CARDIONET AMBULATORY ECG MONITOR WITH ARRHYTHMIA DETECTION, MODEL CN1001
K012241 · Cardionet, Inc. · DSI · Feb 1, 2002 · Cardiovascular
Device Facts
Record ID
K012241
Device Name
CARDIONET AMBULATORY ECG MONITOR WITH ARRHYTHMIA DETECTION, MODEL CN1001
Applicant
Cardionet, Inc.
Product Code
DSI · Cardiovascular
Decision Date
Feb 1, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1025
Device Class
Class 2
Indications for Use
1. Patients who have demonstrated a need for cardiac monitoring and are at low risk of developing primary ventricular fibrillation or sustained ventricular tachycardia. 2. Patients with dizziness or lightheadedness 3. Patients with palpitations 4. Patients with syncope of unknown etiology 5. Patients who require monitoring for non life-threatening arrhythmias. such as atrial fibrillation, other supra-ventricular arrhythmias, evaluation of various bradyarrhythmias and intermittent bundle branch block. This includes post operative monitoring for these rhythms 6. Patients recovering from coronary artery bypass graft (CABG) surgery who require monitoring for arrhythmias 7. Patients requiring monitoring for arrhythmias inducing co-morbid conditions such as hyperthyroidism or chronic lung disease 8. Patients with obstructive sleep apnea to evaluate possible nocturnal arrhythmias 9. Patients requiring arrhythmia evaluation for etiology of stroke or transient cerebral ischemia, possibly secondary to atrial fibrillation 10. Data from the device may be used by another device to analyze, measure or report QT interval. The device is not intended to sound any alarms for QT interval changes.
Device Story
CardioNet Arrhythmia Detector is a modification of the CardioNet Ambulatory ECG Monitor (K003707) adding arrhythmia detection. System comprises patient-worn Sensor, Monitor, and charging Base. Sensor acquires 3-channel ECG via skin electrodes; conditions, filters, digitizes, and transmits signal via RF to Monitor. Monitor performs analysis using Mortara ECG algorithm; stores 24 hours of data; transmits data to Monitoring Center via integrated cellular modem or phone line. Trained clinical personnel at Monitoring Center review data. Device provides pacer pulse detection, leads-off detection, and event markers. Used in ambulatory settings; patient-worn. Benefits include remote arrhythmia evaluation for patients with syncope, palpitations, or post-surgical recovery. Output allows clinicians to identify cardiac rhythm abnormalities and inform clinical decision-making.
Clinical Evidence
Bench testing only. Device performance validated against ANSI/AAMI EC38:1998 (Ambulatory electrocardiographs) and ANSI/AAMI EC57:1998 (Testing and reporting performance results of cardiac rhythm and ST segment algorithms). Biocompatibility testing conducted per ISO 10993 for skin contact > 30 days. Electrical safety verified per UL2601 and ANSI/AAMI EC1-1993. No clinical trial data presented.
Indicated for patients requiring cardiac monitoring for non-life-threatening arrhythmias (e.g., AFib, supra-ventricular arrhythmias, bradyarrhythmias, bundle branch block). Population includes patients with dizziness, palpitations, syncope, post-CABG recovery, hyperthyroidism, chronic lung disease, or obstructive sleep apnea. Contraindicated for patients at high risk of primary ventricular fibrillation or sustained ventricular tachycardia.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
{0}------------------------------------------------
K012241
# FEB 01 2
# 510(k) Summary
# Date: 10/30/2001
### Submitter Information
CardioNet, Inc. Attn: Mr. Donald V. Canal Vice President RAQA 510 Market Street San Diego, Ca 92101
### 1. Name of Device
Trade/Proprietary Name:
Model 1001 - CardioNet ECG Monitor with Arrhythmia Detection.
Ambulatory ECG Monitor and Arrhythmia Common/Usual Name: detector and alarm.
CFR §870.2920 'Telephone Electrocardiograph Classification Name: Transmitter and Receiver', CFR §870.2800 'Medical Magnetic Tape Recorder'. CFR §870.1025 Procode DSI 'Arrhythmia Detector and Alarm'.
Class III Preamendments Device for which FDA Class: has not yet called for PMAs.
