THE QUANTECH FASTRAQ AUTOMATED ANALYZER, AND THE QUANTECH PREPAQ TOTAL HCG TEST CARTRIDGE
Applicant
Quantech , Ltd.
Product Code
NAL · Clinical Chemistry
Decision Date
Jan 18, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1155
Device Class
Class 2
Indications for Use
The Quantech PrePaQ Total hCG Test Cartridge is intended for use with the FasTraQ Automated Analyzer to provide rapid, quantitative measurement of human chorionic gonadotropin (hCG) in whole blood for the determination of pregnancy. The Analyzer and hCG Cartridge combine ease of use and rapid turnaround time with laboratory-quality performance and reliability. The FasTraQ System is designed for use by non-laboratory medical professionals in emergency departments, STAT or central laboratories.
Device Story
The FasTraQ System is an automated, point-of-care diagnostic platform for quantitative hCG measurement in whole blood. It utilizes PrePaQ cartridges containing a solid-phase monoclonal antibody immobilized on a gold surface. The device employs surface plasmon resonance (SPR) to perform a two-site sandwich immunoassay; an anti-alpha hCG polyclonal antibody conjugated to alkaline phosphatase binds the antigen, followed by an enhancing solution to increase surface mass. The system includes a microprocessor for instrument control, data acquisition, and data reduction. Operated by non-laboratory medical professionals in emergency departments or STAT labs, the system provides rapid quantitative results. These results assist clinicians in the early detection of pregnancy, facilitating timely clinical decision-making in acute care settings.
Clinical Evidence
Clinical performance evaluated via normal range testing and patient sample correlation. Normal range study (n=apparently healthy) showed 100% agreement with the predicate device for pregnancy determination (>25 mIU/mL). Patient sample correlation (n=not specified) against the Bayer Centaur yielded a correlation coefficient of 0.97 (slope 0.67, intercept 1.47); correlation against Abbott AxSym yielded 0.93 (slope 0.89, intercept -1.7992). Bench testing included dilution linearity (111% recovery), accuracy (105% recovery), analytical sensitivity (2.5 mIU/mL), and precision (interassay CV 5.9-10.6%). No high-dose hook effect observed up to 100,000 mIU/mL.
Technological Characteristics
In vitro sandwich immunoassay using surface plasmon resonance (SPR). Components: PrePaQ cartridge with gold-surface immobilized monoclonal antibody and alkaline phosphatase-conjugated polyclonal antibody. System: Microprocessor-controlled automated analyzer. Connectivity: Not specified. Sterilization: Not specified. Software: Microprocessor-based instrument control and data reduction.
Indications for Use
Indicated for the quantitative measurement of hCG in whole blood to aid in the determination of pregnancy. Intended for use by non-laboratory medical professionals in emergency departments, STAT, or central laboratories.
Regulatory Classification
Identification
A human chorionic gonadotropin (HCG) test system is a device intended for the early detection of pregnancy is intended to measure HCG, a placental hormone, in plasma or urine. A human chorionic goadotropin (HCG) test system is a device intended for any uses other than early detection of pregnancy (such as an aid in the diagnosis, prognosis, and management of treatment of persons with certain tumors or carcinomas) is intended to measure HCG, a placental hormone, in plasma or urine.
Predicate Devices
AVDIA Centaur (K97xxxx)
Reference Devices
Abbott AxSym
Dade Behring RxL
Submission Summary (Full Text)
{1}------------------------------------------------
# JAN 1 8 2002
# QUANTECH, Ltd. Total hCG Assay on the FasTraQ System Premarket Notification
Safety and Effectiveness Summary
(01299
#### Company Information
Quantech Ltd. 815 Northwest Parkway Suite 100 Eagan, MN 55024 (612) 647-6370 Thomas Witty, Ph.D. - Executive Vice President, Research and Development
#### Contact Information
Robin J. Hellen, M.S. Hellen Professional Services (818) 709-5646
### Product Name
Classification Name: Human Chorionic Gonadotroponin (hCG) Test System, Class II Test Trade Name: Quantech PrePaQ Total B-hCG Cartridge PrePaQ Total ß-hCG Test Test Common Name:
Quantech FasTraQ System Trade Name: System Common Name: FasTraQ System
#### CLIA Categorization
We believe the Quantech FasTraQ Automated Analyzer and PrePaQ Total β-hCG assy to be moderately categorized based on previous classifications of medical devices and analogous tests.
