K001080 · Aesculap, Inc. · KWP · Jun 29, 2000 · Orthopedic
Device Facts
Record ID
K001080
Device Name
SPINE SYSTEM EVOLUTION
Applicant
Aesculap, Inc.
Product Code
KWP · Orthopedic
Decision Date
Jun 29, 2000
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3050
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Spine System Evolution is a pedicle screw system indicated for the treatment of severe spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion. In addition, Spine System Evolution is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, spinal tumor, and failed previous fusion (pseudarthrosis). When used as a posterior, non-cervical, non-pedicle screw fixation system, Spine System Evolution is intended for hook fixation from T1 to the ilium / sacrum. The non-pedicle screw indications are degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (fracture or dislocation), spinal stenosis, deformities or curvatures (scoliosis, kyphosis and lordosis), tumor, pseudoarthrosis, and/or failed previous fusion.
Device Story
Spine System Evolution is a low-profile, top-loading posterior spinal fixation system; consists of single-use Ti6Al4V implants including pedicle screws, sacral plates, rods, rod connectors, linking plates, and hooks; utilizes conical nut connection for all components. Used by surgeons in clinical settings to build spinal constructs for stabilization and fusion in thoracic, lumbar, and sacral spine. Provides immobilization of spinal segments; intended for removal after solid fusion attainment. Benefits patients by stabilizing unstable or deformed spinal segments.
Clinical Evidence
No clinical data provided. Safety and efficacy established via bench testing in accordance with applicable ISO/ASTM standards.
Indicated for skeletally mature patients requiring spinal fusion for severe spondylolisthesis (Grades 3-4, L5-S1), degenerative spondylolisthesis with neurological impairment, fracture, dislocation, scoliosis, spinal tumor, pseudarthrosis, or failed previous fusion. Also indicated for degenerative disc disease, spinal stenosis, kyphosis, and lordosis when used as a non-pedicle screw fixation system (T1 to ilium/sacrum).
Regulatory Classification
Identification
A spinal interlaminal fixation orthosis is a device intended to be implanted made of an alloy, such as stainless steel, that consists of various hooks and a posteriorly placed compression or distraction rod. The device is implanted, usually across three adjacent vertebrae, to straighten and immobilize the spine to allow bone grafts to unite and fuse the vertebrae together. The device is used primarily in the treatment of scoliosis (a lateral curvature of the spine), but it also may be used in the treatment of fracture or dislocation of the spine, grades 3 and 4 of spondylolisthesis (a dislocation of the spinal column), and lower back syndrome.
{1}------------------------------------------------
Spine System® Evolution
# JUN 2 9 2000
Koo1080
Page 1 of 2
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
in Accordance with SMDA of 1990
### SPINE SYSTEM® EVOLUTION
April 3, 2000
- Aesculap®, Inc. COMPANY: 1000 Gateway Blvd. So. San Francisco, CA 94080
- Lia S. Jones, Regulatory Associate CONTACT: 650-624-5073 (phone) 650-589-3007 (fax) lia.jones@aesculap.com (email)
- Spine System® Evolution TRADE NAME:
- COMMON NAME: Posterior Spinal Fixation System
- DEVICE CLASS: Class II
- PRODUCT CODE(S): KWP, MNH, MNI
- CLASSIFICATION(S): 888.3050 Spinal Interlaminal Fixation Orthosis 888.3070 - Spondylolisthesis Spinal Fixation Device System 888.3070 - Pedicle Screw Spinal System
- Orthopedic Devices Branch REVIEW PANEL: Division of General and Restorative Devices
### DEVICE DESCRIPTION
Spine System Evolution is a multiple component system comprised of a variety of single-use, non-sterile implants (Ti6Al4V acc. to ISO 5832/3) that allow the surgeon to build a spinal construct in order to stabilize and promote fusion in the thoracic, lumbar and sacral spine. It is a low-profile, top-loading system that utilizes one connection (conical nut) for all implants, such as standard and polyaxial pedicle screws, sacral plates and screws, rods and rod connectors, linking plates, lateral connectors, and various hook styles and configurations.
