K012238 · Philips Medical Systems North America Co. · JAK · Oct 1, 2001 · Radiology
Device Facts
Record ID
K012238
Device Name
DOSERIGHT
Applicant
Philips Medical Systems North America Co.
Product Code
JAK · Radiology
Decision Date
Oct 1, 2001
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Indications for Use
DoseRight is intended to be used for automatic dose regulation during CT scanning for the purpose of obtaining the desired image quality at the lowest possible dose. DoseRight is a software option to the CT Vision family of CT systems. These systems are whole body Computed Tomography (CT) systems which are diagnostic X-ray systems intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission of data from the same axial plane taken at different angles. They includes signal analysis and display equipment, patient and equipment supports, component parts and accessories which are used in combination with signal and image processing software to facilitate the relative localization of anatomy with gray-scale representation of density relative to water utilizing Houndsfield indices with or without contrast mediums. They are used for the display, storage and analysis of digital diagnostic CT images. They are intended for use by a physician in the diagnosis and planning phases of patient conditions and treatment.
Device Story
DoseRight is a software option for Philips CT Vision whole-body scanners; enables automatic dose regulation. Input: patient scanogram (scanned projection radiograph). Operation: calculates required mAs/KV combination based on patient diameter to achieve user-defined image quality; recommends scan parameters to user. Output: recommended mAs/KV settings displayed on console. Usage: clinical setting; operated by physicians/technicians. Workflow: user performs scanogram; system suggests parameters; user may accept or override with standard protocol. Benefit: optimizes image quality while minimizing patient radiation exposure.
Clinical Evidence
Bench testing only. No clinical data provided. Safety and effectiveness supported by hazard analysis and compliance with established performance standards for CT systems.
Technological Characteristics
Software option for CT Vision family scanners. Operates via scanogram-based calculation of mAs/KV parameters. Complies with UL-2601 and IEC-601-1 electrical safety standards and 21 CFR 1020.30/1020.33 radiation performance standards.
Indications for Use
Indicated for patients undergoing whole body Computed Tomography (CT) scanning where automatic dose regulation is desired to achieve pre-determined image quality at the lowest possible radiation dose.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
SmartScan option to the HiSpeed CT/i CT System (K964746)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the Philips logo. The logo consists of the word "PHILIPS" in bold, sans-serif font, placed above a shield-like shape. Inside the shield, there are wavy lines representing radio waves, and stars are scattered around the waves. The logo is black and white.
Image /page/0/Picture/1 description: The image shows the word "PHILIPS" in bold, black letters against a white background. The font is sans-serif and appears to be slightly distressed, giving it a vintage or worn look. The letters are large and fill most of the frame, making the word easily readable.
OCT = 1 2001
## Philips Medical Systems
KO12238
# 510(K) Summary
In accordance with the requirements of the Safe Medical Device Act, Philips Medical Systems North America Company herewith submits a 510(K) summary of safety and effectiveness for the following device.
| SUBMITTER NAME / ADDRESS: | Philips Medical Systems North America Company<br>710 Bridgeport Avenue<br>Shelton, CT 06484-0917 |
|---------------------------|--------------------------------------------------------------------------------------------------|
| CONTACT PERSON / TEL NO: | Peter Altman, Director of Regulatory Affairs<br>Tel No: (203) 926-7031 |
| DATE SUMMARY PREPARED: | July 16, 2000 |
| ESTABLISHMENT NO .: | 1217116 |
| CLASSIFICATION NAME: | Computed Tomography X-ray System<br>(21 CFR 892.1750; Class II; Tier 2; 90JAK) |
| COMMON/USUAL NAME: | Option for CT X-ray System |
| TRADE/PROPRIETARY NAME: | DoseRight |
| PREDICATE DEVICE(S): | SmartScan option to the HiSpeed CT/I CT System:<br>manufactured by GE Medical Systems |
### DEVICE DESCRIPTION:
DoseRight is a software option to the Philips CT Vision product family of computed tomography systems. The CT Vision products are whole body scanners. The CT Vision product family is comprised of the following systems. Each product has been cleared for commercial distribution via the referenced 510(k) submission.
- · Philips CT Secura (previously named Tomoscan AV-NT, ref: K991278)
- · Philips CT Secura MV (ref: K000819)
- · Philips CT Aura (previously named Tomoscan CS, ref: K982631)
DoseRight is used to enable automatic dose regulation during scanning for the purpose of obtaining the desired image quality at the lowest possible dose. Using DoseRight, the dose required to attain a pre-determined image quality is calculated from the scanogram (scanned projection radiograph) of the patient and resultant scan parameters are then recommended to the user for the scan protocol. The user performs the scanogram in the normal way. At the comletion of the scanogram, and with DoseRight enabled, a mAs/KV combination is displayed to the user, which is the recommended combination to use in the scan protocol to achieve a pra-determined user defined image quality relative to patient diameter. However, if the user so chooses, this recommendation can be ignored and the user can proceed with the standard protocol.
> Phillios Medical Systems North America Company 472 Wheelers Farm Road P.O. Box 3828 Milford, Connecticus 06460 03 \ 926-767 Fax: (203) 783-3060
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows a logo for Philips. The logo is a shield shape with the word "PHILIPS" written at the top. Inside the shield, there are four stars arranged in a square pattern, with a wavy line in the center. The logo is black and white.
