← Product Code [JJY](/productcode/JJY) · K012468

# MODIFICATION TO N/T PROTEIN CONTROL SL (K012468)

_Dade Behring, Inc. · JJY · Sep 4, 2001 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K012468

## Device Facts

- **Applicant:** Dade Behring, Inc.
- **Product Code:** [JJY](/productcode/JJY.md)
- **Decision Date:** Sep 4, 2001
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1660
- **Device Class:** Class 1
- **Review Panel:** Clinical Chemistry

## Indications for Use

N/T Protein Controls SL/L, M, and H are for use as accuracy and precision assayed controls in the determination of the following human serum proteins by immunonephelometry with BN™ Systems: IgG, IgG, IgA, IgM, C3c, C4, Transferrin, in Innonophobionou y ma2-macroglobulin, Haptoglobin, α--acid glycoprotein, Prealbumin, Albumin, c.y antill Jpon) @2 maor J.g/L-chain lambda & kappa, β2-microglobulin, soluble Transferrin Receptor (STFR), Ferritin, IgE, and Total protein; and by immunoturbidimetry with the TurbiTimeSystem: IgG, IgA, IgM, C3c, C4, Transferrin, Albumin, Haptoglobin, α4acid glycoprotein.

## Device Story

N/T Protein Control SL is a liquid quality control material derived from human serum with added stabilizers and preservatives. It serves as an assayed control to verify the accuracy and precision of human serum protein measurements. The device is used in clinical laboratory environments by trained personnel. It is processed on BN Systems (via immunonephelometry) and the TurbiTimeSystem (via immunoturbidimetry). By comparing measured values against the control's assigned ranges, healthcare providers assess the performance of their analytical systems, ensuring reliable diagnostic results for various serum proteins, which supports clinical decision-making regarding patient protein levels.

## Clinical Evidence

Bench testing only. Stability was evaluated according to manufacturer protocols, confirming the control is stable at +2° to +8° C as originally packaged and for at least 14 days at +2° to +8° C once opened.

## Technological Characteristics

Liquid control material prepared from human serum with stabilizers and preservatives. Designed for use with immunonephelometry (BN Systems) and immunoturbidimetry (TurbiTimeSystem) platforms. No specific materials standards or software algorithms are described.

## Regulatory Identification

A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.

## Predicate Devices

- N/T Protein Control SL ([K002852](/device/K002852.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# SEP = 4 2001

## 510(k) Summary For N/T Protein Control SL

### Manufacture's Name, Address, Telephone, and Contact Person, Date of 1. Preparation:

Manufacturer:

Dade Behring Marburg GmbH Emil-von-Behring Str. 76 Marburg/Germany

Contact Information:

Dade Behring Inc. Glasgow Site P.O. Box 6101 Newark, Delaware 19714 Attn: Rebecca S. Ayash Tel: 302-631-6276

Preparation date:

July 31, 2001

#### Device Name/ Classification: 2.

N/T Protein Control SL:

Quality Control Material (assayed)

Classification Number:

Class I (862.1660)

#### Identification of the Legally Marketed Device: 3.

N/T Protein Control SL (K002852)

#### Device Description: 4.

N/T Protein Control SL is a liquid control prepared from human serum with sfabilizers and preservative. It is intended to be used as an accuracy and precision control for the presentation of human serum proteins by immunonephelometry with BN™ Systems and by immunoturbidimetry with the TurbiTimeSystem.

#### 5. Device Intended Use:

N/T Protein Controls SL/L, M, and H are for use as accuracy and precision assayed controls in the determination of the following human serum proteins by immunonephelometry with BN™ Systems: IgG, IgG7, IgA, IgM, C3c, C4, Transferrin, in Innonophobionou y ma2-macroglobulin, Haptoglobin, α--acid glycoprotein, Prealbumin, Albumin, c.y antill Jpon) @2 maor J.g/L-chain lambda & kappa, β2-microglobulin, soluble Transferrin Receptor (STFR), Ferritin, IgE, and Total protein; and by immunoturbidimetry with the TurbiTimeSystem: IgG, IgA, IgM, C3c, C4, Transferrin, Albumin, Haptoglobin, α4acid glycoprotein.

Image /page/0/Picture/22 description: The image shows the word "CONFIDENTIAL" in large, bold, black letters. The letters are closely spaced and fill most of the frame. The background is plain white, providing a stark contrast to the bold text.

000027

{1}------------------------------------------------

### Medical device to which equivalence is claimed and comparison information: 6.

The modified N/T Protein Control SL is substantially equivalent in intended use to N/T The modified NYT T Totell Control SE to Sassianaly Se Sales N/T Control SL, like the Protein Control OL (K002002) ourrolly marketsused as quality control material to current NYT Frotein Oontrol OE is interface to be accuracy in BN™ Systems and the TurbiTimeSystem.

#### Device Performance Characteristics: 7.

## Stability:

Stability.
Stability was evaluated according to Dade Behring protocols and the control was found to Stablity was evaluation as overling to D to +8° C, as originally packaged and for at least 14 days at +2° to +8° C, once opened.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features an abstract image of an eagle with three lines representing its wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

SEP = 4 2001

Ms. Rebecca S. Ayash Director, Regulatory Affairs Dade Behring, Inc. Glasgow Site, PO Box 6101 Newark, DE 19714

K012468 Re:

> Trade/Device Name: N/T Protein Control SL Regulation Number: 21 CFR 862.1660 Regulatory Class: I, reserved Product Code: JJY Dated: July 31, 2001 Received: August 2, 2001

Dear Ms. Ayash:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{3}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and v additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Steven Sutman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Dade Behring Inc.
N/T Protein Control SL
510(k) Notification

# Indications for Use Statement

N/T Protein Control SL Device Name:

## Indications for Use:

N/T Protein Controls SL/L, M, and H are for use as accuracy and precision N/ 1 Toteln Controls of the following human serum proteins by immunonephelometry with BN™ Systems: IgG, IgG14, IgA, IgM, C3c, C4, immanonopholomosiy wantitrypsin, α2-macroglobulin, Haptoglobin, α;-acid Translermir, Albumin, or antilippensin, Ceruloplasmin, RbP, Ig/L-chain lambda & giyooprotein, 1 Toalbumin, Huble Transferrin Receptor (STTR), Ferritin, IgE, and Kappa, p2-morogiobailin, obluble idimetry with the TurbiTimeSystem: IgG, IgA, IgM, C3c, C4, Transferrin, Albumin, Haptoglobin, α+-acid glycoprotein.

(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)

Over-The-Counter-Use (Optional Format 1-2-96)

Kesia Alexander for Liam Cooper
(Division Sign-Off)

510(k) Number

000025

---

**Source:** [https://fda.innolitics.com/device/K012468](https://fda.innolitics.com/device/K012468)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
