K011860 · Keystone Medical · FOZ · Jun 27, 2001 · General Hospital
Device Facts
Record ID
K011860
Device Name
KEYST0NE MEDICAL PICC CATHETER
Applicant
Keystone Medical
Product Code
FOZ · General Hospital
Decision Date
Jun 27, 2001
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5200
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The devices are intended for use when central venous catheterization or intravenous therapy for the administration of fluids, medications and/or nutritional therapy is prescribed. The use of the device for these indications is less than 30 days.
Device Story
Single or double lumen intravenous catheters; intended for central venous access; used for administration of fluids, medications, or nutritional therapy. Device is provided in a kit containing legally marketed or exempt components. Used in clinical settings by healthcare professionals. Provides venous access for short-term therapy (less than 30 days).
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing previously submitted for the predicate device (K953811).
Technological Characteristics
Single or double lumen intravenous catheters. Form factor: catheter kit. Sterilization: not specified. Connectivity: none. Software: none.
Indications for Use
Indicated for patients requiring central venous catheterization or intravenous therapy for administration of fluids, medications, or nutritional therapy for durations less than 30 days.
Regulatory Classification
Identification
An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.
K971085 — SOFT TIP MULTI-LUMEN CENTRAL VENOUS CATHETER · B.Braun Medical, Inc. · Jun 2, 1997
K013182 — TRIPLE LUMEN CENTRAL VENOUS CATHETER · Cook, Inc. · Jul 1, 2002
K090102 — NIAGARA, NIAGARA SLIM-CATH, AND BREVIA SHORT-TERM CURVED EXTENSION HEMODIALYSIS CATHETERS · C.R. Bard, Inc. · Feb 12, 2009
K042504 — CATHY STERILE DISPOSABLE I.V. CANNULA WITH CATHETER · Myco Medical Supplies, Inc. · Jan 19, 2006
K964048 — FIRST MIDCATH CATHETER KIT · Becton Dickinson Vascular Access, Inc. · Jan 6, 1997
Submission Summary (Full Text)
{0}------------------------------------------------
Keystone Medical PICC Catheters 510(k) Summary
# JUN 2 7 2001
Attachment 1
### 510(K) SUMMARY
This summary of 510(k) -safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
The assigned 5|0(k) number is:
#### Submitter's Identification: 1.
Tony Madison, President Keystone Medical 1338 Michael Way Lansdale, PA. 19446
Date Summary Prepared: May 29, 2001
#### 2. Name of the Device:
Keystone Medical Peripherally Inserted Central Catheter (PICC)
#### 3. Predicate Device Information:
These devices are substantially equivalent to devices currently marketed by Medical Components, Inc. Lansdale, PA under K#953811. Medical Components will continue to manufacture these devices.
#### 4. Device Description:
Single or double lumen intravenous catheters. Devices will be assembled in a kit that includes components that are either legally marketed; exempt from premarket requirements; or "grandfathered".
#### 5. Intended Use:
The devices are intended for use when central venous catheterization or intravenous therapy for the administration of fluids, medications and/or nutritional therapy is prescribed. The use of the device for these indications is less than 30 days.
#### 6. Comparison to Predicate Devices:
The Keystone Medical Lumen Catheter is identical to the predicate.
The Keystone device will differ only in that the Keystone Medical name and logo will appear on the unit labels, box labels, and instructions for use. Keystone will also develop it's own sell sheets for these devices in the future.
{1}------------------------------------------------
ll 8660
### Attachment 1
### Discussion of Non-Clinical Tests Performed for Determination of 7. Substantial Equivalence are as follows:
Non clinical (bench) testing was completed to support K#953811, and was included with this 510(k) submission
#### Discussion of Clinical Tests Performed: 8.
Clinical testing was not completed.
#### 9. Conclusion
The Keystone Medical PICC catheters are safe and effective for their intended use.
{2}------------------------------------------------
Image /page/2/Picture/2 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle or bird-like symbol with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol. The text is in all capital letters and is evenly spaced around the circle.
JUN 2 7 2001
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Keystone Medical C/O Mr. Alan P. Schwartz Executive Vice President MDI Consultants, Incorporated 55 Northern Boulevard, Suite 200 Great Neck, New York 11021
Re : K011860 Trade/Device Name: Keystone Medical PICC Catheter Regulation Number: 880.5200 Requlatory Class: II Product Code: FOZ Dated: June 12, 2001 Received: June 14, 2001
Dear Mr. Schwartz:
We have reviewed your Section 510(k) notification of intent to we have reviewed referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code_of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements
{3}------------------------------------------------
Page 2 - Mr. Schwartz
concerning your device in the Federal Register. Please note: this response to your premarket notification submission does entb response or obligation you might have under sections 531 not arress and of the Act for devices under the Electronic chrough 312 or on control provisions, or other Federal laws or requlations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA described in your sia equivalence of your device to a legally marketed predicate device results in a classification for your marketed predicate with your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in regaraciasmostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact che office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). J Other general information on your responsibilities under the Act may be Information on your color of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address
"http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Timothy A. Ulatowski
Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Keystone Medical PICC Catheter Indications for Use
Attachment 3
of イ Page
510(k) Number: _______________________________________________________________________________________________________________________________________________________________
Device Name: Keystone Medical PICC Catheters
### Indications for Use:
The devices are intended for use when central venous catheterization or intravenous therapy for the administration of fluids, medications and/or nutritional therapy is prescribed. The use of the device for these indications is less than 30 days.
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
OR
Over-The-Counter Use
Petuna Cacciste
(Division Sign-Off) Division of Dental, Infection Control, and General Hospital Devices 4011840 1510(k) Number -
(Optional Format 1-2-96)
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.