# Substantial Equivalence
The Quantech FasTraQ Automated Analyzer and PrePaQ Total ß-hCG assay are substantially equivalent to the AVDIA Centaur (predicate device) marketed by Bayer Corporation since 1997 as well as other commercially available devices and assays (i.e., Abbott AxSym and Dade Behring RxL).
#### Intended Use
The FasTraQ Automated Analyzer in coniunction with the PrePaQ Total hCG Cartridge is intended for use as an aid in the early detection of pregnancy where rapid, quantitative results in the emergency department or where other rapid diagnostics are required. The assay is to be used with patient whole blood samples.
#### Device Description
The PrePaQ Cartridge in coniunction with the FasTraQ Automated Analyer is use to quantitate hCG in whole blood using surface plasmon resonance (SPR). The Quantech PrePaQ Total ß-hCG assay is based on the principle of two site, or sandwich immunoassay in combination with SPR surface mass measurement. Each PrePaQ cartridge contains a solid phase ß-hCG monoclonal antibody immobilized onto a gold surface. An anti-alpha hCG polyclonal antibody conjugated to alkaline phosphatase is attached to the antigen to form the sandwich. An enhancing solution is used to further increase the mass on the surface.
{2}------------------------------------------------
### Comparison of Technological Characteristics
The Quantech PrePaQ Total ß-hCG assay is similar to the Bayer Centaur Total ß-hCG assay as follows. Both assays are in vitro sandwich immunometric assays used for quantitative measurement to total ß-hCG. Although, the Quantech assay uses whole blood, both assays report the results in mIU/ml of serum or plasma. Both assays use a ß-hCG monoclonal antibody coated to a solid support. Both systems utilize a microprocessor for instrument control, data acquisition and data reduction.
# Summary of Non-Clinical Performance Data
Dilution Linearity/Parallelism - The parallelism study was conducted to evaluate the linearity of the Quantech Total ß-hCG Assav. Whole blood samples were separately spiked with intact hCG and serially diluted with corresponding unspiked whole blood. The average percent of expected was 111%.
Recovery - Accuracy of the Quantech Total ß-hCG Assay was calculated from test results as the percentage of added analyte, corrected for endogenous analyte, recovered by the assay. After correcting for endogenous hCG content, the average recovery was 105%.
Analytical Sensitivity - Multiple duplicates of whole blood zero samples were assayed to determine the minimum quantity of hCG detectable by the Quantech Assay. The average SPR signal shift plus two standard deviations (2 S.D.) was calculated and translated into a dose. The calculated analytical sensitivity of the Quantech hCG assay is 2.5 mlU/mL.
Precision - The INTERASSAY precision was determined by evaluating three pools in triplicate on different days. The mean hCG concentrations (with % C.V.) were 57.1 (10.6%), 216.9 (8.1%), and 553.8 (5.9%) mlU/mL for the low, medium and high pools, respectively.
TOTAL IMPRECISION was determined from the interassay data, and is a combination of results from multiple runs on multiple days. The mean hCG concentrations (with % C.V.) were 57.2 (13.6%), 216.9 (12.8%), and 553.7 (9.0%) mlU/mL for the low, medium and high pools, respectively.
Interfering Substances - Physiological interference was evaluated by spiking a whole blood pools with hCG and hemoglobin, bilirubin and triglycerides at levels ten times the highest expected physiological concentration. The percent recovery of hCG was determined to be acceptable in all three solutions and no interference was noted by the endogenous substances in the Quantech PrePaQ Total β-hCG assay.
