MODIFICATION TO EXACTECH ACUMATCH INTEGRATED HIP SYSTEM, M-SERIES HIGH OFF-SET NECK SEGMENT, M-SERIES LOW OFF-SET NECK S
K010120 · Exactech, Inc. · LPH · Feb 9, 2001 · Orthopedic
Device Facts
Record ID
K010120
Device Name
MODIFICATION TO EXACTECH ACUMATCH INTEGRATED HIP SYSTEM, M-SERIES HIGH OFF-SET NECK SEGMENT, M-SERIES LOW OFF-SET NECK S
Applicant
Exactech, Inc.
Product Code
LPH · Orthopedic
Decision Date
Feb 9, 2001
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3358
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
AcuMatch M-Series components are indicated for use in skeletally mature individuals undergoing primary surgery for total hip replacement due to osteoarthritis, rheumatoid arthritis, osteonecrosis, ankylosing spondylitis, congenital hip dysplasia, post-traumatic degenerative problems of the hip, and for treatment of proximal femoral fractures where prosthetic replacement is determined by the surgeon as the preferred treatment. Components of the M-Series are also potentially indicated for revision of failed previous reconstructions where sufficient bone stock is present and to restore mobility resulting from previous fusion. The AcuMatch M-Series components are indicated for press-fit and cemented applications.
Device Story
AcuMatch M-Series is a modular femoral stem system for total hip arthroplasty; comprises proximal neck, metaphyseal, and diaphyseal segments secured with a locking screw. Components are titanium alloy. High and low off-set neck segments provide anatomical sizing options. Used in OR by orthopedic surgeons for press-fit or cemented fixation. Device restores hip joint function and mobility in patients with degenerative disease or fractures. Surgeon selects appropriate neck offset based on patient anatomy; modularity allows intraoperative customization. Benefits include improved fit and stability compared to standard components.
Clinical Evidence
Bench testing only. Finite Element Analysis was performed to compare the performance of the proposed high and low off-set neck segments against the predicate M-Series neck design under anticipated in-vivo loading conditions. Results confirmed that the proposed devices exhibit adequate performance characteristics comparable to the predicate.
Technological Characteristics
Modular femoral stem system; titanium alloy construction. Components include proximal neck, metaphyseal, and diaphyseal segments with locking screw. Design features high and low off-set neck variants. Intended for press-fit or cemented application. No software or electronic components.
Indications for Use
Indicated for skeletally mature patients undergoing primary total hip replacement for osteoarthritis, osteonecrosis, congenital hip dysplasia, rheumatoid arthritis, ankylosing spondylitis, post-traumatic degenerative hip problems, or proximal femoral fractures. Also indicated for revision of failed previous reconstructions with sufficient bone stock. Contraindicated in patients with active infection, insufficient bone stock for fixation, neuromuscular disorders preventing hip joint control, or patients whose weight, age, or activity level suggests early system failure.
Regulatory Classification
Identification
A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.
{1}------------------------------------------------
lo120
2320 NW 66TH COUR1 GAINESVILLE, FL 3265.
352-377-1140 FAX 352-378-2617
# FEB - 9 2001
Exactech® AcuMatch™ Integrated Hip System M-Series Femoral Stem Component High & Low Off-Set Neck Segments
### 510(k) Summary of Safety and Effectiveness
Sponsor:
Exactech® Inc. 2320 N.W. 66th Court Gainesville, Florida 32653
Phone: (352) - 377 - 1140 Fax: (352) - 378 - 2617
FDA Establishment Number 1038671
Gary J. Miller, Ph.D. Contact: Executive V.P. of Research and Development
Date:
January 10, 2001
Section 4 Page 1 of 4
{2}------------------------------------------------
# Exactech® AcuMatch™ Integrated Hip System M-Series Femoral Stem Component High & Low Off-Set Neck Segments
### 510(k) Summary of Safety and Effectiveness
| Trade Name: | Exactech® AcuMatch M-Series<br>High Off-Set Neck<br>Low Off-Set Neck |
|-----------------------|--------------------------------------------------------------------------------------------|
| Common Name: | Total Hip Prosthesis Femoral Component |
| Classification Name: | Prothesis, Hip, Semi-Constrained, Metal/Polymer,<br>Porous, Uncemented (Femoral Component) |
| Product Code: | LPH |
| Device Class: | II |
| Classification Panel: | Orthopedic |
#### Legally Marketed Devices for Substantial Equivalence Comparison:
| Model | Manufacturer | 510(k) |
|--------------|---------------------------|----------|
| M-Series | Exactech Inc. | #K993736 |
| S-Rom | Joint Medical<br>Products | |
| Impact | Biomet | |
| Mallory Head | Biomet | |
| Link | Link America | |
The AcuMatch M-Series high and low off-set neck segments are made of similar materials and are of a similar design to other legally marketed modular femoral components. Most notably, the proposed neck segments are equivalent in materials and design to the Exactech's predicate M-Series neck segments. M-Series components are also similar to the "S-ROM" by Joint Medical Products Corporation, the Biomet "Impact", and "Mallory-Head" and the "Link MP" by Link America.
