The Apex Modular Hip Stem is intended for use as the femoral component of a primary total hip replacement. This femoral hip stem is intended for uncemented fixation and single use implantation. This prosthesis may be used for the following conditions, as appropriate: - Non-inflammatory degenerative joint disease, including osteoarthritis and . avascular necrosis: - . Rheumatoid arthritis: - Correction of functional deformity; . - Congenital dislocation: . - Revision procedures where other treatments or devices have failed; . - . Femoral neck and trochanteric fractures of the proximal femur.
Device Story
Apex Modular Hip Stem is a modular femoral component for uncemented total hip arthroplasty. System comprises three modular components: porous-coated femoral stem, modular neck, and modular head. Surgeon selects combination of stem, neck, and head sizes to match patient anatomy; modularity allows adjustment of length and lateral offset. Device implanted by orthopedic surgeons in clinical/hospital settings. Titanium alloy stem/neck and cobalt chromium alloy head provide structural support for hip joint. Assembly testing of neck-stem connection and fatigue testing performed to ensure mechanical integrity. Device intended to restore hip function and mobility in patients with degenerative joint disease, arthritis, or fractures.
Clinical Evidence
Bench testing only. No clinical data provided. Performance testing included assembly testing of the neck-stem modular connection and fatigue testing of the assembled device. Coating performance testing was completed per FDA guidance.
Technological Characteristics
Modular femoral hip prosthesis. Materials: Titanium alloy (Ti6Al4V) stem and neck; wrought cobalt chromium alloy head; unalloyed titanium plasma spray coating on proximal metaphyseal region. Design features: modular neck, straight fluted distal stem, distal slots, proximal steps. Uncemented fixation. No software or electronic components.
Indications for Use
Indicated for patients requiring primary total hip replacement or revision procedures due to non-inflammatory degenerative joint disease (osteoarthritis, avascular necrosis), rheumatoid arthritis, functional deformity, congenital dislocation, or femoral neck/trochanteric fractures.
Regulatory Classification
Identification
A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.
{0}------------------------------------------------
510(K) Summary K000788
## Apex Modular™ Hip Stem
May 5, 2000
1. Submitter: Apex Surgical™, LLC 39 Taunton Green Suite 202 Taunton, MA 02780
Contact: Edward J. Cheal, Ph.D. Managing Director (508) 884-9611 (voice) (508) 884-9816 (fax)
#### 2. Device Name
Proprietary Name: Apex Modular™ Hip Stem Common Name: Hip prosthesis, uncemented Classification Name: Hip joint metal/polymer/metal semi-constrained porouscoated uncemented prosthesis Regulatory Class: Class II per 21 CFR §888.3358
#### 3. Intended Use
The Apex Modular Hip Stem is intended for use as the femoral component of a primary total hip replacement. This femoral hip stem is intended for uncemented fixation and single use implantation. This prosthesis may be used for the following conditions, as appropriate:
- Non-inflammatory degenerative joint disease, including osteoarthritis and . avascular necrosis:
- . Rheumatoid arthritis:
- Correction of functional deformity; .
- Congenital dislocation: .
- Revision procedures where other treatments or devices have failed; .
- . Femoral neck and trochanteric fractures of the proximal femur.
#### 4. Device Description
The Apex Modular Hip Stem consists of three modular components, with various sizes available for each component: the porous coated femoral stem, a modular neck that connects to the proximal end of the femoral stem, and a modular head that connects to the tapered trunion on the neck. This configuration allows the user to choose a combination of stem, neck, and head components to appropriately fit the anatomy of the patient. The various neck sizes allow for several length and lateral offset options for a given stem size. The modular heads are available in standard diameters so as to fit appropriate commercially available acetabular components of the surgeons' choosing. Several offset options are also available for the heads to allow further refinement of the lengths and offsets. The Apex Modular Hip Stem may be used in conjunction with the Link® SPII® Acetabular Cup for total hip arthroplasty.
The femoral stem and neck components are manufactured from titanium alloy, and the head components are manufactured from wrought cobalt chromium alloy, the same materials used to manufacture the predicate hip stems and heads, respectively. The proximal metaphyseal region of each size femoral stem is circumferentially coated with
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for Apex Surgical. The logo consists of a stylized letter A on the left, followed by the words "Apex Surgical" in a bold, sans-serif font. The letters are all black, and there is a small trademark symbol to the right of the word "Surgical".
unalloyed titanium applied by plasma spray. The alignment pin in the femoral stem is manufactured from wrought cobalt chromium alloy.
