# REPROCESSED LASER PROBE (K012682)

_Sterilmed, Inc. · GEX · Nov 8, 2001 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K012682

## Device Facts

- **Applicant:** Sterilmed, Inc.
- **Product Code:** [GEX](/productcode/GEX.md)
- **Decision Date:** Nov 8, 2001
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

Reprocessed laser probes are part of a laser delivery system and are used in ophthalmic procedures where laser energy is the mode of treatment.

## Device Story

Reprocessed laser probes function as accessories to ophthalmic laser delivery systems; consist of handpiece, strain relief, fiber optic cable, and connector; may include laser light pipe. Device connects to laser console to deliver energy to treatment site via probe tip. Used by clinicians in ophthalmic surgical settings. Reprocessing restores used OEM probes to functional state; manufacturing process includes cleaning, sterilization, and packaging validation. Output is delivery of laser energy for therapeutic ophthalmic procedures. Benefits include providing functional laser delivery tools for surgery.

## Clinical Evidence

Bench testing only. Representative samples underwent functional testing. Process validation performed for cleaning, sterilization, and packaging. Manufacturing process includes visual and functional testing of all units.

## Technological Characteristics

Accessory to laser delivery system; consists of handpiece, strain relief, fiber optic cable, and connector. Energy source is external laser console. Sterilization performed as part of reprocessing. No software or electronic components integral to the probe itself.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- HGM Inc.'s Fleischman-Swartz Endo-Ocular Probe ([K840590](/device/K840590.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K012682

# SECTION 2. SUMMARY AND CERTIFICATION

| 510(k) Summary<br>A.              | NOV 0 8 2001                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
|-----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter:                        | SterilMed, Inc.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Contact Person:                   | Patrick Fleischhacker<br>11400 73rd Avenue North<br>Minneapolis, MN 55369<br>Ph: 888-856-4870<br>Fax: 763-488-3350                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Date Prepared:                    | August 10, 2001                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Trade Name:                       | SterilMed Reprocessed Laser Probes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Classification Name:              | Laser, Ophthalmic                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Classification Number:            | Class II, 21 CFR 886.4390                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Product Code:                     | HQF                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Predicate Device(s):              | SterilMed's reprocessed laser probes are substantially<br>equivalent to HGM Inc.'s Fleischman-Swartz Endo-Ocular<br>Probe (K840590), and to the counterpart devices from the<br>original manufacturers.                                                                                                                                                                                                                                                                                                                                                                        |
| Device Description:               | Laser probes are an accessory used as part of a laser<br>delivery system. They consist of a probe handpiece, strain<br>relief, fiber optic cable, and connector. A laser light pipe<br>may or may not be part of the probe. The probes must be<br>connected to a laser console in order to deliver energy. The<br>laser probe delivers the energy to the treatment site through<br>an opening at the probe tip. Note, this submission is only in<br>regard to the laser probe itself. It does not pertain to any<br>other part of a laser delivery system such as the console. |
| Intended Use:                     | Reprocessed laser probes are part of a laser delivery system<br>and are used in ophthalmic procedures where laser energy<br>is the mode of treatment.                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Functional and<br>Safety Testing: | Representative samples of reprocessed laser probes<br>underwent bench testing to demonstrate appropriate                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |

.

{1}------------------------------------------------

012682 2/2

functional characteristics. Process validation testing was done to validate the cleaning and sterilization procedures as well as the device packaging. In addition, the manufacturing process includes visual and functional testing of all products produced.

#### Conclusion:

SterilMed's reprocessed laser probes are substantially equivalent to HGM Inc.'s Fleischman-Swartz Endo-Ocular Probe (K840590), and to the counterpart devices from the original manufacturers. This conclusion is based upon the fact that these devices are essentially identical to the predicate device in terms of functional design, indications for use, and principles of operation.

{2}------------------------------------------------

Image /page/2/Picture/12 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, in the center. Encircling the caduceus are the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement. The text is in all caps and is evenly spaced around the circle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

NOV 0 82001

Mr. Patrick Fleischhacker Vice President Regulatory and Quality Control SterilMed, Inc. 11400 73rd Avenue, North Minneapolis, Minnesota 55369

Re: K012682

Trade Name: Reprocessed Laser Probe Regulation Number: 886.4390 Regulation Name: Laser, ophthalmic Regulatory Class: Class II Product Code: GEX; HQF Dated: August 10, 2001 Received: August 14, 2001

Dear Mr. Fleischhacker:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

### Page 2 - Mr. Patrick Fleischhacker

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Susan Walker, MD

Image /page/3/Picture/5 description: The image contains a black and white drawing of a signature. The signature appears to be cursive and somewhat stylized. The first letter is a large loop, and the rest of the signature is a series of connected strokes.

Celia M. Witten, Ph.D., M.D. S Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

KOi2682

## Indications for Use Page

Device Name: Reprocessed Laser probes

#### Indications for Use:

These devices are reprocessed laser probes from various original equipment manufacturers (OEMs). Reprocessed laser probes are part of a laser delivery system and are used in ophthalmic procedures where laser energy is the mode of treatment.

Concurrence of CDRH, Office of Device Evaluation (ODE)

Susan Walker

(Division Sign- Off) Division of Ge eral, Restorative and Neurological Devices

510(k) Number

K0126sf2

---

**Source:** [https://fda.innolitics.com/device/K012682](https://fda.innolitics.com/device/K012682)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com