← Product Code [MJN](/productcode/MJN) · K011656

# HYPERFORM OCCLUSION BALLOON SYSTEM, MODEL 104-4470, 104-4471, 104-4770, 104-4771 (K011656)

_Micro Therapeutics, Inc. · MJN · Jun 20, 2001 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K011656

## Device Facts

- **Applicant:** Micro Therapeutics, Inc.
- **Product Code:** [MJN](/productcode/MJN.md)
- **Decision Date:** Jun 20, 2001
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.4450
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The MTI Occlusion Balloon Catheter is indicated for use in the blood vessels of the peripheral and neuro vasculature where temporary occlusion is desired. The MTI Occlusion Balloon Catheter offers a vessel selective technique of temporary vascular occlusion, which is useful in selectively stopping or controlling blood flow.

## Device Story

HyperForm™ Occlusion Balloon Catheter is a single-lumen tapered catheter with a non-detachable, low-inflation-pressure compliant balloon at the distal end. Used in peripheral and neuro vasculature for temporary blood flow occlusion. Operated by physicians in clinical settings. Requires MTI 0.010" guidewire for tracking and to occlude the catheter shaft lumen for balloon inflation. Features two platinum markers for angiographic visualization and hydrophilic coating for navigation. Supplied sterile for single use. Provides vessel-selective flow control, aiding clinical procedures requiring temporary occlusion.

## Clinical Evidence

Bench testing only. Biocompatibility verified per ISO 10993-1 for external communicating, blood-contact, short-duration (<24 hrs) devices. Performance testing conducted per ISO 10555 (Parts 1 and 4) including dimensional verification, balloon compliance/integrity, tensile strength, torque strength, flexibility, trackability, and coating integrity.

## Technological Characteristics

Single-lumen tapered catheter; non-detachable compliant balloon; platinum markers for visualization; hydrophilic coating. Materials verified for biocompatibility (ISO 10993-1). Performance tested per ISO 10555. Sterile, single-use. Mechanical operation via guidewire-assisted inflation.

## Regulatory Identification

A vascular clamp is a surgical instrument used to occlude a blood vessel temporarily.

## Predicate Devices

- MTI Equinox™ Occlusion Balloon Catheter

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## JUN 2 0 2001

| Micro Therapeutics, Inc. |                                       |
|--------------------------|---------------------------------------|
| Special 510(k):          | HyperForm™ Occlusion Balloon Catheter |

# Attachment 6

|                                    | 510(k) Summary                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |  |
|------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Prepared May 24, 2001              |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |  |
| TRADE NAME                         | Unknown                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |  |
| GENERIC NAME                       | Occlusion Balloon Catheter                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |  |
| CLASSIFICATION                     | Class II (21 CFR 870.0)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |  |
| SUBMITTED BY                       | Micro Therapeutics, Inc.<br>2 Goodyear<br>Irvine, CA 92618                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |  |
| CONTACT                            | Eben Gordon<br>Regulatory Affairs<br>(949) 837-3700                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |  |
| PREDICATE DEVICE                   | MTI Equinox™ Occlusion Balloon Catheter                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |  |
| DEVICE DESCRIPTION                 | The HyperForm™ Occlusion Balloon Catheter is a single lumen tapered catheter with a non-detachable low inflation pressure compliant balloon attached to the distal end of the catheter. The catheter is designed to track over the MTI 0.010" guidewire, and requires insertion of the guidewire to occlude the catheter shaft lumen to allow inflation of the balloon. Two platinum markers provide angiographic visualization of the balloon length and facilitate intravascular placement of the balloon prior to inflation. The catheter shaft is hydrophilically coated to assist catheter advancement within the vasculature. The HyperForm catheter is supplied sterile for single use as a system, which includes the required 0.010" guidewire.                                     |  |
| INDICATIONS FOR USE                | The MTI Occlusion Balloon Catheter is indicated for use in the blood vessels of the peripheral and neuro vasculature where temporary occlusion is desired. The MTI Occlusion Balloon Catheter offers a vessel selective technique of temporary vascular occlusion, which is useful in selectively stopping or controlling blood flow.                                                                                                                                                                                                                                                                                                                                                                                                                                                        |  |
| TESTING                            | Biocompatibility of the HyperForm catheter was verified in accordance with ISO 10993-1, Biological Evaluation of Medical Devices. Test results confirmed biocompatibility of the HyperForm catheter when tested as an external communicating, blood contact, short duration (<24 hrs.) device.<br>Performance testing of the HyperForm catheter was conducted in accordance with ISO 10555 Sterile, single use intravascular catheters- Parts 1 and 4. Tests included dimensional verification, balloon compliance and integrity, catheter tensile strength, torque strength, flexibility, trackability, and coating integrity. Test results demonstrate that the device meets or exceeds the requirements of these standards and performs substantially equivalent to the predicate device. |  |
| SUMMARY OF SUBSTANTIAL EQUIVALENCE | The MTI HyperForm™ Occlusion Balloon Catheter is substantially equivalent to the predicate device in intended use and principles of operation.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |  |

{1}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/1/Picture/1 description: The image shows a seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circle around a symbol. The symbol consists of three stylized human profiles facing right, with three curved lines above them. The seal is black and white.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

