K011812 · Cook, Inc. · NHC · Nov 8, 2001 · Neurology
Device Facts
Record ID
K011812
Device Name
SPECTRUM VENTRICULAR CATHETER
Applicant
Cook, Inc.
Product Code
NHC · Neurology
Decision Date
Nov 8, 2001
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4100
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The SPECTRUM® Ventricular Catheter is intended for obtaining access to a ventricular cavity of the brain for short-term use to externally drain fluid for the purpose of relieving elevated intracranial pressure or fluid volume. The device will be supplied sterile and is intended for one-time use.
Device Story
The SPECTRUM® Ventricular Catheter is a 9 Fr, 33 cm long, closed-end catheter with 16 sideports and depth markings at 1 cm increments. It is impregnated with minocycline and rifampin (approx. 4-5 mg) to reduce the risk of catheter-related infection. The device includes a pre-loaded stainless steel tunneling trocar and proximal fittings to facilitate placement. It is used by clinicians in a hospital setting to access the brain's ventricular cavity for external drainage. The antimicrobial coating provides a prophylactic benefit against infection during the short-term duration of use. The device is supplied sterile for one-time use.
Clinical Evidence
Prospective, randomized, multicenter clinical trial comparing the SPECTRUM® Ventricular Catheter to a standard non-coated 9 Fr Cook Ventricular Catheter. 149 patients received the SPECTRUM catheter and 139 patients received the control. Average duration of catheter placement was 8.5 ± 5.8 days (control) and 8.2 ± 6.9 days (treatment). Results demonstrated the efficacy of the antimicrobial coating in reducing the incidence of catheter-related infection.
Technological Characteristics
9 Fr, 33 cm length, closed-end catheter with 16 sideports. Material impregnated with minocycline and rifampin. Includes stainless steel tunneling trocar and proximal fittings. Biocompatibility testing performed per standard requirements (cytotoxicity, sensitization, implantation, etc.).
Indications for Use
Indicated for patients requiring short-term external ventricular drainage to relieve elevated intracranial pressure or fluid volume. For prescription use only.
Regulatory Classification
Identification
A ventricular catheter is a device used to gain access to the cavities of the brain for injection of material into, or removal of material from, the brain.
Predicate Devices
Cook Incorporated's Ventricular Catheter
Cook SPECTRUM® Silicone Foley Catheter
Submission Summary (Full Text)
{0}------------------------------------------------
SPECTRUM® Ventricular Catheter - 510(k) Premarket Notification November 1, 2001
### NOV 0 82001
#### 510(K) SUMMARY 9
Merry Lee Bain Merry Lee Bain
Vice President/Director Regulatory Affairs & Clinical Services Cook Incorporated 925 South Curry Pike Bloomington, Indiana, 47402 (812) 339-2235
Trade Name: Common/Usual Name: Proposed Classification: SPECTRUM® Ventricular Catheter SPECTRUM - Ventheter, External Drainage Catheter Ventricular Catheter, External Shunt and Components 21 CFR Part 882.5550 (84JXG) Class II
The SPECTRUM® Ventricular Catheter is intended for obtaining access to a ventricular The SPECTRUM" Ventricular Cather is mielled for the puid first the pulpose of relieving
cavity of the brain for short-terra use to externally drain for the public sterile and cavity of the brain for short-term use to externally dram had bor will be supplied sterile and is
elevated intracranial pressure or fluid volume. The device will be supplied intended for one-time use.
intended for one-time use.
The SPECTRUM® Ventricular Catheter is comparable in terms of intended use and The SPECTRUM® Ventricular Catilete Ventricular catheters, including Cook
technological characteristics to predicate Ventricular catheters, including Cook 110 5 = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = Incorporated's Ventricular Catheter and the Cook SPECTRUM //ABRM Catheter for SPECTRUM® Silicone Foley Catheter and the Cook of CCTRON Problem of the Security of Component
intravascular use, the SPECTRUM® Venticious and rifempin intravascular use, the SF BCTROIA - and rifampin.
The SPECTRUM® Ventricular Catheter is a 9 Fr catheter nominally 33 cm in length.
