DEN250057 · Cardiosense, Inc. · SIF · May 22, 2026 · Cardiovascular
Device Facts
Record ID
DEN250057
Device Name
PCWP Analysis Software
Applicant
Cardiosense, Inc.
Product Code
SIF · Cardiovascular
Decision Date
May 22, 2026
Decision
DENG
Submission Type
Direct
Regulation
21 CFR 870.1150
Device Class
Class 2
Attributes
AI/ML, PCCP
Indications for Use
PCWP Analysis Software is intended to noninvasively estimate pulmonary capillary wedge pressure (PCWP) and identify patients with a PCWP above or below 18mmHg, for patients diagnosed with heart failure with reduced ejection fraction (HFrEF) and New York Heart Association (NYHA) Functional Class II, III, IV symptoms. The hemodynamic data will be used by qualified healthcare professionals in conjunction with other standard of care parameters to evaluate heart failure patients. The results of PCWP Analysis Software are intended to be used under the supervision of a qualified healthcare professional in conjunction with the patient's clinical history, symptoms, and other diagnostic tests, as well as the qualified healthcare professional's judgment. Patient management decisions should not be made solely on the results of the PCWP Analysis Software. The device is only for prescription use.
Device Story
PCWP Analysis Software is a prescription-only software device that noninvasively estimates pulmonary capillary wedge pressure (PCWP) to assist in heart failure management. It processes non-invasive physiologic signals to provide hemodynamic estimates; these outputs are intended for adjunctive use by qualified healthcare professionals alongside clinical history, symptoms, and standard diagnostic tests. The device is not intended to independently direct therapy or estimate systemic vascular blood pressure. It is designed to identify patients with PCWP above or below 18mmHg. By providing noninvasive hemodynamic insights, the device supports clinical decision-making for patients with HFrEF (NYHA Class II-IV). The system includes a Predetermined Change Control Plan (PCCP) for future software modifications.
Clinical Evidence
Clinical performance testing is required as a special control. The device must demonstrate agreement with reference measures across the clinically relevant physiological range using performance metrics with 95% confidence intervals. Data must be representative of the intended use population, including demographic subgroups (race, ethnicity, sex, age) and relevant clinical characteristics. The study must report rates of indeterminate output and demonstrate consistency across data sources and quality levels.
Technological Characteristics
Software-based system for estimation of cardiac pressures. Utilizes software algorithms to analyze non-invasive physiologic signals. Requires software verification, validation, and hazard analysis. Includes specifications for incoming sensor data quality control and characterization of hardware acquisition impacts. Subject to special controls regarding clinical performance, usability, and postmarket monitoring.
Indications for Use
Indicated for patients diagnosed with heart failure with reduced ejection fraction (HFrEF) and NYHA Functional Class II, III, or IV symptoms to noninvasively estimate pulmonary capillary wedge pressure (PCWP) and identify PCWP status relative to 18mmHg threshold.
Regulatory Classification
Identification
PCWP Analysis Software is a prescription-only device intended to noninvasively estimate pulmonary capillary wedge pressure (PCWP) and identify patients with a PCWP above or below 18mmHg, for patients diagnosed with heart failure with reduced ejection fraction (HFrEF) and New York Heart Association (NYHA) Functional Class II, III, IV symptoms. The device uses software algorithms to analyze one or more non-invasive physiologic signals or parameters to estimate cardiac pressure. It is intended for adjunctive use by qualified healthcare professionals in conjunction with other standard of care parameters and is not intended to independently direct therapy or estimate systemic, vascular blood pressure.
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 22, 2026
Cardiosense, Inc.
Arezou Azar
Chief Regulatory and Compliance Officer
400 N Aberdeen St.
Suite 1050
Chicago, Illinois 60642
Re: DEN250057
Trade/Device Name: PCWP Analysis Software
Regulation Number: 21 CFR 870.1150
Regulation Name: Software device system for estimation of cardiac pressures
Regulatory Class: Class II
Product Code: SIF
Dated: November 17, 2025
Received: November 17, 2025
Dear Arezou Azar:
The Center for Devices and Radiological Health (CDRH) of the Food and Drug Administration (FDA) has completed its review of your De Novo request for classification of the PCWP Analysis Software, a prescription device under 21 CFR Part 801.109 with the following indications for use:
PCWP Analysis Software is intended to noninvasively estimate pulmonary capillary wedge pressure (PCWP) and identify patients with a PCWP above or below 18mmHg, for patients diagnosed with heart failure with reduced ejection fraction (HFrEF) and New York Heart Association (NYHA) Functional Class II, III, IV symptoms. The hemodynamic data will be used by qualified healthcare professionals in conjunction with other standard of care parameters to evaluate heart failure patients.
