← Product Code [KFM](/productcode/KFM) · K012325

# CARDIOVENTION CORX SYSTEM, MODEL FG 0001 (K012325)

_Cardiovention, Inc. · KFM · Apr 9, 2002 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K012325

## Device Facts

- **Applicant:** Cardiovention, Inc.
- **Product Code:** [KFM](/productcode/KFM.md)
- **Decision Date:** Apr 9, 2002
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.4360
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The CardioVention CORx System is intended to be used in surgical procedures requiring extracorporeal gas exchange support. This device can be used with a legally marketed thermal regulating system. The device is indicated for use in procedures requiring a maximum blood flow rate of six liters/minute and lasting up to six hours. The CORx System is intended for use with Medtronic's Bio-medicus Bio-Console.

## Device Story

Extracorporeal hemodynamic and gas exchange support system; consists of active venous air removal device (defoamer), centrifugal blood pump, membrane oxygenator, AirVac Controller, and magnetic pump adapter. Operates by centralizing venous line air for active removal; centrifugal pump uses eight-vaned impeller driven by Medtronic Biomedicus Bio-Console. Used in surgical settings by perfusionists/clinicians. AirVac Controller powers air detection/removal. Provides gas transfer via hollow fiber oxygenator. Benefits patient by maintaining hemodynamic stability and gas exchange during bypass; reduces entrained air risks. Output monitored by clinicians to manage perfusion parameters during surgery.

## Clinical Evidence

Bench testing only. Evaluated blood/gas pathway integrity (6 hours/10 minutes), oxygen/CO2 transfer rates, transmembrane pressure drop, blood cell damage (hemolysis, platelet/WBC counts), and air handling (bubble counts). Results showed performance equal to or better than predicate devices over 6-hour duration. No clinical data provided.

## Technological Characteristics

System includes hollow fiber oxygenator (polypropylene), eight-vaned centrifugal impeller pump, and active venous defoamer. Magnetic coupling to external console. Designed for 6-hour use at 6 L/min flow. Sterile, single-use disposable. No software-based control algorithms described; AirVac Controller provides active air removal.

## Regulatory Identification

A nonroller-type cardiopulmonary and circulatory bypass blood pump is a prescription device that uses a method other than revolving rollers to pump the blood through an extracorporeal circuit for periods lasting less than 6 hours for the purpose of providing either:(i) Full or partial cardiopulmonary bypass (i.e., circuit includes an oxygenator) during open surgical procedures on the heart or great vessels; or(ii) Temporary circulatory bypass for diversion of flow around a planned disruption of the circulatory pathway necessary for open surgical procedures on the aorta or vena cava.

## Special Controls

*Classification* —Class II (special controls). The special controls for this device are:(i) Non-clinical performance testing must perform as intended over the intended duration of use and demonstrate the following: Operating parameters, dynamic blood damage, heat generation, air entrapment, mechanical integrity, and durability/reliability;
(ii) The patient-contacting components of the device must be demonstrated to be biocompatible;
(iii) Sterility and shelf life testing must demonstrate the sterility of patient-contacting components and the shelf life of these components; and
(iv) Labeling must include information regarding the duration of use, and a detailed summary of the device- and procedure-related complications pertinent to use of the device.
(b)
*Nonroller-type temporary ventricular support blood pump* —(1)*Identification.* A nonroller-type temporary ventricular support blood pump is a prescription device that uses any method resulting in blood propulsion to provide the temporary ventricular assistance required for support of the systemic and/or pulmonary circulations during periods when there is ongoing or anticipated hemodynamic instability due to immediately reversible alterations in ventricular myocardial function resulting from mechanical or physiologic causes. Duration of use would be less than 6 hours.(2)
*Classification.* Class III (premarket approval).(c)
*Date premarket approval application (PMA) or notice of completion of product development protocol (PDP) is required.* A PMA or notice of completion of a PDP is required to be filed with FDA on or before September 8, 2015, for any nonroller-type temporary ventricular support blood pump that was in commercial distribution before May 28, 1976, or that has, on or before September 8, 2015, been found to be substantially equivalent to any nonroller-type temporary ventricular support blood pump that was in commercial distribution before May 28, 1976. Any other nonroller-type temporary ventricular support blood pump shall have an approved PMA or declared completed PDP in effect before being placed in commercial distribution.

