DEN230035 · Dental Monitoring Sas · SBC · May 17, 2024 · Dental
Device Facts
Record ID
DEN230035
Device Name
DentalMonitoring
Applicant
Dental Monitoring Sas
Product Code
SBC · Dental
Decision Date
May 17, 2024
Decision
DENG
Submission Type
Direct
Regulation
21 CFR 872.1770
Device Class
Class 2
Attributes
AI/ML, Real-World Evidence, Pediatric
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
DEN230035 · May 17, 2024
DentalMonitoring
Dental Monitoring Sas
Retrospective patient intraoral scans (Scans)
Retrospective clinical study used to evaluate the qualitative performance (sensitivity/specificity) of the device for various orthodontic monitoring parameters.
Patients undergoing orthodontic treatment (aligners, brackets, retainers); Sample Size: Varies by indication (e.g., 5,064 for dental plaque, 870 for gingival recession); Number of Sites: 15 US orthodontists
Reference Method (panel of three independent orthodontists)
Sensitivity and Specificity for qualitative orthodontic parameters
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Dental plaque / food residue
Machine learning and locked neural network algorithms
Archwire and auxiliaries study (21-003): 730 results
—
Updated 3D Model
Machine learning and locked neural network algorithms
—
Mean Absolute Error (MAE): 0.10 (95% CI: 0.093, 0.103)
—
—
Updated 3D Model study (21-004): 250 patients
—
Indications for Use
DentalMonitoring is a medical device software using image processing algorithms to analyze pictures of the oral cavity (hereinafter Scans). Scans are taken using the DM App, a smartphone, and the manufacturer's proprietary hardware products. Scans are taken by the patient, a non-healthcare professional, or a healthcare professional. The Scan is taken in healthcare facilities, such as a dental practice, or in a non-healthcare environment, such as the patient's own home. For some clinical parameters, DentalMonitoring requires a 3D Model. The product is designed to assist healthcare professionals in remotely monitoring orthodontic treatments and treatment progress. The results of DentalMonitoring are intended to be used as an aid in diagnosis and monitoring, not on a stand-alone basis for clinical decision-making. DentalMonitoring is indicated for use for patients over the age of 6 and reports results solely on permanent teeth. DentalMonitoring can monitor the following clinical parameters: - oral hygiene: dental plaque / food residue; - soft tissue statement: gingival recession, black triangle; - dental statement: closure of extraction space, tooth wear; - alignment: closure of all anterior spaces; and - dental occlusion: - in 2D Monitoring: midline deviation, overbite/open bite, overjet - in 3D Monitoring: canine class, midline deviation, overbite/open bite, overjet Additionally, the following clinical parameters specific to orthodontic treatment types or phases can be monitored using DentalMonitoring: - for aligner treatments: tracking (seat/unseat), attachment loss, button loss; - for braces: bracket debonding, tie loss, self-ligating clips, passive archwire and auxiliaries; and - for thermoformed retainers: tracking (seat/unseat). Based on an initial 3D Model provided by a healthcare professional, DentalMonitoring can also provide healthcare professionals with 3D Models representative of the patient's dentition and treatment progress. This device is a prescription device and is not intended for over-the-counter use.
Device Story
Device uses smartphone-captured intraoral images (Scans) and proprietary hardware; patient or clinician captures images at home or clinic. Software processes images via algorithms to monitor orthodontic progress; provides 3D models from initial 3D input. Output presented to healthcare professionals as aid for monitoring oral hygiene, soft tissue, dental status, alignment, and occlusion. Assists in tracking aligner/braces/retainer treatment phases. Not for standalone diagnosis; supports clinical decision-making by providing remote monitoring data; benefits patients by enabling remote treatment oversight and reducing in-office visits.
Clinical Evidence
Clinical performance data required to demonstrate device performs as intended in the target population, specifically evaluating accuracy and precision of diagnostic and monitoring algorithms.
Technological Characteristics
Standalone software with embedded AI (ML and locked neural networks). Hardware includes DM Cheek Retractor (biocompatible per ISO 10993-1, ISO 7405) and DM ScanBox. Connectivity via smartphone app and cloud-based Data Analysis Platform. Operates on 2D image processing and 3D model generation/comparison. Software verification/validation performed per FDA 2005 guidance. Cybersecurity aligns with FDA 2020 guidance and Section 524B of the FD&C Act.
Indications for Use
Indicated for patients over age 6 with permanent teeth undergoing orthodontic treatment. Used to monitor oral hygiene, soft tissue, dental status, alignment, occlusion, and treatment-specific parameters (aligners, braces, retainers). Contraindicated for patients under age 6 or those with systemic connective tissue conditions preventing adequate buccal cavity opening.
