DEN240031 · Prapela, Inc. · QVY · Apr 4, 2025 · General Hospital
Device Facts
Record ID
DEN240031
Device Name
Prapela SVS hospital bassinet pad (model P01)
Applicant
Prapela, Inc.
Product Code
QVY · General Hospital
Decision Date
Apr 4, 2025
Decision
DENG
Submission Type
Direct
Regulation
21 CFR 880.5151
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Prapela SVS hospital bassinet pad is indicated as adjunctive non-pharmacological therapy in newborns ≥37 weeks gestational age exposed prenatally to opioids with neonatal opioid withdrawal syndrome (NOWS). The Prapela SVS hospital bassinet pad is indicated for prescription use only.
Device Story
Prapela SVS hospital bassinet pad (model P01) provides adjunctive non-pharmacological therapy for newborns with NOWS. Device consists of a reusable hospital mattress pad with integrated mechanical vibration mechanism. Used in hospital settings; operated by clinical staff. Provides gentle vibrations to infant; intended to assist in managing withdrawal symptoms. Requires proper placement in compatible hospital bassinets to avoid entrapment or strangulation risks. Safety relies on controlled vibration output and auditory sound levels. Clinical benefit derived from non-pharmacological symptom management.
Clinical Evidence
Evidence includes two clinical studies. Study 1 (JAMA Pediatr 2023): Prospective RCT (n=181) comparing SVS to standard of care. Post-hoc analysis of 32 morphine-responsive infants showed SVS reduced length of treatment and cumulative morphine dose. Study 2 (PLoS One 2017): Observational study (n=26) evaluating physiological parameters (HR, respiratory rate, temperature, O2 saturation). No adverse events reported; infants maintained thermoregulation and oxygenation. Data supports safety and effectiveness for the intended population.
Technological Characteristics
Reusable mattress pad with embedded transducer; foam construction. Operates via stochastic vibrotactile stimulation (SVS). Dimensions designed for standard hospital bassinets. Biocompatible per ISO 10993-5, -10, -23. Electrical safety/EMC per IEC 60601-1 and IEC 60601-1-2. Firmware-controlled feedback loop; potentiometer sets Displacement Root Mean Square (DRMS) amplitude. Non-sterile; validated for cleaning/disinfection.
Indications for Use
Indicated for newborns ≥37 weeks gestational age with neonatal opioid withdrawal syndrome (NOWS) following prenatal opioid exposure. Used as adjunctive non-pharmacological therapy.
Regulatory Classification
Identification
The Prapela SVS hospital bassinet pad (model P01) is a non-sterile, reusable, hospital mattress pad with an embedded transducer that produces small-input stochastic stimulations (vibrations) that resonate through the foam. It is indicated as adjunctive non-pharmacological therapy in newborns ≥37 weeks gestational age exposed prenatally to opioids with neonatal opioid withdrawal syndrome (NOWS).
Special Controls
In combination with the general controls of the FD&C Act, the therapeutic vibrational mattress pad is subject to the following special controls:
Submission Summary (Full Text)
{0}
DE NOVO CLASSIFICATION REQUEST FOR
PRAPELA SVS HOSPITAL BASSINET PAD (MODEL P01)
REGULATORY INFORMATION
FDA identifies this generic type of device as:
Therapeutic vibrational mattress pad. A therapeutic vibrational mattress pad is a reusable hospital mattress pad with integrated mechanical vibrations to be used as adjunctive therapy for neonates and infants.
NEW REGULATION NUMBER: 21 CFR 880.5151
CLASSIFICATION: Class II
PRODUCT CODE: QVY
BACKGROUND
DEVICE NAME: Prapela SVS hospital bassinet pad (model P01)
SUBMISSION NUMBER: DEN240031
DATE DE NOVO RECEIVED: June 17, 2024
SPONSOR INFORMATION:
Prapela, Inc.
2 Main ST.
STE 15-219
Biddeford, Maine 04005
INDICATIONS FOR USE
The Prapela SVS hospital bassinet pad is indicated as adjunctive non-pharmacological therapy in newborns ≥37 weeks gestational age exposed prenatally to opioids with neonatal opioid withdrawal syndrome (NOWS).
The Prapela SVS hospital bassinet pad is indicated for prescription use only.
LIMITATIONS
The sale, distribution, and use of the Prapela SVS hospital bassinet pad are restricted to prescription use in accordance with 21 CFR 801.109.