### PURPOSE OF PREMARKET NOTIFICATION
This submission seeks to expand the current product line for Aesculap's Spine System Evolution (K980484, K982914) with new and modified components. There are no changes to the intended use, material composition or fundamental scientific technology.
{2}------------------------------------------------
Spine System® Evolution
K001080
Page 2 of 2
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
in Accordance with SMDA of 1990
### SPINE SYSTEM® EVOLUTION
April 3, 2000
#### INDICATIONS FOR USE
Spine System Evolution is a pedicle screw system indicated for the treatment of severe spondvlolisthesis (Grades 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion.
In addition. Spine System Evolution is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, spinal tumor, and failed previous fusion (pseudarthrosis).
When used as a posterior, non-cervical, non-pedicle screw fixation system, Spine System Evolution is intended for hook fixation from T1 to the ilium / sacrum. The non-pedicle screw indications are degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (fracture or dislocation), spinal stenosis, deformities or curvatures (scoliosis, kyphosis and lordosis), tumor, pseudoarthrosis, and/or failed previous fusion.
#### PERFORMANCE DATA
No performance standards have been promulgated under Section 514 of the Food, Drug and Cosmetic Act for this device system. The implants presented in this 510(k), however, were put through various test methods in accordance to applicable ISO / ASTM standards in order to establish their safety and efficacy.
#### SUBSTANTIAL EQUIVALENCE
The new and modified components described in this premarket notification are substantially equivalent in their intended use, material composition, labeling, design and basic operating principles to those in Aesculap's current Spine System Evolution (K980484, K982914), as well as the following competitor spinal systems:
- ISOBAR®, ISOLOCK® Spinal System (K992738, K990118, K990721) .
- ISOLA / VSP System (K984350) .
- Mirage™ Spinal System (K951846) .
- Xia™ Spinal System (K992792, K984251) .
{3}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, represented by three curved lines.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
## JUN 2 9 2000
Ms. Lia S. Jones Regulatory Associate Aesculap, Incorporated 1000 Gateway Boulevard South San Francisco, California 94080-7028
Re: K001080
Trade Name: Spinal System® Evolution Regulatory Class: II Product Code: KWP, MNH and MNI Dated: April 3, 2000 Received: April 4, 2000
Dear Ms. Jones:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
Page 2- Ms. Lia S. Jones
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its tollfree number (800) 638-2041 or at (301) 443-6597, or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Rene M. Jujan
Say Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Page 1 of 1
## INDICATIONS FOR USE STATEMENT
510(k) Number (if known):
K001080
Device Name:
Spine System® Evolution
Indication for Use:
Spine System Evolution is a pedicle screw system indicated for the treatment of Spine System Evolution 13 a pealor below by of the L5-S1 vertebra in skeletally mature
severe spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebra in skeletaly the the severe spondylolistiesis (Orados o and 1) of the staff having implants attached to the patients receiving fusion by autogenous bene gran handly of the implants after the attainment of a solid fusion.
In addition, Spine System Evolution is intended to provide immobilization and in addition, Opine Oystem Evelation stablifzation of spinal segments in one and chronic instabilities or deformities of the sticative in the treatment of the following about and onlylolisthesis with objective ന്തിവാറ്റ്, fumbal and Sacrar Spine: dogeno, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis).
When used as a posterior, non-cervical, non-pedicle screw fixation system, Spine When used as a postenor, non och fixation from T1 to the illium / sacrum. The System Evolution is Inchart for need maalth in the disease (defined as back pain non-peatcle screw indication of the disc confirmed by first confirmed by first and of useogenic studies), spondylolisthesis, trauma (fracture or dislocation), spinal radiographic studies), Spondylolistics, kyphosis and lordosis), tumor, pseudoarthrosis, and/or failed previous fusion.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Ninel Vang
(Division Sign-Off)
or Over-the-Counter Use
Division of Seral Restorative Devices 510(k) Numb:: 100 1080
(per 21 CFR 801.109)
(Optional Format 3-10-98)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.