Regulatory Classification
Identification
Dental Image Analyzer. A dental image analyzer is a prescription home use device that uses software intended to collect and analyze patient-specific, optical camera-based, intraoral images. The analyses are provided to dental health care professionals as an aid to diagnosis of oral health conditions and/or to monitor treatment progress.
Special Controls
In combination with the general controls of the FD&C Act. the Dental Image Analyzer is subject to the following special controls:
(1) The patient-contacting components of the device must be demonstrated to be biocompatible.
(2) Non-clinical performance testing data must demonstrate that the device performs as intended under anticipated conditions of use and must include full specifications of the hardware requirements and testing to demonstrate the specified hardware ensures adequate data for validated and accurate measurements; (3) Clinical performance data must demonstrate the accuracy and precision of the diagnostic and monitoring device algorithms to perform as intended under anticipated conditions of use.
(4) Usability testing must demonstrate that the user can correctly use the device, based solely on reading the directions for use.
(5) Software verification, validation, and hazard analysis must be performed.
(6) Labeling must include:
(i) Information on the patient population for which the device has been demonstrated to be effective.
(ii) Information on how the device operates during the typical course of treatment
(iii) A detailed summary of the device technical parameters, including compatible operating systems, hardware, and/or accessories.
(iv) Instructions for cleaning of any reusable components.
(v) Warning that the device is not intended to be used independently for diagnosis and that the device does not replace clinical decision making.
- (7) Patient labeling must be provided and must include:
(i) Relevant contraindications, warnings, precautions, adverse effects/complications, and when to seek help from a dental healthcare professional.
(ii) Information on how the device operates during the typical course of treatment.
(iii) A detailed summary of the device technical parameters, including compatible operating systems, hardware, and/or accessories.
(iv) Information on the patient population for which there is clinical evidence of effectiveness.
- (v) Instructions for cleaning of any reusable components.
Submission Summary (Full Text)
{0}------------------------------------------------
## DE NOVO CLASSIFICATION REQUEST FOR
#### DentalMonitoring
#### REGULATORY INFORMATION
FDA identifies this generic type of device as:
Dental Image Analyzer. A dental image analyzer is a prescription home use device that uses software intended to collect and analyze patient-specific. optical camera-based, intraoral images. The analyses are provided to dental health care professionals as an aid to diagnosis of oral health conditions and/or to monitor treatment progress.
NEW REGULATION NUMBER: 872.1770
CLASSIFICATION: Class II
PRODUCT CODE: SBC
#### BACKGROUND
DEVICE NAME: DentalMonitoring
SUBMISSION NUMBER: DEN230035
DATE DE NOVO RECEIVED: 05/04/2023
#### SPONSOR INFORMATION:
Dental Monitoring SAS Isabelle Puraye 75 rue de Tocqueville Paris 75017
#### INDICATIONS FOR USE
DentalMonitoring is a medical device software using image processing algorithms to analyze pictures of the oral cavity (hereinafter Scans). Scans are taken using the DM App, a smartphone, and the manufacturer's proprietary hardware products. Scans are taken by the patient, a non-healthcare professional, or a healthcare professional. The Scan is taken in healthcare facilities, such as a dental practice, or in a non-healthcare environment, such as the patient's own home. For some clinical parameters. DentalMonitoring requires a 3D Model. The product is designed to assist healthcare professionals in remotely monitoring orthodontic treatments and treatment progress. The results of DentalMonitoring are intended to be used as an aid in diagnosis and monitoring, not on a stand-alone basis for clinical decision-making. DentalMonitoring is indicated for use for patients over the age of 6 and reports results solely on permanent teeth.
DentalMonitoring can monitor the following clinical parameters:
- . oral hygiene: dental plaque / food residue;
- 트 soft tissue statement: gingival recession, black triangle;
- . dental statement: closure of extraction space, tooth wear;
- . alignment: closure of all anterior spaces; and
- . dental occlusion:
- o in 2D Monitoring: midline deviation, overbite/open bite, overjet
{1}------------------------------------------------
- o in 3D Monitoring: canine class, midline deviation, overbite/open bite, overjet
Additionally, the following clinical parameters specific to orthodontic treatment types or phases can be monitored using DentalMonitoring:
- for aligner treatments: tracking (seat/unseat), attachment loss, button loss; .
