DEN250016 · Essilor Of America, Inc. · QUR · Sep 25, 2025 · Ophthalmic
Device Facts
Record ID
DEN250016
Device Name
Essilor® Stellest®
Applicant
Essilor Of America, Inc.
Product Code
QUR · Ophthalmic
Decision Date
Sep 25, 2025
Decision
DENG
Submission Type
Direct
Regulation
21 CFR 886.5845
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence, Pediatric
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
DEN250016 · Sep 25, 2025
Essilor® Stellest®
Essilor Of America, Inc.
Published literature on myopia rebound effects
The sponsor provided a summary of existing literature regarding rebound effects to supplement the clinical trial data and assess the risk of accelerated myopia progression following treatment discontinuation.
myopia rebound; literature review; post-treatment safety
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Summary of rebound literature; Literature review
Pediatric patients with myopia
Not applicable for this study
Myopia progression after cessation of treatment
Indications for Use
The Essilor® Stellest® lens is indicated for the correction of myopia with and without astigmatism and for slowing the progression of myopia in children with non-diseased eyes, who, at the initiation of treatment, are aged 6-12 years and have spherical equivalent refraction of -0.75 D to -4.50 D with astigmatism up to 1.50 D.
Device Story
Prescription spectacle lenses featuring concentric rings of aspheric lenslets; lenslets focus light in front of retina to create peripheral myopic defocus; central 9mm zone remains clear for refractive correction; 11 concentric rings of varying power; polycarbonate material with anti-reflective coating; mounted in standard spectacle frames; worn by pediatric patients; prescribed/fitted by eye care professionals; slows eye elongation; reduces myopia progression rate; benefits include potential prevention of high-myopia-related sight-threatening complications; lower risk profile compared to contact lenses.
Clinical Evidence
Pivotal multicenter, randomized, controlled, double-masked trial (FIN-3101) with 149 subjects (6-12 years). Primary endpoints: change in cycloplegic spherical equivalent refraction (cSER) and axial length at 24 months. Results: 71% reduction in myopia progression (0.64 D difference, 95% CI: 0.50-0.79 D) and 53% reduction in axial elongation (-0.24 mm difference, 95% CI: -0.29 to -0.19 mm) compared to single-vision control. Safety profile acceptable; no significant lens-related adverse events; BCVA stable.
Technological Characteristics
Prescription spectacle lenses featuring additional physical optical design elements. Requires impact resistance testing per 21 CFR 801.410. Optical characterization, durability, and performance testing required to verify technical specifications. Lenses are mounted in spectacle frames.
Indications for Use
Indicated for correction of myopia (with/without astigmatism) and slowing myopia progression in children aged 6-12 years with non-diseased eyes; spherical equivalent refraction -0.75 D to -4.50 D; astigmatism up to 1.50 D.
Regulatory Classification
Identification
Prescription spectacle lenses to reduce the progression of myopia consist of spectacle lenses with additional physical optical design lens elements. In addition to optical correction of myopic refractive error, these lenses are intended to be used by patients who have myopia to reduce the rate of myopia progression. The Essilor® Stellest® features concentric rings of aspheric lenslets (small lenses) which focus portions of the incoming light in front of the retina, causing blur on the retina, and is indicated for the correction of myopia with and without astigmatism and for slowing the progression of myopia in children with non-diseased eyes, who, at the initiation of treatment, are aged 6-12 years and have spherical equivalent refraction of -0.75 D to -4.50 D with astigmatism up to 1.50 D.
Special Controls
In combination with the general controls of the FD&C Act, the prescription spectacle lenses to reduce the progression of myopia are subject to the following special controls:
Submission Summary (Full Text)
{1}
DE NOVO CLASSIFICATION REQUEST FOR
ESSILOR® STELLEST®
REGULATORY INFORMATION
FDA identifies this generic type of device as:
Prescription spectacle lenses to reduce the progression of myopia. Prescription spectacle lenses to reduce the progression of myopia consist of spectacle lenses with additional physical optical design lens elements. In addition to optical correction of myopic refractive error, these lenses are intended to be used by patients who have myopia to reduce the rate of myopia progression. The lenses are mounted within a spectacle frame classified under § 886.5842.
NEW REGULATION NUMBER: 21 CFR 886.5845
CLASSIFICATION: Class II
PRODUCT CODE: QUR
BACKGROUND
DEVICE NAME: Essilor® Stellest®
SUBMISSION NUMBER: DEN250016
DATE DE NOVO RECEIVED: April 28, 2025
SPONSOR INFORMATION:
Essilor of America, Inc.