#### JUN 2 0 2001

Mr Eben Gordon Mr Eben Gordon
Director of Regulatory Affairs and Quality Assurance Micro Therapeutics, Inc. 2 Goodyear Irvine, CA 92618

Re: K011656

K011656
Trade Name: MTI HyperForm™ Occlusion Balloon Catheter
Trade Name: MTI HyperForm a Regulation Number: 870.4450 Regulation Class: II (two) Product Code: MJN Dated: May 24, 2001 Received: May 29, 2001

Dear Mr. Gordon:

We have reviewed your Section 510(k) notification of intent to market the device referenced
 in the subscription of the layies is gubstentially equivalent (for the indication We have reviewed your Section 510(c) notification of the indications for use indications for use
above and we have determined the devices marked in interstate commerce above and we have determined the device is subsantany of an arter and stated in the enclosure) to legally marketed predical Device Amendments, or to devices that
prior to May 28, 1976, the enactment date of the Federal Food. Drug and prior to May 28, 1976, the enatment uite the provisions of the Federal Food, Drug, and
have been reclassified in accordance with the provisions of the Federal contra have been reclassified in accordance with the provisions of a subject to the general controls
Cosmetic Act (Act). You may, therefore, market the Act include requirements fo Cosmetic Act (Act). You may, therefore, market incurrents and include requirements for annual
provisions of the Act. The general controls provisions of the Actively and pro provisions of the Act. The general controls provisions of the Ace 10 for and prohibitions against
registration, listing of devices, good manufacturing practice, labeling, an misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III
) in the subscription of the stars of any as additional controls. Existing maj If your device is classified (see above) into entrely studions controls. Existing major regulations
(Premarket Approval), it may be subject to such additions Title 2 . Part (Premarket Approval), it may be subject to such adultions, Title 21, Parts 800 to 895.
affecting your device can be found in the Code of Federal Regulations, Title 21, Parts affecting your device can be lound in the counds compliance with the Current Good
A substantially equivalent determination assumes compliance with ( A substantially equivalent determination asstilles complains in System Regulation (QS) for 
Manufacturing Practice requirements, as set forth in the Quality System Reciodic O Manufacturing Practice regulation (2) CFR Part 820) and that, through periodic QS
Medical Devices: General regulation (2) CFR Part 820) and that, through periodic of Medical Devices: General regulation (ZTCA) will verify such assumptions. Failure to
inspections, the Food and Drug Administration (FDA) will verify such assumptions, FDA may inspections, the Food and Drug Administration in Proxy Jonadition, FDA may publish
comply with the GMP regulation may resimination in the Forester this comply with the GMP regulation may result new Federal Register. Please note: this further announcements concerning your device in the reast affect any obligation you might
response to your premarket notification submission does not a Flectronic Product response to your premarket nothleadel studies under the Electronic Product 
have under sections 531 through 542 of the Act for devices under the Electronic Product 
 Sales an have under sections 531 through 542 of the Federal laws or regulations.
Radiation Control provisions, or other Federal laws or regulations.

{2}------------------------------------------------

Page 2 - Mr Eben Gordon

This letter will allow you to begin marketing your device as described in your 510(k) premarket I his letter will anow you to begin maneting your device of your device to a legally marketed .
notification. The FDA finding of substantial equivalence of your device to notification. The FDA inding of substantial equiveland thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and If you desire specific advice for your do rise contact the Office of Compliance at additionally 809.10 for in Vitto draghostions on the promotion and advertising of your device, (301) 394-4648. Addinonally, for questions of (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general entitled, "Wisoranding by relevelo to prema act may be obtained from the Division of Small
information on your responsibilities under the Act and 2011 - (201) 412 (6597 - at information on your responsibilities and its tool more (800) 638-2041 or (301) 443-6597 or at its Maindracturers / Issiblance as a catalog as a dsmamain.html".

Sincerely yours,

V. Dale Tell

James E. Dillard Director Division of Cardiovascular and Respiratory Devices Office of Device Evaluation Center for Devices and Radiological Health

{3}------------------------------------------------

|  | Attachment 2 |
|--|--------------|
|--|--------------|

|                           | Indications for Use Statement |
|---------------------------|-------------------------------|
| 510(k) Number (if known): | KO11656                       |

Device Name:

### MTI HyperForm™ Occlusion Balloon Catheter

Indications for Use: The MTI HyperForm™ Occlusion Balloon Catheter issue The MIT Tryper of the blood vessels of the peripheral and Indicated for use in the semporary occlusion is desired. The MTI HyperForm™ Occlusion Balloon Catheter offers a MITI Hyperrorm---------------------------------------------------------------------------------------------------------------------------------------------------------------vesser selective technique vir stopping or controlling blood flow.

#### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

|                   |                      | Concurrence of CDRH, Office of Device Evaluation (ODE) |                       |
|-------------------|----------------------|--------------------------------------------------------|-----------------------|
|                   |                      |                                                        |                       |
| Prescription Use_ | OR                   |                                                        | Over the Counter Use_ |
|                   | (Per 21 CFR 801.109) |                                                        |                       |
|                   |                      | VICES                                                  | Page 18 of 100        |

---

**Source:** [https://fda.innolitics.com/device/K011656](https://fda.innolitics.com/device/K011656)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