- . for braces: bracket debonding, tie loss, self-ligating clips, passive archwire and auxiliaries; and
- for thermoformed retainers: tracking (seat/unseat). I
Based on an initial 3D Model provided by a healthcare professional. DentalMonitoring can also provide healthcare professionals with 3D Models representative of the patient's dentition and treatment progress. This device is a prescription device and is not intended for over-the-counter use
#### LIMITATIONS
DentalMonitoring should not be used on patients under the age of 6.
Dental Monitoring should not be used by patients presenting systemic conditions affective connective tissues not allowing them to open/close their buccal cavity sufficiently to acquire acceptable Scans.
The following population should be assisted by a third party to perform Scans:
- . Children up to 12;
- . Adults or children visually impaired;
- . Adults or children hearing impaired:
- . Or any condition that might prevent the patient from adopting the right position to take a Scan.
DentalMonitoring is not intended to replace standard practices for treatment. Final clinical decisions remain the sole responsibility of the healthcare professional. In order to establish a diagnosis. further examinations are required according to the current standard of care, such as dental radiographs and/or tactile examinations with instrumentation.
DentalMonitoring results are limited only to elements visible in the input Scans. DentalMonitoring does not provide results on non-visible elements such as potential interproximal and lingual cavities. DentalMonitoring does not provide results on lingual surfaces.
#### PLEASE REFER TO THE LABELING FOR A COMPLETE LIST OF WARNINGS. PRECAUTIONS AND CONTRAINDICATIONS.
#### DEVICE DESCRIPTION
DentalMonitoring is a standalone software with embedded artificial intelligence comprising machine learning and locked neural network algorithms. The product is composed of these parts: the Dashboard, the DM App along with a DM Cheek Retractor and DM ScanBox, and the Data Analysis Platform.
{2}------------------------------------------------
Image /page/2/Figure/0 description: This image shows a diagram of a 4-step process. The first step is patient profile set-up, which is represented by a computer screen. The second step is intraoral pictures acquisition, which is represented by a smartphone. The third step is analysis of acquired intraoral pictures, which is represented by a tooth and magnifying glass. The fourth step is communication of results, which is represented by a smartphone.
The four steps are detailed below:
- 1. Patient profile set-up: The healthcare professional sets up the patient profile through a web- based interface, Dashboard, accessible at www.dental-monitoring.com/doctor. During the patient profile set-up, the healthcare professional is prompted to set up a Protocol in order to select the clinical parameters they wish to monitor in accordance with the patient's treatment. During the set-up, the healthcare professional also determines the appropriate frequency at which the patient should use the DM App to take pictures of their intraoral cavity.
- 2. Intraoral pictures acquisition: A set of pictures of the patient's intraoral cavity, hereinafter referred to as a Scan, is taken using the DM App along with a DM Cheek Retractor and DM ScanBox. The DM Cheek Retractor is an intraoral retraction device for the cheeks and lips to allow for sufficient space for image capture. The DM ScanBox is an extraoral holding apparatus for the smartphone that attaches to the DM Cheek Retractor. The DM App guides the user through the appropriate steps in order to obtain a complete Scan. This DM App is a mobile application installed on a smartphone; thus, the pictures are captured through the smartphone's main camera allowing the procedure to be totally non-invasive to the patient.
{3}------------------------------------------------
- 3. Analysis of acquired intraoral pictures: The Scan is processed through the Data Analysis Platform. The Data Analysis Platform includes a technical processing phase and clinical processing phase, the latter being a clinical analysis to determine if any event has occurred within the clinical parameters the healthcare professional has set up to be monitored. The Data Analysis Platform uses AI comprising image processing algorithms and neural networks.
- 4. Communication of results of the analysis performed through the Data Analysis Platform are communicated to the healthcare professional through the Dashboard on one hand, and to the patient through the DM App on the other hand. Results are shared with the healthcare professional in an exhaustive fashion, providing them with detailed information. As for the patient, the results are communicated in accordance with the rules defined by the healthcare professional in the Protocol applied to the patient.
DentalMonitoring enables HCPs to adapt the use of the product according to their needs depending on each patient's orthodontic treatment. DentalMonitoring comprises two types of monitoring: 2D Monitoring and 3D Monitoring. Some clinical parameters are specific to either 2D Monitoring, or 3D Monitoring.
#### SUMMARY OF NONCLINICAL/BENCH STUDIES
#### BIOCOMPATIBILITY/MATERIALS
The DM Cheek Retractor were evaluated per ISO 10993-1 (2020): "Biological evaluation of medical devices Part 1: evaluation and testing with a risk management" and ISO 7405-2018 - Dentistry - evaluation of biocompatibility of medical devices used in dentistry. The proposed device passed cytotoxicity tests according to ISO 10993-5, skin sensitization according to ISO 10993-10, and irritation or intracutaneous reactivity according to ISO 10993-23.