13455 Branchview Lane
Dallas, Texas 75234
INDICATIONS FOR USE
The Essilor® Stellest® is indicated as follows:
The Essilor® Stellest® lens is indicated for the correction of myopia with and without astigmatism and for slowing the progression of myopia in children with non-diseased eyes, who, at the initiation of treatment, are aged 6-12 years and have spherical equivalent refraction of -0.75 D to -4.50 D with astigmatism up to 1.50 D.
LIMITATIONS
{2}
The sale, distribution, and use of the Essilor® Stellest® spectacle lens is restricted to prescription use in accordance with 21 CFR 801.109.
The following warnings and precautions apply to the Essilor® Stellest® spectacle lenses:
**Warnings:**
As spectacles are manufactured based on a precise prescription, the eye care professional should ensure they are correctly fitted and appropriately assessed on the patient.
The Essilor® Stellest® spectacle lens, like any ophthalmic lens, may require an initial adaptation period, typically under one week. During this period, it is advisable to avoid high-impact activities where altered vision could present a risk. If adaptation takes longer or causes significant issues, it is recommended the wearer consults their eye care professional.
The lenses used in the FIN-3101 clinical trial were equipped with a Crizal Easy Pro coating, which does not provide any tint or filtering capabilities.
Patients should be advised of the following warnings pertaining to Essilor® Stellest® spectacle lenses:
- The effectiveness of the lens to slow myopia progression was not studied with any tints or filters. Speak to your doctor before adding any tint or filters to the lenses, as it is possible these may impact device effectiveness.
**Precautions:**
At this point, potential rebound effects (i.e., accelerated myopia progression following discontinuation of Essilor® Stellest® spectacle lens wear compared to single vision lenses) have not been evaluated in U.S. patients.
The effects of Essilor® Stellest® spectacle lenses on reading skill development and on peripheral visual function—including obstacle detection in the periphery—are unknown.
**DEVICE DESCRIPTION**
The Essilor® Stellest® is a prescription only (Rx only) device that is a pair of spectacle lenses fitted with the patient's refractive prescription. The unique features of this device are concentric rings of aspheric lenslets (small lenses) which focus portions of the incoming light in front of the retina, causing blur on the retina. The figure below shows a single lens where the 9mm diameter central zone is free from lenslets and 11 concentric lenslet rings are visible. Each ring has a slightly different focusing power. The device is designed to slow the elongation of the eye and a provide a reduction in the progression of myopia in myopic children.
De Novo Summary (DEN250016)
Page 2 of 16
{3}

Figure 1: A single Essilor® Stellest® spectacle lens is pictured above, showing where the central zone is free from lenslets and 11 concentric lenslet rings are visible.
The Essilor® Stellest® spectacle lenses are intended to be worn by a pediatric patient in a spectacle frame, as eyeglasses are worn. There is no specific frame that comes with the Essilor® Stellest® lenses. Rather, the Essilor® Stellest® lenses are mounted into a chosen spectacle frame, based on patient preference in conjunction with their healthcare provider or optician. The Essilor® Stellest® lenses are made from polycarbonate with Crizal® Easy Pro coating, an anti-reflectance coating, and are not tinted. Labeling for the Essilor® Stellest® device is provided to both the healthcare provider/optician and the pediatric patient/caregiver(s).
One example of Essilor® Stellest® spectacle lenses, mounted into a spectacle frame as eyeglasses are, is shown below.
De Novo Summary (DEN250016)
{4}

Figure 2: Photograph of an example spectacle frame (eyeglasses) that contain Essilor® Stellest® spectacle lenses
# SUMMARY OF NONCLINICAL/BENCH STUDIES
# PERFORMANCE TESTING - BENCH
Non-clinical testing for the Essilor® Stellest® spectacle lenses included the following:
Optical power
Spectral transmittance
- Adhesion/reflectance testing
- Abrasion testing
Prescription Ophthalmic Lenses Recommendations
- Requirements for uncut finished lenses
- Impact resistance testing
Aging simulation test
Lenslet dimensional testing
Lenslet power testing
Spectral transmittance
Lenslet dimensional testing
Lenslet power testing
- Abrasion testing
De Novo Summary (DEN250016)
{5}
- Adhesion/reflectance testing
- Impact resistance testing
Prescription Ophthalmic Lenses Recommendations
- Requirements for uncut finished lenses
- Aging simulation test
- Wearer simulation at the hands of the wearer
Table 1 below provides a detailed summary of the non-clinical bench and optical testing performed for the Essilor® Stellest® spectacle lenses.