{1}
Newborns meeting the following criteria should not be treated:
- Born less than 37 weeks
- Has a clinically significant congenital abnormality
- Has a clinically significant fetal anomaly
- Has hydrocephalus or intraventricular hemorrhage >grade 2
- Has a seizure disorder not related to drug withdrawal
- Has a clinically significant cardiac shunt
- Has anemia (hemoglobin<8g/dL)
- Has hypoxic ischemia encephalopathy
- Has respiratory failure that requires invasive ventilatory support
- Has methicillin-resistant Staphylococcus aureus (MRSA) or receiving treatment for bacterial or viral conditions
The device is intended to be used in a professional healthcare setting with temperature-controlled environments, including but not limited to neonatal care step-down units, hospital patient rooms (next to mother's bedside), and offsite specialty centers treatment centers with mother/child patient rooms that provide treatment for newborns diagnosed with Neonatal Opioid Withdrawal Syndrome (NOWS)/Neonatal Abstinence Syndrome (NAS). The device should not be used in isolated oxygen rich environments or in or near any radiological imaging device.
PLEASE REFER TO THE LABELING FOR A COMPLETE LIST OF WARNINGS, PRECAUTIONS AND CONTRAINDICATIONS.
## DEVICE DESCRIPTION
The Prapela Stochastic Vibrotactile Stimulation (SVS) hospital bassinet pad (model P01) is a non-sterile, reusable, mattress pad that has a transducer embedded in the foam of the mattress. The transducer is located in the center of the pad and produces small-input stochastic stimulations that resonate through the foam of the pad. The pad comes with a hand-held control unit that toggles the vibrations on and off. When the vibrations are turned on, the stimulation cycles in 3 hours ON/OFF cycle. The mattress pad is designed to fit most standard hospital grade bassinets.
De Novo Summary (DEN240031)
Page 2 of 9
{2}

## SUMMARY OF NONCLINICAL/BENCH STUDIES
### BIOCOMPATIBILITY/MATERIALS
The Prapela SVS hospital bassinet pad (model P01) is classified as intact skin-contacting with a prolonged contact duration. Biocompatibility was evaluated in accordance with ISO 10993-1:2018 Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process and FDA Guidance Use of International Standard ISO 10993-1 “Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process.” Biocompatibility testing was conducted per the following standards to support the biological safety of the device:
- ISO 10993-5:2009 Biological evaluation of medical devices – Part 5: Tests for in vitro cytotoxicity
- ISO 10993-10:2021 Biological evaluation of medical devices – Part 10: Tests for skin Sensitization
- ISO 10993-23:2021 Biological evaluation of medical devices – Part 23: Tests for irritation
The results of these evaluations support that the biological risks of the device have been adequately mitigated.
### REPROCESSING, STERILITY & SHELF LIFE
The Prapela SVS hospital bassinet pad (model P01) is not provided sterile. Performance testing after accelerated aging confirmed the Prapela SVS hospital bassinet pad (model P01) maintained device performance over the 6-month labeled shelf life. Reprocessing and disinfection instructions for the end user were provided and were evaluated per the 2015 FDA guidance Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling under clinical worst-case conditions. Cleaning and disinfection validation studies were provided and met the recommendations of the guidance.
De Novo Summary (DEN240031)
{3}
De Novo Summary (DEN240031)
Page 4 of 9
# ELECTROMAGNETIC CAPABILITY & ELECTROMAGNETIC SAFETY
The Prapela SVS hospital bassinet pad (model P01) conforms to the requirements for electrical safety and electromagnetic compatibility (EMC) of the following standards:
- IEC 60601-1-1:2005 Medical Electrical Equipment - Part 1: General requirements for basic safety and essential performance
- IEC 60601-1-2:2014+A1:2020, Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility – Requirements and tests.
# HUMAN FACTORS
Prapela conducted human factors testing by recruiting 18 clinicians located in neonatal care step-down units for NOWS diagnosed newborns to assess their ability to set up, utilize, and clean-up & reprocess the mattress pad. The testing was conducted based on FDA 2016 guidance *Applying Human Factors and Usability Engineering to Medical Devices*. The participants were successful in setting up and using the Prapela SVS hospital bassinet pad (model P01) with neonates. Major tasks included assembling the device in the bassinet basket, turning on the device, cleaning the pad, and disassembling and storing the device. The human factors testing report supports the participants were able use the device for the intended use within the expected use environments.