### 2. Predicate Devices
The predicate devices selected are as follows:
- 1. CardioNet Ambulatory ECG Monitor, cleared by FDA under 510(k) number K003707; 870.2800 "Medical Magnetic Tape Recorder". Procode: MWJ "Electrocardiograph, Ambulatory (Without analysis)":
- 2. Cardiac Telecom Cardiac TeleCom HeartLink II ECG Arrhythmia detector and Alarm System- Manufactured by Cardiac Telecom, Inc., cleared by FDA under 510(k) number K982803; 870.1025 Procode DSI 'ECG Arrhythmia detector and alarm".
#### 3. Device Description ·
The CardioNet Arrhythmia Detector which is also referred to as the subject device, is a modification to the CardioNet Ambulatory ECG monitor cleared by FDA under 510(k) number K003707, in May 2001. The subject device includes the addition of an ECG analysis capability that allows detection of cardiac arrhythmia.
{1}------------------------------------------------
As described in the previous 510(k) the subject device is comprised of three components: 1) a patient-worn Sensor, 2) a Monitor and 3) a charging Base.
The Sensor acquires the ECG signal from the patient's body. It is connected to electrodes attached to the patient's skin. It is suspended in place by a neck strap. The ECG signal is conditioned, filtered, digitized and transmitted to the Monitor via Radio Frequency bi-directional communications. The Sensor utilizes a disposable AAA battery.
The Monitor receives, analyzes, stores and transmits the ECG data. The Monitor has an LCD display and handles the user interaction for the system. The Monitor has an integrated cellular modem to allow communications with and without physical connection to a phone line. The Monitor has a re-chargeable battery. When an ECG threshold is exceeded or if the patient initiates a call, the ECG data is transmitted to the CardioNet Monitoring Center where trained clinical personnel review the data. The Monitor Center can also request data from the Monitor, the Monitor will transmit ECG data are requested. All ECG data for the past 24 hours is resident on the Monitor and can be transmitted to the Monitoring Center as requested by Monitoring Center Personnel. The subject device utilizes the proven Mortara ECG analysis algorithm that is currently used in the Quinton Q-tel telemetry system, 510(k) number K003576, 807.1025, Procode DSI, Class III; and the Datex-Ohmeda CS/3 telemetry system, 510(k) number K000882, 807.1025, Procode DSI, Class III. The Quinton Q-tel telemetry system and the Datex-Ohmeda telemetry system are both Class III Preamendments devices for which FDA has not yet requested PMAs.
The Base provides power, RS232, and telephone communication connectivity to allow telephone transmission of ECG data when in the Monitor is charging. The Base does not include any software; it is a passive component that provides input to the Monitor.
### 4. Indications For Use
The indications for use for the subject device is as follows:
- 1. Patients who have demonstrated a need for cardiac monitoring and are at low risk of developing primary ventricular fibrillation or sustained ventricular tachycardia.
- 2. Patients with dizziness or lightheadedness
- 3. Patients with palpitations
- 4. Patients with syncope of unknown etiology
- 5. Patients who require monitoring for non life-threatening arrhythmias, such as atrial fibrillation, other supra-ventricular arrhythmias, evaluation of various bradvarrhythmias and intermittent bundle branch block. This includes post operative monitoring for these rhythms
- 6. Patients recovering from coronary artery bypass graft (CABG) surgery who require monitoring for arrhythmias
- 7 . Patients requiring monitoring for arrhythmias inducing co-morbid conditions such as hyperthyroidism or chronic lung disease
- 8. Patients with obstructive sleep apnea to evaluate possible nocturnal arrhythmias
{2}------------------------------------------------
- 9. Patients requiring arrhythmia evaluation for etiology of stroke or transient cerebral ischemia, possibly secondary to atrial fibrillation
- 10. Data from the device may be used by another device to analyze, measure or report QT interval. The device is not intended to sound any alarms for QT changes.