# PERFORMANCE TESTING - BENCH
Table 1: Summary of Performance Testing - Bench Studies
| Test | Purpose | Method | Acceptance Criteria | Results |
| --- | --- | --- | --- | --- |
| Optical Power | Verify that the device meets the ISO 8980-1 specifications. | Based on ISO 8980-1: Ophthalmic optics – Uncut finished spectacle lenses – Part 1: Specifications for single-vision and multifocal lenses | ISO 8980-1:2017: Ophthalmic optics – Uncut finished spectacle lenses – Part 1: Specifications for single-vision and multifocal lenses | Passed |
| Spectral Transmittance | Evaluate the spectral transmittance of the device. | Based on ISO 8980-3: 2022 Ophthalmic optics — Uncut finished spectacle lenses — Part 3: Transmittance specifications and test methods. | ISO 8980-3: 2022 Ophthalmic optics — Uncut finished spectacle lenses — Part 3: Transmittance specifications and test methods. | Passed |
| Adhesion/Reflectance | Evaluate the device for reflectance and adhesion | Based on ISO 8980-4: 2006 Ophthalmic optics — Uncut finished spectacle lenses — Specifications and test methods for anti-reflective coatings | ISO 8980-4: 2006 Ophthalmic optics — Uncut finished spectacle lenses — Specifications and test methods for anti-reflective coatings | Passed |
| Abrasion | Evaluate the resistance of the lens surface to abrasion damage during normal use | Based on ISO 8980-5: 2005 Ophthalmic optics — Uncut finished spectacle lenses - Minimum requirements for spectacle lens surfaces claimed to be abrasion-resistant | ISO 8980-5: 2005 Ophthalmic optics — Uncut finished spectacle lenses — Minimum requirements for spectacle lens surfaces claimed to be abrasion-resistant | Passed |
| Prescription Ophthalmic Lenses Recommendations | Verify that the device conforms to the ANSI Z80-1. | Based on ANSI Z80-1: 2020 - for Ophthalmics – Prescription | ANSI Z80-1: 2020 - for Ophthalmics – Prescription | Passed |
De Novo Summary (DEN250016)
{6}
De Novo Summary (DEN250016)
Page 6 of 16
| | | Ophthalmic Lenses – Recommendations | Ophthalmic Lenses – Recommendations | |
| --- | --- | --- | --- | --- |
| Requirements for uncut finished lenses | Verify that the device meets the requirements for uncut lenses | Based on ISO 14889:2013 Ophthalmic optics — Spectacle lenses — Fundamental requirements for uncut finished lenses | ISO 14889:2013 Ophthalmic optics — Spectacle lenses — Fundamental requirements for uncut finished lenses | Passed |
| Impact Resistance | Verify the ability of the device not to break and injure patient eye and/or skin when put under standard usage impact | Based on 21 CFR 801.410 | 21 CFR 801.410 | Passed |
| Lenslet dimensional testing | Verify that the lenslets rings are the correct size and contain the correct number of lenslets | Optical testing performed by deflectometry was conducted where the lens is located between a screen and a camera. A sinusoidal grid pattern is displayed on the screen. The image of this pattern through the lens is acquired by the camera and processed to provide power and geometrical measurements. | Each ring has the correct number of lenslets and the distance from the center of the optic to each lenslet is within ±0.3 mm of the specification for that ring. | Passed |
| Lenslet power testing | Verify that the lenslet optical performance is as specified | Optical testing performed by deflectometry was conducted where the lens is located between a screen and a camera. A sinusoidal grid pattern is displayed on the screen. The image of this pattern through the lens is acquired by the camera and processed to provide power and geometrical measurements. | Average lenslet power around ring was within -0.5/+1.0 D of specification; Standard deviation around ring was ±0.25 D; max-min power within a ring was ≤ 0.65 D for rings 1-8 and ≤ 0.75 D for rings 9-11 | Passed |
| Aging simulation test | Evaluate optics of the micro-lens and lens of the finished lens, Anti-reflective (AR) adhesion, AR transmittance, and impact resistance after simulated 3 years of storage of uncut lens | Optics of the finished lens: ISO 8980-1 Optics of the lenslet: See “Lenslet power testing” row above. Lenslet design: See “Lenslet dimensional testing” row above. | Optics of the finished lens: ISO 8980-1 Optics of the lenslet: See “Lenslet power testing” row above. Lenslet design: See “Lenslet dimensional testing” row above. | Passed |
{7}
| | prior to fabricating the finished product (glasses) | AR adhesion and transmittance: ISO 8980-4
Impact resistance: 21 CFR 801.410 | AR adhesion and transmittance: ISO 8980-4
Impact resistance: 21 CFR 801.410 | |
| --- | --- | --- | --- | --- |
| Wearer simulation at the hands of the wearer | Evaluate the optical power of the lens and lenslets of the finished lens after simulated 2 years of wear. | Optical power of the lens and lenslets after Q-Sun exposure.
