PJ.S COMFORT PORTABLE ELECTRIC BREAST PUMP

K012275 · Limerick, Inc. · HGX · Jan 7, 2002 · Obstetrics/Gynecology

Device Facts

Record IDK012275
Device NamePJ.S COMFORT PORTABLE ELECTRIC BREAST PUMP
ApplicantLimerick, Inc.
Product CodeHGX · Obstetrics/Gynecology
Decision DateJan 7, 2002
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 884.5160
Device ClassClass 2

Intended Use

The intended use of the electrically powered (diaphragm-type) suction device is to express milk from the breast.

Device Story

Portable electric breast pump; diaphragm-type pump with vacuum regulator; battery or line-powered (12V DC). Inputs: manual breast cup placement. Operation: electric sensor controls intermittent suction (30-45 cycles); automatic vacuum release when set level reached. Safety: bacterial filter with optical sensor; automatic motor shut-off if milk enters filter to prevent valve damage. Output: expressed breast milk into collection cup. Used by lactating women in home or clinical settings. Benefits: facilitates milk expression; protects pump mechanism from contamination via automatic shut-off.

Clinical Evidence

No clinical data. Safety and effectiveness supported by bench testing per FDA November 1993 Draft Reviewer Guidance for Electrical, Mechanical, and Environmental Performance Requirements.

Technological Characteristics

Diaphragm-type pump; 12V DC power (external transformer or battery pack); vacuum range 150-220 mm Hg; includes bacterial filter with optical sensor; silicone breast cups and tubing; dimensions 7 x 12 x 3.5 inches; weight 4.5 lbs. Materials: food-contact approved per 21 CFR 176, 177, 178.

Indications for Use

Indicated for lactating women to express milk from the breast.

