← Product Code [OLW](/productcode/OLW) · K011834

# BIS ENGINE (K011834)

_Aspect Medical Systems, Inc. · OLW · Jul 10, 2001 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K011834

## Device Facts

- **Applicant:** Aspect Medical Systems, Inc.
- **Product Code:** [OLW](/productcode/OLW.md)
- **Decision Date:** Jul 10, 2001
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 882.1400
- **Device Class:** Class 2
- **Review Panel:** Neurology

## Indications for Use

Intended to monitor the state of the brain by data acquisition of EEG signals in the intensive care unit, operating room and for clinical research. The BIS, a processed EBG variable, may be used as an aid in monitoring the effects of certain anesthetic agents.

## Device Story

BIS Engine is a printed circuit board (PCB) component for OEM EEG monitors; acquires EEG signals; processes signals using proprietary BIS technology; outputs processed EEG variable (BIS) to assist clinicians in monitoring anesthetic agent effects; used in ICU, OR, and clinical research; operated by healthcare professionals; provides real-time brain state monitoring to guide clinical decision-making regarding anesthesia.

## Clinical Evidence

No clinical data. Substantial equivalence supported by risk analysis and software validation testing.

## Technological Characteristics

Printed circuit board (PCB) component; 4-channel EEG signal processing; BIS algorithm; same hardware/electrical/mechanical design as predicate; software-based modification for increased channel capacity.

## Regulatory Identification

An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.

## Predicate Devices

- BIS Engine ([K002837](/device/K002837.md))
- A-1000 EEG Monitor with 4 channel amplifier (DSC-4) (K923043A)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## 510(k) SUMMARY

## Date Prepared: June 11, 2001

4011834

#### Company Name and Address

Aspect Medical Systems, Inc. 141 Needham St. Newton, MA 02464

| Contact Person: | Christine M. Vozella                           |
|-----------------|------------------------------------------------|
|                 | Director, Regulatory Affairs/Quality Assurance |
| Device Name     |                                                |

| Proprietary Name: | BIS Engine (PCB component in an EEG Monitor) |
|-------------------|----------------------------------------------|
| Common Name:      | EEG Monitor                                  |

#### Classification

Electroencephalograph (EEG) monitors and their software have been classified by the Neurological Devices Panel as Class II devices (21 CFR 882.1400)

#### Predicate Device

| BIS Engine                                                | K002837, FDA cleared September 19, 2000 |
|-----------------------------------------------------------|-----------------------------------------|
| A-1000 EEG Monitor<br>with 4 channel amplifier<br>(DSC-4) | K923043A, FDA cleared December 3, 1992  |

#### Device Description

The Aspect Medical Systems, Inc. BIS Engine - 4 channel support provides the means for incorporating Aspect's proprietary BIS technology into OEM (original equipment manufacturer's, i.e. our business partner's) finished devices. It is a small printed circuit board (PCB) that can either reside inside the OEM finished device or is re-designed for smaller size and packaged in a housing that will connect to the OEM finished device.

The fundamental scientific technology has not changed. The BIS technology remains the same. The BIS Engine - 4 channel support (subject of this 510(k)) has the same basic function, and same operating principal as the Predicate Device.

Only the software is changing. More specifically, the only difference is that the BIS Engine (subject of this 510(k)) can process up to 4 channels of EEG, compared to the Predicate Device, which can process up to 2 channels of EEG. The BIS processed parameter will only be calculated when in 2 channel maximum mode.

{1}------------------------------------------------

#### Intended Use

Intended to monitor the state of the brain by data acquisition of BEG signals in the intensive care unit, operating room and for clinical research.

The BIS, a processed EBG variable, may be used as an aid in monitoring the effects of certain anesthetic agents.

### Summary of Technological Characteristics Compared to Predicate Device

The BIS Engine - up to 4 channel support is substantially equivalent to the Predicate device (BIS Engine - up to 2 channel support).

#### Similarities

The fundamental technology has not changed. The indications for use are the same. The BIS technology remains the same. The BIS algorithm is the same. The BIS Engine has the same parameters, same operating principle, and same signal processing design. The hardware design is the same. The electrical and mechanical designs are the same.

#### Minor Differences

Only the software is changing. More specifically, the only difference is that the BIS Engine (subject of this 510(k)) can process a maximum of 4 channels of EEG, compared to the Predicate Device, which can process a maximum of 2 channels of EEG. The BIS processed parameter will only be calculated when in 2 channel maximum mode.

The following analysis and validation was performed:

- 1) Risk analysis
- Software validation 2)

Results of risk analysis: There are no additional hazards introduced by the BIS Engine - 4 channel that are severe enough to warrant tracking on the risk management record.

Results of validation: The applicable testing was completed (the modified BIS Engine will be a component to a finished device owned by our business partners, and as such, there are no patient or user safety concerns due to the BIS Engine in and of itself).

Results show all tests are acceptable.

The BIS Engine - 4 channel support is substantially equivalent to the Predicate Device.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with its wings spread, symbolizing protection and care. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle, indicating the department's name and national affiliation. The logo is presented in black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -W066-G609 Silver Spring, MD 20993-0002

Ms. Christine M. Vozella Director, Regulatory Affairs/Quality Assurance Aspect Medical Systems, Inc. 141 Needham Street Newton, Massachusetts 02464

Re: K011834 Trade/Device Name: BIS Engine Regulation Number: 21 CFR 882.1400 Regulation Name: Electroencephalograph Regulatory Class: II Product Code: OLW, OMC, ORT Dated (Date on orig SE ltr): June 11, 2001 Received (Date on orig SE ltr): June 12, 2001 APR - 9 2012

Dear Ms. Vozella:

This letter corrects our substantially equivalent letter of July 10, 2001.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Ms. Christine M. Vozella

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Kesia Alexander

Image /page/3/Picture/7 description: The image shows a signature in black ink on a white background. The signature appears to be stylized and cursive, with a prominent loop extending downwards and a smaller loop or curve at the top right. The overall impression is that of a quick, fluid, and practiced signature.

Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# Indications for Use Statement

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

| 510(k) Number (if known) | K011834                                                                                                                                                                                                                                                                 |
|--------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name              | Aspect Medical Systems, Inc. BIS Engine (4 channel support)                                                                                                                                                                                                             |
| Indications for Use      | Intended to monitor the state of the brain by data acquisition of EEG signals in the intensive care unit, operating room and for clinical research.<br>The BIS, a processed EEG variable, may be used as an aid in monitoring the effects of certain anesthetic agents. |

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED

Concurrence of CDRH, Office of Device Evaluation (ODE)

![Signature](signature.png)
(Division Sign-Off)

Division of General, Restorative and Neurological Devices

| 510(k) Number                         | K011834                                |
|---------------------------------------|----------------------------------------|
| Prescription Use (Per 21 CFR 801.109) | <img alt="Checkmark" src="check.png"/> |
| OR Over-The-Counter Use               | _________________                      |

ﺳ

---

**Source:** [https://fda.innolitics.com/device/K011834](https://fda.innolitics.com/device/K011834)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
