THE TRABECULAR METAL TECHNOLOGY ACETABULAR AUGMENT SYSTEM, MODELS 4893-XX-YY (ZIMMER), 02-212-05XX1 ( IMPLEX)

K012507 · Implex Corp. · LPH · Aug 30, 2001 · Orthopedic

Device Facts

Record IDK012507
Device NameTHE TRABECULAR METAL TECHNOLOGY ACETABULAR AUGMENT SYSTEM, MODELS 4893-XX-YY (ZIMMER), 02-212-05XX1 ( IMPLEX)
ApplicantImplex Corp.
Product CodeLPH · Orthopedic
Decision DateAug 30, 2001
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3358
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Trabecular Metal Technology Acetabular Augment System is intended to provide the orthopedic surgeon with a prosthetic alternative to structural allograft in cases of segmental acetabular deficiencies.

Device Story

The Trabecular Metal Technology (TMT) Acetabular Augment System is a prosthetic device used in total hip replacement surgery to address segmental acetabular deficiencies. The device features a truncated hemispherical geometry with an inner diameter 3mm larger than the corresponding acetabular cup. It is available in outer diameters ranging from 40 to 70 mm and thicknesses from 5mm to 30mm. The augment includes portals for adjunct fixation using Continuum Bone Screws (5mm and 6.5mm) or Zimmer Trilogy and HGP II 6.5 mm bone screws. These portals also serve as a containment system for bone graft in cementless applications. The device is intended for use by orthopedic surgeons in a clinical/surgical setting to provide a prosthetic alternative to structural allograft, potentially improving patient outcomes in complex acetabular reconstruction.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Truncated hemispherical geometry; Trabecular Metal material; sizes 40-70mm OD; 5-30mm thickness; includes portals for bone screw fixation (Continuum, Zimmer Trilogy, HGP II); cementless application; non-software device.

Indications for Use

Indicated for orthopedic surgeons performing total hip replacement in patients with segmental acetabular deficiencies as a prosthetic alternative to structural allograft.

Regulatory Classification

Identification

A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS The Trabecular Metal Technology Acetabular Augment System | Submitter Name: | Implex Corp. | |------------------------|------------------------------------------------------------------------------------| | Submitter Address: | 80 Commerce Drive<br>Allendale, New Jersey 07401-1600 | | Contact Person: | Robert A. Poggie, Ph.D. | | Phone Number: | (201) 818-1800 | | Fax Number: | (201) 995-9722 | | Date Prepared: | August 2, 2001 | | Device Trade Name: | The Trabecular Metal Technology Acetabular Augment<br>System | | Device Common<br>Name: | Acetabular augmentation devices for total hip replacement<br>acetabular components | | Classification Number: | 21 CFR § 888.3358 | The term "substantial equivalence" as used in this Substantial 510(k) notification is limited to the definition of Equivalence: substantial equivalence found in the Federal Food, Drug and Cosmetic Act, as amended and as applied under 21 CFR 807, Subpart E under which a device can be marketed without premarket approval or A determination of substantial reclassification. equivalency under this notification is not intended to have any bearing whatsoever on the resolution of patent infringement suits or any other patent matters. No statements related to, or in support of substantial equivalence herein shall be construed as an admission aqainst interest under the US Patent Laws or their application by the courts. The TMT Acetabular Augment possesses a truncated Device Description: hemispherical geometry with an inner diameter 3mm larger than the major diameter of the corresponding acetabular cup. The TMT Augment is available in OD sizes 40 to 70 mm in 2 mm increments, and each size is offered in various thicknesses, 5mm to 30mm in 5 mm The TMT Acetabular Augment includes increments. portals that allow the use of Continuum Bone Screws, 5mm and 6.5mm, and Zimmer Trilogy and HGP II 6.5 mm bone screws for adjunct fixation. In addition, the portals also provide a containment system for bone graft in cementless applications. {1}------------------------------------------------ <012507 2 OF 2 ## 510(k) Summary (Continued) Indications for Use: The Trabecular Metal Technology Acetabular Augment System is intended to provide the orthopedic surgeon with a prosthetic alternative to structural allograft in cases of segmental acetabular deficiencies. Conclusion: The Trabecular Metal Technology Acetabular Augment System is substantially equivalent to the identified předicate devices. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines beneath them. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 AUG 3 0 2001 Robert A. Poggie, Ph.D. Director of Applied Research Implex Corporation 80 Commerce Drive Allendale, New Jersey 07401-1600 Re: K012507 Trade Name: The Trabecular Metal Technology Acetabular Augmentation System Regulation Number: 888.3358 Regulatory Class: II Product Codes: LPH Dated: August 3, 2001 Received: August 6, 2001 Dear Dr. Poggie: We have reviewed your Section 510(k) notification of intent to market the device referenced wo have and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce suiced in also are, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls Cosisions of the Act. The general controls provisions of the Act include requirements for annual provisions of its in the concess, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations (Trenditor Inproval), in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to mspotions, and regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {3}------------------------------------------------ Page 2 - Robert A. Poggie, Ph.D. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed nredicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html". Sincerely yours, Mark N Wilkerson Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Devices Evaluation Center for Devices and Radiological Devices Enclosure {4}------------------------------------------------ 510(k) Number (if known) : Device Name: The Trabecular Metal Technology Acetabular Augment System Indications For Use: The Trabecular Metal Technology Acetabular Augment System is intended to provide the orthopedic surgeon with a prosthetic alternative to structural allograft in cases of segmental acetabular deficiencies. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) | Concurrence of CDRH; Office of Device Evaluation (ODE) | <img alt="signature" src="signature.png"/> | |--------------------------------------------------------|-----------------------------------------------------------| | | (Division Sign-Off) | | | Division of General, Restorative and Neurological Devices | | 510(k) Number | K012507 | | Prescription Use (Per 21 CFR 801.109) | Over-The-Counter Use | | | (Optional Format 1-2-96) |
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...