The results of PCWP Analysis Software are intended to be used under the supervision of a qualified healthcare professional in conjunction with the patient's clinical history, symptoms, and other diagnostic tests, as well as the qualified healthcare professional's judgment. Patient management decisions should not be made solely on the results of the PCWP Analysis Software. The device is only for prescription use.
FDA concludes that this device should be classified into Class II. This order, therefore, classifies the PCWP Analysis Software, and substantially equivalent devices of this generic type, into Class II under the generic name software device system for estimation of cardiac pressures.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
DEN250057 - Arezou Azar
Page 2
FDA identifies this generic type of device as:
**Software device system for estimation of cardiac pressures.** A software device system for estimation of cardiac pressures is a prescription-only device that uses software algorithms to analyze one or more non-invasive physiologic signals or parameters to estimate a cardiac pressure (e.g. pulmonary capillary wedge pressure, left ventricular end diastolic pressure, or pulmonary artery pressure). The device system may contain hardware sensors. This device system is intended for adjunctive use with patient information and is not intended to independently direct therapy. This device system is not intended to estimate systemic, vascular blood pressure.
Section 513(f)(2) of the Food, Drug and Cosmetic Act (the FD&C Act) was amended by section 607 of the Food and Drug Administration Safety and Innovation Act (FDASIA) on July 9, 2012. This law provides two options for De Novo classification. First, any person who receives a "not substantially equivalent" (NSE) determination in response to a 510(k) for a device that has not been previously classified under the Act may request FDA to make a risk-based classification of the device under section 513(a)(1) of the Act. On December 13, 2016, the 21st Century Cures Act removed a requirement that a De Novo request be submitted within 30 days of receiving an NSE determination. Alternatively, any person who determines that there is no legally marketed device upon which to base a determination of substantial equivalence may request FDA to make a risk-based classification of the device under section 513(a)(1) of the Act without first submitting a 510(k). FDA shall, within 120 days of receiving such a request, classify the device. This classification shall be the initial classification of the device. Within 30 days after the issuance of an order classifying the device, FDA must publish a notice in the Federal Register announcing the classification.
On November 17, 2025, FDA received your De Novo requesting classification of the PCWP Analysis Software. The request was submitted under section 513(f)(2) of the FD&C Act. In order to classify the PCWP Analysis Software into class I or II, it is necessary that the proposed class have sufficient regulatory controls to provide reasonable assurance of the safety and effectiveness of the device for its intended use. After review of the information submitted in the De Novo request, FDA has determined that, for the previously stated indications for use, the PCWP Analysis Software can be classified in class II with the establishment of special controls for class II. FDA believes that class II (special) controls provide reasonable assurance of the safety and effectiveness of the device type. The identified risks and mitigation measures associated with the device type are summarized in the following table:
| Risks to Health | Mitigation Measures |
| --- | --- |
| Delayed or incorrect treatment due to user misinterpretation or overreliance | Usability assessment Labeling |
| Inaccurate pressure estimation leading to incorrect treatment or diagnosis | Clinical performance testing Software verification, validation, and hazard analysis Labeling Postmarket monitoring |
| Delayed or incorrect treatment due to no output, indeterminate output, or erroneous output as a result of software malfunction or algorithm processing error | Clinical performance testing Software verification, validation, and hazard analysis Labeling Postmarket monitoring |
{2}
DEN250057 - Arezou Azar
Page 3
In combination with the general controls of the FD&C Act, the software device system for estimation of cardiac pressures is subject to the following special controls:
(1) Clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following must be met:
(i) Agreement of the measure(s) with the reference measure(s) must be assessed across the clinically relevant physiological range as well as the full physiological range using performance metrics relevant to the device output. All performance metrics must include 95% confidence intervals;
(ii) The data set must be representative of the intended use population for the device. Any selection criteria or limitations of the samples must be fully described and justified; The study must include representation across key demographic subgroups (race, ethnicity, sex, age) and relevant clinical characteristics; and
(iii) Data must be provided within the clinical study or equivalent datasets to demonstrate the consistency of the output and representativeness of the range of data sources and data quality likely to be encountered in the intended use population and relevant use conditions in the intended use environment. The study must report any rate(s) of indeterminate output across demographic groups.
