OPTETRAK TOTAL KNEE SYSTEM, STRAIGHT STEM EXTENSIONS - ADDITIONAL SIZES
Applicant
Exactech, Inc.
Product Code
JWH · Orthopedic
Decision Date
Jul 31, 2001
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3560
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The OPTETRAK® Total Knee Systems are indicated for use in skeletally mature individuals undergoing primary surgery for total knee replacement due to osteoarthritis, osteonecrosis, rheumatoid arthritis and/or post-traumatic degenerative problems. They are also indicated for revision of failed previous reconstructions where sufficient bone stock and soft tissue integrity are present. CAUTION: For cemented use only in the USA.
Device Story
Optetrak Total Knee System Stem Extensions are orthopedic implants used in total knee replacement surgery. These extensions are designed to provide additional stability and fixation for the femoral or tibial components. They are manufactured from Titanium alloy (ASTM F-1472) and feature a Morse taper connection. The device is intended for use by orthopedic surgeons in a hospital or surgical center setting. The extensions are implanted during primary or revision knee arthroplasty to address bone stock deficiencies or to enhance component stability. The surgeon selects the appropriate length and diameter to match the patient's anatomy. The device is for cemented use only in the USA. By providing a range of sizes, the system allows for better customization to the patient's specific anatomical needs, potentially improving the longevity and stability of the knee reconstruction.
Clinical Evidence
No clinical data. Bench testing only. Analytical analysis demonstrated that stem extension strength is independent of length and diameter, supporting the conclusion that the additional sizes perform similarly to the existing, predicate Optetrak stem extensions.
Technological Characteristics
Material: Titanium alloy (ASTM F-1472). Design: Morse taper connection with flutes. Sterilization: Gamma radiation (Cobalt 60), 25-37 kGy, SAL 10^-6. Packaging: Vacuum-sealed high barrier bags (Nylon/EVOH/Polyethylene) and Tyvek pouches.
Indications for Use
Indicated for skeletally mature patients undergoing primary total knee replacement for osteoarthritis, osteonecrosis, rheumatoid arthritis, or post-traumatic degeneration, and for revision of failed reconstructions with sufficient bone stock and soft tissue integrity. Contraindicated in patients with active infection, insufficient bone stock or soft tissue integrity, mental/neuromuscular disorders preventing joint control, or conditions where weight/age/activity level risk extreme loads and early failure.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
{1}------------------------------------------------
KO12251
2320 NW 66TH COURT GAINESVILLE, FL 32653
352-377-1140 FAX 352-378-2617
## JUL 3 1 2001
# Exactech® Optetrak ™ Total Knee System Stem Extensions - Additional Sizes
# 510(k) Summary of Safety and Effectiveness Special 510(k)
Exactech® Inc. Sponsor: 2320 N.W. 66th Court Gainesville, Florida 32653
| Phone: | (352) - 377 - 1140 |
|--------|--------------------|
| Fax: | (352) - 378 - 2617 |
FDA Establishment Number 1038671
Robert Paxson Contact: Director of Engineering & Development
Date:
June 22, 2001
Section 4 Page 1 of 4
{2}------------------------------------------------
# Exactech® Optetrak ™ Total Knee System Stem Extensions - Additional Sizes
# 510(k) Summary of Safety and Effectiveness Special 510(k)
### Classifications / Proprietary Names:
Classification Name:
Prosthesis, knee, patellofemorotibial, semi-contrained, cemented, polymer/metal polymer
Trade / Proprietary Model Names:
Optetrak Total Knee System Stem Extensions - Additional Sizes
JWH Product Code: 888.3560 C.F.R. Section:
II Device Class:
Classification Panel: Orthopedic
# Legally Marketed Devices for Substantial Equivalence Comparison:
| Model | Manufacturer | 510(k) Number |
|--------------------------------------------|----------------------|----------------------|
| Optetrak Total Knee<br>Optetrak Total Knee | Exactech<br>Exactech | #K933610<br>#K954208 |
| Nexgen | Zimmer | -------- |
| Maxim | Biomet | -------- |
| Durcon Total Stabilizer | Howmedica | -------- |
### Device Information:
### INDICATIONS
The OPTETRAK® Total Knee Systems are indicated for use in skeletally mature individuals undergoing primary surgery for total knee replacement due to osteoarthritis, marradals and organis id arthritis and/or post-traumatic degenerative problems. They ore also indicated for revision of failed previous reconstructions where sufficient bone stock and soft tissue integrity are present.
CAUTION: For cemented use only in the USA.
{3}------------------------------------------------
## Exactech® Optetrak ™ Total Knee System Stem Extensions - Additional Sizes
## 510(k) Summary of Safety and Effectiveness Special 510(k)
#### CONTRAINDICATIONS
The OPTETRAK® Total Knee Systems are contraindicated in patients with active infection, patients without sufficient bone stock to allow appropriate insertion and fixation of the prosthesis, patients without sufficient soft tissue integrity to provide adequate stability, and in patients with either mental or neuromuscular disorders that do not allow control of the knee joint, and in patients whose weight, age, or activity level might cause extreme loads and early failure of the system.
