THE SALICEPT ORAL PATCH

K012126 · Carrington Laboratories, Inc. · OLR · Oct 3, 2001 · SU

Device Facts

Record IDK012126
Device NameTHE SALICEPT ORAL PATCH
ApplicantCarrington Laboratories, Inc.
Product CodeOLR · SU
Decision DateOct 3, 2001
DecisionSESE
Submission TypeTraditional
Device ClassClass U
AttributesTherapeutic

Indications for Use

The SaliCept Oral Patch is recommended for use in extraction sites and management of alveolar osteitis (dry socket). The SaliCept Oral Patch is also intended for the management of all types of wounds, injuries and ulcers of the oral mucosa. It is intended for professional use only, and will be marketed to dentists and other health professionals.

Device Story

SaliCept Oral Patch is a freeze-dried hydrogel dressing; composed of Acemannan Hydrogel (Aloe vera L. derivative), hydroxyethyl cellulose, polyvinylpyrrolidone, benzethonium chloride, and simethicone. Pliable, white/off-white, cotton-like texture; 1 cm diameter, 0.5 cm thick. Applied by dentists/health professionals to oral extraction sites or mucosal lesions. Upon contact with oral environment, patch slowly reverts to gel state; adheres to tissue; provides physical barrier protecting affected area from irritation; relieves pain. Benefits include reduced incidence of alveolar osteitis post-extraction and pain relief for aphthous ulcers. No complex electronics or software involved.

Clinical Evidence

Multiple clinical trials conducted. Aphthous ulcer studies (n=37, n=90, n=155) demonstrated significant pain reduction within 2-20 minutes compared to controls (Orabase-Plain, negative control). Alveolar osteitis study (n=1064 extraction sites) showed significantly lower incidence of dry socket (1.1%) compared to clindamycin-soaked Gelfoam (7.6%, p-value not specified but noted as significant). In vitro study confirmed high affinity of Acemannan Hydrogel for human buccal mucosal scrapings and gingival fibroblasts.

Technological Characteristics

Freeze-dried hydrogel dressing. Ingredients: Acemannan Hydrogel, hydroxyethyl cellulose, polyvinylpyrrolidone, benzethonium chloride, simethicone. Form factor: 1 cm diameter, 0.5 cm thick pledgets. Mechanism: Adhesion to oral mucosa via physical barrier protection; slow reversion to gel state in oral environment. No energy source, connectivity, or software.

Indications for Use

Indicated for patients requiring management of tooth extraction sites, alveolar osteitis (dry socket), and various oral mucosal wounds, injuries, or ulcers. For professional use by dentists and health professionals.