## Predicate Devices

- Optimin Hollow Fiber Oxygenator (Cobe)
- Sarns Delphin Centrifugal Blood Pump (Terumo Sarns)
- HVR2200 Hard Shell Venous Reservoir (Cobe)

## Reference Devices

- Medtronic Biomedicus Bio-Console ([K941921](/device/K941921.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# APR = 9 2002

CardioVention CORx System 510(k) Notification

#### CONFIDENTIAL

KO12325

## CardioVention CORx System

# 510(k) Summary

This 510(k) summary is being submitted in accordance with the requirements of 21 CFR 808.87 and 807.92.

Date Prepared: June 16, 2001

ll.

510(k) number:

#### l. 510(k) Submitter Information

Tessa Yamut Name: Director of Quality Assurance / Regulatory Title: Affairs 3045 Stender Way Address: Santa Clara, CA 95054 Same as above Contact Person: Preparation Date of the June 16, 2001 510(k) Summary: Device Information CardioVention CORx System Device Name: CardioVention CORx System Trade Name: Oxygenator Common / Usual Centrifuqal Blood Pump Name: Venous Defoamer Classification Name: Cardiopulmonary Bypass Oxygenator Blood Pump, Non-roller type, Cardiopulmonary Bypass Cardiopulmonary Bypass Defoamer Class III Preamendments device for which FDA has Device Classification: not yet called for premarket approval applications.

510(k) Summary

{1}------------------------------------------------

#### III. Device Description

The CardioVention CORx System is an extracorporeal hemodynamic and gas exchange support system for extracorporeal perfusion. The CardioVention CORx System consists of an active venous air removal device (defoamer), a centrifugal blood pump, a membrane oxygenator, an AirVac Controller and a pump adapter. The AirVac Controller powers the detection and facilitates removal of entrained air in the air handling system (defoamer) of the disposable unit. The pump adapter, which is magnetic, allows the use of the Medtronic Biomedicus Bio-Console (k941921) with the CORx System. The Medtronic's Biomedicus Bio-Console drives the CORx System's centrifugal pump.

The following components of the CORx System disposable are provided sterile, for single use only:

| Device                 | Description                                                                                                                              |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------|
| Defoamer (Air Removal) | A device that replaces the air removal<br>function of venous reservoirs by centralizing<br>venous line air and actively removing the air |
| Centrifugal Blood Pump | An eight-vaned impeller (centrifugal) pump                                                                                               |
| Oxygenator             | A hollow fiber gas transfer device utilizing<br>polypropylene, double-mat, commercially<br>available fiber                               |

#### IV. Predicate Devices

The CardioVention CORx System, which consists of an active venous air removal device (defoamer), a centrifugal blood pump, a membrane oxygenator, an AirVac Controller and a pump adapter, is substantially equivalent to a combination of three predicate devices.

The Predicate Device for the CardioVention CORx System Oxygenator is:

| Name:         | Optimin Hollow Fiber Oxygenator |
|---------------|---------------------------------|
| Manufacturer: | Cobe                            |
| Status:       | Post-enactment                  |

510(k) Summary

{2}------------------------------------------------

The Predicate Device for the CardioVention CORx System Centrifugal Pump, with the pump adaptor for use with Medtronic's Biomedicus Bio-Console is:

Name: Manufacturer: Status:

Sarns Delphin Centrifugal Blood Pump Terumo Sarns Post-enactment

The predicate device for the CardioVention CORx System Defoamer, with the AirVac Controller, is:

Name:

Manufacturer: Status:

HVR2200 Hard Shell Venous Reservoir (defoamer) Cobe Post-enactment

#### V. Indications for Use

The CardioVention CORx System is intended to be used in surgical procedures requiring extracorporeal gas exchange support. This device can be used with a legally marketed thermal requlating system. The device is indicated for use in procedures requiring a maximum blood flow rate of six liters/minute and lasting up to six hours.