#### DEVICE DESCRIPTION
Like the original Optetrak stem extensions, the proposed Optetrak Straight Stem Extension additional sizes are composed of Titanium alloy (ASTM F-1472). The proposed stem extensions have the same Morse taper call out and flute that the existing stem extensions offer. The design incorporates the same design features as the predicate Optetrak stem extensions over new lengths and diameters to create a complete range of sizes for the surgeon. The proposed Straight Stem Extension Line Extensions are fully compatible with the existing Optetrak Tibial Trapezoid trays (#K933610) and the Constrained Condlyar Femoral Component and bushing (#K954208).
| Packaging Materials | |
|---------------------|---------------------------------------------------------------------|
| Material | Composition |
| Vacuum Bag | High Barrier – 3 Mil, 75 gauge<br>Nylon with EVOH 2.25 Polyethylene |
| Polymer Bags | 3 Mil Polyethylene |
| Inner/Outer Pouches | 48GAPET/.002LDPE |
| Pouch Seals | 1073B Uncoated Tyvek® |
| Box | Heavy weight cardboard |
| Outer Shrink-Wrap | Clear, Light-Weight Plastic |
| Shipping Cartons | Heavy-weight Corrugated Cardboard |
#### PACKAGING MATERIALS
#### STERILIZATION INFORMATION
Method: Gamma radiation (Cobalt 60 source) Dose: 25 - 37 kGy Sterility Assurance Level (SAL): 10-6
{4}------------------------------------------------
# Exactech® Optetrak ™ Total Knee System Stem Extensions - Additional Sizes
## 510(k) Summary of Safety and Effectiveness Special 510(k)
### PERFORMANCE DATA
The design assurance reviews for the proposed products did not indicate the need for additional testing since the parts were so comparable to the existing Exactech stem extensions and no new failure modes were identified.
According to analytical analysis the proposed stem extension components showed that the stem extension strength was independent of length and diameter. Therefore, the proposed stem extensions should have the similar clinical results to the existing Exactech straight stem extensions.
We conclude that the new Optetrak Stem Extensions are substantially equivalent to other devices legally marketed in the United States, most notably Exactech's predicate Optetrak products.
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the text "Public Health Service" in a simple, sans-serif font. The text is arranged on a single line, with each word clearly legible. The overall impression is clean and straightforward, suggesting an official or institutional context.
Image /page/4/Picture/2 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three swooping lines representing its wings. The eagle is positioned in the center of a circular border. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper portion of the circle.
JUL 3 1 2001
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Robert Paxson Director of Engineering & Development Exactech. Inc. 2320 N.W. 66th Court Gainsville, Florida 32653
Re: K012251
Trade Name: Optetrak Total Knee System Regulatory Class: II Regulation Number: 888.3560 Product Code: JWH Dated: July 17, 2001 Received: July 18, 2001
Dear Dr. Paxson:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general control provisions of the Act. The general control provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
{6}------------------------------------------------
### Page 2 - Mr. Robert Paxson
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and 11 you desire specific darices), please contact the Office of Compliance at additionally 609110 for questions on the promotion and advertising of your device, (2017) 59 + 105 + 1 = 0 ffice of Compliance at (301) 594-4639. Also, please note the regulation predise ounder the Oriety of eference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597, or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
R.M.M.h.l.u.u.m.o. P.R.
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
### Exactech® Optetrak Total Knee System Straight Stem Extensions - Additional Sizes
### Indications for Use
510(k) Number:
K. 2-25|
Device Name:
Optetrak Total Knee System Straight Stem Extensions
### INDICATIONS
The OPTETRAK® Total Knee Systems are indicated for use in skeletally mature individuals undergoing primary surgery for total knee replacement due to osteoarthritis, osteonecrosis, rheumatoid arthritis and/or post-traumatic degenerative problems. They are also indicated for revision of failed previous reconstructions where sufficient bone stock and soft tissue integrity are present.
CAUTION: For cemented use only in the USA.
### CONTRAINDICATIONS
The OPTETRAK® Total Knee Systems are contraindicated in patients with active infection, patients without sufficient bone stock to allow appropriate insertion and fixation of the prosthesis, patients without sufficient soft tissue integrity to provide adequate stability, and in patients with either mental or neuromuscular disorders that do not allow control of the knee joint, and in patients whose weight, age, or activity level might cause extreme loads and early failure of the system.
smhheeein2
(Division Sign-Off) Division of General, Restorative and Neurological Devices
510(k) Number K012251
Please do not write below this line - use another page if needed.
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
or
Over the Counter Use
Section 3 Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.