VERTIS PNT CONTROL UNIT, VERTIS PNT SAFEGUIDE KIDS, MODELS CU100, SGXXX
K011702 · Vertis Neuroscience, Inc. · NHI · Dec 21, 2001 · Neurology
Device Facts
Record ID
K011702
Device Name
VERTIS PNT CONTROL UNIT, VERTIS PNT SAFEGUIDE KIDS, MODELS CU100, SGXXX
Applicant
Vertis Neuroscience, Inc.
Product Code
NHI · Neurology
Decision Date
Dec 21, 2001
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5890
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Percutaneous Neuromodulation Therapy (PNT) is indicated for the symptomatic relief and management of chronic or intractable low back pain and/or as an adjunctive treatment in the management of post-surgical low back pain and post-trauma low back pain.
Device Story
Vertis PNT System delivers percutaneous electrical stimulation for pain management. System components: software-driven, 5-channel, AC-powered control unit; sterile needle electrodes (Safeguides); patient cable. Used in clinical environments by physicians (e.g., physical medicine and rehabilitation) or by physical therapists under physician order; not for patient self-use. Control unit generates electrical stimulus delivered via needle electrodes to target nerves. Output allows clinicians to manage chronic/intractable low back pain. Benefits include symptomatic relief and adjunctive treatment for post-surgical/post-trauma pain.
Clinical Evidence
Prospective clinical study conducted under IDE. Trial duration: November 2000 to March 2001. Study involved 215 PNT therapy sessions with 2150 total electrode placements. Results demonstrated safety and effectiveness of the therapy for the indicated pain management.
Technological Characteristics
5-channel, AC-powered nerve stimulator. Includes sterile needle electrodes and patient cable. Conforms to ANSI/AAMI NS4, ES1, NS15; IEC 60601-1, 60601-1-2, 60601-1-4, 60601-2-10; EN 1441. Sterilization via radiation (ANSI/AAMI/ISO 11137). Software-driven control unit.
Indications for Use
Indicated for patients with chronic or intractable low back pain, including post-surgical and post-trauma low back pain, requiring symptomatic relief or adjunctive management.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{0}------------------------------------------------
## Appendix 25 - revised 510(k) Summary of Safety and Effectiveness
Class II
DEC 2 1 2001
## Vertis Neuroscience, Inc. Percutaneous Neuromodulation Therapy (PNT) ™ Control Unit and Accessories
#### General Information
K011702
Classification
Trade Name
Percutaneous Neuromodulation Therapy (PNT) Nerve Stimulation System
Submitter
Contact
Vertis Neuroscience, Inc. 2101 Fourth Avenue, Suite 200 Seattle, Washington, USA 98121
Lori Glastetter Vice President, Regulatory Affairs/ Quality Assurance
#### Substantially Equivalent Devices
| Manufacturer | Substantially Equivalent devices | 510(k) |
|----------------------------------------|-------------------------------------------------------------------------------------|--------------------|
| Empi, Inc.<br>St. Paul, MN | EPIX Tens Device System<br>Model EPIX VT<br>Model EPIX XL | K970203<br>K951903 |
| Rehabilicare, Inc.<br>New Brighton, MN | SMP-PlusTM<br>Model 4930 | K982410 |
| Chattanooga Group, Inc.<br>Hixson, TN | Intelect ® Legend Stim<br>Model IFC2 (catalog #INT002) | K924666 |
| Medtronic, Inc<br>Minneapolis, MN | Mattrix® Neurostimulation System<br>Receiver model: 3272<br>Transmitter model: 3210 | K982902 |
| TECA Corporation<br>Pleasantville, NY | TECA Disposable Monopolar Needles<br>902-DMFxx-TP series<br>892-DMGxx-TP series | K973442 |
| Medtronic, Inc.<br>Minneapolis, MN | DMNTM Disposable Monopolar Needle<br>Electrodes<br>DMFxx series<br>DMNxx series | K950314 |
| Medtronic, Inc.<br>Minneapolis, MN | Resume II Lead<br>Model 3587a | K915540 |
#### Intended use
Percutaneous Neuromodulation Therapy (PNT) is indicated for the symptomatic relief and management of chronic or intractable low back pain and/or as an adjunctive treatment in the management of post-surgical low back pain and post-trauma low back pain.
{1}------------------------------------------------
. ...
{2}------------------------------------------------
## Device Description
The Vertis PNT System is designed for delivering percutaneous electrical stimulation (termed: Percutaneous Neuromodulation Therapy - PNT ). The Vertis PNT System is intended to be used in pain management by a physician (e.g., I NT Uystein is intendousical medicine and rehabilitation physicians) or on the order of a physician (e.g., by a physical therapist) in a clinic environment. It is not intended for patient use. The device system includes 3 major components:
- the Vertis PNT Control Unit a software-driven, five channel, AC . powered nerve stimulator which generates the electrical stimulus;
- the sterile Safeguides which are sterile, needle electrodes, and ●
- the Patient Cable which interconnects the PNT Control Unit to the . electrodes
## Bench/Animal Testing
Extensive bench and animal was conducted on the Vertis PNT System and included performance and safety testing. The Vertis PNT System conforms to applicable performanted on the technical references and FDA-recognized consensus standards noted below. Additionally, the PNT Control Unit software was validated per recognized validation techniques.
Results: All testing of the products yielded acceptable results prior to commercial distribution.
