTP System and allows them to be performed on the FOCAL Workstation.
Summary of Non-Clinical Testing: The testing performed by CMS and our results can be found in Section 9 of this submittal.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the edge. In the center of the seal is an abstract image of an eagle.
Food and Drug Administration 200 Corporate Boulevard Rockville MD 20850
DEC 0 5 2001
Mr. Michael A. Parsons Director, Quality Assurance and Regulatory Affairs Computerized Medical Systems, Inc. TM 1145 Corporate Lake Drive ST. LOUIS MO 63132
### Re: K013112
Trade/Device Name: FOCAL CT Simulation (CT Sim) Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II
Product Code: 90 MUJ Dated: September 14, 2001 Received: September 18, 2001
Dear Mr. Parsons:
We have reviewed your Section 510(k) premarket notification of intent to market the device w one review and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### Statement of Indication for Use
510(k) Number: K013112
Device Name: FOCAL Workstation with CT Simulation Capability
Indication for use: The FOCAL workstation software is a computer software package intended to be used as an accessory to a radiation treatment planning system.
FOCAL Sim is intended to permit CT simulation to be performed on the FOCAL workstation. The CT scan is read into the radiation treatment planning system and then sent to the FOCAL workstation. On FOCAL, the user is able to identify patient isocenters, place treatment beams and identify beam modifiers (blocks, wedges, etc.). This information is then passed back to the radiation treatment planning system for storage and documentation of the resultant treatment plan and calculation of patient dose based on this information. The resultant plan is to be evaluated, modified as necessary, approved and delivered by qualified medical personnel.
> Concurrence of the Center for Devices and Radiological Health, Office of Device Evaluation (ODE)
Prescription Use
✓
OR per 21 CFR 801.109
Over the Counter Use _________________________________________________________________________________________________________________________________________________________
Dave Brion
(Division Sign Off)
Division of Reprodustila, Abdominal,
and Radioteland Daviess K013112
510(k) Number...
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.