K012591 · Wrightmedicaltechnologyinc · HRY · Nov 2, 2001 · Orthopedic
Device Facts
Record ID
K012591
Device Name
ADVANCE UNICONDYLAR KNEE SYSTEM
Applicant
Wrightmedicaltechnologyinc
Product Code
HRY · Orthopedic
Decision Date
Nov 2, 2001
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 888.3530
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Indicated for use in knee arthroplasty for reduction or relief of pain and/or improved knee function in skeletally mature patients with the following: 1) noninflammatory degenerative joint disease including osteoarthritis, traumatic arthritis, or avascular necrosis; 2) inflammatory degenerative joint disease including rheumatoid arthritis; 3) correction of functional deformity; 4) revision procedures where other treatments or devices have failed; and 5) treatment of fractures that are unmanageable using other techniques. Unicondylar knee system is indicated for patients with unicompartmental joint disease secondary to the above indications with or without valgus, varus, or flexion deformities where all ligaments are intact. The ADVANCE® Unicondylar Knee System components are for single use only. The ADVANCE® Unicondylar Knee System components are for cemented use only.
Device Story
The ADVANCE® Unicondylar Knee System is a cemented, non-constrained/semi-constrained femorotibial prosthesis used in knee arthroplasty. It consists of anatomically shaped, nonporous Cobalt-Chromium (Co-Cr) femoral components and all-polyethylene (UHMWPE) tibial components. The femoral component is available in four sizes with left/right configurations; the tibial component is available in four universal sizes. The device is intended for single-use implantation by orthopedic surgeons to relieve pain and improve knee function in patients with unicompartmental joint disease. It functions as a mechanical replacement for the joint surface, secured via bone cement.
Indicated for skeletally mature patients requiring knee arthroplasty for unicompartmental joint disease (osteoarthritis, traumatic arthritis, avascular necrosis, rheumatoid arthritis), functional deformity, revision, or unmanageable fractures. Requires intact ligaments; may be used with valgus, varus, or flexion deformities.
Regulatory Classification
Identification
A knee joint femorotibial metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace part of a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that consist of a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component made of ultra-high molecular weight polyethylene and is limited to those prostheses intended for use with bone cement (§ 888.3027).
Submission Summary (Full Text)
{0}------------------------------------------------
K0/259/
NOV 0 2 2001
## 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS
In accordance with the Food and Drug Administration Rule to implement provisions of the Safe Medical Devices Act of 1990 and in conformance with 21 CRF 807, this information serves as a Summary of Safety and Effectiveness for the use of the ADVANCE® Unicondylar Knee System.
Submitted By:
Date:
Contact Person:
Proprietary Name:
Common Name:
Classification Name and Reference:
Wright Medical Technology, Inc.
August 9, 2001
Ehab M. Esmail Manager, Regulatory Affairs
ADVANCE® Unicondylar Knee System
UNICONDYLAR KNEE SYSTEM
21 CFR 888.3520 Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer -Class II
21 CFR 888.3530 Prosthesis, Knee, Femorotibial, Semi-Constrained, Cemented, Metal/Polymer -Class II
Device Product Code and Panel Code:
Orthopedics/87/ HSX, HRY
### DEVICE INFORMATION
#### A. INTENDED USE
Indicated for use in knee arthroplasty for reduction or relief of pain and/or improved knee function in skeletally mature patients with the following:
- 1) noninflammatory degenerative joint disease including osteoarthritis, traunanc arthritis, or avascular necrosis;
- 2) inflammatory degenerative joint disease including rheumatoid arthuitis;
- 3) correction of functional deformity;
- 4) revision procedures where other treatments or devices have failed; and
- treatment of fractures that are unmanageable using other techniques. 5)
Image /page/0/Picture/27 description: The image shows the logo for Wright Medical Technology, Inc. The logo features a stylized "W" above the word "WRIGHT" in a bold, sans-serif font. Below "WRIGHT" is the text "MEDICAL TECHNOLOGY, INC." in a smaller font. There is also some text on the left side of the image, but it is too blurry to read.
Image /page/0/Picture/28 description: The image shows a black and white logo. The logo appears to be a globe with a star on the upper right side. The globe is surrounded by a circle.
IOFD
51000 SUMMARY
KNIGHT
AL TECHNOLOGY, INC.
{1}------------------------------------------------
K012591
2 of 2
Unicondylar knee system is indicated for patients with unicompartmental joint disease secondary to the above indications with or without valgus, varus, or flexion deformities where all ligaments are intact.
The ADVANCE® Unicondylar Knee System components are for single use only.
The ADVANCE® Unicondylar Knee System components are for cemented use only.
