The FM-1 NO Blender provides user set concentrations of inhaled Nitric Oxide (NO), in a balance of nitrogen, mixed into a user settable constant flow of oxygen gas that is being delivered to a patient. The intended use for the FM-1 NO Blender is as a back up to a primary nitric oxide delivery system or for use with a self-inflating manual resuscitator bag. This intended use includes applications within a medical facility. The FM-1 NO blender is not intended for use as a primary NO delivery system for long term use.
Device Story
FM-1 NO Blender acts as a backup nitric oxide (NO) delivery system; mixes NO (in nitrogen balance) with oxygen gas at user-set concentrations/flows. Used in medical facilities; operated by clinicians. Device serves as a secondary source when primary systems fail or during manual resuscitation. Input: NO gas source and oxygen gas source. Output: Controlled NO/O2 gas mixture. Benefit: Ensures continuity of NO therapy during primary system downtime or manual ventilation. Device is a mechanical blender; no software or AI components.
Clinical Evidence
Bench testing only. Validation included FTIR testing of PEEK material, performance characteristics testing (5-15 lpm), validation with self-inflating manual resuscitator bags, concentration profile validation, regulator check function, failure testing, drop testing, and packaging/shipping validation. No clinical data presented.
Technological Characteristics
Mechanical gas blender. Capillary tubes constructed of PEEK. Designed to comply with CGA 626 (NO gas connections) and IEC 601-1 (general electrical safety). No software or electronic control algorithms.
Indications for Use
Indicated for patients requiring inhaled Nitric Oxide (NO) therapy when the primary delivery system is unavailable or for use with a self-inflating manual resuscitator bag in a medical facility setting. Not for long-term use as a primary delivery system.
Regulatory Classification
Identification
The nitric oxide administration apparatus is a device used to add nitric oxide to gases that are to be breathed by a patient. The nitric oxide administration apparatus is to be used in conjunction with a ventilator or other breathing gas administration system.
Special Controls
The special control for this device is FDA's “Guidance Document for Premarket Notification Submissions for Nitric Oxide Administration Apparatus, Nitric Oxide Analyzer, and Nitrogen Dioxide Analyzer.”
The special control developed by the agency is a guidance document, entitled Guidance Document for Premarket Notification Submissions for Nitric Oxide Delivery Apparatus, Nitric Oxide Analyzer and Nitrogen Dioxide Analyzer. This guidance document identifies the risks associated with these types of devices and contains information that will help manufacturers address those risks. This document is available on FDA's website at http://www.fda.gov/cdrh/ode/1157.pdf.
{0}------------------------------------------------
**Fadasis**
**Medical,**
*Inc.*
JUL 2 0 2001
7601-B Murphy Drive, Middleton, W1 53562 Phone (608) 831-0025, ext. 276, FAX (608) 831-2202
| Subject: | 510(k) Summary of Safety and Effectiveness Information for the<br>Fadasis Medical FM-1 NO Blender |
|-------------------|---------------------------------------------------------------------------------------------------|
| Proprietary Name: | Fadasis Medical FM-1 NO Blender |
| Common Name: | Nitric Oxide Administration Apparatus - Back-up System |
| Classification: | Class II, 21CFR868.5165, MRO |
| Panel: | Anesthesiology |
| Contact Person: | Raymond Riddle, Vice President, Regulatory Affairs |
R. Riddle 6/12/01
The 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 1992.
The Fadasis Medical FM-1 NO Blender is substantially equivalent to the Fadasis FM-1 NO Blender, The 1 addis Incontaint I 1 1 rith 510(k) premarket notification number K003665. The purpose of this Which was clouded by I DA whan 2000 (EEK, rather than stainless steel, in the capillary tubes.
Indications for use: The FM-1 NO Blender provides user set concentrations of inhaled Nitric Oxide (NO), in a balance of nitrogen, mixed into a user settable constant flow of oxygen gas that is being delivered to a patient. The intended use for the FM-1 NO Blender is as a back up to a primary nitric oxide delivery pation. The intended use when a primary delivery device cannot be used. In this capacity, it system or tor site inflating manual resuscitator bag. This intended use includes applications within a can or used with a set manages of a medical facility. The FM-1 NO blender is not intended for use modions a primary NO delivery system for long term use. (Note: These indications for use are unchanged form K003665.
The Fadasis Medical FM-1 NO Blender was designed to comply with the limited applicable portions of the following:
- 1. CGA 626: Medical NO Gas Connections.
- 2. IEC 601-1: Medical Electrical Equipment (for general requirements).
- FDA Guidance Document for Premarket Notification Submissions for Nitric Oxide Delivery 3. Apparatus, Nitric Oxide Analyzer and Nitrogen Dioxide Analyzer.
The materials selected were primarily based upon the Datex-Ohmeda INOvent Delivery System. New validation and verification completed included the FTIR Testing of PEEK. Validation and verification for 510(k) K003665 included FM-1 NO Blender 5-15 lpm Performance Characteristics Testing, FM-1 NO Blender Validation with Self Inflating Manual Resuscitator Bags, FM-1 NO Blender Validation of the Concentration Profile When Used with Self Inflating Manual, Resuscitator Bags, FM-1 NO Blender Regulator Check Function, FM-1 NO Blender 5-15 Ipm Failure Testing, FM-1 NO Blender Material Compatibility Information, FM-1 NO Blender Drop Test and FM-1 NO Blender Packaging and Shipping Validation. All testing indicated the FM-1 NO Blender met its design input specifications, design output specifications, hazard analysis and risk control requirements.
