MONOCHROME PERFECTLY FLAT PANEL DISPLAYS, ME SERIES
Applicant
Totoku Electric Co., Ltd.
Product Code
LLZ · Radiology
Decision Date
Jul 31, 2001
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Indications for Use
The ME Series Medical Displays are intended for use with Picture Archiving Communication Systems (PACS) for medical imaging applications by physicians.
Device Story
ME Series Medical Displays are monochrome CRT-based monitors for medical imaging. Input signals include analog video (0.714Vp-p) and synchronization signals via BNC or D-sub 15p connectors. Devices transform electrical video signals into visual images on a flat panel display. Used in clinical environments by physicians for viewing PACS images. Output is a high-resolution monochrome medical image. Calibration is performed via software or an optional external photosensor (X-Rite DTP92). Benefits include high-resolution diagnostic visualization of medical images.
Clinical Evidence
No clinical data. Substantial equivalence is based on bench testing and comparison of technical specifications (luminance, resolution, scanning frequency, and safety/EMC compliance) against the predicate device.
Technological Characteristics
Monochrome CRT display; P45 or P4 luminescent material; analog video input (0.714Vp-p); BNC/D-sub 15p connectivity; 1200x1600 to 2048x2560 pixel resolution; 250-500MHz image clock; software-based luminance calibration with optional X-Rite DTP92 photosensor; power 100-240V AC; safety standards UL2601-1, CSA No.601-1, IEC60601-1.
Indications for Use
Indicated for use by physicians for medical imaging applications within Picture Archiving Communication Systems (PACS).
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{1}------------------------------------------------
K0/2099
## JUL 3 1 2001
510(k) Summary as required by 807.92
### 1. Company Identification
Totoku Electric Co., Ltd. 300 Oya, Ueda-shi, Nagano-ken , 386-0192, Japan Tel : 011-81-268-34-5484 Fax : 011-81-268-34-5565
## 2. Official Correspondent
Mikio Hasegawa (Mr.) Manager Design Group
#### 3. Date of Submission
June 7, 2001
#### 4. Device Trade name
Monochrome perfectly flat panel displays, ME Series
#### 5. Common Name
Monitor, display, workstation, and others
#### 6. Classification
Medical displays were classified in Class II per 21 CFR 890.2050.
#### 7. Predicate Device
BARCO MGD 221 2 MegaPixel Diagnostic Display, manufactured by BARCO NV / Display Systems (K000923). Comparison of the principal characteristics of the three devices which are pertinent to clinical performance is shown in Appendix 1.
#### 8. Description of Device
The ME Series Medical Displays are displays for medical use. This product complies with Radiation performance standards (21 CFR Chapter J, Sub-chapter J).
#### 9. Intended Use
The ME Series Medical Displays are intended for use with Picture Archiving Communication Systems (PACS) for medical imaging applications by physicians.
{2}------------------------------------------------
10. Explanation of ME Series
ME Series consists of the following models.
ME221f (Model No. MDT 2205A) ME222f (Model No. MDT 2210A) ME223f (Model No. MDT 2212A) ME224f (Model No. MDT 2210A) ME521f (Model No. MDT 2209A) ME311L (Model No. MDL 2102A)
Comparison of specifications is shown in Appendix 2.
11. Compliance standards
Medical safety: UL2601-1, CSA No.601-1, IEC60601-1, MDD/CE (EN60601-1)
EMC: MDD/CE (EN60601-1-2), FCC-A, VCCI-A, IEC60601-2
Others: DHHS
{3}------------------------------------------------
1. The state of the state of the states of the states
.