Lens optical power: focimeter.
Lenslet optical power: See “Lenslet power testing” row above. | Optical power of the lens: ISO 8980-1 and Z80.1
Optical power of the lenslet: See “Lenslet power testing” row above. | Passed |
# BIOCOMPATIBILITY
The Essilor® Stellest® spectacle lenses have no tissue contact when worn, as they are mounted within spectacle frames and are worn as eyeglasses. Spectacle frames used with Essilor® Stellest® lenses are class I (510(k) exempt) under 21 CFR 886.5842, product code HQZ. Because the lenses themselves have no tissue contact with the patient when worn and no additional biocompatibility evaluation is needed for class I spectacle frames, biocompatibility testing was not needed to support the safety of the subject device.
However, cytotoxicity testing per ISO 10993-5:2009 was performed on the Essilor® Stellest® spectacle lenses. Results support that the Essilor® Stellest® lenses were not cytotoxic.
# SUMMARY OF CLINICAL INFORMATION
There were two studies conducted to evaluate the safety and effectiveness of the Essilor® Stellest® spectacle lenses. The FIN-3101 study evaluated the safety and effectiveness of the device as compared to a single vision control lens. The FIN-3102 study, a sub-study of the FIN-3101 study, evaluated visual performance through a variety of clinical tests and evaluations. An assessment was also provided that assessed the potential for rebound (an increase in myopia progression after discontinuing treatment) based on currently available information for the device.
# FIN-3101 Clinical Study:
# Overview:
Clinical performance data were collected from a pivotal clinical study (the FIN-3101 study).
The FIN-3101 study was a multicenter, randomized, controlled, prospective, two-arm, parallel-group, double-masked clinical trial. The study was conducted at 9 investigational sites in the U.S. The primary objective was to evaluate the effectiveness of the Essilor® Stellest® spectacle lens in slowing myopia progression compared to single vision control lenses. The study enrolled
De Novo Summary (DEN250016)
{8}
175 participants between the ages of 6-12 years at initial fitting. Participants were consented for up to 3 years.
Key inclusion criteria included the following:
- Participants between 6 and 12 years of age;
- spherical equivalent refraction by manifest refraction between -0.75 and -4.50 diopters in each eye;
- astigmatism, if present, not exceeding 1.50 diopters;
- the difference in spherical equivalent refraction between the two eyes (anisometropia) by manifest refraction not exceeding 1.00 diopter;
- best corrected visual acuity in each eye equal to or better than $+0.10$ logMAR (equivalent to 20/25 Snellen).
Key exclusion criteria included the following:
- any history of myopia control interventions such as atropine, orthokeratology, or multifocal contact lenses;
- strabismus by cover test at near or distance while wearing correction;
- amblyopia;
- any ocular or systemic condition known to affect refractive status;
- current use of systemic or topical medications known to significantly affect pupil size, accommodation, or refractive state;
- current or anticipated use of growth hormones.
The FIN-3101 study participants were randomized equally between test and control groups. The study was conducted across nine investigational sites located throughout the United States, with each site recruiting between 7 and 28 subjects. The study enrolled 175 subjects, of whom 159 were randomized to test (77) and control (82) groups. Of the 175 subjects, 149 were dispensed study lenses. Of the 149 dispensed subjects, $91\%$ (135/149) completed the 24-month visit, with completion rates of $93\%$ (69/74) for the test group and $88\%$ (66/75) for the control group. The dispensed population consisted of $52.3\%$ males and $47.7\%$ females, with a mean age of 10.2 years at baseline (range: 6.6 to 12.97 years). The largest racial group was White $(63.8\%)$ , followed by East/Southeast Asian $(20.1\%)$ and Black/African-American $(19.5\%)$ , with Hispanic or Latino ethnicity representing $22.8\%$ of participants (see Table 2).