| | CardioNet Arrhythmia<br>Detector<br>(Subject device) | CardioNet Ambulatory<br>ECG Monitor | Cardiac Telecom<br>HeartLink II |
|-------------------------------------------|--------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | CardioNet, Inc. | CardioNet, Inc. | Cardiac Telecom, Inc. |
| 510k number | Subject Device<br>Class III DSI<br>870.1025 | K003707<br>Class II MWJ | K982803<br>Class III<br>Preamendments<br>device, DSI<br>870.1025 |
| Maximum storage<br>capacity. | 24 hours | 24 hours | Last 30 minutes of<br>data |
| Includes<br>transtelephonic<br>capability | Yes | Yes | Yes |
| Event Recording<br>time. | Marker inserted into full disclosure data. All data is available for 24 hours at the request of Monitoring Center. | +/- 5 minutes from Patient Marker in event mode. Patient Marker inserted into full disclosure data in Holter mode. | Marker inserted into full disclosure data. All data is available for 30 minutes at the request of the Monitoring Center. |
| Pacer Pulse<br>Detection and<br>reporting | Yes | Yes | Yes |
| Leads off detection | Yes | Yes | Yes |
| Power input/<br>Battery Type | Sensor - 1 AAA,<br>Monitor - Li lon,<br>rechargeable<br>Base - 6 volts | Sensor - 1 AAA,<br>Monitor - Li Ion,<br>rechargeable<br>Base - 6 volts | Sensor - NiCad<br>Tower PC - 120 V AC |
| Enclosure | Sensor - Molded TPE,<br>Monitor - Molded Plastic | Sensor - Molded TPE,<br>Monitor - Molded<br>Plastic | Molded plastic |
| A to D sampling<br>rate (samples/sec) | 250 | 250 | Not Specified in<br>available literature |
| Resolution (A/D<br>conversion bits) | 12 bit | 12 bit | Not Specified in<br>available literature |
| Number of<br>channels | 3 | 3 | 1 |
| Number of<br>Electrodes | 3 | 3 | 3 |
| Storage Type<br>(digital or Tape) | Digital flash non-<br>removable | Digital flash non-<br>removable | Hard disk |
| | CardioNet Arrhythmia<br>Detector<br>(Subject device) | CardioNet Ambulatory<br>ECG Monitor | Cardiac Telecom<br>HeartLink II |
| Input impedance | >1 Mohm | >1 Mohm | 2.5 Mohm @10 Hz |
| Frequency<br>Response | 0.5 Hz up to 40 Hz | 0.5 Hz up to 40 Hz | Not Specified in<br>available literature |
| Communication<br>Means | Cellular Phone,<br>PSTN phone,<br>RS232, audio Coupled | Cellular Phone,<br>PSTN phone,<br>RS232, audio Coupled | PSTN telephone<br>connection |
| Operating<br>temperature range | Sensor: 20°C to 45°C<br>Monitor: 0°C to 45°C | Sensor: 20°C to 45°C<br>Monitor: 0°C to 45°C | 25+/- 10 degrees C |
| Storage<br>temperature range | -20°C to 65°C | -20°C to 65°C | Not Specified in<br>available literature |
| Relative Humidity | 5% to 95% | 5% to 95% | 50%+/-30% |
| Dimensions | Sensor -2" x 12" x 0.5"<br>Monitor - 4.5" x 3.5" x 1.5" | Sensor -2" x 12" x 0.5"<br>Monitor - 4.5" x 3.5" x<br>1.5" | Sensor - 3. 5" x 1.25"<br>x 4.5 "<br>Tower PC - 17" x 5" x<br>12" |
| Weight | Sensor 1.8 ounces<br>Monitor 10 ounces | Sensor 1.8 ounces<br>Monitor 10 ounces | Sensor 10.5 ounces<br>Monitor 10 lbs. |
| RF Modulation | 902 - 928 MHz with<br>frequency hopping<br>(30 foot range) | 902 - 928 MHz with<br>frequency hopping (30<br>foot range) | 902-928 MHz<br>No frequency hopping<br>(300 foot range) |
| LCD display | Yes with touch screen | Yes | Yes on the Tower PC |
# 5. Comparison to Predicate Devices
{3}------------------------------------------------
# 6. Non-Clinical Performance Test Summary
There are no applicable device-specific quidance documents for the subject device. The CardioNet Ambulatory ECG Monitor meets or exceeds the applicable requirements in ANSI/AAMI/ISO EC38 'Ambulatory electrocardiographs', 1998. The ECG analysis algorithm and the ECG viewing station meets the requirements of ANSI/AAMI EC57:1998 "Testing and reporting performance results of cardiac rhythm and ST segment algorithms". Note: the CardioNet Arrhythmia detector does not generate alarms for ST segment changes.