#### INTENDED USE
DoseRight is intended to be used for automatic dose regulation during CT scanning for the purpose of obtaining the desired image quality at the lowest possible dose.
DoseRight is a software option to the CT Vision family of CT systems. These systems are whole body Computed Tomography (CT) systems which are diagnostic X-ray systems intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission of data from the same axial plane taken at different angles. They includes signal analysis and display equipment, patient and equipment supports, component parts and accessories which are used in combination with signal and image processing software to facilitate the relative localization of anatomy with gray-scale representation of density relative to water utilizing Houndsfield indices with or without contrast mediums. They are used for the display, storage and analysis of digital diagnostic CT images. They are intended for use by a physician in the diagnosis and planning phases of patient conditions and treatment.
#### SUBSTANTIAL EQUIVALENCE INFORMATION:
The DoseRight option is considered comparable and substantially equivalent to the following predicate device.
- · SmartScan Option used in GE Medical System's HiSpeed CT/i CT System (ref: K964746)
The above-mentioned CT system has been cleared for commercial distribution via the referenced 510(k) submission.
#### SAFETY INFORMATION:
The DoseRight option introduces no new safety issues to the CT Vision family of CT systems other then those already known with these systems in which this type option is used. Computed tomography is a mature technology with which industry and users have many years of experience. These devices must comply with the appropriate sections of the Radiation Control for Health and Safety Act. The DoseRight option as part of the CT Vision family of CT systems and its associated labeling complies with the applicable requirements of the Federal X-ray Performance standards 21CFR 1020.30, 1020.33.
The Philips CT Vision family of CT systems with which the DoseRight option is used are designed to comply with the requirements of Underwriters Laboratories (UL) Standard for Safety of Medical Electrical Equipment (UL-2601) and be classified by Underwriters Laboratories or an equivalent test laboratory. They are also designed to comply with the requirements of IEC-601-1 (Medical Electrical Equipment).
The results of the hazard analysis, combined with the appropriate preventive measures taken indicate that the device is of minor level of concern as per the August 29, 1991 issue of the "Reviewer's Guidance for Computer Controlled Medical Devices Undergoing 510(k) Review".
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a seal for the U.S. Department of Health & Human Services. The seal features an abstract eagle with three stripes on its back, symbolizing the department's mission to promote health and well-being. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
d and Drug Administration 00 Corporate Boulevard ockville MD 20650
OCT - 1 2001
Mr. Frank Gianelli Senior Regulatory Affairs Specialist Philips Medical Systems North America 710 Bridgeport Avenue P.O. Box 860 SHELTON CT 06484-0917
Re: K012238 Trade/Device Name: DoseRight_ Option for CT X-ray System Regulation Number: 21 CFR 892.1750 Regulation Name: Computed Tomography X-ray System Regulatory Class: II Product Code: 90 JAK Dated: July 16, 2001 Received: July 17, 2001
Dear Mr. Gianelli:
We have reviewed your Section 510(k) premarket notification of intent to market the device We nave reviewed your Section 3 ro(t) premium is substantially equivalent (for the indications
referenced above and have determined the device is substantially equivalent (fo ferenced above and nave decemined the arroed predicate devices marketed in interstate for use stated in the encrosule) to regally manat date of the Medical Device Ameradments, or to commerce prior in May 26, 1776, the enaother with the provisions of the Federal Food, Drug, devices that have occh recuire approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The You may, dietelore, market the device, but jeer of and of engistration, listing of general controls provisions of the rict labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), IT your device is classined (500 a00 v) ... Existing major regulations affecting your device can be it may be sunfect to additional condois. Extrems, Title 21, Parts 800 to 898. In addition, FDA may found in the Ood of I occal Regaring your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean I rease of advised that I Dr 3 rossance or our device complies with other requirements of the Act that I DA has made a decorminations administered by other Federal agencies. You must of any Federal Statutes and regulations annualing, but not limited to: registration and listing comply with and not 110. 5 requences Part 801); good manufacturing practice requirements as set (21 ℃FR Part 807), laborang (21 Cegulation (21 CFR Part 820); and if applicable, the electronic forth in the quality systems (QD) rogusations 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name :
Indications For Use :
DoseRight is intended to be used for automatic dose regulation during CT scanning for the purpose of obtaining the desired image quality at the lowest possible dose.
DoseRight is a software option to the CT Vision family of CT systems. These systems are whole body Computed Tomography (CT) systems which are diagnostic X-ray systems intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission of data from the same axial plane taken at different angles. They includes signal analysis and display equipment, patient and equipment supports, component parts and accessories which are used in combination with signal and image processing software to facilitate the relative localization of analomy with gray-scale representation of density relative to water utilizing Houndsfield indices with or without contrast mediums. They are used for the display, storage and analysis of digital diagnostic CT images. They are intended for use by a physician in the diagnosis and planning phases of patient conditions and treatment.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use |
|------------------------|
| ( Per 21 CFR 801.109 ) |
Over-The-Counter Use
Nancy C. Brogdon
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K012238
OR
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.