#### SOFTWARE
DentalMonitoring has a Modate Level of Concern (LOC). Appropriate software documentation for was provided in compliance with the FDA guidance document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" (issued in May 2005). Verification and validation (V&V) activities were performed and described at the unit and system level.
#### CYBER SECURITY
DentalMonitoring approach to cybersecurity aligns with FDA's 2020 guidance titled, "Guidance on Cybersecurity For Medical Devices" The device also conforms to the cybersecurity requirements identified in Section 524B to the FD&C Act.
#### PERFORMANCE TESTING - BENCH
#### AI/ML standalone testing
Standalone performance of each network and algorithm was evaluated. Neural networks and algorithms were classified into three families according to the outputs they provide. A unique protocol was established per family in order to perform the evaluation, with acceptance criteria adapted to each neural network and algorithm. DentalMonitoring comprises technical networks and algorithms as well as clinical neural networks and algorithms. The standalone testing showed acceptable performance of critical
{4}------------------------------------------------
features to demonstrate the device's ability to perform both it's clinical and technological functions.
## SUMMARY OF CLINICAL INFORMATION
The Sponsor provided data from a Clinical Investigation Program to support the clinical performance of DentalMonitoring.
| Study type | Stud<br>y<br>code | Study name | Overview of targeted<br>indications | List<br>(indication;<br>monitoring plan)<br>couples targeted |
|---------------|-------------------|------------------------------|---------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Retrospective | 21-001 | Retrospective | Qualitative indications<br>with Reference Method<br>compatible with<br>DentalMonitoring picture<br>set review | (dental plaque / food<br>residue: 2D)<br>(gingival recession: 2D)<br>(black triangle: 2D)<br>(closure of extraction<br>space: 2D)<br>(closure of all anterior<br>spaces: 2D)<br>(tracking (seat/unseat):<br>2D)<br>(attachment loss: 2D)<br>(button loss: 2D)<br>(bracket debonding: 2D)<br>(tie loss: 2D)<br>(self-ligating clips: 2D) |
| Prospective | 21-002 | Occlusion | Occlusion indications | (canine class: 3D)<br>(midline deviation: 2D)<br>(midline deviation: 3D)<br>(overbite/open bite: 2D)<br>(overbite/open bite: 3D)<br>(overjet: 2D)<br>(overjet: 3D) |
| Prospective | 21-003 | Archwire<br>&<br>Auxiliaries | Passive archwire and<br>auxiliaries indication | (passive archwire and<br>auxiliaries: 2D)<br>(passive archwire and<br>auxiliaries: 3D) |
| Prospective | 21-004 | Updated 3D<br>Model | Updated 3D Model<br>indication | (Updated 3D Model: 3D) |
| DentalMonitoring Clinical Investigation Program | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
#### Retrospective study
DentalMonitoring evaluated the qualitative indications that were able to have the Reference Method used solely DentalMonitoring picture sets (= Scan). The study was performed using Scans retrospectively collected and involved a total of 15 US orthodontists.
The Reference Method results were generated as follow:
{5}------------------------------------------------
- Scans were reviewed by a panel of three independent orthodontists with an assessment done per tooth ్ల / interdental space.
- In case of non-dominant results, a consensus was established by the same group of three ਼ independent orthodontists. Dominant results and result of the consensus review were merged in order to generate the Reference Method initial outcome.
- -Per indication, results presenting discrepancies between the Reference Method and the Candidate Method, i.e., False Positive results and False Negative results, were reviewed by one of the orthodontists among the 15 that took part in the study.
Data was collected in order to qualify the patient demographics.
