MUCHECK - MONITOR UNIT VALIDATION PROGRAM, MODEL V4.1
K012227 · Oncology Data Systems, Inc. · IYE · Oct 1, 2001 · Radiology
Device Facts
Record ID
K012227
Device Name
MUCHECK - MONITOR UNIT VALIDATION PROGRAM, MODEL V4.1
Applicant
Oncology Data Systems, Inc.
Product Code
IYE · Radiology
Decision Date
Oct 1, 2001
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5050
Device Class
Class 2
Attributes
Software as a Medical Device
Intended Use
The intended use for the muCheck - Monitor Unit Validation Program is the same as for the predicate devices: to calculate a monitor unit or dose for the purpose of validating a monitor unit or dose previously calculated by a primary radiation treatment planning system or hand calculation. The intended use is as a quality assurance tool only and not as a treatment planning device.
Device Story
muCheck is a standalone software program for IBM-compatible PCs in a Windows environment; functions as a secondary check for radiation therapy monitor unit and dose calculations. Operates independently of radiation treatment planning systems; does not connect to or control radiation hardware. Used by medical dosimetrists or radiation therapy staff as a quality assurance tool to validate primary calculations. Input involves manual entry of treatment parameters; software performs independent dose/monitor unit calculations. Output provides a secondary verification value; allows clinicians to compare results against primary planning system outputs. Benefits patient safety by ensuring accuracy of radiation delivery protocols; serves as a secondary check recommended by Task Group 40.
Clinical Evidence
No clinical data. Bench testing only; verification and validation test plans conducted in accordance with GMP guidelines. System specifications met; results compared against primary treatment planning systems and hand calculations, showing close correlation.
Technological Characteristics
Windows-based software for IBM-compatible PC; mouse and keyboard interface. Standalone operation; no hardware connectivity. Performs independent monitor unit and dose calculations for radiation therapy quality assurance.
Indications for Use
Indicated for use as a quality assurance tool to verify monitor unit or dose calculations performed by a primary radiation treatment planning system in radiation therapy departments.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
K980904 — MUCHECK MONITOR UNIT VALIDATION PROGRAM · Medical Dosimetry Services, Inc. · Jun 8, 1998
K091602 — MUCHECK V8 · Oncology Data Systems, Inc. · Jun 18, 2009
K992689 — MU PLUS · Rad Therapy Solutions, Inc. · Feb 10, 2000
K061152 — MUCHECK V7.0 · Oncology Data Systems, Inc. · May 10, 2006
K102583 — MUCHECK, VERSION 9.0 · Oncology Data Systems, Inc. · Nov 2, 2010
Submission Summary (Full Text)
{0}------------------------------------------------
# 510(k) Safety and Effectiveness Summary
K012227
1/2
Submitter:
Oncology Data Systems, Inc. 12401 Riverview Rd. Oklahoma City, Ok. 73173 tel. (405) 691-2770 fax (405) 691-5158
muCheck - Monitor Unit Validation Program
Contact: Date:
Gregory G. Miller July 13, 2001
Trade Name:
Common Name: Monitor Unit Validation Program
Classification Panel: Radiology
Classification Name:
Medical Charged Particle Radiation Therapy System(Accessory) 21 CFR 892.5050 (class II)
| Performance Standards: | none established under section 514 |
|--------------------------|-----------------------------------------------------------------|
| Substantial Equivalence: | mucheck – Monitor Unit Validation(K980904)<br>RadCalc (K010464) |
### Description:
The muCheck Monitor Unit Validation Program is a software program that is designed to operate on an IBM compatible personal computer in a Windows environment. It has been designed to operate on a stand alone mode independent of any radiation treatment planning system. It does not connect to or control any radiation hardware device. MuCheck performs monitor unit and dose calculations to verify the monitor unit and dose calculated by the primary radiation treatment planning system.
# Substantial Equivalence Summary:
### Intended Use:
The intended use for the muCheck - Monitor Unit Validation Program is the same as for the predicate devices: to calculate a monitor unit or dose for the purpose of validating a monitor unit or dose previously calculated by a primary radiation treatment planning system or hand calculation. The intended use is as a quality assurance tool only and not as a treatment planning device.
{1}------------------------------------------------
1-3 K0122227
2/2
In a radiation therapy department quality assurance is an important part of patient care. The ability to provide a secondary check for the primary monitor unit calculation is part of good treatment protocol as well being a recommendation by Task Group 40. Mucheck provides this very important quality assurance function.
# Safety and Effectiveness:
The staff at Oncology Data Systems includes a certified medical dosimetrist with over 23 years of clinical experience. The computer programming and design has been provided by a systems analyst with over 23 years of experience in the design and development of systems. The combined expertise as well as conformance to the GMP regulations helped to insure that the finished product is safe and effective to use.
A comprehensive users manual available as a hard copy as well as on-line, provides extensive documentation for the user. Initial system startup and training is provided as part of the service provided by Oncology Data Systems.
### Technological Characteristics:
The technological characteristics are the same as for the predicate devices. MuCheck was designed to operate in a windows environment using both mouse and keyboard.
#### Non-clinical tests:
Verification and validation test plans were completed in accordance with Oncology Data Systems procedures and GMP guidelines. A Hazard Analysis was completed and hazards were resolved as appropriate. All system specifications were met and testing performed to demonstrate substantial equivalence. The non-clinical tests were conducted using a treatment planning system or hand calculations and muCheck. The test results all matched very closely which supports the claim of substantial equivalence. See Figure 6.0 in section 6 for comparison summary.
### Summary of Clinical Testing:
Clinical testing was not required to demonstrate substantial equivalence or safety and effectiveness.
#### Conclusions:
Based upon the technological characteristics, intended use, and non-clinical tests. muCheck is substantially equivalent to the predicate device. The documentation submitted for review supports this claim.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular emblem with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol consisting of three curved lines that resemble a human figure or a flame. The logo is black and white.
Food and Drug Administration 0 Corporate Boulevard Rockville MD 20850
OCT - 1 2001
Mr. Gregory Miller President Oncology Data System, Inc. 12401 S. Riverview Road OKLAHOMA CITY OK 73173 Re: K012227
Trade/Device Name: muCheck-Monitor Unit Validation Program Regulation Number: 21 CFR 892.5050 Regulation Name: Medical Charged-Particle Radiation Therapy System
Regulatory Class: II Product Code: 90 IYE Dated: July 13, 2001 Received: July 16, 2001
Dear Mr. Miller:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Number (if known): KD12227
Device Name: Mucheck- Monitor Unit Validation Program
Indications For Use:
The muCheck - Monitor Unit Validation Program (the device under review) verifies the monitor unit or dose calculated by the primary treatment planning system. It serves as quality assurance as part of good treatment protocol to have a second means to verify the accuracy of the primary calculation.
The Intensity Modulated Radiotherapy and Diode corrections features added to the I the imensity Modulated Nationers of the program to verify dose calculations performed by the primary treatment planning system.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
Prescription Use L (Per 21 CFR 801.109)
OR
Over-The-Counter Use
(Optional Format 1-2-96)
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.