The CORx System is intended for use with Medtronic's Bio-medicus Bio-Console.

### Comparison to Predicate Devices

The CardioVention CORx System is used for patient support and duplicates the functionality of three separate disposable predicate devices. A summary comparison of the intended uses and generic specifications between the CORx System and the predicate systems is provided on the following page. For many of the plastic components, the CardioVention CORx System utilizes the same materials as the predicate devices.

The largest surface area component of the CORx System, the microporous membrane material, is the identical material as the predicate membrane oxygenator. Both the predicate and CORx System membrane oxygenators are blood outside the fiber, hollow fiber designs. The predicate oxygenator contains an integral heat exchanger while the CORx System is designed to work with ancillary heating / cooling products.

510(k) Summarv

{3}------------------------------------------------

Both the predicate centrifugal blood pump and the centrifugal blood pump component of the CORx System are of the "impeller" type and incorporate similar materials.

#### VI. Summary of Physical Characterization / Integrity and Performance Testing

The table below summarizes the tests and results that were performed on the CORx System to ensure that it conforms to specificationsand is as safe and effective as the predicate devices:

| Test Type                                   | Purpose                                                                                                                               | Summary of Results                                                                                                                                                                                                                                             |
|---------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Blood pathway<br>integrity- 6 hours         | To ensure that the blood pathway<br>does not leak under worst case<br>conditions                                                      | The blood pathway was<br>exposed to pressures of 750<br>mmHg (1.5X rated maximum<br>pressure) for 6 hours with no<br>failures of integrity detected.                                                                                                           |
| Gas Pathway<br>Integrity- 10<br>minutes     | To ensure that the gas pathway<br>does not leak under worst case<br>conditions                                                        | The gas pathway was<br>exposed to pressures of 30<br>mmHg (1.5X rated maximum<br>pressure) for 10 minutes with<br>no failures of integrity<br>detected.                                                                                                        |
| Blood Pathway<br>Volume                     | To determine the static priming<br>volume of the blood pathway                                                                        | The volume of the blood<br>pathway was determined to<br>be 328 mls.                                                                                                                                                                                            |
| Oxygen Transfer<br>Rate- 6 hours            | To compare oxygen transfer and<br>outlet saturations at minimum,<br>nominal, and maximum rated flow<br>rates to the predicate device. | Oxygen transfer was greater<br>than or equal to the predicate<br>device at all blood/gas flow<br>rates over 6 hours. Arterial<br>saturation (outlet) was 100%<br>at all conditions and<br>timepoints.  No loss in<br>performance occurred over<br>the 6 hours. |
| Carbon Dioxide<br>Transfer Rate- 6<br>hours | To compare oxygen transfer and<br>outlet saturations at minimum,<br>nominal, and maximum rated flow<br>rates to the predicate device. | Carbon dioxide transfer was<br>greater than or equal to the<br>predicate device at all<br>blood/gas flow rates over 6<br>hours. No loss in<br>performance occurred over<br>the 6 hours.                                                                        |

510(k) Summary

{4}------------------------------------------------

| Transmembrane<br>pressure drop- 6<br>hours | To compare transmembrane<br>pressure drop at minimum,<br>nominal, and maximum rated flow<br>rates to the predicate device. | Transmembrane pressure<br>drop was lower than the<br>predicate device.  No<br>clinically significant pressure<br>increases occurred over the 6<br>hours.                                               |
|--------------------------------------------|----------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Blood Cell<br>Damage                       | To compare blood cell damage to<br>the predicate devices at their<br>respective maximum rated flow<br>rates.               | Blood cell damage (plasma<br>free hemoglobin, index of<br>hemolysis, modified index of<br>hemolysis, platelet count,<br>white blood cell count) was<br>equal to or less than the<br>predicate devices. |
| Air Handling                               | To compare venous air handling<br>performance to the predicate<br>devices.                                                 | Arterial return bubble counts<br>were equal or less than the<br>predicate devices.                                                                                                                     |

Based on a descriptive comparison of the CORx System and its predicate systems and the performance data summarized above, the CORx System is substantially equivalent to the currently marketed predicate systems for its intended use.

