K011800 · Vanguard Medical Concepts, Inc. · NLU · Sep 27, 2001 · Gastroenterology, Urology
Device Facts
Record ID
K011800
Device Name
VANGUARD REPROCESSED ELECTRIC BIOPSY FORCEPS
Applicant
Vanguard Medical Concepts, Inc.
Product Code
NLU · Gastroenterology, Urology
Decision Date
Sep 27, 2001
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.4300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
When used with a compatible electrosurgical unit, endoscope and patient grounding pad, electric biopsy forceps are intended for electrocautery and removal of polyps and/or tissue within the gastrointestinal tract. The forceps are prescription devices intended for a single patient use only.
Device Story
Vanguard Reprocessed Electric Biopsy Forceps are reprocessed versions of OEM devices (Boston Scientific Microvasive Radial Jaw 3 and C.R. Bard Precisor). Device consists of flexible, electrically isolated sheath with distal grasping cups controlled by proximal handle. Used via endoscope operating channel (min 2.8 mm) to collect tissue samples or remove polyps. Electrocautery function requires connection to compatible electrosurgical unit and patient grounding pad. Operated by physicians in clinical settings. Vanguard collects used devices, cleans, inspects, tests, repackages, and sterilizes them for return to healthcare facilities. Reprocessing restores device to original OEM specifications, allowing for safe clinical use in tissue removal.
Clinical Evidence
No clinical data. Bench testing only. Validation included decontamination and cleaning efficacy, packaging integrity, sterilization validation, functional/performance testing, shelf-life testing, and biocompatibility testing.
Technological Characteristics
Flexible, electrically isolated sheath with distal grasping cups and proximal handle. Monopolar electrocautery capability. Dimensions and materials are identical to OEM predicate devices. Sterilization performed as part of the reprocessing cycle.
Indications for Use
Indicated for patients requiring electrocautery and removal of polyps or tissue within the gastrointestinal tract via endoscopy. Prescription use only.
Regulatory Classification
Identification
An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.
Predicate Devices
Boston Scientific Microvasive Radial Jaw 3 Electric Biopsy Forceps (K860366)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol consisting of three overlapping profiles facing to the right, representing people.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
1 2004 NOV
Heather Crawford, RAC Director of Regulatory Affairs Vanguard Medical Concepts, Inc 5307 Great Oak Drive LAKELAND FL 33815
K011800 - Supplemental Validation Submission Re: Trade/Device Name: See Enclosed List Regulation Number: 21 CFR §876.4300 Regulation Name: Endoscopic electrosurgical unit and accessories Regulatory Class: II Product Code: NLU Dated: September 7, 2001 Received: September 10, 2001
Dear Ms. Crawford:
The above-referenced premarket notification (510(k)) was cleared by the Office of Device Evaluation (ODE) on September 10, 2001. We have received your supplemental validation data as required for reprocessed single-use devices by the Medical Device User Fee and Modernization Act of 2002. After reviewing your supplemental validation data, we have determined the devices listed in the enclosure accompanying this letter are substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market these devices, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your devices are classified (see above) into either class II (Special Controls) or class III (PMA) they may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your devices in the Federal Register.
{1}------------------------------------------------
#### Page 2 - Ms. Heather Crawford
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your devices comply with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's applicable requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
The FDA finding of substantial equivalence of your devices to legally marketed predicate devices results in classification for your devices and thus, permits you to legally market the devices. This letter will allow you to continue marketing the devices listed in the enclosure accompanying this letter.
If you desire specific advice for your devices on our labeling regulation (21 CFR Part 801). please contact the Office of Compliance at 240-276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely vours.
Nancy Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## ·Attachment 1
The following list is the models used in this submission.