Table 2: Participant Demographics - Participants dispensed study lenses.
| Variable | | Total | Test | Control |
| --- | --- | --- | --- | --- |
| No. of Subjects | | 149 | 74 | 75 |
| Participant | 6 yrs old | 4 (2.7) | 2 (2.7) | 2 (2.7) |
| Integer Age | 7 yrs old | 10 (6.7) | 6 (8.1) | 4 (5.3) |
| (n(%)) | 8 yrs old | 19 (12.8) | 8 (10.8) | 11 (14.7) |
| | 9 yrs old | 30 (20.1) | 16 (21.6) | 14 (18.7) |
De Novo Summary (DEN250016)
{9}
| Variable | | Total | Test | Control |
| --- | --- | --- | --- | --- |
| | 10 yrs old | 35 (23.5) | 16 (21.6) | 19 (25.3) |
| | 11 yrs old | 35 (23.5) | 16 (21.6) | 19 (25.3) |
| | 12 yrs old | 16 (10.7) | 10 (13.5) | 6 (8.0) |
| | Mean (SD) | 9.7 (1.52) | 9.7 (1.59) | 9.7 (1.47) |
| Sex (n(%)) | Male | 78 (52.3) | 34 (45.9) | 44 (58.7) |
| | Female | 71 (47.7) | 40 (54.1) | 31 (41.3) |
| Ethnicity (n(%)) | Hispanic or Latino | 34 (22.8) | 18 (24.3) | 16 (21.3) |
| | Not Hispanic | 115 (77.2) | 56 (75.7) | 59 (78.7) |
| Race* (n(%)) | White | 95 (63.8) | 46 (62.2) | 49 (65.3) |
| | East/Southeast Asian | 30 (20.1) | 14 (18.9) | 16 (21.3) |
| | Black/African American | 29 (19.5) | 14 (18.9) | 15 (20.0) |
| | South Asian | 4 (2.7) | 2 (2.7) | 2 (2.7) |
| | American Indian or Alaska Native | 2 (1.3) | 0 (0) | 2 (2.7) |
| | Native Hawaiian or Other Pacific Islander | 1 (0.7) | 1 (1.4) | 0 (0) |
* Proportions do not sum to 100% as subjects selecting more than one race will be counted multiple times.
The co-primary effectiveness endpoints were the change in spherical equivalent refraction from baseline, as measured by cycloplegic auto refraction, and axial length change from baseline. The pre-specified hypotheses for the primary co-effectiveness endpoints were that the changes in cycloplegic spherical equivalent refraction (SER) and axial length from baseline are significantly less with the Essilor® Stellest® lens compared with a single vision lens. The protocol pre-specified the evaluation of the co-primary effectiveness endpoints at 36 months, with a planned analysis at 24 months. The final analysis was based on the 24-month data.
Safety outcomes assessed included adverse events, best-corrected visual acuity (BCVA), and patient reported symptoms, problems, and complaints.
## Clinical study results:
## Effectiveness Outcomes
The FIN-3101 study demonstrated acceptable effectiveness of the Essilor® Stellest® spectacle lens in slowing myopia progression compared to single vision control lenses. For spherical equivalent refraction (SER), the study showed a 71% reduction in myopia progression at 24 months, with a calculated difference of 0.64 D (95% CI: 0.50 to 0.79 D) between test and control groups based on the Mixed effect model. For axial length, the test group showed a 53% reduction in axial elongation at 24 months, with a model-calculated difference of -0.24 mm (95% CI: -0.29 to -0.19 mm) compared to controls. Additionally, 56% of test eyes showed less than 0.25 D of myopia progression at 24 months compared to 19% of control eyes. The least-squares
De Novo Summary (DEN250016)
{10}
mean cycloplegic refractive error and axial length changeover 2-years are shown below in Table 3.
Table 3: Cycloplegic refractive error and axial length change over 2 years in the FIN-3101 randomized, controlled clinical trial. Difference calculated as Test minus Control. (LS Mean - multiple imputation - Intent-To-Treat (ITT) (159 subjects))
| | Arm | LS Mean (SE) | 95% CI | Difference in LS Means (SE) | 95% CI of Difference | P-value | % Control |
| --- | --- | --- | --- | --- | --- | --- | --- |
| cSER | Essilor® Stellest® | -0.25 D (0.05) | -0.35 to -0.16 | 0.64 (0.07) | 0.50 to 0.79 | <0.0001 | 71% |
| | Control | -0.90 D (0.05) | -1.00 to -0.79 | - | - | | |
| Axial Length | Essilor® Stellest® | 0.21 mm (0.02) | 0.17 to 0.25 | -0.24 (0.03) | -0.29 to -0.19 | <0.0001 | 53% |
| | Control | 0.45 mm (0.02) | 0.42 to 0.49 | - | - | | |
LS-Means: least-square means, ITT: Intent-to-Treat, SE: standard error, CI: confidence interval.