Hook Effect - Samples well beyond the standard curve range were assayed. No high dose hook effect was observed. Therefore, the Quantech Total ß-hCG Assay does not give erroneously low results for grossly elevated samples up to at least 100,000 mlU/mL.
{3}------------------------------------------------
# Safety and Effectiveness Summary
# Summary of Clinical Performance Data
Normal Range - Testing of apparently healthy individuals demonstrated that the Quantech Total 3-hCG Assay and the predicate device perform similarly at and below the accepted value for determination of pregnancy (greater than 25 mlU/ml). The 95% normal range of the Quantech hCG assay was 8.2 mlU/ml compared to 2.8 mlU/ml for the Centaur hCG assay.
The Quantech device demonstrates no false positive results with these apparently healthy individuals and is in 100% agreement with the predicate device.
Patient Sample Correlation - Results from human samples with values distributed throughout the quantitative range of the Quantech hCG assay, were compared with those obtained with a commercially available methods. The internal paired sample comparison resulted in a correlation coefficient of 0.97 (slope = 0.67. intercept = 1.47) with the Bayer Centaur (predicate device). The external paired sample comparison resulted in a correlation coefficient of 0.93 (slope 0.89, intercept -1.7992) with the Abbott AxSym.
#### Conclusions Drawn From Performance Tests
The Quantech Total ß-hCG Assay provides results which are internally accurate, unaffected by ordinary variation of sample matrix and equivalent to the results obtained using the approved device in a valid laboratory setting.
Additionally, both clinically-based studies (normal range, patient correlation) demonstrated essential equivalence between the devices as measured by their correlation and the degree to which assay results are linearly related to one another over a broad range of values. Likewise, the normal range evaluation provided empirical evidence that the assay value is similar for both devices, and in agreement with published data.
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image shows a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the edge. Inside the circle is a stylized image of an eagle with its wings spread. The eagle is facing to the right.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Quantech Limited c/o Ms. Robin J. Hellen, M.S. Hellen Professional Services 9418 Lasaine Avenue Northridge, CA 91325
# 'JAN 1 8 2002
k012943 Re: .
Trade/Device Name: The Quantech FasTraQ Automated Analyzer, and The Quantech PrePaQ Total hCG Test Cartridge. Regulation Number: 21 CFR 862.1155 Regulation Name: Human chorionic gonadotropin (HCG) test system Regulatory Class: Class II Product Code: NAL Dated: November 16, 2001 Received: November 19, 2001
Dear Ms. Hellen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
Page 2 -
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed nredicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and 1 additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entified, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/odrh/dsma/dsmamain.html".
Sincerely yours,
Steven Butman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# PART A - 510(k) Information
| IV. | Statement for Indications for Use |
|-----|-----------------------------------|
|-----|-----------------------------------|
| 510(k) Number (if known): | K012943 |
|---------------------------|------------------------------------|
| Device Name: | Quantech FasTraQ Automated System: |
| | FasTraQ Automated Analyzer and the |
| | PrePaQ Total hCG Test Cartridge |
Indications for Use:
The Quantech PrePaQ Total hCG Test Cartridge is intended for use with the
FasTraQ Automated Analyzer to provide rapid, quantitative measurement of
human chorionic gonadotropin (hCG) in whole blood for the determination
of pregnancy. The Analyzer and hCG Cartridge combine ease of use and
rapid turnaround time with laboratory-quality performance and reliability.
The FasTraQ System is designed for use by non-laboratory medical
professionals in emergency departments, STAT or central laboratories.

| (Division Sign Off) | |
|---------------------|--------------------|
| Division of C | Laboratory Devices |
| 510(k) Numb | K012943 |
Concurrence of the CDRH, Office of Device Evaluation (ODE)
| Prescription Use: <img alt="Check" src="check.png"/> | OR | Over the Counter Use: __________ |
|------------------------------------------------------|----|----------------------------------|
|------------------------------------------------------|----|----------------------------------|
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.