# SOFTWARE
The Prapela SVS hospital bassinet pad (model P01) contains software to control the feedback loop for vibration delivery. The firmware is responsible for device startup/initialization, control of ON/OFF function, and generation of a signal that is passed to device hardware for vibration delivery. The firmware generates a random signal in the specified frequency range that is passed to a potentiometer on the device circuit board. The overall amplitude of the pad surface vibration, referred to as Displacement Root Mean Square (DRMS), is set by the potentiometer. There are no other user interactions with the firmware. Software verification, validation, and analysis was performed, and the results demonstrate that the software functions as intended, and the required specifications are met. The device's software has a Basic Documentation Level.
A cybersecurity evaluation was not needed as the device does not connect to external networks or devices.
# PERFORMANCE TESTING – BENCH
The functionality of the Prapela SVS hospital bassinet pad (model P01) was evaluated using multiple performance tests. The following table is a summary of the functional tests that were performed:
| Test Name | Purpose | Method | Acceptance Criteria | Results |
| --- | --- | --- | --- | --- |
| Vibration Displacement and Frequency | Evaluate the vibrational | (1) (4) | (1) (4) | All results were passing. |
{4}
| Test Name | Purpose | Method | Acceptance Criteria | Results |
| --- | --- | --- | --- | --- |
| | functionality of the device | a) (4) | b) (4) | |
| Effect of patient weight on stimulation | Ensure that the shaking from the vibrations is not excessive for the infant or subdued by the weight of the infant | | | All results were passing. |
De Novo Summary (DEN240031)
{5}
| Test Name | Purpose | Method | Acceptance Criteria | Results |
| --- | --- | --- | --- | --- |
| Dimensional Testing | Evaluate the dimensions of the pad and ensure gaps are not formed on the edges of the compatible bassinets. | (b) (4) | (b) (4) | All results were passing. |
| Stimulation Uniformity Test | Evaluate the uniformity of the stimulation across the mattress | | | All results were passing. |
| Auditory Sound Level | To evaluate the sound pressure levels produced by the device and ensure that they are within acceptable limits. | | | All results were passing. |
De Novo Summary (DEN240031)
{6}
De Novo Summary (DEN240031)
Page 7 of 9
# SUMMARY OF CLINICAL INFORMATION
Data from two clinical studies conducted with the Prapela SVS hospital bassinet pad were submitted to provide reasonable assurance of the safety and effectiveness of the device:
## Study 1:
Bloch-Salisbury E, Wilson JD, Rodriguez N, et al. Efficacy of a Vibrating Crib Mattress to Reduce Pharmacologic Treatment in Opioid-Exposed Newborns: A Randomized Clinical Trial. JAMA Pediatr. Jul 1 2023;177(7):665-674. doi:10.1001/jamapediatrics.2023.1077
This study was a prospective, randomized controlled trial assessing the following primary endpoints: administration of morphine treatment, cumulative morphine dose, and hospital length of stay. A total of 181 opioid-exposed newborns completed the study of which 94 infants were treated with SVS, and 87 infants received standard of care treatment. Of the 181 infants, 60 infants received morphine treatment. The authors carried out a post-hoc analysis on 32 infants receiving and responding to morphine treatment, comparing SVS treatment (n = 17) and standard of care treatment (n = 15). Results of the post-hoc analysis showed improvement in length of treatment and cumulative morphine dose administered in patients that were found to be responsive to morphine treatment and received SVS treatment.
## Study 2:
Zuzarte I, Indic P, Barton B, Paydarfar D, Bednarek F, Bloch-Salisbury E. Vibrotactile stimulation: A non-pharmacological intervention for opioid-exposed newborns. PLoS One. 2017;12(4):e0175981. DOI:10.1371/journal.pone.0175981
This study enrolled 26 opioid-exposed newborns and evaluated the movement activity, heart rate, respiratory rate, axillary temperature, and the blood oxygen saturation levels. While these clinical endpoints were not appropriate to provide a reasonable assurance of effectiveness for the indications for use of this device, information on adverse events, such as those related to thermoregulation and blood-oxygen desaturation levels, were also collected as an endpoint of the study. No adverse events were reported in the study and the infants were reported to be able to maintain their body temperatures and oxygen levels throughout the study. Therefore, these results provided additional support for the safety of the device.