Age-based subgroup analysis showed treatment effectiveness across the three age subgroups (6-8 years old, 9-10 years old, and 11-12 years old), with 24-month SER reductions of $89\%$ , $62\%$ , and $72\%$ respectively, representing mean differences of 1.08 D, 0.56 D, and 0.55 D. Similarly, both baseline myopia severity groups demonstrated a treatment effect, with lower myopia subjects $(-0.75$ to $-2.50$ D) experiencing a $71\%$ reduction (0.56 D difference) and higher myopia subjects $(-2.75$ to $-4.50$ D) showing a $69\%$ reduction (0.72 D difference) at 24 months. When analyzed by wearing time, both groups wearing less than 12 hours per day and those wearing 12 or more hours per day showed treatment effects of $75\%$ and $69\%$ respectively.
In the FIN-3101 study, safety information was collected when subjects were questioned by study investigators at each of the follow-up visits about possible symptoms, problems or complaints with the spectacles. Table 4 provides details on the incidence of symptoms, problems, and complaints in the test and control subjects. The "Other symptoms" category in Table 4 includes symptoms such as itchy eyes, squinting, stinging sensations, very infrequent dizziness, rare shadows and afterimages when looking at bright light. None of these symptoms, problems, and complaints noted in Table 4 occurred at substantially higher rates in the test arm compared to the control arm.
Twenty-one ocular adverse events were reported in eleven subjects (nine in test group, twelve in control group). None of the events were classified as lens-related or significant by study investigators. The most common ocular adverse event reported was conjunctivitis (bacterial, allergic, or unknown). Best corrected visual acuity remained stable throughout the study, with no subjects in either group showing a worsening of two or more lines compared to baseline. These results are summarized in Table 5.
The test lens demonstrated acceptable safety and tolerability in the FIN-3101 study. Best corrected visual acuity remained stable throughout the study, with no subjects in either group showing a worsening of two or more lines compared to baseline.
De Novo Summary (DEN250016)
{11}
Table 1: Summary of symptoms, problems and complaints - Participants dispensed study lenses.
| | Test | Control |
| --- | --- | --- |
| No. of Subjects | 74 | 75 |
| 1. Symptoms | | |
| i. Headache | 2 (3%) | 4 (5%) |
| ii. Halos | 1 (1%) | 2 (3%) |
| iii. Glare | 0 (0%) | 0 (0%) |
| iv. Blurred vision | | |
| • Distance | 5 (7%) | 9 (12%) |
| • Near | 2 (3%) | 2 (3%) |
| • General | 7 (9%) | 6 (8%) |
| v. Double vision | 1 (1%) | 0 (0%) |
| vi. Other | 3 (4%) | 7 (9%) |
| 2. Other | | |
| i. Problems with frames | 10 (14%) | 10 (13%) |
| ii. Problems with lenses | 4 (5%) | 8 (11%) |
| iii. Other | 0 (0%) | 2 (3%) |
Table 5: Best corrected distance visual acuity by visit and treatment (n(%))- Participants dispensed study lenses
| | Baseline | | 6-Month | | 12-Month | | 18-Month | | 24-Month | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | Test | Control | Test | Control | Test | Control | Test | Control | Test | Control |
| No. of Eyes | 148 | 150 | 148 | 140 | 146 | 138 | 140 | 130 | 138 | 132 |
| >20/20 | 45 (30%) | 60 (40%) | 72 (49%) | 66 (47%) | 66 (45%) | 54 (39%) | 64 (46%) | 63 (48%) | 69 (50%) | 65 (49%) |
| 20/20 | 92 (62%) | 81 (54%) | 69 (47%) | 65 (46%) | 77 (53%) | 74 (54%) | 73 (52%) | 62 (48%) | 65 (47%) | 65 (49%) |
| 20/25 | 11 (7%) | 9 (6%) | 7 (5%) | 9 (6%) | 3 (2%) | 9 (7%) | 3 (2%) | 5 (4%) | 4 (3%) | 2 (2%) |
| 20/30 | 0 (0%) | 0 (0%) | 0 (0%) | 0 (0%) | 0 (0%) | 1 (1%) | 0 (0%) | 0 (0%) | 0 (0%) | 0 (0%) |
| <20/30 | 0 (0%) | 0 (0%) | 0 (0%) | 0 (0%) | 0 (0%) | 0 (0%) | 0 (0%) | 0 (0%) | 0 (0%) | 0 (0%) |
NOTE: Data in the FIN-3101 study was reported in logMAR. The following is the conversion from logMAR to Snellen reported in the table above:
<0.00 = >20/20
0.00 to 0.04 = 20/20 (20/20 - 20/21.9)
0.06 to 0.12 = 20/25 (20/23 - 20/26.4)
0.14 to 0.20 = 20/30 (20/27.6 - 20/31.7)
>0.20 = <20/30
De Novo Summary (DEN250016)
{12}
FIN-3102 sub-study summary:
In the FIN-3102 sub-study to evaluate safety risks related to the design of the device, visual performance tests including, reading speed, contrast sensitivity, and effects of glare on visual acuity, were conducted on 88 of the 149 dispensed subjects from the FIN-3101 study. The visual performance tests used in this sub-study were not adequately validated to be used for the intended patient population (children aged 6 to 12 years). Thus, there is significant uncertainty in the interpretation of the results of these visual performance tests.