{6}------------------------------------------------
## Results per indication
Dental plaque / food residue
| Characteristic | # (%) |
|--------------------------|-------------|
| Age group | |
| 12- 22 years of age | 367 (49.5%) |
| ≥ 22 years of age | 4 (0.6%) |
| ≤12 years old | 369 (49.8%) |
| Unknown | 1 (0.1%) |
| Location | |
| Central | 420 (56.7%) |
| East Coast | 209 (28.2%) |
| US - Other | 5 (0.7%) |
| West Coast | 60 (8.1%) |
| Western | 30 (4.0%) |
| Out of US | 17 (2.3%) |
| Treatment type | |
| Aligners | 415 (56.0%) |
| Brackets - Ceramic | 9 (1.2%) |
| Brackets - Self-ligating | 128 (17.3%) |
| Brackets - Traditional | 135 (18.2%) |
| Mixed | 32 (4.3%) |
| None | 5 (0.7%) |
| Retainers | 17 (2.3%) |
#### Final comparison between DentalMonitoring and Reference
| | | Final comparison between DentalMonitoring and Reference<br>Reference Result | | |
|--------------|------------------------------------|-----------------------------------------------------------------------------|----------|-------|
| | | Positive (Noticeable or Slight) | Negative | Total |
| | Positive (Noticeable or<br>Slight) | 1,874 | 356 | 2,230 |
| DM<br>Result | Negative | 362 | 2,472 | 2,834 |
| | Total | 2,236 | 2,828 | 5,064 |
## Final study results for dental plaque / food residue
| | Result | 95% CI Lower<br>Bound | 95% CI Upper<br>Bound |
|-------------|--------|-----------------------|-----------------------|
| Sensitivity | 82.5% | 79.9% | 84.8% |
| Specificity | 83.2% | 80.5% | 85.5% |
Scans were taken using the following phone manufacturers: Apple, Samsung, Google, Motorola, LG, OnePlus, Huawei, Xiaomi, Essential Phone
{7}------------------------------------------------
## Gingival recession
| Characteristic | # (%) |
|--------------------------|-------------|
| Age group | |
| 12- 22 years of age | 205 (51.3%) |
| ≥ 22 years of age | 194 (48.5%) |
| ≤12 years old | 1 (0.2%) |
| Location | |
| Central | 236 (59%) |
| East Coast | 96 (24%) |
| West Coast | 30 (7.5%) |
| US - Other | 5 (1.2%) |
| Out of US | 12 (3%) |
| Western | 21 (5.3%) |
| Treatment type | |
| Aligners | 205 (51.2%) |
| Retainers | 17 (4.2%) |
| Brackets - Ceramic | 7 (1.8%) |
| Brackets - Self-ligating | 68 (17%) |
| Brackets - Traditional | 82 (20.5%) |
| Mixed treatment | 20 (5%) |
| No treatment | 1 (0.3%) |
# Final comparison between DentalMonitoring and Reference
| | | Reference Result | | |
|--------------|----------|------------------|----------|-------|
| | | Positive | Negative | Total |
| DM<br>Result | Positive | 77 | 18 | 95 |
| DM<br>Result | Negative | 19 | 756 | 775 |
| DM<br>Result | Total | 96 | 774 | 870 |
## Final study results for gingival recession
| | Result | 95% CI Lower<br>Bound | 95% CI Upper<br>Bound |
|-------------|--------|-----------------------|-----------------------|
| Sensitivity | 79.9% | 70.3% | 86.9% |
| Specificity | 97.4% | 95.8% | 98.4% |
Scans were taken using the following phone manufacturers: Apple, Samsung, Google, Motorola, LG, Huawei, Essential Phone, Kyocera, OnePlus, and Wiko.
{8}------------------------------------------------
## Black triangle
| Characteristic | # (%) |
|--------------------------|-------------|
| Age group | |
| 12 - 22 years of age | 233 (62%) |
| ≥ 22 years of age | 140 (37.2%) |
| ≤ 12 years old | 3 (0.8%) |
| Location | |
| Central | 231 (61.4%) |
| East Coast | 97 (25.8%) |
| West Coast | 23 (6.1%) |
| Out of US | 7 (1.9%) |
| Western | 14 (3.7%) |
| US-Other | 4 (1.1%) |
| Treatment type | |
| Aligners | 166 (44.2%) |
| Retainers | 12 (3.2%) |
| Brackets - Ceramic | 9 (2.4%) |
| Brackets - Self-ligating | 79 (21%) |
| Brackets - Traditional | 87 (23.1%) |
| Mixed treatment | 21 (5.6%) |
| No treatment | 2 (0.5%) |
## Final comparison between DentalMonitoring and Reference
| | Reference Result | | | |
|--------------|------------------|----------|----------|-------|
| DM Result | | Positive | Negative | Total |
| | Positive | 179 | 12 | 191 |
| DM<br>Result | Negative | 40 | 906 | 946 |
| | Total | 219 | 918 | 1,137 |
#### Final study results for black triangle
| | Result | 95% CI Lower<br>Bound | 95% CI Upper<br>Bound |
|-------------|--------|-----------------------|-----------------------|
| Sensitivity | 81.0% | 73.9% | 86.5% |
| Specificity | 98.4% | 96.9% | 99.2% |
Scans were taken using the following phone manufacturers: Apple, Samsung, Google, Motorola, LG, OnePlus, Huawei, and Xiaomi.