Regulatory Classification

Identification

Intended as a physical barrier for temporary protection of oral mucosal tissue and to provide pain relief. Unclassified status per SE to MGQ.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{1}------------------------------------------------ # 510 (k) SUMMARY The SaliCept™ Oral Patch The following represents the summary for premarket notification of The SaliCept Oral Patch, a product to be marketed by Carrington Laboratories, Inc., located at 2001 Walnut Hill Lane, Irving, TX 75038, (972) 518-1300. | CONTACT PERSON: | Kenneth M. Yates, D.V.M.<br>Vice President, Research and<br>Development/Regulatory Affairs<br>(972) 650-7312 | |---------------------------------------|--------------------------------------------------------------------------------------------------------------| | DEVICE OR PROPRIETARY NAME: | The SaliCept™ Oral Patch | | CLASSIFICATION NAME: | Hydrogel Wound Dressing | | COMMON NAME: | Hydrogel Wound Dressing | | ESTABLISHMENT<br>REGISTRATION NUMBER: | 16125446 | | DATE OF SUMMARY: | June 19, 2001 | #### CLASSIFICATION OF THE DEVICE: Unclassified under 21 CFR 8728.4022, hydrogel wound dressing and burn dressing. The predicate device, K953423, The Carrington™ Patch, is an unclassified device. #### DEVICE DESCRIPTION: The SaliCept Oral Patch is a freeze-dried gel that contains Acemannan Hydrogel," a product obtained from the clear inner gel of Aloe vera L. Other ingredients include hydroxyethyl cellulose, polyvinylpyrrolidone, benzethonium chloride, simethicone. It is pliable, white to off-white, with a texture similar to that of finely woven cotton. The SaliCept Oral Patch is packaged in blister cards of six pledgets per card, two blister cards per carton, for a total of 12 pledgets per carton. Each pledget is approximately 1 cm in diameter and 0.5 cm thick. #### INTENDED USE: The SaliCept Oral Patch is recommended for use in extraction sites and management of alveolar osteitis (dry socket). The SaliCept Oral Patch is also intended for the management of all types of wounds, injuries and ulcers of the oral mucosa. It is intended for professional use only, and will be marketed to dentists and other health professionals. #### CLAIMS: The SaliCept Oral Patch relieves pain by adhering to and protecting affected tissues from further irritation. It is safe if swallowed. {2}------------------------------------------------ # TECHNOLOGICAL CHARACTERISTICS: The SaliCept Oral Patch is a freeze-dried hydrogel that adheres to oral mucosa. It slowly reverts to the gel state in the oral environment while it adheres to and protects tissues. ## NON-CLINICAL DATA: An in vitro study was conducted in which both cultured human buccal mucosal scrapings and cultured human gingival fibroblasts were exposed to Acemannan Hydrogel that had been labeled with fluorescein-5-isothiocyanate I (FITC). A high affinity of FITC-labeled Acemannan Hydrogel for both cell cultures was demonstrated, whereas FITC alone did not appreciably stain either cell type. Of significance was the finding that FITC-labeled Acemannan Hydrogel had a high affinity for keratin. ### CLINICAL DATA: Thirty-seven patients with oral aphthous ulcers applied either Orabase-Plain or the SaliCept Oral Patch to their lesions. Both groups experienced significant reduction in pain within 20 minutes, with the pain relief lasting longer than the 1-hour observation period (Plemons J, Rees T, Binnie W, Wright J: Pain relief evaluation of Carrasyn® Hydrogel for recurrent aphthous stomatitis. Abstract presentation, 73th General Session and Exhibition of the International Association for Dental Research, Singapore, 1995). A clinical trial in which 90 patients with oral aphthous ulcers was conducted. Patients applied either Orabase-Plain, Carrington's Carrasyn Hydrogel Dressing in gel form, or Carrington's SaliCept Oral Patch to their ulcers. The SaliCept Oral Patch group experienced significant reduction in pain within 2 minutes of application (Plemons JM, Rees TD, Binnie WH, Wright JM, Guo I, Hall, JE: Evaluation of acemannan in the treatment of recurrent aphthous stomatitis. Wounds 1994;6(2):40-45). A clinical trial compared reduction in discomfort in 155 patients with oral aphthous ulcers. Patients applied either the SaliCept Oral Patch, a cyanoacrylate bioadhesive, or a negative control to their lesions. Both the SaliCept Oral Patch group and the cyanoacrylate bioadhesive group experienced significant reduction in pain vs negative control. There was no difference in pain reduction between the two active groups. A clinical trial compared the incidence of alveolar osteitis (AO, dry socket) in tooth extraction sites in which either clindamycin-soaked Gelfoam or the SaliCept Oral Patch was immediately placed post-extraction. Analysis revealed that 78 of 975 (8.0%) of mandibular 3d molar sites in the Gelfoam group developed AO whereas 11 of 958 (1.1%) mandibular 3rd molar sites developed AO in the SaliCept Oral Patch group. When all extraction sites were analyzed, it was shown that 78 of 1031 (7.6%) extraction sites in the Gelfoam group developed AO whereas 12 of 1064 (1.1%) extraction sites developed AO in the SaliCept Oral Patch group. The difference between groups was significant. {3}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or a bird in flight. The logo is black and white. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 OCT - 3 2001 Mr. Kenneth M. Yates Vice President, Research and Development Carrington Laboratories, Incorporated 2001 Walnut Hill Lane Irving, Texas, 75038 Re: K012126 Trade/Device Name: The SaliCept™ Oral Patch Regulation Number: None Regulation Name: Hydrogel Wound Dressing Regulatory Class: Unclassified Product Code: MGQ Dated: June 19, 2001 Received: July 6, 2001 Dear Mr. Yates: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. {4}------------------------------------------------ Page 2 - Mr. Yates You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Timothy A. Ulatowski Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ Page 1 of 1 510(k) Number: K012126 Device Name: The SaliCept™ Oral Patch Indications for Use: Recommended for use in tooth extraction sites and management of alveolar osteitis (dry socket). The SaliCept Oral Patch is also intended for the management of all types of oral wounds, injuries and ulcers of the oral mucosa. The Salicept Oral Patch relieves pain by adhering to and protecting affected tissues from further irritation. (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use OR Over-The-Counter Use Susan Rinn (Division Sign-Off) Division of Dental, Infection Control, and General Hospital Device oldid 510(k) Number -------
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...