OSI MEDICAL DOLPHIN STAND-ALONE PULSE OXIMETER & DIGITAL DOLPHIN DISPOSABLE OPTICAL SENDOR
K012626 · Osi Medical, Inc. · DQA · Nov 2, 2001 · Cardiovascular
Device Facts
Record ID
K012626
Device Name
OSI MEDICAL DOLPHIN STAND-ALONE PULSE OXIMETER & DIGITAL DOLPHIN DISPOSABLE OPTICAL SENDOR
Applicant
Osi Medical, Inc.
Product Code
DQA · Cardiovascular
Decision Date
Nov 2, 2001
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2700
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The OSI Medical Dolphin Stand-Alone Pulse Oximeter and Accessories and indicated for the continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (measured by an SpO2 sensor). The OSI Medical Dolphin Pulse Oximeter and Accessories are indicated for use with adult, pediatric and neonatal patients in hospital, clinical, and home environments.
Device Story
Portable stand-alone pulse oximeter (Model 2100) monitors SpO2 and pulse rate; inputs signals from LEDs within oximetry sensors via connecting cable; sensor measures light absorption of blood using two LEDs; monitor calculates ratio of oxygenated to total hemoglobin; displays plethysmographic waveform, SpO2, pulse rate, alarm limits, trends, and status messages; used by clinicians or patients in hospital, clinical, or home settings; provides real-time physiological data to assist in patient monitoring and clinical decision-making; benefits include noninvasive assessment of oxygenation status.
Clinical Evidence
Bench testing only. Compliance with standards including ISO 9919:1992, EN 865:1997, and FDA Pulse Oximeter Guidance (1992). Biocompatibility and performance validation performed for disposable sensors.
Technological Characteristics
Portable stand-alone monitor; 10cm x 27.5cm x 25cm; 4kg; powered by 100-240 Vac or sealed lead-acid battery. Uses two-wavelength LED oximetry sensing. Complies with IEC 601-1, ISO 9919, EN 865, and ASTM 1415. Disposable and reusable sensor options. Connectivity via proprietary extension cable.
Indications for Use
Indicated for continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate in adult, pediatric, and neonatal patients in hospital, clinical, and home environments.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
Predicate Devices
OSI Medical Dolphin Stand-Alone Pulse Oximeter, Model No. 2100 and Accessories (K002036)
Submission Summary (Full Text)
{0}------------------------------------------------
K012626
## NOV 0 2 2001
August 8, 2001
| Subject: | Summary of Safety and Effectiveness Information for the OSI Medical Dolphin Stand-<br>Alone Pulse Oximeter and Accessories. |
|-----------------|-----------------------------------------------------------------------------------------------------------------------------|
| Proprietary: | OSI Medical Dolphin Stand-Alone Pulse Oximeter, Model No. 2100 and Accessories. |
| Common: | Oximeter. |
| Classification: | Oximeter Class II - 21 CFR 870.2700 – 74 DQA |
The 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of the SMDA 1990 and 1992.
The OSI Medical Dolphin Pulse Oximeter and Accessories is substantially equivalent to the following currently marketed device(s):
- OSI Medical Dolphin Stand-Alone Pulse Oximeter, Model No. 2100 and Accessories #K002036. .
The OSI Medical Dolphin Pulse Oximeter and Accessories is a portable stand-alone device, connecting cable, and oximetry sensor(s) to noninvasively calculate the functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate for adult, pediatric and neonatal patients.
The monitor consists of a screen that displays the pulse plethysmographic waveform, the pulse rate, SpO2 value, the high and low SpO2 and pulse rate alarm limits, alarms, trends and status messages. It contains the electronic hardware and software that receives and calculates the signals from the LEDs within the sensor to determine the functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate, and provide for the connection to the connecting cable.
The OSI Medical Dolphin Pulse Oximeter is available in one configuration as a portable stand-alone oximeter that is 10 cm / 3.94 inches high, 27.5 cm / 10.83 inches wide, 25 cm / 9.84 inches deep and weighs about 4 kg / 8.8 lbs. The unit is powered either with a voltage input of 100-240 Vac, 50-60 Hz or with a sealed lead-acid battery with an operating time of approximately 4 hours based upon 2 Ampere hour battery (200mA OEM Module, 300mA System Module) and a charge time of about 4.5 hours to 80% capacity.
The extension cable connects between the monitor and oximetry sensor(s) and transfers LED drive power and the calibration drive to the oximetry sensor from the monitor receives the detector signal from the oximetry sensor.
The extension cable is available in one configuration and is approximately 8 feet / 2.44m in length, and the sensor(s) are approximately 18 inches / 45.72 cm in length.
The sensor(s) measure light absorption of blood from two light emitting diodes (LED's). Oxygen saturated blood absorbs light differently as compared to unsaturated blood. The amount of light absorbed by the blood is used to calculate the ratio of oxygenated hemoglobin to total hemoglobin in arterial blood.