Both studies were conducted in a hospital environment on newborns with prenatal opioid exposure at a gestational age of greater than or equal to 37 weeks and receiving morphine treatment. Baseline characteristics such as sex, race, ethnicity, gestational age, or birth weight were not evaluated by the sponsor for potential association with safety and effectiveness outcomes. The studies were not specifically powered for any relevant subgroups, and no additional subgroup analyses were performed.
## Pediatric Extrapolation
Pediatric extrapolation was not needed to support the intended use population because the pediatric age ranges in the clinical study are the same as the pediatric age range in the intended use population.
## RISKS TO HEALTH
{7}
The table below identifies the risks to health that may be associated with the use of a therapeutic vibrational mattress pad.
| Risks to Health | Mitigation Measures |
| --- | --- |
| Sleep disruption or hearing loss due to inappropriate vibrations or high auditory sound levels | Clinical performance data
Software verification, validation, and hazard analysis
Non-clinical performance testing |
| Ineffective treatment from worsening of signs and symptoms during use of the device or return of symptoms after discontinuation of the device, and inappropriate placement of infant on pad | Clinical performance data
Software verification, validation, and hazard analysis
Labeling |
| Inappropriate use or inadequate securement of pad in bassinet leading to injury by:
• strangulation by cords or straps
• entrapment in gaps around pad/against the bassinet | Non-clinical performance testing
Labeling |
| Electrical shock or burns | Electrical safety testing |
| Interference with other medical devices | Electromagnetic compatibility (EMC) testing |
| Adverse tissue reaction | Biocompatibility evaluation |
| Infection | Labeling |
## SPECIAL CONTROLS
In combination with the general controls of the FD&C Act, the therapeutic vibrational mattress pad is subject to the following special controls:
1. Clinical performance data must include information from adverse events and performance endpoints in support of the indications for use in the intended patient population under clinically relevant use scenarios.
2. Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and environment(s) of use. Testing must include:
- (i) Verification and validation of the vibration mechanism output of the device;
- (ii) Verification and validation of auditory sound levels of the device; and
- (iii) Dimensional testing to demonstrate compatibility with compatible neonatal hospital beds/bassinets.
3. Performance testing must demonstrate electrical safety and electromagnetic compatibility (EMC) of the device in the intended use environment.
4. Software verification, validation, and hazard analysis must be performed for any software components of the device.
5. The patient-contacting components of the device must be biocompatible.
De Novo Summary (DEN240031)
{8}
(6) Labeling must include:
(i) Instructions to ensure proper fit and infant placement;
(ii) A list of compatible neonatal hospital beds/bassinets;
(iii) Instructions for cleaning and disinfection; and
(iv) Warnings related to suffocation and strangulation risks related to straps and cords.
## BENEFIT-RISK DETERMINATION
### Summary of Benefits
Based on the clinical evidence provided, the Prapela SVS hospital bassinet pad (model P01) demonstrated probable benefit as a non-pharmacological adjunctive treatment option to decrease the length of treatment and cumulative doses of morphine in NOWS patients that are found to be responsive to morphine treatment. This population presents a challenge to treat given NAS/NOWS symptomology. In addition, the sponsor provided qualitative data to support a probable benefit that the device can reduce the burden on health care providers providing supportive care for opioid exposed neonates and infants by reducing the time spent holding and rocking the baby.
### Summary of Risks
While no adverse events were observed in the clinical performance data, risks associated with the device include sleep disruption and hearing loss due to inappropriate vibrations or high auditory sound levels, ineffective treatment, inappropriate use or inadequate securement of the pad in the bassinet, strangulation due to the length of the electrical cord or entrapment (i.e., patient may roll over and get stuck in corners of the pad), electrical shock or burns, adverse tissue reaction, and infection due to the reusable nature of the device. These risks have been adequately mitigated through clinical performance data, non-clinical performance testing, EMC and electrical safety testing, biocompatibility evaluation, and labeling.
### Patient Perspectives
This submission did not include specific information on patient perspectives for this device.
### Benefit/Risk Conclusion
In conclusion, given the available information above, the data support that, for the indications for use stated above, the probable benefits outweigh the probable risks for the Prapela SVS hospital bassinet pad (model P01). The device provides benefits and the risks can be mitigated by the use of general and special controls.
## CONCLUSION
The De Novo request for the Prapela SVS hospital bassinet pad (model P01) is granted and the device is classified as follows:
Product Code: QVY
Device Type: Therapeutic vibrational mattress pad
Regulation Number: 21 CFR 880.5151
Class: II
De Novo Summary (DEN240031)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.