### Section 4 Page 2 of 4
{3}------------------------------------------------
# Exactech® AcuMatch™ Integrated Hip System M-Series Femoral Stem Component High & Low Off-Set Neck Segments
### 510(k) Summary of Safety and Effectiveness
#### Description:
The Exactech Modular Stem is a four-piece system consisting of a proximal neck segment, metaphyseal segment, diaphyseal segment and a locking screw. All of the components are interchangeable, therefore allowing for many sizing combinations to meet varying anatomical situations. The components are composed of titanium alloy. Design changes were made to the original neck segments cleared through premarket notification #K993736:
- A. High Offset Neck Segments
- Material was added to the lateral radius of all neck segments. 1.
- 2. A new "-5 Neck Segment" model was added.
#### B. Low Offset Neck Segments
A new "Low Offset" Neck version of neck is proposed. These models represent a reduction in horizontal offset of 5mm as compared to the predicate design (#K993736):
- Low Offset -- 5 Neck Segment 1.
- Low Offset Standard (+0) Neck Segment 2.
- 3. Low Offset +10 Neck Segment
#### Intended Use:
AcuMatch M-Series components are indicated for use in skeletally mature individuals undergoing primary surgery for total hip replacement due to osteoarthritis, rheumatoid arthritis, osteonecrosis, ankylosing spondylitis, congenital hip dysplasia, post-traumatic degenerative problems of the hip, and for treatment of proximal femoral fractures where prosthetic replacement is determined by the surgeon as the preferred treatment. Components of the M-Series are also potentially indicated for revision of failed previous reconstructions where sufficient bone stock is present and to restore mobility resulting from previous fusion.
The AcuMatch M-Series components are indicated for press-fit and cemented applications.
{4}------------------------------------------------
# Exactech® AcuMatch™ Integrated Hip System M-Series Femoral Stem Component High & Low Off-Set Neck Segments
# 510(k) Summary of Safety and Effectiveness
### Contraindications:
AcuMatch M-Series components are contraindicated in patients with active infection, Aculviation M-36Ttes components are to allow appropriate insertion and fixation of the patients without surficient bolle stock of the mot allow control of the hip joint, and in prosulchis, in nedromascural enetivity level would cause the surgeon to expect early failure of the system.
### Performance Data Summary:
Finite Element Analysis was conducted to verify that the performance of the proposed I high and low off-set neck segments would be adequate for anticipated in-vivo loading. The results showed that the proposed devices have performance characteristics I he results blowed than Exactech's predicate M-Series neck design.
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, rendered in a simple, bold line drawing.
FEB = 9 2001
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Lisa Simpson Regulatory Representative Exactech 2320 NW 66" Court Gainesville, Florida 32653
Re: K010120
Trade Name: Exactech® Integrated Hip System
AcuMatch™ M-series Femoral Stem Component High and Low Offset Neck Segments Regulatory Class: II Product Code: LPH Dated: January 10, 2001
Received: January 16, 2001
Dear Ms. Simpson:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
{6}------------------------------------------------
Page 2 - Ms. Lisa Simpson
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597, or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Mark N Mulkerson
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# Exactech® AcuMatch™ Integrated Hip System M-Series Femoral Stem Component
#### Indications for Use
510(k) Number:
K0/0/20
Device Name:
Exactech Integrated Hip System AcuMatch M-Series Femoral Stem Component High Off-Set Neck Segment Low Off-Set Neck Segment
#### Indications for use:
AcuMatch M-Series Femoral Components are indicated for use in skeletally mature individuals undergoing primary surgery for total hip replacement due to osteoarthritis, osteonecrosis, congenital hip dysplasia, rheumatoid arthritis, ankylosing spondylitis, and/or post-traumatic degenerative problems. M-Series components are also potentially indicated for revision of failed previous reconstructions where sufficient bone stock is present.
AcuMatch M-Series components are intended to be used in press-fit and cemented applications.
#### Contraindications:
AcuMatch M-Series components are contraindicated in patients with active infection, patients without sufficient bone stock to allow appropriate insertion and fixation of the prosthesis, in neuromuscular disorders that do not allow control of the hip joint, and in patients whose weight, age, or activity level would cause the surgeon to expect early failure of the system.
Mark N Melkerson
Division Sign-Off) Division of General, R rative and Neurological Devi
< 010120
Please do not write below this line - use another page if needed.
510(k) Number
time - use another page if needed.
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
12
or
Over the Counter Use
Counter Use No
Section 3 Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.