K012633 · Respironics, Inc. · BTT · Feb 28, 2002 · Anesthesiology
Device Facts
Record ID
K012633
Device Name
RESPIRONICS HEATED HUMIDIFIER
Applicant
Respironics, Inc.
Product Code
BTT · Anesthesiology
Decision Date
Feb 28, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5450
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Respironics Heated Humidifier is an accessory for CPAP and Bi-level systems to provide moisture to the patient circuit. It is intended for use with adult patients, in the home or hospital/institutional environment, who use mask-applied CPAP or Bi-level therapy for the treatment of Obstructive Sleep Apnea.
Device Story
Respironics Heated Humidifier is a heated Passover-type respiratory gas humidifier; designed to warm and add moisture to breathing gases for CPAP or Bi-level therapy. Device consists of a thermoplastic enclosure containing control electronics and a metal heater plate; humidification chamber slides onto heater plate. Operated by user via top panel controls to adjust heat settings (off to 65°C). Used in home or hospital settings by patients undergoing treatment for Obstructive Sleep Apnea. By adding heated moisture to the breathing circuit, the device relieves airway drying and irritation associated with CPAP/Bi-level therapy, potentially improving patient comfort and compliance.
Clinical Evidence
No clinical data. Safety and effectiveness were demonstrated through non-clinical bench testing, including mechanical, electrical, and thermal safety, environmental conditions, electromagnetic compatibility, functional verification, and system validation.
Technological Characteristics
Heated Passover-type humidifier. Thermoplastic enclosure; metal heater plate. Dimensions: 5.5" x 6.5" x 5.25". Weight: 1.24-2.5 lbs. Energy source: AC power. User controls: 5 settings. Temperature range: off to 65°C. Complies with IEC 60601-1 and UL 2601-1 safety standards.
Indications for Use
Indicated for adult patients using mask-applied CPAP or Bi-level therapy for the treatment of Obstructive Sleep Apnea to provide moisture to the patient circuit and relieve airway drying/irritation.
Regulatory Classification
Identification
A respiratory gas humidifier is a device that is intended to add moisture to, and sometimes to warm, the breathing gases for administration to a patient. Cascade, gas, heated, and prefilled humidifiers are included in this generic type of device.
Predicate Devices
Respironics REMstar Plus CPAP System/REMstar Heated Humidifier (K010263)
{0}------------------------------------------------
# FEB 2 8 2002
K012633
# FECTIVE
| Official Contact | Zita A. Yurko<br>Manager, Regulatory Affairs/Product Assurance<br>Respironics, Inc.<br>1001 Murry Ridge Lane<br>Murrysville, PA 15668 |
|--------------------------|---------------------------------------------------------------------------------------------------------------------------------------|
| Classification Reference | 21 CFR 868.5450 |
| Product Code | Respiratory gas humidifier (BTT) |
| Common/Usual Name | HUMIDIFIER, RESPIRATORY GAS, (DIRECT PATIENT INTERFACE) |
| Proprietary Name | Respironics Heated Humidifier |
| Predicate Device(s) | Respironics REMstar Plus CPAP System/REMstar Heated Humidifier<br>(K010263) |
| Reason for submission | Fisher & Paykel, HC100 Respiratory Humidifier (K915460)<br>New design. |
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92:
### (a)(1) - (a)(3) Refer to information above and concluding this summary
#### (a)(4) Description of the Device
The Respironics Heated Humidifier is a Respiratory Gas Humidifier (heated Passover type) according to 21 CFR 868.5450. Heat is used to provide an evaporated water content to dry breathing gases.
The Respironics Heated Humidifier has a thermoplastic enclosure with dimensions of 5.5 in. high x 6.5 in. wide x 5.25 in. deep and weighs 1.24 lbs. (without the chamber fitted), 1.63 lbs. (with the chamber fitted), and 2.5 lbs. (with the chamber fitted and filled with water). The unit is compromised of a metal heater plate that is controlled from control electronics, contained in a plastic enclosure that is directly supplied with AC power. The heater plate is positioned in the front of the unit. The humidification chamber slides onto the heater plate and is held in place by a rim on the enclosure. The unit controls are located on the top panel.