Regulatory Classification

Identification

A powered breast pump in an electrically powered suction device used to express milk from the breast.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## JAN 0 7 2002 Name of Company: Limerick Inc. Premarket Notification - 510(k) r remantel Notification - Scomfort® Portable Electric Breast Pump EXHIBIT #1 Page 1 Of 3 ## 510(K) SUMMARY This summary of 510 (k) safety and effectiveness information is being submitted This summary of of of to (k) oursty and 21 CFR §807.92. The assigned 510(k) number is: _150120 צו 201201 - Submitter's Identification: 1. Limerick Inc. 903 North San Fernando Blvd., Suite 5 Burbank, California 91504 Date Summary Prepared: March 27, 2001 #### Name of the Device: 2. pi·s comfort® Portable Electric Breast Pump #### Predicate Devices Information: 3. K#850705, White River Portable Electric Breast Pump, White Rive, Division of Natural Technologies, Inc. K971543, Luxmatic 12-15 ISC Portable electric Breast Pump, KaWeCo GmbH #### Device Description: র্ব . pirs comfort® Portable Electric Breast Pump (suction device with a compressor unit) contains a diaphragm-type pump with a vacuum regulator (between 150mm unity oontaine a clapmelling a filter, soft silicone tube and plastic tubing. An accessory kit includes silicone breast cups and a collection cup. The device is battery and line-powered for 12V DC by an external transformer or utilizes a 12V-battery pack. The device is controlled by an electric sensor with intermittent suction at approximately 30-45 cycles. The bacterial filter contains an optical sensor, with an automatic shut off. The motor stops when liquid is sucked into the bacterial. The pump can be used with the Single Breast Pump Kit #6166-3 or the Double Breast Pump Kit #6166-0. Dimensions are 7 x 12 x 3 .5 inches and the weight is 4.5 lbs. {1}------------------------------------------------ K012275 Name of Company: Limerick Inc. Premarket Notification - 510(k) Name of Devise: The pj•scomfort® Portable Electric Breast Pump ### EXHIBIT #1 Page 2 of 3 Suction is controlled by the pump automatically. It will automatically be released when the adjusted vacuum is reached. An electronic sensor is built-in for automatic motor shut off: If milk flows into the bacterial filter, the motor stops automatically at once and the inside valves from the pump cannot be damaged through the milk. The bacterial filter should be changed regularly. #### 5. Intended Use: The intended use of the electrically powered (diaphragm-type) suction device is to express milk from the breast. #### Comparison to Predicate Devices: 6. Table of Comparison to Legally Marketed Device: The following is a comparison chart outlining differences and similarities between pj.s comfort® Portable Electric Breast Pump, White River Portable Electric Breast Pump and Luxmatic 12-15 ISC Portable Electric Breast Pump: | PARAMETER | pj's comfort | WHITE RIVER (K#850705) | LUXMATIC (K#971543) | |--------------------------|---------------|------------------------|---------------------| | Pump Type | Diaphragm | Same | Same | | Vacuum Control Cycle | Yes | Same | Same | | Adjustable Suction Range | 150-220 mm Hg | Not available | Same | | Vacuum Gauge | Yes | Same | Same | | Bacteria Filter | Yes | Not available | Same | | Collection Bottles | Yes | Same | Same | | Single Patient Use | Yes | Same | Same | | Indications For Use | Same | Same | Same | | Timer | Yes | Not available | Not available | | Weight | 4.5 lbs. | 10 lbs. | 4.5 lbs. | {2}------------------------------------------------ KO12275 Name of Company: Limerick Inc. Premarket Notification - 510(k) Name of Devise: The pj•scomfort® Portable Electric Breast Pump ### EXHIBIT #1 Page 3 of 3 ### Discussion of Non-Clinical Tests Performed for Determination of 7. Substantial Equivalence is as follows: - Testing information demonstrating safety and effectiveness of the pj•s comfort® a) Portable Electric Breast Pump in the intended environment of use is supported by testing that was conducted in accordance with the FDA November 1993 Draft "Reviewer Guidance for Premarket Notification Submissions", DCRND, which outlined Electrical, Mechanical and Environmental Performance Requirements. - Patient-contacting material includes both skin contacted material (breast cups) b) and milk contacting materials. 21 CFR Parts 176, 177 and 178 were reviewed in order to ascertain materials approved for food contact. Materials are approved for food contact. #### Discussion of Clinical Tests Performed: 8. Non-Applicable #### 9. Conclusions: The pirs comfort® Portable Electric Breast Pump has the same intended use, similar design and technology as the White River Portable Electric Breast Pump and the Luxmatic 12-15 ISC Portable Electric Breast Pump. As our comparison chart indicates, as well as our testing data, the pjrs comfort® Portable Electric Breast Pump raises no new questions of safety or effectiveness. Thus, when compared to the predicated device, the pj.s comfort® Portable Electric Breast Pump does not incorporate any significant changes in intended use, method or operation, material or design that could affect safety or effectiveness. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular emblem with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is a stylized image of three wavy lines that resemble a caduceus, a symbol often associated with healthcare. ood and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JAN 0 7 2002 Ms. Patricia A. Kelly President Limerick, Inc. 903 N. San Fernando Boulevard Suite 5 BURBANK CA 91504-4327 Re: K012275 Trade/Device Name: pj.s comfort® Portable Electric Breast Pump Regulation Number: 21 CFR 884.5160 Regulation Name: Powered breast pump Regulatory Class: II Product Code: 85 HGX Dated: November 9, 2001 Received: November 13, 2001 Dear Ms. Kelly: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050. {4}------------------------------------------------ This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter: | 8xx.1xxx | (301) 594-4591 | |----------------------------------|----------------| | 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 | | 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 | | 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 | | Other | (301) 594-4692 | Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, Nancy C. Brogdon Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ Name of Company: Limerick Inc. Namo of Golmpacation - 510(k) 0 122 25 Name of Devise: The pj.scomfort® Portable Electric Breast Pump ### EXHIBIT B # INDICATIONS FOR USE 510(K) Number: 1012275 The pj•scomfort® Portable Electric Breast Pump Device Name: The intended use of the electrically powered (diaphragm-Indication for Use: type) suction device is to express milk from the breast of lactating women. Concurrence of CDRH, Office of Device Evaluation (ODE) | Prescription Use | or | Over-the-Counter Use | x | |----------------------|----|----------------------|---| | (Per 21 CFR 801.109) | | | | Nancy Brogdon (Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices | 510(k) Number | K012275 | |---------------|---------| |---------------|---------|
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...