16 sidenticular Cather stime with neminally 16 sidenors, and has The SPECTRUM® Ventricular Catherer is a 711 cansonally 16 sideports, and has
The catheter has a closed-end configuration with admin of placement. The The catheter has a closed-end contriguration was notest of placement. The markings at 1 cm increments to and in determing depth of provincial combination of
SPECTRUM® Ventricular Catheter is impregnated with an antimicrobial of SPECTRUM® Ventricular Catherer is impreglated with artuated infection during
minocycline and rifampin which may reduce the risk of eatheter is minocycline and rifampin which inay reduce are minocycline on the catheter is use. Based on HPCC analysis, the average amount of ifamino of tifampin on the catheter is
approximately 5 mg (159 Φg/cm), and the average amount of the SPECTRUM® approximately 4 mg (116 @g/cm), and alcaretes supplied with the SPECTRUM®
approximately 4 mg (116 @g/cm). I components supplied with the SPECTRUM approximately 4 mg (116 @g/cm). Components steel stylet, a stainless steel tumneling
Ventricular Catheter include a pre-loaded stainless steel tunneling Ventricular Catheter include a pre-loaded stailitess steel by rough bear
trocar, and proximal fittings, which are included to facilitate placement and use of the Ventricular Catheter.
Koll812
{1}------------------------------------------------
SPECTRUM® Ventricular Catheter - 510(k) Premarket Notification November 1, 2001
The SPECTRUM® Ventricular Catheter is similar with respect to indications for use,
10(k) The SPECTRUM" Ventricular Catheter is summar with simply of section 510(k)
materials and physical construction to predicate devices in terms of section 510(k) materials and physical construction to predicate usevices in retime equivalence.
substantial equivalency, and has undergone testing to support substantial equivalence. substantial equivalent determination under the Federal Food, Drug, and Cosmetic
The substantially equivalent determination under the Federal Food, Drug, and Cosmetic The substantially equivalent determination under une i second and the substant and intended to have and Act IS not intended in other patent matters.
The device will be manufactured according to specified process controls, undergoing The device will be manufactured according to specifical process consembly of the good of the processing, stermization and putching. Of Cook Incorporated.
The SPECTRUM® Ventricular Catheter has undergone biocompatibility testing (Dermal
) in the may of Bear Mass begal from with Histopathology, The SPECTRUM® Venticity, 7 Day Muscle Implantation with Histopathology,
Sensitization, Cytotoxicity, 7 Day Muscles, Genotoxicity, and a T Sensitization, Cytotoxicity, 7 Day Muscle Inplaniansis, Genotoxicity, and a Two Week
Intracutaneous Toxicity, Systemic Toxicity, Hemolysis, Genotoxicity, and a Two of Intracutaneous Toxicity, Systems Toxical performance testing, HPLC analysis, zone of
Brain Implantation Study), physical performance testing, Results of this tes Brain Implantation Study), physical performatice evaluation. Results of this testing
inhibition testing, susceptibility testing, and clinical evaluation. Results of this te inhibition testing, susceptibility testing, and emiled of the intended use.
provide reasonable assurance of device performance for its intended use.
#### Clinical Study
Clinical Study
To evaluate efficacy of the Cook SPECTRUM® Ventricular Catherer clinical To evaluate efficacy of the Cook SPEC I RUM - Vennondized, multicenter clinical
incidence of catheter-related infection, a prollective, and randomized, multicented to receive incidence of catheter-related intection, a prospective, and randomly assigned to receive
trial was conducted in which patients were enrolled and randomly or a 9 Fr Co trial was conducted in which patients were enrouled alle ransonary of a 9 Fr Cook
either a standard non-coated 9 Fr Cook Venticular Catherts available for either a standard non-coated 9 FF Cook Ventricular Cather (288 patients available for SPECTRUM® Ventricular Cathere (readleth). "Or atterer and 139 patients received
follow-up, 149 received the SPECTRUM® venthere placement was 8 ± 5,8 days in follow-up, 149 received the SPEC I KOive of catherer placement was 8.5 ± 5.8 days in
the control catheter. The average duration of catheren was 8.5 ± 5.8 days in the control cathere. The average duration of cantect proven. Results showed that patients
the treatment group, and 8.2 ± 6.9 days in the control group. Results showed that pa the treatment group, and 8.2 ± 6.9 days in the control groups and receiving the SPECTRUM * Ventricular Catheter had sighthoundy of ************************************************************************************************************* respectively).