## 510(k) Summary

Page 5 of 5

{5}------------------------------------------------

### Class III Summary and Certification

CardioVention, Inc. certifies to the best of our knowledge that a reasonable search of all information known or otherwise available about the types and causes of safety or information known of other the arbeen reported for the nonroller-type cardiopulmonary encouvenous problems that here ertify that we are aware of the types of complications to which the device is susceptible and that, to the best of our knowledge, the following which the afthe types and causes of safety or effectiveness problems about the nonroller-type cardiopulmonary bypass blood pump is complete and accurate.

| Problem            | Potential Causes                                           |
|--------------------|------------------------------------------------------------|
| Blood Leak         | Compromised physical integrity                             |
| Inadequate pumping | Bearing / seal failure, decoupling                         |
| Biocompatibility   | Material selection                                         |
| Blood damage       | Heat generation, bearing / seal failure,<br>sheer stresses |

The section below includes citations of the materials upon which the above summary is based:

- Kolff J, RN Ankney, D Wurzel, R Devineni. Centrifugal pump failures. J Extra Corpor 트 Technol 1996 Sep; 28(3)118-22.
- Curtis JJ, TM Boley, JT Walls, TL Demmy, and RA Schmaltz. Frequency of seal . disruption with Sarns centrifugal pump in postcardiotomy circulatory assist. Artif Organs 1994 Mar; 18(3):225-7.
- Curtis JJ, JT Walls, CC Wagner-Mann, RA Schmaltz, TL Demmy, CA NcKenney, FA 트 Mann. Centrifugal pumps: description of devices and surgical techniques.
- Crane TN. Causes of cardiopulmonary bypass accidents 1996. 트 http://www.perfusion.com/perfusion/articles/general/9610-accidents.html

Tessa Yamut Director of QA/RA

uyanut

Date of Submission:

July 20, 2001

510(k) Number:

{6}------------------------------------------------

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

APR = 9 2002

Ms. Tessa Yamut Director of Quality Assurance/Regulatory Affairs CardioVention, Inc. 3045 Stender Way Santa Clara, CA 95054

Re: K012325

Trade Name: CardioVention CORx System, Model FG0001 Regulation Number: 21 CFR 870.4350, 870.4360, 870.4230 Regulation Name: Oxygenator, Pump, Defoamer Regulatory Class: Class III (three) Product Code: DTZ, KFM, DTP Dated: November 7, 2001 Received: November 9, 2001

#### Dear Ms. Yamut:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{7}------------------------------------------------

Page 2 - Ms. Tessa Yamut

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

L. DeRatt, M.D.

Bram D. Zuckerman, M.D. Acting Director Division of Cardiovascular and Respiratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{8}------------------------------------------------

#### INDICATIONS FOR USE STATEMENT IV.

510(k) Number (if known): _k @/ 232 5

Device Name: CardioVention CORx System

Indications for Use:

The CardioVention CORx System is intended to be used in surgical procedures requiring extracorporeal gas exchange support. This device can be used with a legally marketed thermal regulating system. The device is indicated for use in procedures requiring a maximum blood flow rate of six liters/minute and lasting up to six hours.

The CORx System is intended for use with the Medtronic Bio-medicus Bio-Console.

✓

21 CFR 801.109)

Division of Cardiovascular & Respiratory Devices
510(k) Number K012325

---

**Source:** [https://fda.innolitics.com/device/K012325](https://fda.innolitics.com/device/K012325)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
