← Product Code [LLZ](/productcode/LLZ) · K012439

# DENTALEYE 2 DENTAL IMAGE MANAGEMENT SYSTEM (K012439)

_Radco Data AB · LLZ · Feb 6, 2002 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K012439

## Device Facts

- **Applicant:** Radco Data AB
- **Product Code:** [LLZ](/productcode/LLZ.md)
- **Decision Date:** Feb 6, 2002
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

Indicated for acquisition, storage, display, and manipulation of Dental Images. Intended uses in the dental industry include the following: Acquiring and handling of radiographs and color images: Viewing and manipulation for insurance claims adjudication Viewing and manipulation for diagnostic purposes Viewing for patient education and consultation including cosmetic imaging. When used for diagnostic purposes, the patient population will be the general public, and the diseases/conditions that the device will be used to diagnose are: dental caries, periodontal disease and bone loss, tooth fractures, jaw misalignment, and other diseases and conditions that are encountered by general practitioners and specialists in the dental care field.

## Device Story

DentalEye 2 is a dental image management system used by dental practitioners. It acquires, stores, displays, and manipulates dental radiographs and color images. The system operates on a standard PC platform (Windows 95, 98, 2000, or NT) using a Studio PCTV video capture card. It allows clinicians to view and manipulate images for diagnostic purposes, insurance claims, and patient education. By facilitating image management, it assists in the diagnosis of dental conditions such as caries, periodontal disease, and fractures, potentially improving clinical decision-making and patient consultation.

## Clinical Evidence

Bench testing only. Software testing and validation were performed according to internal company protocols. Results confirmed that device performance satisfies design intent.

## Technological Characteristics

PC-based software system; utilizes Studio PCTV video capture hardware; compatible with Windows 95, 98, 2000, and NT; supports acquisition, storage, display, and manipulation of dental radiographs and color images.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- TAU TigerView ([K955237](/device/K955237.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K012439

## EXHIBIT 2

FEB 0 6 2002

## Radco Data AB Anestagatan 29 SE-163 53 Spånga, SWEDEN Phone (011) 46 (0)8 5221 84 00 Fax (011) 46 (0) 73 920 84 10 Contact: Anders Berggren, President and CEO July 30, 2001

## 510(k) Summary of Safety and Effectiveness

- 1. Identification of the Device: Proprietary-Trade Name: "DentalEye 2" Dental Image Management System Classification Name: 90 LLZ Common/Usual Name: Dental Image Management System
- 2. Equivalent legally marketed device: This product is similar in design and identical in function to the TAU TigerView, K955237
- 3. Indications for Use (intended use): Indicated for acquisition, storage, display, and manipulation of Dental Images
- 4. Description of the device: Intended uses in the dental industry include the following:

Acquiring and handling of radiographs and color images:

Viewing and manipulation for insurance claims adjudication

Viewing and manipulation for diagnostic purposes

Viewing for patient education and consultation including cosmetic imaging. When used for diagnostic purposes, the patient population will be the general public, and the diseases/conditions that the device will be used to diagnose are: dental caries, periodontal disease and bone loss, tooth fractures, jaw misalignment, and other diseases and conditions that are encountered by general practitioners and specialists in the dental care field.

| Comparison Areas                  | TAU TigerView, K955237                                                         | "DentalEye 2" Dental Image Management System |
|-----------------------------------|--------------------------------------------------------------------------------|----------------------------------------------|
| Indications for use               | Indicated for acquisition, storage, display, and manipulation of Dental Images | SAME                                         |
| Host platform                     | Standard PC platform with Windows NT                                           | SAME, can also use Windows 95, 98, and 2000  |
| Video capture                     | Matrox Meteor II                                                               | Studio PCTV                                  |
| Receive images from other systems | YES                                                                            | SAME                                         |

- 5. Safety and Effectiveness, comparison to predicate device:

{1}------------------------------------------------

- 6. Testing information and Conclusion
r ssting information and Ochelable: "
In all material respects, the "DentalEye 2" Dental Image Management System is In an material respocts, and Tiger View, K955237. Testing was performed substantially equirtations procedures. Software testing and validation were according to internal company protocols established before testing was conducted. done according to which by designated technical professionals before software I est results were results support the conclusion that actual device performance satisfies the design intent.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal is circular and contains the words "HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized design of three overlapping shapes that resemble an abstract caduceus or a bird in flight. The image is in black and white.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

FEB 0 6 2002

DentalEve AB % Mr. Daniel Kamm P.E. Regulatory Engineer Kamm & Associates P.O. Box 7007 Deerfield, IL 60015

Re: K012439

Trade/Device Name: DentalEye 2, Dental Image Management System Regulation Number: 21 CFR 872.1800 Regulation Name: Extraoral source x-ray system Regulatory Class: II Product Code: 90 MUH Dated: December 11, 2001 Received: December 12, 2001

Dear Mr. Kamm:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050,

{3}------------------------------------------------

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed noutication. The I Drice results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you don't the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

| 8xx.1xxx                         | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4654 |
| Other                            | (301) 594-4692 |

Additionally, for questions on the promotion and advertising of your device, please contact the Additionally) , 101 questions of (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on of relocution to promazive the Act may be obtained from the Division of Small Manufacturers, your responsibilities and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Nancy C. Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## j) Indications for Use

510(k) Number

5 ro(if) Names.

Acquiring and handling of dental radiographs and color images Viewing and manipulation for insurance claims adjudication. Viewing and manipulation for diagnostic purposes Viewing for patient education and consultation including cosmetic imaging.

Concurrence of CDRH, Office of Device Evaluation (ODE)

ﻬﻢ ﻓﻲ ﺍﻟﻤﺮﺍﺟﻊ Prescription Use_

Over the Counter Use_ OR (Per 21 CFR 801.109)

David G. Leggott

---

(Division Sign-Off)

Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number

---

**Source:** [https://fda.innolitics.com/device/K012439](https://fda.innolitics.com/device/K012439)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