| Item | ME221f/4 | MGD221 | | ME221f | ME222f | ME223f*1 | ME224f*2 | ME521f | NOTE |
|--------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------|
| 510(k) Number | Not known | K000923 | Item | | | | | | |
| Display area | Horizontal 300mm, vertical 400mm | 300mm x 400mm | 10(k) Number | Not known | Not known | Not known | Not known | Not known | *1 ME223f is identical to ME221f except for CRT focus. |
| Luminescent material | P45 (standard) or P4 (optional) | B4-L8 | Display area | Horizontal 300mm, vertical 400mm | Horizontal 400mm, vertical 300mm | Horizontal 300mm, vertical 400mm | Horizontal 400mm, vertical 300mm | Horizontal 302mm, vertical 378mm | |
| Input signal | Video signal - analog 0.714Vp-p/75 $\Omega$<br>Synchronizing signal - 1.0-5.5Vp-p/3.9k $\Omega$<br>Synchro input - Separate/composite/<br>composite video.<br>Connector - BNC and SD-sub 15p (VGA<br>connection, Gch for video) | • Signal inputs<br>Video input connector type: BNC<br>Video input voltage: 0.7 Vpp<br>Termination: 75 $\Omega$<br>• Synchronization inputs<br>Inputs provided: CS or HS and VS<br>Input connector type: BNC<br>Termination: 75 $\Omega$<br>Sync input voltage: nominal 0.5Vpp | Luminescent material | P45 (standard) or P4 (optional) | P45 (standard) or P4 (optional) | P45 (standard) or P4 (optional) | P45 (standard) or P4 (optional) | P45 (standard) or P4 (optional) | |
| Maximum display pixels | 1200 dots x 1600 line | 1280 dots x 1600 line | Input signal | Video signal - analog 0.714Vp-p/75Ω<br>Synchronizing signal - 1.0-5.5Vp-p/3.9kΩ<br>Synchro input - Separate/composite/composite video.<br>Connector - BNC and SD-sub 15p (VGA connection, Gch for video) | Video signal - analog 0.714Vp-p/75Ω<br>Synchronizing signal - 1.0-5.5Vp-p/3.9kΩ<br>Synchro input - Separate/composite/composite video.<br>Connector - BNC and SD-sub 15p (VGA connection, Gch for video) | Video signal - analog 0.714Vp-p/75Ω<br>Synchronizing signal - 1.0-5.5Vp-p/3.9kΩ<br>Synchro input - Separate/composite/composite video.<br>Connector - BNC and SD-sub 15p (VGA connection, Gch for video) | Video signal - analog 0.714Vp-p/75Ω<br>Synchronizing signal - 1.0-5.5Vp-p/3.9kΩ<br>Synchro input - Separate/composite/composite video.<br>Connector - BNC and SD-sub 15p (VGA connection, Gch for video) | Video signal - analog 0.714Vp-p/50Ω or 75Ω (optional)<br>Synchronizing signal - TTL level, negative<br>Synchro input - Separate.<br>Connector- BNC | *2 ME224f is identical to ME223f except for luminance. |
| Scanning frequency | Horizontal 30k-127kHz, vertical 50-180Hz | Horizontal 80-140kHz, Vertical 48-150Hz | Maximum display pixels | 1200 dots x 1600 line | 1600 dots x 1200 line | 1200 dots x 1600 line | 1600 dots x 1200 line | 2048 dots x 2560 line | |
| Maximum image clock | 250MHz | 250MHz | Scanning frequency | Horizontal 30k-127kHz, vertical 50-180Hz | Horizontal 30k-110kHz, vertical 50-180Hz | Horizontal 30k-127kHz, vertical 50-180Hz | Horizontal 30k-110kHz, vertical 50-180Hz | Horizontal 186kHz, vertical 71Hz (fixed) | |
| Maximum luminance | 410 cd/m2 min. (P4 luminescent material)<br>for both partial white (20%) and total white<br>(100%) | Calibrated: 222cd/m2; 65fL<br>Peak: 360cd/m2; 105fL | Maximum image clock | 250MHz | 250MHz | 250MHz | 250MHz | 500MHz | |
| Luminance calibration | Software (standard item)<br>Photosensor (optional item) – DTP92<br>(X-Lite) | Frontal X-Rite DTP92 | Maximum luminance | 410 cd/m² min. (P4 luminescent material) for both partial white (20%) and total white (100%) | 600 cd/m² min. (P45 luminescent material) for both partial white (20%) and total white (100%) | 800 cd/m² min. (P4 luminescent material) for both partial white (20%) and total white (100%) | 600 cd/m² min. (P45 luminescent material) for both partial white (20%) and total white (100%) | 600 cd/m² min. (P45 luminescent material) for both partial white (20%) and total white (100%) | |
| Serial communication connector | D-sub 9p x 2 | 1 input, 1output<br>9600 baud, RS232<br>SUB-D9 male/female connector<br>Operational at save state | Luminance calibration | Software (standard item) | Software (standard item) | Software (standard item) | Software (standard item) | Software (standard item) | |
| Agency standards | Medical safety: UL2601-1, CSA No.601-1,<br>IEC 60601-1, MDD/CE (EN60601-1)<br>Others: DHHS | Medical safety: UL2601-1, cUL2601-1,<br>IEC 60601-1, CE (EN60601-1)<br>Others: DHHS | | Photosensor (optional item) - DTP92 (X-Lite) | Photosensor (optional item) - DTP92 (X-Lite) | Photosensor (optional item) - DTP92 (X-Lite) | Photosensor (optional item) - DTP92 (X-Lite) | Photosensor (optional item) - DTP92 (X-Lite) | |