{2}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle-like symbol with three curved lines representing the body and wings. The symbol is enclosed within a circle, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper half of the circle.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
### NOV 0 82001
Ms. April Lavender, RAC Vice President, Regulatory Affairs Cook, Inc. 925 South Curry Pike P.O. Box 489 Bloomington, Indiana 47402
Re: K011812
: K011812
Trade/Device Name: SPECTRUM® Vectricular Catheter FraderDon Number: 21 CFR 882.4100 Regulation Name: Ventricular Catheter Regulatory Class: Class II Product Code: NHC Dated: September 5, 2001 Received: September 10, 2001
Dear Ms. Lavender:
We have reviewed your Section 510(k) premarket notification of intent to market the device We have reviewed your Section > IQK premaise is substantially equivalent (for the indications
referenced above and have determined the device is substanted in interstate referenced above and have decimined the nevice to redicate devices marketed in interstate
for use stated in the enclosure) to legally marketed produce Amendments, for use stated in the enclosure) to the enactment date of the Medical Device Amendments, or to
commerce prior to May 28, 1976, the enactment date of the Federal Food. Drug, commerce prior to May 28, 1976, the enactiven use of the Federal Food, Drug, devices that have been reclassified in accordance with as proval application (PMA).
and Cosmetic Act (Act) that do not require approval control control sonovisions of the Act and Cosmetic Act (Act) that do not require approval on the general controls por isions of the Act. The
You may, therefore, market the device, subject to the general control You may, therefore, market the device, subject of the genting for annual registration, listing of
general controls provisions of the Act include requirements misbranding and general controls provisions of the Act mender requirements seems of the one of the adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classified (see above) mito entire) and to regular one affecting your device can
may be subject to such additional controls. Existing major regulations aff may be subject to such additional controls: Existing major regals o 898. In addion, FDA may
be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. be found in the Code of Federal Regulations, This DV, Children
Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that FDA's issuance of a substantial other requirements of the Act
that FDA has made a determination that your device complies with other requirements Yo that FDA has made a determination inal your dereces by other Federal agencies. You must
or any Federal statutes and regulations administered by other to: registration and li or any Federal statutes and regulations and limited to: registration and listing (21
comply with all the Act's requirements, including, but limited to: reguirements as set comply with all the Act 5 tequirements, including on intecturing matice requirements as set
CFR Part 807); labeling (2) CFR Part 801); good manufacturing practice requirement CFR Part 807); labeling (21 CFR Parl ovi ); good manutetan:) gtd. if applicable, the electronic
forth in the quality systems (QS) regulation (21 CFR Part 820); and 1050. forth in the quality systems (QS) regulation (21 -51 R Parveller) - 1 CFR 1000-1050.
product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
## Page 2 - Ms. April Lavender, RAC
This letter will allow you to begin marketing your device as described in your Section 510(k)
The articles of the RD A. First Finding of substantial equivalence of your devic This letter will allow you to begin marketing your device of your device of your device of your device to a legally
premarket notification. The in a cleasification for your d Premarket notification. The FDA finding of substantial equivalence of your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and If you desire specific advice for your device on our labelling regulation of the Office of
additionally 21 CFR Part 1150 - Additionally, for questions on the promotion and ad additionally 21 CFR Part 809.10 for in xitten devices), promotions on the promotion of a Compliance at (301) 594-4659. Additionally, for questions of Compliance of College of College of College your device, please contact the Office of Complaatics and fiestion" (21CFR Part 807.97).
regulation entitled, "Misbranding by reference to premaitet morification it a regulation entitled, "Misbranding by responsibilities under the Act may be obtained from the Other general information on your responsionities and Consumer Assistance at its toll-free number
Division of Small Manufacturers, International and Consumer Assistance at it Division of Small Manufacturers, Inc at its Internet address
(800) 638-2041 or (301) 443-6597 or at its Internet an . ht (800) 638-2041 or (301) 445-0597 or at its manam. html
Sincerely yours,
Sincerely yours,
Mark N Millman
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# NOV 0 8 2001
510(k) Number (if known): K01XXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXX
K.011 812
Device Name: _SPECTRUM® Ventricular Catheter
Indications For Use:
The SPECTRUM® Ventricular Catheter is intended for l he SPECTHolm "None" ventricular cavity of the brain for
obtaining access to a ventricular cavity of the purpose of obtaining access to a von
short-term use to externally drain fluid for the purpose of short-term use to oxtomally
relieving elevated intracranial pressure or fluid volume. The relieving elevated intracturing intended for one-time
device will be supplied sterile and is intended for one-time use.
(Please do not write below this line - continue on another page if needed)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Susan Walker
(Division Sign-on)
Division of General, Restorative Division of General Devices
510(k) Number -
Prescription Use (Per 21 CFR 801.109) OR
Over-The-Counter Use_
11812
(Optional Format 1-2-96
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.