Indicated for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions: 1. non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, ankylosis, protrusio acetabuli, and painful hip dysplasia; 2. inflammatory degenerative joint disease such as rheumatoid arthritis; 3. correction of functional deformity; 4. revision procedures where other treatments or devices have failed; and, treatment of nonunion, femoral neck and trochanteric fractures of the proximal 5. femur with head involvement that are unmanageable using other techniques.
Device Story
Modular revision hip prosthesis; comprises distal stem, extension adapter, proximal body, and modular neck. Distal stems available in short, medium, long lengths with cutting flanges for diaphyseal fixation; proximal bodies available in seven sizes; modular necks offer various anteversion/retroversion and varus/valgus options. Components secured via morse taper and fixation screws. Used by orthopedic surgeons in clinical settings for total hip arthroplasty. Provides structural support and joint function restoration for patients with degenerative joint disease or complex fractures. Benefits include improved hip function and pain relief through modular customization to patient anatomy.
Clinical Evidence
No clinical data provided; substantial equivalence supported by design features, materials data, and bench testing results.
Technological Characteristics
Modular hip prosthesis; components include distal stem, extension adapter, proximal body, and modular neck. Distal stems feature cutting flanges for primary fixation; secured via morse taper and fixation screws. Materials not specified by standard. Modular design allows for anatomical adjustment (anteversion/retroversion, varus/valgus).
Indications for Use
Indicated for skeletally mature patients undergoing total hip arthroplasty for non-inflammatory/inflammatory degenerative joint disease (osteoarthritis, avascular necrosis, ankylosis, protrusio acetabuli, hip dysplasia, rheumatoid arthritis), functional deformity correction, revision procedures, or treatment of nonunion, femoral neck, and trochanteric fractures of the proximal femur with head involvement.
Regulatory Classification
Identification
A hip joint femoral (hemi-hip) metallic cemented or uncemented prosthesis is a device intended to be implanted to replace a portion of the hip joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum. This generic type of device includes designs which are intended to be fixed to the bone with bone cement (§ 888.3027) as well as designs which have large window-like holes in the stem of the device and which are intended for use without bone cement. However, in these latter designs, fixation of the device is not achieved by means of bone ingrowth.
Submission Summary (Full Text)
{0}------------------------------------------------
# DEC 1 3 2000
Image /page/0/Picture/1 description: The image shows the logo for Wright Medical Technology, Inc. The logo consists of a stylized "W" on the left, followed by the company name in bold, uppercase letters. Below the company name is the address: 5677 Airline Road, Arlington, TN 38002, and the phone number: 901-867-9971.
## 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS
In accordance with the Food and Drug Administration Rule to implement provisions of the Safe Medical Devices Act of 1990 and in conformance with 21 CRF 807, this information serves as a Summary of Safety and Effectiveness for the use of the PRO-FEMUR R Revision Hip System.
Submitted By:
Date:
Contact Person:
Proprietary Name:
Common Name:
Classification Name and Reference:
Device Product Code and Panel Code:
Wright Medical Technology, Inc.
September 26, 2000
Ehab M. Esmail Senior Regulatory Affairs Associate
PRO-FEMUR R
REVISION HIP SYSTEM
21 CFR 888.3358 Prosthesis, Hip, Semi-Constrained, metal/polymer, Uncemented - Class II Orthopedics/87/LPH
## DEVICE INFORMATION
## A. INTENDED USE
Indicated for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions:
- 1. non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, ankylosis, protrusio acetabuli, and painful hip dysplasia;
- 2. inflammatory degenerative joint disease such as rheumatoid arthritis;
- 3. correction of functional deformity;
- 4. revision procedures where other treatments or devices have failed; and,
- treatment of nonunion, femoral neck and trochanteric fractures of the proximal 5. femur with head involvement that are unmanageable using other techniques.
Image /page/0/Picture/25 description: The image shows the logo for Wright Medical Technology, Inc. The logo features a stylized "W" symbol above the word "WRIGHT" in a bold, sans-serif font. Below "WRIGHT" is the text "MEDICAL TECHNOLOGY, INC." in a smaller, sans-serif font. The logo is black and white.
Image /page/0/Picture/26 description: The image shows a black and white drawing of a globe. The globe is tilted on its axis and shows the continents of Asia, Australia, and parts of Africa. The globe is sitting on a stand.
510(K) SUMMARY
{1}------------------------------------------------
The PRO-FEMUR R Revision Hip System are single use components, intended for use in conjunction with associated ceramic or metal femoral heads as part of an uncemented total hip arthroplasty.