{1}------------------------------------------------
The oximetry sensor is available in a disposable or reusable configuration, and with one configuration for the extension cable (8 feet).
The OSI Medical Dolphin Stand-Alone Pulse Oximeter, Model No. 2100 and Accessories have been designed to comply with the following standards:
- 1. CSA C22.2 No. 601.1
- IEC 601-1, Part 1 and Amendments 1 and 2 2.
- 3. IEC 601-1-1, Part 1
- IEC 601-1-2, Part 1 ধ:
- 5. ISO 9919: 1992
- 6. EN 865: 1997
- 7. FDA Guidance Document for Pulse Oximeters: 9/7/1992
- 8. ASTM 1415:1992, and Draft 10.1
- UL2601-1: Second Edition, 1997 9.
- 10. IEC 6068-2-6
- 11. IEC 6068-2-27
- 12. IEC 6068-2-64
- 13. ISTA Procedure 2A
The OSI Medical Dolphin Stand-Alone Pulse Oximeter with disposable sensors is substantially equivalent The OSI Medical Dolphill Galle-Alone Pulse Online of the OSI Medical Dolphin Stand-Alone Pulse Oximeter, with reusable sensors. Additional testing has been performed for the disposable sensors including biocompatibility and performance validation which has been included in this submission.
# 000012
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized eagle with three curved lines representing its wings, positioned to the right of the text. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle, indicating the department's name and country.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# NOV 0 2 2001
Mr. Jon Werner OSI Medical, Inc. 13801 McCormick Drive Tampa, FL 33626
K012626 Re:
R012020
OSI Medical Dolphin Stand-Alone Pulse Oximeter, Model No. 2100 Regulation Number: 870.2700 Regulation Name: Oximeter Regulatory Class: II (two) Product Code: 74 DQA Dated: August 8, 2001 Received: August 13, 2001
Dear Mr. Werner:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your becamed the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encreated of the enactment date of the Medical Device Amendments, or to conninered pror to may 20, 2017 11:12 accordance with the provisions of the Federal Food, Drug, devices that have boon require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The I ou may, dicrorore, manel a the Act include requirements for annual registration, listing of general obtarvill profice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it 11 70dr de subject to such additional controls. Existing major regulations affecting your device can may or subject to back are as Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean r tease be arribed and i Drimination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{3}------------------------------------------------
#### Page 2 - Mr. Jon Werner
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set OF IT rat 007); adoning (21 CFR Part 820); and if applicable, the electronic roral in and quinn control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) I mis lotter will and my of your e FDA finding of substantial equivalence of your device to a legally market notification: "The stars in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and If you desire specific ad 1309.10 for in vitro diagnostic devices), please contact the Office of additionally 21 OF N Far 804. Additionally, for questions on the promotion and advertising of Compination at (301) 591 the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). regulation onlined, "Thisorana wour responsibilities under the Act may be obtained from the Oiner general information of Small and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
K. Darl Tellem
James F. Dillard III
ames E. Dillard III Director Division of Cardiovascular and Respiratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## NOV 0 2 2001 INDICATIONS FOR INTENDED USE FOR 2100 PULSE OXIMETER
K012626
PR-032100.D
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: OSI Medical Dolphin Stand-Alone Pulse Oximeter, Model No. 2100
August 8, 2001
Indications for Use:
Indications for Use:
The OSI Medical Dolphin Stand-Alone Pulse Oximeter and Accessories and indicated for the continuous The OSI Medical Dolphin Standanon ense oxines of and Aboooolobin (SpO) and pulse rate (measured noninvasive monitoring of functional or use Oximeter and Accessories are inicated for use with adult,
by an SpO2 sensor). The OSI Medical Dolphin Pulse Oximeter and home by an SpO2 sensor). The OSI Medical Dolphil Pulse Oxineler and Acoboonoo and Industral and home environments.
The OSI Medical's sensor(s) are intended/indicated for use only with the OSI Medical Dolphin Stand-Alone Pulse Oximeter, and consist of the following:
Digital Dolphin Disposable Sensors Digital Dolphin Disposable Sensors
Digital Dolphin 210 Reusable Finger Clip intended for adults / pediatrics greater than 66.14 lbs. (30kg). Digital Dolphin 110 Extension Cable
The OSI Medical extension cable is intended/indicated for use only with the OSI Medical sensor(s) and the OSI Medical Dolphin Stand-Alone Pulse Oximeter.
(Please DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) | |
|---------------------------------------------------------------------|---------------------------------------------------------------------------|
| | (Division Sign-Off) |
| | Division of Cardiovascular, Respiratory, and Neurological Devices |
| 510(k) Number: | K012624 |
| Prescription Use <span style="text-decoration: underline;">✓</span> | OR over-the-counter Use <span style="text-decoration: underline;"></span> |
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.