ANSI/AAMI NS4-1985 Transcutaneous Electrical Nerve Stimulators, approved 5-20-86. ANSI/AAMI ES1 1993 - Safe Current Limits for Electromedical Apparatus, approved 12-2-93 ANSI/AMI NS15-1995 Implantable Peripheral Nerve Stimulators, approved 2-1-95 IEC60601-1:1993 Medical Electrical Equipment, Part 1: General Requirements for Safety and Amendments A1:1991, A2:1995. IEC 60601-1-2 First Edition 1993-4 Medical Electrical Equipment, - Part 1: General Requirements for Safety: Electromagnetic Compatibility -Requirements and Tests (tested in accordance to the IEC 61000 series). IEC 60601-1-4:1996 Medical Electrical Equipment, Part 1: General Requirements for Safety, Part 2: Collateral Standard: Programmable electrical medical systems. IEC 60601-2-10:1987 Medical Electrical Equipment, Part 2: Particular Requirements for the Safety of Nerve and Muscle Stimulators. EN (CEN) 1441:1997 Medical Devices - Risk Analysis.
{3}------------------------------------------------
ANSI/AAMI/ISO 15223: 2000 Medical Devices - Symbols to be used with Medical Device Labels, Labeling and Information to be Supplied, approved 3-13-00.
IEC 60601-2-10:1987 Medical Electrical Equipment, Part 2: Particular Requirements for the Safety of Nerve and Muscle Stimulators.
ANSI/AAMI/ISO 11137: 1994 Sterilization of Health Care Products -Requirements for Validation and Routine Control -Radiation Sterilization; using Method 1 as described per AAMI/ISO TIR No. 13409: 1996 Sterilization of Health care Products - Radiation Sterilization -Substantiation of 25kGy as a Sterilization Dose for Small or Infrequent Production Batches.
ANSI/AAMI/ISO 10993-1:1997, Biological Evaluation of Medical Devices -Part 1: Evaluation and Testing.
## Clinical Testing
Under the auspices of an Investigational Device Exemption, Vertis conducted a prospective clinical study evaluating the safety and effectiveness of percutaneous neuromodulation therapy.
The Vertis clinical trial was reviewed and approved in November 2000. The trial commenced shortly thereafter at multiple United States clinical sites. The trial was completed in March 2001. During the multi-week investigational protocol, there were a total of 2150 electrode placements in 215 PNT therapy sessions with patients.
Results: The clinical trial demonstrated the safety and effectiveness of the percutaneous neuromodulation therapy.
#### Summary of Substantial Equivalence
Therefore, due to the similarity of design features, materials, test results, clinical results and indicated use to other predicate devices, Vertis believes the Vertis PNT System does not raise any new safety or effectiveness issues and is substantially equivalent to currently available nerve stimulators, electrodes and accessories that have been determined to be substantially equivalent to devices in commercial distribution prior to May 28, 1976.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized symbol of three curved lines that resemble a person. The symbol is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter of the circle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC 2 1 2001
Lori J. Glastetter Vice President, Regulatory Affairs/Quality Assurance Vertis Neuroscience, Inc. 2101 Fourth Avenue, Suite 200 Seattle, Washington 98121
Re: K011702/S1
rever router System (Vertis PNT Control Unit and Vertis PNT Lumbar Safeguard Kit)
Regulation Number: 21 CFR 882.5890 and 21 CFR 882.1350
Regulation Name: Transcutaneous Electrical Nerve Stimulator for Pain Relief and Needle Electrode Regulatory Class: Class II Product Code: NHI and GXZ Dated: September 24, 2001 Received: September 25, 2001
Dear Ms. Glastetter:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commence prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{5}------------------------------------------------
Page 2 – Ms. Lori Glastetter
CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as setting CHK Part 807); labeling (21 CFR Part 001); good rate 820); and if applicable, the electronic
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable forth in the quality systems (Sections 531-542 of the Act); 21 CFR 1000-1050.
product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 5 10(k) I his letter will anow you to ought manteaning of substantial equivalence of your device your device premarket notification. The PDA miding of baction for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and If you desire specific advice for your acvice disenostic devices), please contact the Office of additionally 21 CFN Fall 607.10 for in the e.for suestions on the promotion and advertising of Compliance at (301) 594 4657. Thursday, Jim 1594-4639. Also, please note the your device, piease comace the Ories of eeee on premarket notification" (21CFF Part 807.97).
regulation entitled, "Misbranding by reference to premarket notification" (21CFF) regulation entitied, "Misorananing of responsibilities under the Act may be obtained from the Other general information on your responsional and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address (600) 050 2017 01 (501) - 1 (501) - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - - 1 - - - - - - - - - - - - - - - - - - - - - - - - - - -
Sipeerely gours,
Celia M. Witten, P Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Page__________________________________________________________________________________________________________________________________________________________________________
510(k) Number (if known): Koll 7-02
Device Name: Vertis PNT System
Indications For Use:
Percutaneous Neuromodulation Therapy (PNT) is indicated for the symptomatic relief and management of chronic or intractable low back pain and/or as an adjunctive treatment in the management of post-surgical low back pain and post-trauma low back pain.
The PNT Control Unit is to be used with a PNT Safeguide Kit - Lumbar.
# (Please Do not WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of General, Restorative and Neurological Devices
(Optional Format 3-10-98)
510(k) Number KD11702
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.