### B. DEVICE DESCRIPTION
The PRO-FEMUR R Revision Hip System is a modular prosthesis comprising of four principal parts:
- · Distal stem with cutting flanges for the diaphyseal fixation
- · Extension Adapter
- · Proximal body for support in the metaphyseal region
- " Modular neck
The PRO-FEMUR R distal stem will be available in three different lengths, namely SHORT, MEDIUM, and LONG, in increments of 1mm from 10mm to 22mm in diameter. The SHORT distal stem (135mm) is straight, and the MEDIUM (175mm) and LONG (215mm) are both anatomically curved. All distal stems are slightly conical in profile and feature cutting flanges that provide excellent primary fixation. The cutting flanges are absent on the anterior and posterior curvature of the stem in order to reduce the possibility of femoral fracture during impaction.
The PRO-FEMUR R Proximal Body will be available in seven different sizes: Extra Small, Small, 4x Standard and Large. The four standard components have the same length but became progressively wider.
The distal stem is secured to the proximal body by morse taper and secondly by a fixation screw.
To further lengthen the implant by either 26 or 52 mm, extension adapters are available. They are positioned between the selected proximal body and distal stem and secured by morse taper and a longer fixation screw.
The proximal body has a specific oblong housing for the insertion of the twelve modular necks which are available in six versions and two lengths: Neutral, antiversion/ retroversion 8° or 15°, varus/ valgus 8°, or combination of anteverted/ retroverted - varus/ valgus (in both short and long lengths)
The design features of PRO-FEMUR R Revision Hip System are substantially equivalent to competitive devices previously cleared for market.
## C. SUBSTANTIAL EQUIVALENCE INFORMATION
The intended use, material, type of interface, and design features of PRO-FEMUR R Revision Hip System are substantially equivalent to the competitive devices previously cleared for market. The safety and effectiveness of the PRO-FEMUR R Revision Hip System are adequately supported by the substantial equivalence information, materials data, and testing results provided within the Premarket Notification.
Image /page/1/Picture/16 description: The image shows the logo for Wright Medical Technology, Inc. The logo features a stylized "W" symbol above the word "WRIGHT" in a bold, sans-serif font. Below the word "WRIGHT" are the words "MEDICAL TECHNOLOGY, INC." in a smaller, sans-serif font.
Image /page/1/Picture/17 description: The image shows a globe with a focus on the continents of Asia and Australia. The globe is positioned at an angle, with the northern hemisphere tilted slightly towards the viewer. The landmasses are shaded to indicate topography, and the oceans are a lighter color. The globe is mounted on a stand, which is partially visible at the bottom of the image.
{2}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, which is a symbol often associated with medicine and healthcare. The caduceus is depicted with three intertwined strands and a wing-like structure at the top. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the caduceus.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
## DEC 1 3 2000
Mr. Ehab M. Esmail Senior Regulatory Affairs Associate Wright Medical Technology Incorporated 5677 Airline Road Arlington, Tennessee 38002
Re: K003016
Trade Name: Pro-Femur R Revision Hip System Regulatory Class: II Product Code: LWJ Dated: September 26, 2000 Received: September 27, 2000
#### Dear Mr. Esmail:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the sured in the encroune) to ceal Device Amendments, or to devices that have been reclassified in enceenhence with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general thereols, mainto and the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations ( remaired in pp. of cound in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
{3}------------------------------------------------
#### Page 2 - Mr. Ehab M. Esmail
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). "Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597, or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Sincerely, yours,
Mark N Mullane
- Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Wright Medical Technology, Inc. The logo consists of a stylized "W" on the left, followed by the company name in a serif font. Below the company name is the address "5677 Airline Road, Arlington, TN 38002" and the phone number "001 867 0071".
### PROFEMUR R REVISION HIP SYSTEM
### INDICATIONS STATEMENT
### Indicated for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions:
- 1) non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, ankylosis, protrusio acetabuli, and painful hip dysplasia;
- 2) inflammatory degenerative joint disease such as rheumatoid arthritis;
- 3) correction of functional deformity;
- 4) revision procedures where other treatments or devices have failed; and,
- 5) treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
#### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of General Restorative Devices .510(k) Number
Prescription Use (Per21 CFR 801.109)
OR
Over-The Counter Use
(Optional Format 1-2-96)
for Mark N. Millhussen
(Division Sign Off)
(Division Sign-Off) Division of General Restorative Devices 510(k) Number.
Image /page/4/Picture/18 description: The image shows the text "KOO301" in what appears to be a handwritten style. Below the text is a logo that looks like a stylized letter "W" or an abstract geometric shape. The logo is in black and is positioned at the bottom center of the image.
Image /page/4/Picture/19 description: The image shows a globe with a focus on Asia and Australia. The globe is a simple, stylized representation, with the continents outlined in dark ink. The surrounding area of the globe is shaded, giving it a three-dimensional appearance. The image is a black and white graphic.
Image /page/4/Picture/20 description: The image shows the logo for Wright Medical Technology, Inc. The logo features a stylized "W" symbol above the word "WRIGHT" in a bold, serif typeface. Below "WRIGHT" are the words "MEDICAL TECHNOLOGY, INC." in a smaller, sans-serif font. The logo is simple and professional, conveying a sense of stability and innovation.
INDICATIONS STATEMENT
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.