#### B. DEVICE DESCRIPTION
The ADVANCE® Unicondylar Knee System contains anatomically shaped, medial, nonporous Co-Cr femoral components. The tibial components will consist of all-poly ultrahigh molecular-weight polyethylene (UHMWPE) tibial implants.
The ADVANCE® Unicondylar Femoral component will be available in four anatomic sizes (to provide coverage of the condyle from posterior to anterior). The anatomic shape of the femoral component necessitates separate left and right configurations.
The ADVANCE® Unicondylar Tibial component will be available in four sizes (universal in shape- no lefts and rights).
The design features of ADVANCE® Unicondylar Knee System are substantially equivalent to the design features of competitive devices previously cleared for market.
## C. SUBSTANTIAL EQUIVALENCE INFORMATION
The intended use, material, type of interface, and design features of ADVANCE® Unicondylar Knee System are substantially equivalent to the predicate devices previously cleared for market. The safety and effectiveness of ADVANCE® Unicondylar Knee System are adequately supported by the substantial equivalence information, materials data, and testing results provided within the Premarket Notification.
Image /page/1/Picture/12 description: The image shows a black and white logo of a globe. The globe is tilted and the continents are visible. The logo is simple and stylized. The bottom of the image has some text, but it is illegible.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle with three lines representing its wings. The eagle is facing to the right.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV 0 2 2001
Mr. Ehab M. Esmail Manager, Regulatory Affairs Wright Medical Technology, Inc. 5677 Airline Road Arlington, Tennessee 38002
Re: K012591
Trade/Device Name: ADVANCE® Unicondylar Knee System Regulation Number: 21 CFR 888.3530 Regulation Name: Knee joint femorotibial metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: HRY Dated: August 9, 2001 Received: August 10, 2001
Dear Mr. Esmail:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your becall be o(s) fee device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate 10 use surved in the entrebate) 77 ttg enactment date of the Medical Device Amendments, or to conninered pror to 1125 2017 11:11 accordance with the provisions of the Federal Food, Drug, de rices that have been recises of require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The I ou may, dierelove, mains of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is elabilitional controls. Existing major regulations affecting your device can may be back of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean r lease of acrised a determination that your device complies with other requirements of the Act that I Drives Internet and regulations administered by other Federal agencies. You must or mry receits all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set Of It Part 607); adoming (21 CFR Part 820); and if applicable, the electronic forth in also quind provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Mr. Ehab M. Esmail
This letter will allow you to begin marketing your device as described in your Section 510(k) This letter will anow you to obgen mailsoning your and equivalence of your device to a legally premits that in the Pro Promating of a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and II you desire specific advice for your dovitro diagnostic devices), please contact the Office of additionally 21 CF R Fart 607.10 for the market on the promotion and advertising of Compliance at (301) 594-4639. Additionally, 18 41-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Tegulation clittined, "Misoranang of responsibilities under the Act may be obtained from the Outler general information on your respectional and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
L. Mark N. Milliman
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows a sequence of handwritten characters and numbers. The sequence starts with the letter 'k', followed by the number '0', then '1', '2', '5', '9', and ends with '1'. The characters are written in a simple, slightly uneven style, typical of handwriting.
NOV 0 2 2001
# ADVANCE® Unicondylar Knee System
## INDICATIONS STATEMENT
Indicated for use in knee arthroplasty for reduction or relief of pain and/or improved Indicated for ass as as a leater patients with the following conditions:
- 1) noninflammatory degenerative joint disease including osteoarthritis, traumatic arthritis, or avascular necrosis;
- 2) inflammatory degenerative joint disease including rheumatoid arthritis;
- 3) correction of functional deformity;
- 3) convoluon of raires where other treatments or devices have failed; and
- 4) freatment of fractures that are unmanageable using other techniques.
Unicondylar knee system is indicated for patients with unicompartmental joint disease secondary to the above indications with or without valgus, varus, or flexion deformities where all ligaments are intact.
The ADVANCE® Unicondylar Knee System components are for single use only.
The ADVANCE® Unicondylar Knee System components are for cemented use only.
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (OD
for Mark A. Millane
Division Sior Division of General Re Devices 510(k) Number
Prescription Use
(Per21 CFR 801.109) X
OR
Over-The Counter Use (Optional Format 1-2-96)
Image /page/4/Picture/21 description: The image shows a black and white drawing of a globe. The globe is tilted and has a dark land mass on the left side. There is a partial ring around the top right of the globe.
INDICATIONS STATEMENT
2012-214 14 11 11 1 1 1 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
Image /page/4/Picture/23 description: The image shows a logo for Wright Medical Technology. The logo features a stylized "W" on the left, followed by the words "WRIGHT MEDICAL TECHNOLOGY" stacked vertically. Below the logo, it says "A Wright Medical Group Company" in smaller font. The logo appears to be for a medical technology company.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.