GIMMI ALPHA Hysteroscopes and Accessories are intended to be used by qualified physicians to provide access, illumination and visualization of the cervical canal and the uterine cavity by a telescopic system introduced into the uterus through the cervix. These generic devices are used to perform diagnostic and surgical procedures.
Device Story
GIMMI ALPHA Hysteroscopes are rigid, panoramic telescopes utilizing rod lens technology; accessories include reusable handle/shaft assemblies and tip assemblies for tissue manipulation (grasping, cutting, dissecting, coagulating, suturing). Used in clinical settings by physicians for gynecological diagnostic and surgical procedures. Devices provide access, illumination, and visualization of the cervical canal and uterine cavity. Benefits include enabling minimally invasive visualization and surgical intervention within the uterus.
Clinical Evidence
Bench testing only. Performance and physical tests conducted to evaluate safety, effectiveness, and reliability. All results conformed to harmonized device standards.
Technological Characteristics
Rigid, panoramic telescopes using rod lens technology. Body contact portions composed of surgical grade stainless steel. Reusable handle, shaft, and tip assemblies. Compliant with relevant ISO/EN/ASTM/AAMI/ANSI/IEC standards.
Indications for Use
Indicated for use by qualified physicians for access, illumination, and visualization of the cervical canal and uterine cavity to perform diagnostic and surgical procedures.
Regulatory Classification
Identification
A hysteroscope is a device used to permit direct viewing of the cervical canal and the uterine cavity by a telescopic system introduced into the uterus through the cervix. It is used to perform diagnostic and surgical procedures other than sterilization. This generic type of device may include obturators and sheaths, instruments used through an operating channel, scope preheaters, light sources and cables, and component parts.
Predicate Devices
Günter Bissinger Medizintechnik
Comeg Endoscopy
Dufner Instrumente GmbH
Henke-Sass Wolf, GmbH
Optus, Inc.
Pilling Weck Group
Wolf
Karl Storz Endoscopy
Submission Summary (Full Text)
{0}------------------------------------------------
4H.
Gimmi GmbH
Abbreviated 510(k)
**MAR 11 2002**
K012869
page 1 of 3
### 510(k) SUMMARY of Safety and Effectiveness 2. GIMMI GmbH
As required by Section 807.92(c)
- 2.1 Submitter: [807.92 (a)(1)] GIMMI GmbH Junkerstr. 1 D-78532 Tuttlingen Germany
Tel. +49-7461 - 9 65 90-0 Fax +49-7461 - 9 65 90-33 e-mail info@gimmi.de
- 2.2 Contact Person: [807.92 (a)(1)] Dagmar S. Mäser Business Support International +31-20-428 95 91 Amstel 320-l Tel. +31-20-428 94 29 1017 AP Amsterdam Fax bsi@xs4all.nl eMail The Netherlands
#### 2.3 Date Summary Prepared: [807.92 (a)(1)} November 30, 2001
- 2.4 Device Names: [807.92 (a)(2)] GIMMI ALPHA Hysteroscopes & Accessories Proprietary
Common
Endoscopic Instruments & Accessories
| Classification Names | Product Codes | CFR Reg'n |
|-------------------------------------------------|---------------|-----------|
| Cannula & Trocar, Suprapubic,<br>Non-Disposable | 78 FBM | 876.5090 |
| Tenaculum, Uterine | 85 HDC | 884.4530 |
| Laparoscope, Gynecologic (and<br>Accessories) | 85 HET | 884.1720 |
| Screw, Fibroid, Gynecological | 85 HHO | 884.4530 |
| Hysteroscope & Accessories | 85 HIH | 884.1690 |
2-1
{1}------------------------------------------------
Instrument, Manual, Specialized (Obstetric-Gynecologic) Endoscope & Accessories
- 2.5 Reason for Submission: New Devices
- 2.6 Predicate Devices: [807.92 (a)(3)] Predicate devices are produced by Günter Bissinger Medizintechnik Comeg Endoscopy Dufner Instrumente GmbH Henke-Sass Wolf, GmbH Optus, Inc. Pilling Weck Group Wolf Karl Storz Endoscopy and a wide range of other manufacturers
#### 2.7 Device Description: [807.92(a)(4)+(6)]
GIMMI ALPHA Hysteroscopes are comprised of rigid, panoramic telescopes using rod lens technology. The body contact portions are composed of surgical grade stainless steel, which is commonly used in medical devices for a wide range of applications and has a long history of biocompatibility for human use.
Laparoscopic and obstetric-gynecologic accessories are composed of reusable handle and shaft assemblies and removable, reusable tip assemblies. Some specialized instruments, such as the gynecologic fibroid screw and uterine tenaculum are one-piece.
The instruments are designed and manufactured specifically for the purpose of manipulating soft tissue structures (grasping, cutting, dissecting, coagulating and suturing).
#### 2.8 Intended Use: 1807.92 (a)(5)}
GIMMI ALPHA Hysteroscopes and Accessories are intended to be used by qualified physicians to provide access, illumination and visualization of the cervical canal and the uterine cavity by a telescopic system introduced into the uterus through the cervix. These generic devices are used to perform diagnostic and surgical procedures.
{2}------------------------------------------------
#### 2.9 Industry Standards/Performance Data: [807.92 (d)] GIMMI certifies compliance with relevant ISO/EN/ASTM/
AAMI/ANSI/IEC and other device-related standards that apply to the manufacture, packaging, labeling, and reprocessing of subject devices including the validation of these processes.
#### 2.10 Summary of Testing
All materials used in the composition of GIMMI ALPHA Hysteroscopes and Accessories were subjected to performance and physical tests to evaluate safety, effectiveness, and reliability of the devices. All results were in conformance with the cited harmonized device standards.
#### 2.11 Information Bearing on the Safety and Effectiveness: [807.92 (b)(3)]
The GIMMI "ALPHA Hysteroscopes & Accessories have the same intended use as predicate devices. They are made of the same material and produced to the same international and FDArecognized standards. Slight modifications in design do not adversely affect the safety and effectiveness of these devices.
In summary, the
- . intended use
- performance attributes .
- materials and .
- basic design
are identical and/or substantially equivalent to SE devices.
The results of design validation raise no new issues of safety and effectiveness.
{3}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR 1 1 2002
GIMMI GmbH % Mr. Dagmar S. Mäser Business Support International Amstel 320-I, 1017 AP Amsterdam THE NETHERLANDS
Re: K012869
Trade/Device Name: Gimmi Alpha Hysteroscopes & Accessories Regulation Number: 21 CFR 884.1690 Regulation Name: Hysteroscopes and accessories Regulatory Class: II Product Code: 85 HIH Dated: December 18, 2001 Received: December 20, 2001
Dear Mr. Mäser:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual recistration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Snogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
GIMMI GmbH
K012869
510(k) Number Device Name
GIMMI ALPHA® Hysteroscopes & Accessories
# INDICATIONS FOR USE
GIMMI ALPHA® Hysteroscopes and Accessories are intended to be used by qualified physicians to provide access, illumination and visualization of the cervical canal and the uterine cavity by a telescopic system introduced into the uterus through the cervix. These generic devices are used to perform diagnostic and surgical procedures.
(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per CFR 801 109)
OR
V
Over-The-Counter Use _
(Optional Format 1-2-96)
David A. Flynn
(Division Si Division and Ra 510(k) Nun
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.