{9}------------------------------------------------
## Closure of extraction space
| Characteristic | # (%) |
|--------------------------|-------------|
| Age group | |
| 12 -22 years of age | 105 (41.3%) |
| ≥ 22 years of age | 5 (2.0%) |
| ≤ 12 years old | 142 (55.9%) |
| Unknown | 2 (0.8%) |
| Location | |
| Central | 137 (53.9%) |
| East Coast | 67 (26.4%) |
| US - Other | 6 (2.4%) |
| West Coast | 25 (9.8%) |
| Western | 11 (4.3%) |
| out of US | 8 (3.2%) |
| Treatment type | |
| Aligners | 125 (49.2%) |
| Brackets - Ceramic | 3 (1.2%) |
| Brackets - Self-ligating | 55 (21.7%) |
| Brackets - Traditional | 44 (17.3%) |
| Mixed treatment | 16 (6.3%) |
| None | 1 (0.4%) |
| Retainers | 10 (3.9%) |
# Final comparison between DentalMonitoring and Reference
| | | Reference Result | | |
|--------------|----------|------------------|----------|-------|
| | | Positive | Negative | Total |
| DM<br>Result | Positive | 198 | 24 | 222 |
| | Negative | 0 | 256 | 256 |
| | Total | 198 | 280 | 478 |
#### Final study results for closure of extraction space
| | Result | 95% CI Lower Bound | 95% CI Upper Bound |
|-------------|--------|--------------------|--------------------|
| Sensitivity | 100.0% | 94.2% | / |
| Specificity | 91.8% | 87.2% | 94.9% |
Scans were taken using the following phone manufacturers: Apple, Samsung, LG, Google, Motorola, Essential Phone, Huawei, and Wiko.
{10}------------------------------------------------
## Tooth wear
| Characteristic | # (%) |
|--------------------------|-------------|
| Age group | |
| 12 - 22 years of age | 200 (54.8%) |
| ≥ 22 years of age | 164 (44.9%) |
| ≤ 12 years old | 1 (0.3%) |
| Location | |
| Central | 232 (63.6%) |
| East Coast | 83 (22.7%) |
| West Coast | 22 (6.0%) |
| Western | 14 (3.8%) |
| Out of US | 11 (3.0%) |
| US-Other | 3 (0.8%) |
| Treatment type | |
| Aligners | 162 (44.4%) |
| Brackets - Ceramic | 5 (1.4%) |
| Brackets - Self-ligating | 74 (20.3%) |
| Brackets - Traditional | 88 (24.1%) |
| Mixed | 1 (0.3%) |
| Mixed treatment | 22 (6%) |
| None | 1 (0.3%) |
| Retainers | 12 (3.3%) |
Final comparison between DentalMonitoring and Reference
| | Reference Result | | | |
|--------------|------------------|----------|-------|-------|
| | Positive | Negative | Total | |
| DM<br>Result | Positive | 89 | 21 | 110 |
| | Negative | 13 | 943 | 956 |
| | Total | 102 | 964 | 1,066 |
## Final study results for tooth wear
| | Result | 95% CI<br>Lower Bound | 95% CI<br>Upper Bound |
|-------------|--------|-----------------------|-----------------------|
| Sensitivity | 84.5% | 74.1% | 91.2% |
| Specificity | 97.0% | 94.9% | 98.2% |
Scans were taken using the following phone manufacturers: Apple, Samsung, LG, Google, Motorola, Huawei, and Essential Phone.
{11}------------------------------------------------
## Closure of all anterior spaces
| Characteristic | # (%) |
|--------------------------|-------------|
| Age group | |
| 12 - 22 years of age | 297 (57.5%) |
| > 22 years of age | 213 (41.3%) |
| ≤ 12 years old | 4 (0.8%) |
| Unknown | 2 (0.4%) |
| Location | |
| Central | 307 (59.5%) |
| East Coast | 131 (25.4%) |
| West Coast | 38 (7.4%) |
| Out of US | 10 (1.9%) |
| US - Other | 4 (0.8%) |
| Western | 26 (5.0%) |
| Treatment type | |
| Aligners | 248 (48.1%) |
| Brackets - Ceramic | 6 (1.2%) |
| Brackets - Self-ligating | 114 (22.1%) |
| Brackets - Traditional | 106 (20.5%) |
| Mixed | 2 (0.4%) |
| Mixed treatment | 21 (4.1%) |
| None | 6 (1.2%) |
| Retainers | 13 (2.5%) |
#### Final comparison between DentalMonitoring and Reference
| | | Reference Result | | |
|--------------|----------|------------------|----------|-------|
| | | Positive | Negative | Total |
| | Positive | 180 | 88 | 268 |
| DM<br>Result | Negative | 3 | 442 | 445 |
| | Total | 183 | 530 | 713 |
#### Final study results for anterior space closure
| | Result | 95% CI Lower<br>Bound | 95% CI Upper<br>Bound |
|-------------|--------|-----------------------|-----------------------|
| Sensitivity | 98.3% | 94.9% | 99.5% |
| Specificity | 83.3% | 79.9% | 86.3% |
Scans were taken using the following phone manufacturers: Apple, Samsung, Google, Motorola, LG, Huawei, T-Mobile, Essential Phone, and Xiaomi.
