← Product Code [HWC](/productcode/HWC) · K011910

# IN-FAST BONE SCREW SYSTEM (K011910)

_American Medical Systems, Inc. · HWC · Jul 13, 2001 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K011910

## Device Facts

- **Applicant:** American Medical Systems, Inc.
- **Product Code:** [HWC](/productcode/HWC.md)
- **Decision Date:** Jul 13, 2001
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3040
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The In-Fast Bone Screw System is intended for soft tissue fixation to the publo bone by means of bone screws with attached suture. The In-Fast System is indicated for cystourethropexy and vaginal sling procedures for the cystouremropexy and vaginary incontinence resulting from urethral hypermobility and/or intrinsic sphincter deficiency.

## Device Story

In-Fast Bone Screw System facilitates soft tissue fixation to pubic bone; consists of transvaginal inserter and bone screws with attached braided suture. Used by surgeons during cystourethropexy and vaginal sling procedures to treat female stress urinary incontinence. Inserter delivers bone screws into pubic bone; attached suture provides fixation for soft tissue. Device modification adds braided suture to existing system; mechanical properties verified via bench testing.

## Clinical Evidence

Bench testing only. Mechanical properties of the braided suture were tested for compatibility with the In-Fast system; suture complies with USP Monograph for Non-Absorbable Sutures.

## Technological Characteristics

System includes transvaginal inserter and bone screws with attached braided suture. Suture complies with USP Monograph for Non-Absorbable Sutures. Mechanical fixation principle.

## Regulatory Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- In-Fast Bone Screw System ([K970292](/device/K970292.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# JUL 1 3 2001

## 510(k) SUMMARY

Submitter's Name: American Medical Systems, Inc. Address: 10700 Bren Road West Minnetonka, MN 55343 Tel: 952-933-4666 Fax: 952-930-6157 Contact Person: Elsa A. Linke June 18, 2001 Date of Summary Preparation: Device Common Name: Bone Screw Inserter & Bone Screws Device Trade Name: In-Fast Bone Screw System Device Classification Name: Fastener, Fixation, Nondegradable, Soft Tissue Classification: Class II Product Code: MBI

Predicate Device:

In-Fast Bone Screw System K970292

### Device Description

The In-Fast Bone Screw System is intended for soft tissue fixation to the public bone by means of bone screws threaded with suture. It consists of a transvaginal inserter and bone screws with attached suture.

### Indications for Use

The In-Fast Bone Screw System is intended for soft tissue fixation to the public bone by means of bone screws with attached suture. The In-Fast System is indicated for cystourethropexy and vaginal sling procedures for the treatment of stress type (female) urinary incontinence resulting from urethral hypermobility and/or intrinsic sphincter deficiency.

### Comparison to Predicate Device

The fundamental scientific technology of the device does not change with this modification. The only material change is the addition of braided suture to the device system.

### [510(k) Summary continued]

{1}------------------------------------------------

### Supporting Information

The mechanical properties of the braided suture have been tested on the bench for compatibility with the In-Fast system. The suture complies with the USP Monograph for Non-Absorbable Sutures.

### Conclusion

The proposed modification is equivalent to the predicate with respect to intended use, technological characteristics, and performance characteristics.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes representing the department's mission. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUL 1 3 2001

Ms. Elsa A. Linke Regulatory Affairs Specialist American Medical Systems, Inc. 10700 Bren Road West Minnetonka, Minnesota 55343

Re: K011910

Device Name: In-Fast Bone Screw System Regulation Number: 21 CFR 888.3040 Regulatory Class: II Product Codes: HWC, MBI, GAS Dated: June 18, 2001 Received: June 19, 2001

Dear Ms. Linke:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic OS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{3}------------------------------------------------

### Page 2 - Ms. Elsa A. Linke

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed nredicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Mark N Milken

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# INDICATIONS FOR USE ENCLOSURE

510(k) Number:

Koi 1916

Device Name:

In-Fast Bone Screw System

Indications for Use:

The In-Fast Bone Screw System is intended for soft tissue fixation to the publo bone by means of bone screws with attached suture. The In-Fast System is indicated for cystourethropexy and vaginal sling procedures for the cystouremropexy and vaginary incontinence resulting from urethral hypermobility and/or intrinsic sphincter deficiency.

# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-off) Division of General and Restorative Devices

510(k) Number __

Prescription Use X (Per 21 CFR801:109)

OR

Over the Counter Use_

4

Mark N. Mulburn

Division Sign-Off) Division of General, Restorative and Neurological Devices

510(k) Number K011916

---

**Source:** [https://fda.innolitics.com/device/K011910](https://fda.innolitics.com/device/K011910)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