{1}------------------------------------------------
Device Name - Proprietary:
Device Name - Common: Device Name - Classification:
Nitric Oxide Mixer Nitric Oxide Administration Apparatus Back-up System
Fadasis Medical FM-1 NO Blender
Fadasis Medcial FM-1 NO Blender
To the best of Fadasis Medical's knowledge,
performance standards have not been promulgated
FDA 510(k) log number K003665
Device Panel:
Anesthesiology
Device Classification:
Class II 21CFR868.5165 Product Code: MRO
by FDA for this device.
Predicate Devices:
Performance Standards:
Owner/Operator Number:
9042308
Establishment Registration Number: 2135137
Facility Information:
Device Background:
Fadasis Medical, Inc. 7601-B Murphy Drive Middleton, WI 53562 Telephone: (608) 831-0025 (608) 831-2202 Fax:
The new FM-1 NO blender is identical to the predicate FM-1 NO blender with the exception of the capillary tube material, which has been changed to PEEK from stainless steel. The use of this material has been validated via FTIR testing, as required by FDA Guidance Document for Premarket Notification Submissions for Nitric Oxide Delivery Apparatus, Nitric Oxide Analyzer and Nitrogen Dioxide Analyzer. The original device performance validation data was obtained using both stainless steel and PEEK capillary tubes. Because of inadequate testing of PEEK, the use of this material was removed from the original 510(k) until further testing was completed, thus prompting this Special 510(k).
{2}------------------------------------------------
Image /page/2/Picture/2 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" printed around the edge. Inside the circle is a stylized symbol that resembles an abstract human figure or a bird in flight, composed of three curved lines.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL 2 0 2001
Mr. Raymond T. Riddle Fadasis Medical, Inc. 7601-B Murphy Drive Middleton, WI 53562
Re: K011874
FM-1 NO Blender Regulation Number: 868.5165 Regulatory Class: II (two) Product Code: 73 MRO Dated: July 3, 2001 Received: July 9, 2001
Dear Mr. Riddle:
We have reviewed your Section 510(k) notification of intent to market the device referenced we nave reviewed your becally in the substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstant commerce stated in the enclosure) to logally markets of the Medical Device Amendments, or to devices that prior to May 20, 1970, the ensounces with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls Cosment Act (Act). Tourmaly, mercore, manovisions of the Act include requirements for annual provisions of the Act. "The goneral consults curing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III If your device is classified (see asono) anto additional controls. Existing major regulations (Fremarket Approval), it they of Sale of Federal Regulations, Title 21, Parts 800 to 895. affecting your do rece can or reamination assumes compliance with the Current Good A substantany equirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to inspections, the rood and Drag rummissault in regulatory action. In addition, FDA may publish comply with the GMT 10garation inde Federal Register. Please note: this response to your premarks octification submission does not affect any obligation you might
{3}------------------------------------------------
Page 2 - Mr. Raymond T. Riddle
have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket I his letter will anow you to begin manoung your device to a legally marketed notification. THE IDA imania of succeantial organ device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFF Part 801 and If you desire specific advice to Your do rice of ca. near contact the Office of Compliance at additionally 809.10 for in vito diagnestions on the promotion and advertising of your device, (301) 594-4040. Adultionally, for quest.chs on 1501) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Illionnation on your responsionals and its toll-free number (800) 638-2041 or (301) 443-6597 or at its Maindracturers / 1551bitters www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
N. Darl. Teth
James E. Dillard III Director Division of Cardiovascular and Respiratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
| 1000<br>510(k) N<br>known)<br>) Number (11 | V All 274<br>1<br>S<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>A<br>and and the state of the same of the comments of the comments of the comments of the comments of the comments of the comments of the comments of the comments of the comments |
|---------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|---------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
Fadasis Medical, Inc. FM-1 NO Blender Device Name:
Indications For Use:
The FM-1 NO Blender provides user set concentrations of inhaled Nitric Oxide (NO), in a balance of nitrogen, mixed into a user settable constant flow of oxygen gas that is being delivered to a batient. The intended use for the FM-I NO Blender is as a back up to a primary nitrics of this associt patient. The intended use for the FMF I (O Dichice is a se a vice cannot be used. In this capacity, system of for short term attenced use when a practitator bag. This intended use includes applications it can be used with a sch-inthaing manual result and the lice if a medical facility. The FM-1 NO blender is not within a modious as a primary NO delivery system for long term use.
Note: These indications for use are unchanged from K003665
## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
18 +
Oake Tell
(Division Sign-Off) Division of Cardiovascular, Respiratory, and Neurological Devices
510(k) Number: -
Prescription Use -(Per 21CFR801.109) OR
Over-The-Counter Use-(Optional Format 1-2-96)
6
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.