Several other ECG related standards have also been considered as design inputs for the CardioNet Ambulatory ECG Monitor: ANSI/AAMI EC13:'Cardiac Monitors, Heart rate meters and alarms', 1992; ANSI/AAMI EC11 "Diagnostic Electrocardiographic devices", 1991; and ANSI/AAMI EC53-1995 "ECG Cable and Lead wires".
{4}------------------------------------------------
The Sensor skin contact materials will be tested to meet the requirements for surface devices, skin contact > 30 days as required in ISO10993 'FDA modified Biocompatibility' table.
The Sensor, Monitor and Base comply with the applicable requirements in UL2601, and ANSI/AAMI EC1-1993, "Safe Current Limits for electromechanical Apparatus: December", 1993.
The applicable communications for the Monitor will meet the applicable requirements in FCC part 15, subpart C, and part 68.
### 7. Substantial Equivalence Conclusion
The CardioNet Arrhythmia detector is equivalent to the predicate devices as supported by the descriptive information and the performance testing demonstrates that it meets the applicable requirements of ANSI/AAMI EC38:1998 'Ambulatory electrocardiographs', the ECG analysis algorithm and the ECG viewing station meet the requirements of ANSI/AAMI EC57:1998 "Testing and reporting performance results of cardiac rhythm and ST segment algorithms".
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of a human figure with outstretched arms, resembling a bird in flight. The figure is composed of three curved lines that create a sense of movement and dynamism.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB - 1 2002
Mr. Jerry Kalieta Director Quality Assurance CardioNet, Inc. 510 Market Street San Diego, CA 92101
Re: K012241
Trade Name: CardioNet Ambulatory ECG Monitor with Arrhythmia Detection Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector and Alarm Regulatory Class: Class III (three) Product Code: DSI Dated: October 31, 2001 Received: November 5, 2001
Dear Mr. Kalieta:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Page 2 - Mr. Jerry Kalieta
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
L
Bram D. Zuckerman, M.D. Acting Director Division of Cardiovascular and Respiratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# Indications for Use Statement
510(k) Number (if known):
Device Name: CardioNet Ambulatory Monitor with Arrhythmia Detection
Indications for Use:
- 1. Patients who have demonstrated a need for cardiac monitoring and are at low risk of developing primary ventricular fibrillation or sustained ventricular tachycardia.
- 2. Patients with dizziness or lightheadedness
- 3. Patients with palpitations
- 4. Patients with syncope of unknown etiology
- 5. Patients who require monitoring for non life-threatening arrhythmias. such as atrial fibrillation, other supra-ventricular arrhythmias, evaluation of various bradyarrhythmias and intermittent bundle branch block. This includes post operative monitoring for these rhythms
- 6. Patients recovering from coronary artery bypass graft (CABG) surgery who require monitoring for arrhythmias
- 7. Patients requiring monitoring for arrhythmias inducing co-morbid conditions such as hyperthyroidism or chronic lung disease
- 8. Patients with obstructive sleep apnea to evaluate possible nocturnal arrhythmias
- 9. Patients requiring arrhythmia evaluation for etiology of stroke or transient cerebral ischemia, possibly secondary to atrial fibrillation
- 10. Data from the device may be used by another device to analyze, measure or report QT interval. The device is not intended to sound any alarms for QT interval changes.
Image /page/7/Figure/14 description: The image is a black and white photograph of an object. The object is long and narrow, and it appears to be made of a solid material. The object is dark in color, and it has a rough surface. The object is positioned horizontally in the image.
(Please Do Not Write Below This Line - Continue On Another Page If Needed) ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Concurrence of CDRH, Office Of Device Evaluation (ODE)
Prescription Use Over-The-Counter Use (Per 21 CFR 801.109)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.