{12}------------------------------------------------
#### Tracking (seat/unseat)
| Characteristic | # (%) |
|------------------------|-------------|
| [12; 22[ years of age | 279 (44.8%) |
| ≥ 22 years of age | 332 (53.3%) |
| ≤ 12 years old | 9 (1.4%) |
| Unknown | 3 (0.5%) |
| Location | |
| Central | 286 (45.9%) |
| East Coast | 192 (30.8%) |
| out of US | 17 (2.7%) |
| US - Other | 8 (1.3%) |
| West Coast | 85 (13.7%) |
| Western | 35 (5.6%) |
| Treatment type | |
| Aligners | 581 (93.3%) |
| Brackets - Traditional | 4 (0.6%) |
| Mixed | 1 (0.2%) |
| Mixed treatment | 9 (1.4%) |
| Retainers | 28 (4.5%) |
Clinical performances are expressed as two different Sensitivities and Specificities since the parameter outputs three levels.
- Sensitivity and Specificity of the product in its ability to detect people with the evaluated 미 condition from people without; and
- Sensitivity and Specificity of the product in its ability to discriminate the evaluated condition l at a slight level from a noticeable level
{13}------------------------------------------------
| | | Reference Result | | |
|--------------|------------------------------------|------------------------------------|----------|-------|
| | | Positive (Noticeable<br>or Slight) | Negative | Total |
| | Positive (Noticeable or<br>Slight) | 1,360 | 165 | 1,525 |
| DM<br>Result | Negative | 101 | 1,697 | 1,798 |
| | Total | 1,461 | 1,862 | 3,323 |
## Detection of patients with the evaluated condition (presence/absence)
| | Result | 95% CI Lower Bound | 95% CI Upper Bound |
|-------------|--------|--------------------|--------------------|
| Sensitivity | 93.2% | 91.3% | 94.7% |
| Specificity | 86.2% | 83.4% | 88.6% |
## Detection of patients with the level of the condition (slight/noticeable)
| | | Reference Result | | |
|--------------|------------|------------------|--------|-------|
| | | Noticeable | Slight | Total |
| DM<br>Result | Noticeable | 193 | 99 | 292 |
| | Slight | 18 | 1,050 | 1,068 |
| | Total | 211 | 1,149 | 1,360 |
| | Result | 95% CI Lower<br>Bound | 95% CI Upper<br>Bound |
|-------------|--------|-----------------------|-----------------------|
| Sensitivity | 91.1% | 85.9% | 94.5% |
| Specificity | 90.5% | 87.7% | 92.7% |
Scans were taken using the following phone manufacturers: Apple, Samsung, Google, LG, Motorola, OnePlus, Huawei, Essential Phone, Wiko, and Xiaomi.
{14}------------------------------------------------
## Attachment loss
| Characteristic | # (%) |
|------------------------|-------------|
| Age group | |
| ]12; 22[ years of age | 146 (41.9%) |
| > 22 years of age | 200 (57.5%) |
| ≤ 12 years old | 1 (0.3%) |
| Unknown | 1 (0.3%) |
| Location | |
| Central | 171 (49.1%) |
| East Coast | 109 (31.3%) |
| West Coast | 39 (11.2%) |
| US - Other | 4 (1.2%) |
| Out of US | 8 (2.3%) |
| Western | 17 (4.9%) |
| Treatment type | |
| Aligners | 330 (94.8%) |
| Retainers | 8 (2.3%) |
| Brackets - Traditional | 4 (1.2%) |
| Mixed treatment | 6 (1.7%) |
## Final comparison between DentalMonitoring and Reference
| | | Reference Result | | |
|--------------|----------|------------------|----------|-------|
| | | Positive | Negative | Total |
| DM<br>Result | Positive | 183 | 0 | 183 |
| | Negative | 3 | 579 | 582 |
| | Total | 186 | 579 | 765 |
## Final study results for attachment loss
| | Result | 95% CI Lower<br>Bound | 95% CI Upper<br>Bound |
|-------------|--------|-----------------------|-----------------------|
| Sensitivity | 98.2% | 94.3% | 99.4% |
| Specificity | 100.0% | 98.7% | |
Scans were taken using the following phone manufacturers: Apple, Samsung, Google, Motorola, LG, Xiaomi, OnePlus, and Essential Phone.
