BANDVIEW SYSTEM

K012103 · Applied Spectral Imaging , Ltd. · LNJ · Sep 14, 2001 · Hematology

Device Facts

Record IDK012103
Device NameBANDVIEW SYSTEM
ApplicantApplied Spectral Imaging , Ltd.
Product CodeLNJ · Hematology
Decision DateSep 14, 2001
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 864.5260
Device ClassClass 2

Indications for Use

The BandView system is intended to be used for karyotyping with real-time microscope images from cultured and stained cell specimens in their metaphase. The system works with bright field and fluorescent samples with all currently applied banding techniques including G-banding and DAPI-banding. All specimens suitable for banding analysis including from amniotic fluid, peripheral blood, chorionic villus, bone marrow, and tissue can be used without limitation to specific diseases. Karyotyping is normally applied for the pre and postnatal diagnosis of birth defects, chromosome abnormalities, genetic diseases (such as Down's syndrome), cancer, and for the follow up of cancer treatment. The BandView system does not locate metaphase spreads; it does not rank the given cells according to quality; it does not automatically classify chromosomes; it does require and relies completely on the operator to manipulate the digitized microscope images.

Device Story

System aids cytogeneticists in karyotype preparation; transfers real-time microscope images of metaphase chromosome spreads to computer; eliminates manual photograph processing/archiving. Operator manually manipulates digitized images using BandView analysis software to assemble karyotypes. Does not automatically locate spreads, rank cell quality, or classify chromosomes. Used in clinical laboratory settings by cytogenetic specialists. Benefits include increased efficiency in data management and document preparation for diagnostic workflows.

Clinical Evidence

No clinical data provided. Bench testing only.

Technological Characteristics

Computer-based image acquisition and analysis system. Compatible with bright field and fluorescent microscopy. Complies with EN 55022:1998, EN 55024:1998, IEC 61000-4-2:1995, and EN-1441 risk analysis standards. Software-assisted manual karyotyping tool.

Indications for Use

Indicated for karyotyping of cultured/stained metaphase cell specimens (amniotic fluid, peripheral blood, chorionic villus, bone marrow, tissue) for pre/postnatal diagnosis of birth defects, chromosome abnormalities, genetic diseases (e.g., Down's syndrome), and cancer/cancer treatment follow-up. No specific disease limitations.

Regulatory Classification

Identification

An automated cell-locating device is a device used to locate blood cells on a peripheral blood smear, allowing the operator to identify and classify each cell according to type. (Peripheral blood is blood circulating in one of the body's extremities, such as the arm.)

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ . # SEP 1 4 2001 ### 510(K) SUMMARY # BandView system # 510(k) Number KO |al03 _______________________________________________________________________________________________________________________________________________________ ### Applicant's Name: Applied Spectral Imaging Ltd. P.O.B. 101 Migdal Ha Emek 10551, Israel Tel: 972-4-6547567 Fax: 972-4-6547507 ## Contact Person: Shoshana Friedman, RAC Push-med Ltd. 117 Ahuzah St. Ra'anana 43373, Israel Tel: 972-9-7718130 Fax: 972-9-7718131 ## Date Prepared: June, 2001 ### Trade Name: BandView system ### Classification Name: Automated cell-locating device # Classification: The FDA has classified Automated cell-locating device as class II devices (product code 88 LNJ, Regulation No. 864.5260) and they are reviewed by the Pathology Panel. {2}------------------------------------------------ #### Predicate Device: - · Ikaros, manufactured by MetaSystems GmbH, cleared under K940240. - · Genevision, manufactured by Applied Imaging Corp. cleared under K902311. #### Performance Standards: No performance standards have been established for such device under Section 514 of the Federal Food, Drug, and Cosmetic Act. However, the BandView system complies with the following voluntary standards: - · EN 55022:1998, Class A: - · EN 55024:1998: - · IEC 61000-4-2:1995 - · EN-1441: Medical devices Risk Analysis. #### Indications: The BandView system is intended to be used for karyotyping with real-time microscope images from cultured and stained cell specimens in their metaphase. The system works with bright field and fluorescent samples with all currently applied banding techniques including G-banding and DAPI-banding. All specimens suitable for banding analysis including from amniotic fluid, peripheral blood, chorionic villus, bone marrow, and tissue can be used without limitation to specific diseases. Karyotyping is normally applied for the pre and postnatal diagnosis of birth defects, chromosome abnormalities, genetic diseases (such as Down's syndrome), cancer, and for the follow up of cancer treatment. The BandView system does not locate metaphase spreads; it does not rank the given cells according to quality; it does not automatically classify chromosomes; it does require and relies completely on the operator to manipulate the digitized microscope images. #### Device Description: The BandView system aids the cytogenetic specialist in the preparation of karyotypes of human cells in their metaphase. By transferring images of chromosome spreads from the microscope to a computer, the labor-intensive manual processing of photographs, data archiving and document preparation is eliminated. Karyotypes are assembled by the operator with the support of the Band View analysis software. The BandView system does not locate metaphase spreads, it does not rank the given cells according to quality; it does not automatically classify chromosomes; it does require and relies completely on the cytogeneticist to manipulate the digitized microscope images. {3}------------------------------------------------ # Substantial Equivalence: Applied Spectral Imaging Ltd. believes that the BandView system is substantially equivalent to its predicate device cited above without raising new safety and/or effectiveness issues. {4}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - ISSA" around the perimeter. Inside the circle is a stylized image of three human profiles facing right, stacked on top of each other. Food and Drug Administration 2098 Gaither Road Rockville MD 20850 Applied Spectral Imaging, Ltd. c/o Ms. Shoshanna Friedman, RAC. Managing Director Push-med Ltd. 117, Ahuza Street Ra'ananna 43373, Israel SEP 1 4 2001 Re: k012103 Trade/Device Name: BandView system Regulation Number: 21 CFR 864.5260 Regulation Name: Automated cell-locating device Regulatory Class: Class II Product Code: LNJ Dated: June 16, 2001 Received: July 5, 2001 Dear Ms. Shoshanna Friedman: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {5}------------------------------------------------ ## Page 2 - This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and 1 11 you active specific arrivitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, (301) 59 + 12 contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small mionifacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html". Sincerely yours, Steven Butman Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ ### INDICATIONS FOR USE # 510(k) Number (if known): Band View system Device Name: Indications for Use: The BandView system is intended to be used for karyotyping with real-time microscope images from cultured and stained cell specimens in their metaphase. The system works with braght field and fluorescent samples with all currently applied banding techniques including G-banding and DAPI-banding. All specimens suitable for banding analysis including from amniotic fluid, peripheral blood, chorionic villus, bone marrow, and tissue can be used without limitation to specific diseases. Karyotyping is normally applied for the pre and postnatal diagnosis of birth defects, chromosome abnormalities, genetic diseases (such as Down's syndrome), cancer, and for the follow up of cancer treatment. > The BandView system does not locate metaphase spreads; it does not rank the given cells according to quality; it does not automatically classify chromosomes; it does require and relies completely on the operator to manipulate the digitized microscope images. Donald A. Pape Division Sign-Off Division of Cliffical Laboratory Devices 0/2/0 510(k) Number (PLEASE DO NOT WRITE BELOW THIS LINE -CONTINUE ON ANOTHER PAGE IF NEEDED) 510(k) Number - - - Prescription Use V (Per 21 CFR 801.109) OR Over the Counter Use
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...