← Product Code [MWJ](/productcode/MWJ) · K011837

# LIFECARD CF 7 DAY (K011837)

_Reynolds Medical , Ltd. · MWJ · Jul 13, 2001 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K011837

## Device Facts

- **Applicant:** Reynolds Medical , Ltd.
- **Product Code:** [MWJ](/productcode/MWJ.md)
- **Decision Date:** Jul 13, 2001
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.2800
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The Reynolds Lifecard CF 7-Day Holter recorder is a modification of the Reynolds Lifecard CF Holter Recorder, K001025. It is indicated when it is desired to record the patient's ambulatory electrocardiogram. It is a portable Holter recorder designed to record the patient's ambulatory electrocardiogram for up to seven days.

## Device Story

Portable Holter recorder; records ambulatory ECG signals for up to seven days using flashcard storage. Device performs no analysis; data is transferred to external Holter analyzers for clinical review. Used in clinical settings; operated by healthcare professionals. Employs data compression system for extended recording duration. Provides continuous ECG data to physicians to assist in cardiac rhythm diagnosis and patient management.

## Clinical Evidence

No clinical data required. Substantial equivalence supported by non-clinical bench testing, including compliance with AAMI EC38, IEC 601-1, and IEC 601-1-2 standards, and software/compression algorithm validation.

## Technological Characteristics

Portable Holter recorder; flashcard-based storage; data compression system for extended recording. Meets AAMI EC38, IEC 601-1, and IEC 601-1-2 standards. Standalone device for data acquisition; no onboard analysis.

## Regulatory Identification

A medical magnetic tape recorder is a device used to record and play back signals from, for example, physiological amplifiers, signal conditioners, or computers.

## Predicate Devices

- Reynolds Lifecard CF Holter recorder ([K001025](/device/K001025.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Summary

Page A 2

JUL 1 3 2001

## Reynolds Medical Ltd. Special 510(k) Lifecard CF 7-Day Holter Recorder

## 510(k) Summary

(1) Submitter Information

Name: Reynolds Medical Ltd.

Address:

1 Harforde Court John Tate Court Hertford, Herts SG13 7NW ENGLAND

Telephone Number: 44-1992-507700

Contact Person: Dr. George Myers (Official Correspondent) Medsys Inc. 377 Route 17 S Hasbrouck Heights, NJ 07604 Telephone 201-727-1703 Fax 201-727-1708

Date Prepared: June 8th, 2001

(2) Name of Device

Trade Name: Lifecard CF 7-Day Holter recorder Common Name: Holter Recorder. Common Name: Recorder, Magnetic Tape, Medical (please note that this device does not use magnetic tape)

(3) Equivalent legally-marketed device.

Reynolds Lifecard CF Holter recorder, K001025

(4) Description

The device is a portable Holter recorder designed to record the patient's ambulatory The device is a portable frecer days (with the appropriate flashcard and battery). The 

{1}------------------------------------------------

use of the modified device as described in its labeling has not changed. The fundamental science technology of the device has not changed.

The unit uses a data compression system to store seven days of ECG data in extended mode. In "standard" mode (up to two days of recording) the unit is identical to the predicate device.

(5) Intended Use

The Lifecard CF is a portable Holter recorder intended to record the patient's ambulatory electrocardiogram for up to seven days (with the appropriate flashcard and announatory clectrocardiograms can then be analyzed by Holter analyzers. The unit does no analysis.

(6) Performance Data

(a) Non-clinical tests

The Lifecard CF 7-Day Holter recorder meets the requirements of AAMI EC38, IEC 601-1 and IEC 601-1-2.

The software has undergone extensive validation testing. The compression algorithm has been separately tested.

(b) Clinical tests

None required.

(c) Conclusions

The Lifecard CF 7-Day Holter recorder is equivalent in safety and efficacy to the legally-marketed predicate device.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three wing-like shapes, symbolizing health and human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUL 1 3 2001

Reynolds Medical, Ltd. George H. Myers, Sc.D. Designated Agent c/o Medsys Inc. 377 Route 17 South Hasbrouck Heights, NJ 07604

Re: K011837

Trade Name: LifeCard CF 7-day Holter Recorder Regulation Number: 870.2800 Regulatory Class: II (Two) Product Code: MWJ Dated: June 7, 2001 Received: June 12, 2001

Dear Dr. Myers:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic OS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have 

{3}------------------------------------------------

Page 2 - George H. Myers, Sc.D.

under sections 531 through 542 of the Act for devices under the Electronic Product Radiation i Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed nontication. The I Drivening sification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and 11 you dealer openits arrivitro diagnostic devices), please contact the Office of Compliance at additionally 607.10 for in This allestions on the promotion and advertising of your device, (2017) 591-1010. Tradition for at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small mormation on your copy at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

James E. Dillard III Director Division of Cardiovascular and Respiratory Devices Office of Device Evaluation Center for Devices and Radiological Health

{4}------------------------------------------------

Page 2

Page 1 1837 510(k) Number (if known):

Indications for Use Form

Device Name: Lifecard CF 7-Day Holter recorder

Indications for Use:

The Reynolds Lifecard CF 7-Day Holter recorder is a modification of the Reynolds Lifecard CF Holter Recorder, K001025. It is indicated when it is desired to record the patient's ambulatory electrocardiogram.

It is a portable Holter recorder designed to record the patient's ambulatory electrocardiogram for up to seven days.

## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use
Use
(Per 21 CFR 810.109)

OR

Over-the-Counter
(Optional Format 1-2-

of Cardiovascula 510(k) Number

(Optional Format 1-2-96)

---

**Source:** [https://fda.innolitics.com/device/K011837](https://fda.innolitics.com/device/K011837)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