### K011800: Vanguard Reprocessed Electric Biopsy Forceps
Microvasive® Radial Jaw 3, Model 1550 Bard Precisor, Model 852 Bard Precisor, Model 853 Bard Precisor, Model 854 Bard Precisor, Model 856
{3}------------------------------------------------
### Indications for Use
4011800 510(k) Number:
Device Name: Vanguard Reprocessed Electric Biopsy Forceps
Indications for Use:
When used with a compatible electrosurgical unit, endoscope and patient grounding pad, electric biopsy forceps are intended for electrocautery and removal of polyps and/or tissue within the gastrointestinal tract. The forceps are prescription devices intended for a single patient use only.
#### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED.)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use V (Per 21 CFR 801.109)
OR
Over-The-Counter Use
(Optional Format 1-2-96)
Nancy C. Brogdon
f Reproductive, Abdor rical Device
iv
{4}------------------------------------------------
SEP 2 7 2001
K011800
Page 1 of 2
## 510(k) Summary of Safety & Effectiveness
:
:
| Submitter | Vanguard Medical Concepts, Inc.<br>5307 Great Oak Drive<br>Lakeland, FL 33815 |
|----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact | Mike Sammon, Ph.D.<br>Director, Research & Development<br>(863) 683-8680, extension 228<br>mikes@safe-reuse.com |
| Date | May 31, 2001 |
| Devices | Trade Names: Vanguard Reprocessed Electric Biopsy Forceps<br>⇒ Boston Scientific Microvasive® Radial Jaw® 3, and<br>⇒ C.R. Bard Precisor® Electric Biopsy Forceps Common Name: Electric biopsy forceps Classification: 21 CFR 876.4300 - Endoscopic Electrosurgical Unit and<br>Accessories - Class II Product Code KGE |
| Predicate<br>Devices | Boston Scientific Microvasive® Radial Jaw® 3 Electric Biopsy Forceps<br>believed to be legally marketed under 510(k) premarket notification<br>K860366 C.R. Bard Precisor® Hot Biopsy Forceps believed to be legally marketed<br>under 510(k) premarket notification K912601 |
| Indications for<br>Use | When used with a compatible electrosurgical unit, endoscope and patient<br>grounding pad, electric biopsy forceps are intended for electrocautery and<br>removal of polyps and/or tissue within the gastrointestinal tract. The forceps<br>are prescription devices intended for a single patient use only. |
| Device<br>Description | Biopsy forceps are devices used to collect tissue samples and/or remove<br>polyps within the gastrointestinal tract via the operating channel of<br>endoscopic instruments.<br><br>Electric biopsy forceps include an electrocautery function when used with a<br>compatible electrosurgical unit.<br><br>The devices consist of flexible, electrically isolated sheaths with distal<br>grasping cups that are controlled by a proximal handle. The forceps are<br>guided by endoscopy through a biopsy channel with a minimum dimension of<br>2.8 mm. Monopolar electrocautery requires electrical connection of the<br>forceps to a compatible electrosurgical unit and use of an appropriate patient<br>grounding pad.<br><br>Vanguard receives previously used electric biopsy forceps from healthcare<br>facilities; cleans, inspects, tests, repackages and sterilizes the forceps; and<br>returns them to the healthcare facility. |
| Technological<br>Characteristics | The technological characteristics of the Vanguard reprocessed electric biopsy<br>forceps are the same as the predicate devices, the original equipment<br>manufacturer (OEM) electric biopsy forceps. The materials and dimensions<br>are unchanged and the physical characteristics, performance specifications,<br>and all other characteristics are essentially identical. |
| Test Data | Decontamination and cleaning, packaging and sterilization validations and<br>functional/performance, shelf life and biocompatibility testing demonstrates<br>that the reprocessed devices perform as intended and are safe and effective. |
| Conclusion | Based on the information provided herein and the 510(k) "Substantial<br>Equivalence" Decision Making Process Chart, we conclude that the Vanguard<br>reprocessed electric biopsy forceps are substantially equivalent to the<br>predicate device, the respective OEM biopsy forceps under the Federal Food,<br>Drug and Cosmetic Act. |
Continued on next page
: :
{5}------------------------------------------------
K011800
Page 2 of 2
# 510(k) Summary of Safety & Effectiveness, Continued
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.