{1}------------------------------------------------
Accessories for the Respironics Heated Humidification chambers, breathing tubing and mounting arrangements.
#### (a)(5) Statement of the Intended Use
The Respironics Heated Humidifier is intended to warm and add moisture to the breathing gases for administration to a patient. It is used for patients who use mask-applied CPAP or Bi-level the a treatment of Obstructive Sleep Apnea. The addition of heated humidification to this therapy relieves the drying and irritating effects on the patient airways, which may arise from use of a CPAP or Bi-level system.
#### (a)(6) Technological Characteristics Summary
The technological characteristics of the Respironics Heated Humidifier are equivalent to the Fisher & Paykel HC100 humidifier predicate device listed above.
The Respironics Heated Humidifier is equivalent in terms of: type (heated Passover humidification) configuration (chamber, breathing tubing, physical design, user controls, energy delivered, humidifier design concept, inlet port and outlet port sizing). Differences from the predicate device include: weight, size, energy type, user controls (5 settings vs. 9 settings), temperature range (off to 65°C) and delivered humidity.
#### (b)(1) Discussion of the Non-Clinical Tests
Non-clinical testing of the Respironics Heated Humidifier has been carried out covering mechanical, electrical and thermal safety, environmental conditions and electromagnetic compatibility, functional verification, software verification, system validation and performance.
The Respironics Heated Humidifier meets the requirements of the IEC 60601-1 Medical Electrical Equipment Part 1: General Requirements for safety and the relevant USA deviations in UL 2601-1.
#### (b)(2) Discussion of the Clinical Tests
Clinical verification studies on the Respironics Heated Humidifier were not required in order to demonstrate safety, effectiveness and performance of the device.
{2}------------------------------------------------
# (b)(3) Conclusions Demonstrating Safety, Effectiveness of Performance
The testing carried out for the Respironics Heated Humiditier indicates that it meets design and performance functional requirements. The proposed device meets the requirements of international and USA medical electrical equipment standards for safety, and key performance and safety requirements from particular standards for humidification systems.
This information indicates that the Respironics Heated Humidifier is equivalent to the predicate device in terms of safety, effectiveness and performance.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, enclosed within a circle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper portion of the circle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB 2 8 2002
Mr. David J. Vanella Respironics, Inc. 1001 Murry Ridge Lane Murrysville, PA 15668
Re: K012633
Respironics Heated Humidifier Regulation Number: 868.5450 Regulation Name: Respiratory Gas Humidifier Regulatory Class: II (two) Product Code: BTT Dated: February 7, 2002 Received: February 8, 2002
Dear Mr. Vanella:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave roviewed your becamed the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use sured in the encreate) ve ttgractment date of the Medical Device Amendments, or to commerce pror to May 20, 1978, are except and encordance with the provisions of the Federal Food, Drug, devices that have been received in avire approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The I ou may, mercere, maniel and act include requirements for annual registration, listing of general controls profice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is olassinod (600 a00 royals. Existing major regulations affecting your device can may or subject to back as a such as a manufacture, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 – Mr. David J. Vanella
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Dan Cattell
Bram D. Zuckerman, M.D. Acting Director Division of Cardiovascular and Respiratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Page _1_of_1
510(k) Number (if known): _ KOJZ 6 3 3
Device Name: Respironics Heated Humidifier
## Intended Use/Indications for Use
The Respironics Heated Humidilier is an accessory for CPAP and Bi-level systems to provide moisture to the patient circuit.
#### Environment of Use/Patient Population
It is intended for use with adult patients, in the home or hospital/institutional environment, who use maskapplied CPAP or Bi-level therapy for the treatment of Obstructive Sleep Apnea.
#### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
U Prescription Use (Per 21 CFR 801.109)
OR
Over-The-Counter Use (Optional Format 1-2-96)
Davull
Division of Cardiovascular & Respiratory Devices
510(k) Number K012633
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.