(2) Software verification, validation, and hazard analysis must be performed. Software documentation must include:
(i) A full characterization of technical parameters of the software, including any algorithms or models used, all inputs and outputs for the software, and the supported patient population;
(ii) Description of the expected impact of all applicable sensor acquisition hardware characteristics on performance and any associated hardware specifications;
(iii) Specification of acceptable incoming sensor data quality control measures; and
(iv) Data documentation describing any training, tuning, or validation datasets used in algorithm development.
(3) Usability assessment must be provided to demonstrate that risk of misinterpretation of the device output is appropriately mitigated.
(4) Labeling must include the following:
(i) A summary of the clinical performance testing with the device, including:
(A) Objective performance measures reported with 95% confidence intervals; and
(B) A description of the patient population studied (including age, sex, race or ethnicity, relevant clinical conditions and comorbidities).
(ii) Information for interpretation of the output must include:
(A) A description of what the device measures and outputs to the user;
(B) Warnings identifying sensor reading acquisition factors that may impact measurement results; and
(C) A warning(s) that the device output is only for adjunctive use.
(5) The device manufacturer must develop and implement a postmarket performance management plan that ensures regular assessment of the generalizability and device performance in the intended patient population in real-world use. The plan must include:
(i) Data collection, analysis methods, and procedures for:
{3}
DEN250057 - Arezou Azar
Page 4
(A) Monitoring relevant performance characteristics and detecting changes in performance;
(B) Identifying sources of performance changes between validation and real-world environment over time; and
(C) Assessing the results from the performance testing on safety and effectiveness;
(ii) Procedures for communicating the device’s current performance to the users.
In addition, this is a prescription device and must comply with 21 CFR 801.109.
Although this letter refers to your product as a device, please be aware that some granted products may instead be combination products. If you have questions on whether your product is a combination product, contact CDRHProductJurisdiction@fda.hhs.gov.
FDA's granting of your De Novo request also included the review and authorization of your Predetermined Change Control Plan (PCCP) titled “Predetermined Change Control Plan (PCCP) for PCWP Analysis Software Rev 1”. Under 21 CFR 807.81(a)(3), a new premarket notification is required if there is a major change or modification in the intended use of a device, or if there is a change or modification in a device that could significantly affect the safety or effectiveness of the device, e.g., a significant change or modification in design, material, chemical composition, energy source, or manufacturing process. Accordingly, if deviations from the authorized PCCP result in a major change or modification in the intended use of the device, or result in a change or modification in the device that could significantly affect the safety or effectiveness of the device, then a new premarket notification would be required consistent with 21 CFR 807.81(a)(3). Failure to submit such a premarket submission would constitute adulteration and misbranding under sections 501(f)(1)(B) and 502(o) of the Act, respectively.
Section 510(m) of the FD&C Act provides that FDA may exempt a class II device from the premarket notification requirements under section 510(k) of the FD&C Act, if FDA determines that premarket notification is not necessary to provide reasonable assurance of the safety and effectiveness of the device type. FDA has determined premarket notification is necessary to provide reasonable assurance of the safety and effectiveness of the device type and, therefore, the device is not exempt from the premarket notification requirements of the FD&C Act. Thus, persons who intend to market this device type must submit a premarket notification containing information on the software device system for estimation of cardiac pressure they intend to market prior to marketing the device.
Please be advised that FDA's decision to grant this De Novo request does not mean that FDA has made a determination that your device complies with other requirements of the FD&C Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the FD&C Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and if applicable, the electronic product radiation control provisions (Sections 531-542 of the FD&C Act; 21 CFR 1000-1050).
{4}
DEN250057 - Arezou Azar
Page 5
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System Rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
A notice announcing this classification order will be published in the Federal Register. A copy of this order and supporting documentation are on file in the Dockets Management Branch (HFA-305), Food and Drug Administration, 5630 Fishers Lane, Room 1061, Rockville, MD 20852 and are available for inspection between 9 a.m. and 4 p.m., Monday through Friday.
As a result of this order, you may immediately market your device as described in the De Novo request, subject to the general control provisions of the FD&C Act and the special controls identified in this order.
For comprehensive regulatory information about medical devices and radiation-emitting products, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
If you have any questions concerning the contents of the letter, please contact Monica Okon at Monica.Okon@fda.hhs.gov.
Sincerely,
HETAL B. ODOBASIC -S
Hetal Odobasic
Director
Division of Cardiac Electrophysiology,
Diagnostics, and Monitoring Devices
Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.