Rebound assessment:
Essilor provided an assessment showing a continuing cumulative treatment effect based on the 24-month clinical study results and also provided a summary of rebound literature available for their device. This information provided preliminary evidence that any potential rebound effect would likely not negate the observed treatment effect for their device but supported the need for a postmarket rebound study to confirm the information provided.
Pediatric Extrapolation:
In this De Novo request, existing clinical data were not leveraged to support the use of the device in a pediatric patient population. The device is indicated for pediatric patients 6-12 years old and those patients were enrolled in the clinical study (see Table 2).
POSTMARKET EVALUATION
A postmarket study will be conducted to evaluate whether there is a “rebound effect.” The 12 month study will evaluate the change in axial length and spherical equivalent refractive error experienced by subjects after cessation of treatment with the Essilor® Stellest® spectacle lenses, as compared to a control group wearing single vision spectacle lenses. The study will also evaluate all observed ocular adverse events.
LABELING
The labeling is sufficient and satisfies the requirements of 21 CFR 801.109 for prescription devices.
There is a Healthcare Professional Fitting Guide as well as a Patient and Caregiver Guide. These two documents provide:
- a description of the device including the optical design elements and description of what lens tint(s) or coating(s) were used in clinical testing
- Description of how the device functions to slow the progression of myopia
- Recommended wearing schedule
- Spectacle fitting considerations
De Novo Summary (DEN250016)
{13}
- A summary of the clinical study performed to support the device’s safety and effectiveness in the intended use population, including a summary of the demographic and racial distribution of the study cohort.
- Warnings and precautions to ensure safe use of the device in the intended patient population.
The Patient and Caregiver Guide accompanies the device to be dispensed.
## RISKS TO HEALTH
The table below identifies the risks to health that may be associated with use of prescription spectacle lenses to reduce the progression of myopia and the measures necessary to mitigate these risks.
| Risks to Health | Mitigation Measures |
| --- | --- |
| Failure to slow the progression of myopia, leading to poorer long-term vision and increased risk of myopia-related ocular disease | Clinical performance testing
Postmarket surveillance
Non-clinical performance testing
Labeling |
| Adverse visual symptoms or impaired visual performance | Clinical performance testing
Labeling |
| Lens breakage leading to eye injury | Non-clinical performance testing |
## SPECIAL CONTROLS
In combination with the general controls of the FD&C Act, the prescription spectacle lenses to reduce the progression of myopia are subject to the following special controls:
(1) Data obtained from premarket clinical performance validation testing, and from postmarket surveillance conducted per a protocol approved by FDA and acquired under anticipated conditions of use, must demonstrate that the device performs as intended when used in the intended patient population, and must evaluate the following, unless FDA determines based on the totality of the information provided for premarket review that data from postmarket surveillance is not required:
(i) Assessment of the change in spherical equivalent refractive error and axial length in the intended patient population as compared to a clinically justified control group. Data must demonstrate the following:
(A) The lower bound of the 95% confidence interval for the difference in mean refractive error as compared to the control is no less than -0.50 diopters spherical equivalent with a corresponding change in axial length as compared to control. If these endpoints are not achieved, an alternative clinical justification must be provided; and
De Novo Summary (DEN250016)
Page 13 of 16
{14}
(B) Clinical performance testing must demonstrate that the device has a continuing cumulative treatment effect in both spherical equivalent refraction and axial length over the duration of the study;
(ii) Assessment of adverse events and visual symptoms, considering the specific design characteristics of the device;
(iii) Assessment of rebound effect after cessation of device use; and
(iv) Assessment of the impact of optical lens design elements on visual performance (e.g., visual acuity and contrast sensitivity) and vision-related activities (e.g., reading and peripheral vision).