| Dimensions and weight<br>(incl. tilt and swivel) | Net, 412 x 578 x 518 mm (W x H x D) 35kg<br>Packed, 560 x 710 x 730 mm (W x H x D)<br>41kg | Unpacked, 400 x 561 x 558 mm, 37.4kg<br>Packed, 630 x 780 x 788 mm, 55.9 kg | Serial communication connector | D-sub 9p x 2 | D-sub 9p x 2 | D-sub 9p x 2 | D-sub 9p x 2 | D-sub 9p x 2 | |
| Power supply | 100-240V AC, 50/60Hz | 100-240V AC, 50/60Hz | Agency standards | Medical safety: UL2601-1, CSA No.601-1, IEC 60601-1, MDD/CE (EN60601-1)<br>Others: DHHS | Medical safety: UL2601-1, CSA No.601-1, IEC 60601-1, MDD/CE (EN60601-1)<br>Others: DHHS | Medical safety: UL2601-1, CSA No.601-1, IEC 60601-1, MDD/CE (EN60601-1)<br>Others: DHHS | Medical safety: UL2601-1, CSA No.601-1, IEC 60601-1, MDD/CE (EN60601-1)<br>Others: DHHS | Medical safety: UL2601-1, CSA No.601-1, IEC 60601-1, MDD/CE (EN60601-1)<br>Others: DHHS | |
| Dimensions and weight (incl. tilt and swivel) | Net, 412 x 578 x 518 mm (W x H x D) 35kg<br>Packed, 560 x 710 x 730 mm (W x H x D) 41kg | Net, 505 x 500 x 515 mm (W x H x D) 35kg<br>Packed, 615 x 655 x 650 mm (W x H x D) 40kg | Net, 412 x 578 x 518 mm (W x H x D) 35kg<br>Packed, 560 x 710 x 730 mm (W x H x D) 41kg | Net, 505 x 500 x 515 mm (W x H x D) 35kg<br>Packed, 615 x 655 x 650 mm (W x H x D) 40kg | Net, 412 x 578 x 552 mm (W x H x D) 35kg<br>Packed, 560 x 710 x 730 mm (W x H x D) 41kg | | | | |
| Power supply | 100-240V AC, 50/60Hz | 100-240V AC, 50/60Hz | 100-240V AC, 50/60Hz | 100-240V AC, 50/60Hz | 100-240V AC, 50/60Hz | | | | |
:
100 100
{4}------------------------------------------------
Appendix 2 Comparison table among application model
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is a stylized image of a caduceus, a symbol often associated with healthcare, featuring a staff with a serpent entwined around it.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL 3 1 2001
Totoku Electric Co., Ltd. % Mr. Shinichi Yamanaka Safety Dept. Cosmos Corporation 319 Akeno, Obata-cho, Watarai-gun Mie-ken, 519-05 JAPAN
Re: K012099 ME Series Medical Display Dated: June 7, 2001 Received: July 5, 2001 Regulatory Class: II 21 CFR 892.2050/Procode: 90 LLZ
Dear Mr. Yamanaka:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the w inave is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices device is substantally cquiraline (tor the materials of the Medical Device Amendments, or to devices matkled in meetstance commerce provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Premarket Approval), it may be subject 11 your device is classince (Sc above) and etaxe to entry vour device can be found in the Code of Federal Regulations, to such and conceller. A substantially equivalent determination assumes compliance with the Current Good Manufacturing I I uc 21, 1 at 3 ove to 099. 11 our of the System Regulation (QS) for Medical Devices: General regulation (21 CFR Part Placitie requirements, as set tour in the Quality Dysterial regultures Administration (FDA) will verify such assumptions. Failure o comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements to comply with the Federal Register. Please note: this response to your premarket notification submission does not concenting your device in the Poccal Regions 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding I his when while he of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro 11 you deste spothe arrice to your corress of Compliance at (301) 594-4639. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your common "Misolanding of Ferences to premation in the Division of Small Manufacturers Assistance at its toll-free number (800) 1638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely vours
Nancy C. bogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure (s)
{6}------------------------------------------------
KO12099 510(k) Number (If known): Not known
Device Name: Monochrome perfectly flat panel display, ME Series
Indications for Use:
The ME Series Medical Displays are intended for use with Picture Archiving Communication Systems (PACS) for medical imaging applications by physicians.
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
# Concurrence of CDRH, Office of Device Evaluation
Prescription Use
OR Over-The-Counter Use
(Optional Format 1-2-96)
Nancy Broghon
(Division Sign-Off)
Division of Reproductive. Abdominal.
and Radiologi 510(k) Numbe
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.