### 5. Predicate Device Comparison
Substantial equivalence is claimed to the S-ROM® Femoral Stem and the Biomet Modular Hip System. The table below compares the features and characteristics of the Apex Modular Hip Stem to these predicate devices.
| | Apex Modular<br>Hip Stem | S-ROM®<br>Femoral Stem<br>(K913231, K934412,<br>and K954935) | Biomet Modular<br>Hip System<br>(K921274) |
|---------------------------------------------------------------|--------------------------|--------------------------------------------------------------|-------------------------------------------|
| INTENDED USE | | | |
| Primary and revision hip<br>replacement, non-<br>cemented use | Yes | Yes | Yes |
| DESIGN | | | |
| Plasma spray coating | Yes | No (sintered<br>beads) | Yes |
| Proximal coating (only) | Yes | Yes | Yes |
| Modular head | Yes | Yes | Yes |
| Modular stem | Yes - modular<br>neck | Yes - modular<br>sleeve | Yes - modular<br>distal stem |
| Straight distal stem | Yes | Yes | Yes |
| Fluted stem | Yes | Yes | Yes |
| Distal slot(s) | Yes - 1 or 2<br>slots | Yes - 1 slot | Yes - 1 slot |
| Proximal steps | Yes | Yes | No |
| MATERIALS | | | |
| Titanium alloy (Ti6Al4V)<br>stem and neck | Yes | Yes | Yes |
| Cobalt chromium alloy<br>head | Yes | Yes | Yes |
| Titanium porous coating | Yes - unalloyed | Yes - unalloyed | Yes - Ti6Al4V |
The two most significant differences between the Apex Modular hip stem and the two predicate hip stems relate to the design of the stem modularity (modular neck versus modular stem and sleeve for the S-ROM® or modular distal stem for the Biomet Modular) and the porous coating (plasma sprayed CP titanium versus sintered bead CP titanium for the S-ROM® or plasma sprayed titanium alloy for the Biomet Modular). Performance testing of the modular stem has been completed as per the relevant FDA guidance documents, including assembly testing of the neck-stem modular connection and fatigue testing of the assembled device. Performance testing of the plasma sprayed unalloyed (CP) titanium coating was completed as per the relevant FDA guidance documents with data in the referenced Device Master File.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three stripes forming its body and wings. The eagle is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
4 2000 AUG
Edward J. Cheal, Ph.D. Managing Director Apex Surgical, LLC 39 Taunton Green Suite 202 Taunton, Massachusetts 02780
Re: K000788
Trade Name: Apex Modular™ Hip Stem Regulatory Class: II Product Code: LPH Dated: May 11, 2000 Received: May 12, 2000
Dear Dr. Cheal:
We have reviewed your Section 510(k) notification of intent to market the device referenced we have reviewed your bected. B r (x) xe is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the stated in the encreater) to teal Device Amendments, or to devices that have been reclassified in chactifient unto of the Federal Food, Drug, and Cosmetic Act (Act): You may, therefore, market the device, subject to the general control provisions of the Act. The general thereolo, incribed of the Act include requirements for annual registration, listing of devices, condor provisions of the or the , labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations (1 remarket ripproval), in the Code of Federal Regulations, Title 21, Parts 800 to 895. arretmig Jour as a revealent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to mopolitish the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
{3}------------------------------------------------
Page 2 - Edward J. Cheal, Ph.D.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597, or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Donna R. Lochner.
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications For Use Device Name: Apex Modular™ Hip Stem K 000788
The Apex Modular™ Hip Stem is intended for use as the femoral component of a The Apcx Moudlar - Tip Stom to infernoral hip stem is intended for uncemented fixation primary total hip replacement. This forthering in a prosthesis may be used for the following conditions, as appropriate:
- Non-inflammatory degenerative joint disease, including osteoarthritis and . avascular necrosis;
- Rheumatoid arthritis; ●
- Correction of functional deformity; �
- Congenital dislocation; .
- Revision procedures where other treatments or devices have failed; .
- Femoral neck and trochanteric fractures of the proximal femur.
(Please do not write below this line - Continue on another page if necessary) Concurrence of CDRH, Office of Device Evaluation (ODE)
Donna R. Vochner
510(k) Number
Prescription Use_ X (Per 21 CFR §801.109) OR
Over-the-Counter Use
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.