(2) Non-clinical performance testing data must demonstrate that the device performs as intended under anticipated conditions of use. The following testing must be provided:
(i) Impact resistance testing requirements as required by 21 CFR 801.410;
(ii) Optical and bench testing of the critical parameters, including:
(A) Optical characterization testing;
(B) Durability testing; and
(C) Performance testing to verify technical specifications.
(3) Labeling for the healthcare professional; and labeling for the patient and/or caregiver that accompanies the device to be dispensed, must include the following:
(i) A description of the optical design elements;
(ii) The recommended wearing schedule;
(iii) For patient and/or caregiver labeling, fitting considerations from the patient and/or caregiver perspective;
(iv) For healthcare professional labeling, instructions for fitting and positioning of the lenses in the frame relative to the primary gaze position;
(v) A description of what lens tint(s) or coating(s) were used in clinical testing and a warning that the effectiveness of the lenses was not studied with other coating(s) or tint(s);
(vi) A summary of the visual and clinical performance testing obtained with the device; and
(vii) A detailed summary of relevant postmarket surveillance data collected, including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
## BENEFIT-RISK DETERMINATION
The probable risks of the Essilor® Stellest® device are based on data collected in the FIN-3101 clinical study described above as well as risks generally associated with device based on the device characteristics and design. The risk of symptoms based on the reports obtained by open-ended questions in the clinical trial include visual halos and blurred vision.
The probable benefits of the device are also based on clinical effectiveness data collected in the clinical study as described above. The 2-year clinical study results show that the device significantly slows the progression of myopia in the test group as compared to the control group (-0.64 D SER and 0.24mm AL). The benefit of slowing myopia progression lies in its potential to prevent the sight threatening complications of myopia, particularly high myopia. The Essilor®
De Novo Summary (DEN250016)
{15}
Stellest® spectacle lenses slow the progression of myopia as compared to a control group in the FIN-3101 study. This can help to mitigate risk factors for pathologic myopia: high refractive error and axial elongation.
Additional factors evaluated in determining probable risks and benefits for the Essilor® Stellest® spectacle lenses included consideration of the impact of the optical components of the Essilor® Stellest® lenses to interfere with reading, learning to read, and detection of obstacles in peripheral vision. Although there remains significant uncertainty regarding the risks associated with these vision-related activities, the overall benefit/risk associated with the use of the device was deemed acceptable based on available safety data, the effectiveness outcomes of the clinical trial, the preliminary assessment of a potential rebound effect, as well as the specific lens design, which included a 9mm clear central zone. A postmarket surveillance study was determined to be necessary to reduce the uncertainty regarding the potential for rebound following cessation of treatment with the Stellest® spectacle lenses.
The only currently FDA-approved intervention for reduction in myopia progression is a contact lens for daily wear which is approved for children aged 8 to 12 years. The Essilor® Stellest® lenses are indicated for children aged 6 to 12 years old. Thus, Essilor® Stellest® will meet an unmet need among children aged 6 to 7 years. In addition, the Essilor® Stellest® lenses are a lower risk device compared to contact lenses, since spectacle lenses do not carry the risk of eye infections such as microbial keratitis. This wider age range and lower risk provides probable benefits to young children at risk of myopia progression.
## Patient Perspectives
A symptoms questionnaire was administered during the pivotal clinical trial. However, there was a lack of evidence to show that the questionnaire measured what it was intended in the trial. Therefore, the data were deemed uninterpretable and could not be used in the benefit-risk assessment.
## Benefit/Risk Conclusion
In conclusion, given the available information above, for the following indication statement:
The Essilor® Stellest® lens is indicated for the correction of myopia with and without astigmatism and for slowing the progression of myopia in children with non-diseased eyes, who, at the initiation of treatment, are aged 6-12 years and have spherical equivalent refraction of -0.75 D to -4.50 D with astigmatism up to 1.50 D.
The probable benefits outweigh the probable risks for the Essilor® Stellest® spectacle lenses. The device provides benefits and the risks can be mitigated by the use of general controls and the identified special controls.
## CONCLUSION
The De Novo request for the Essilor® Stellest® is granted and the device is classified as follows:
De Novo Summary (DEN250016)
Page 15 of 16
{16}
Product Code: QUR
Device Type: Prescription spectacle lenses to reduce the progression of myopia
Regulation Number: 21 CFR 886.5845
Class: II
De Novo Summary (DEN250016)
Page 16 of 16
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.