{15}------------------------------------------------
## Button loss
| Characteristic | # (%) |
|------------------------|-------------|
| Age group | |
| ]12; 22[ years of age | 200 (52.5%) |
| > 22 years of age | 158 (41.5%) |
| ≤ 12 years old | 20 (5.2%) |
| Unknown | 3 (0.8%) |
| Location | |
| Central | 138 (36.2%) |
| East Coast | 125 (32.8%) |
| West Coast | 92 (24.1%) |
| Out of US | 4 (1%) |
| Western | 16 (4.2%) |
| US - Other | 6 (1.6%) |
| Treatment type | |
| Aligners | 370 (97.1%) |
| Retainers | 1 (0.3%) |
| Brackets - Ceramic | 1 (0.3%) |
| Brackets - Traditional | 3 (0.8%) |
| Mixed treatment | 6 (1.6%) |
## Final comparison between DentalMonitoring and Reference
| | | Reference Result | | |
|--------------|----------|------------------|----------|-------|
| | | Positive | Negative | Total |
| DM<br>Result | Positive | 127 | 5 | 132 |
| | Negative | 2 | 525 | 527 |
| | Total | 129 | 530 | 659 |
## Final study results for button loss
| | Result | 95% CI Lower<br>Bound | 95% CI Upper<br>Bound |
|-------------|--------|-----------------------|-----------------------|
| Sensitivity | 98.4% | 94.0% | 99.6% |
| Specificity | 99.0% | 96.9% | 99.7% |
Scans were taken using the following phone manufacturers: Apple, Samsung, LG, Motorola, Google, and Huawei.
{16}------------------------------------------------
## Bracket debonding
| Characteristic | # (%) |
|--------------------------|-------------|
| Age group | |
| ]12; 22[ years of age | 270 (71.3%) |
| ≥ 22 years of age | 107 (28.2%) |
| ≤ 12 years old | 2 (0.5%) |
| Location | |
| Central | 291 (76.8%) |
| East Coast | 60 (15.8%) |
| West Coast | 16 (4.2%) |
| Out of US | 1 (0.3%) |
| Western | 11 (2.9%) |
| Treatment type | |
| Aligners | 1 (0.3%) |
| Brackets - Ceramic | 10 (2.5%) |
| Brackets - Self-ligating | 171 (45.1%) |
| Brackets - Traditional | 150 (39.6%) |
| Mixed treatment | 45 (11.9%) |
| None | 1 (0.3%) |
| Retainers | 1 (0.3%) |
## Final comparison between DentalMonitoring and Reference
| | | Reference Result | | |
|--------------|----------|------------------|----------|-------|
| | | Positive | Negative | Total |
| DM<br>Result | Positive | 127 | 2 | 129 |
| | Negative | 2 | 528 | 530 |
| | Total | 129 | 530 | 659 |
## Final study results for bracket debonding
| | Result | 95% CI Lower<br>Bound | 95% CI Upper<br>Bound |
|-------------|--------|-----------------------|-----------------------|
| Sensitivity | 98.4% | 93.8% | 99.6% |
| Specificity | 99.6% | 98.5% | 99.9% |
Scans were taken using the following phone manufacturers: Apple, Samsung, Google, LG, Motorola, One Plus.
{17}------------------------------------------------
## Tie loss
| Characteristic | # (%) |
|--------------------------|-------------|
| <b>Age group</b> | |
| ]12; 22[ years of age | 240 (77.4%) |
| $\geq$ 22 years of age | 66 (21.3%) |
| $\leq$ 12 years old | 4 (1.3%) |
| <b>Location</b> | |
| Central | 255 (82.2%) |
| East Coast | 39 (12.6%) |
| West Coast | 10 (3.2%) |
| Western | 3 (1.0%) |
| Out of US | 3 (1.0%) |
| <b>Treatment type</b> | |
| Aligners | 1 (0.3%) |
| Brackets - Ceramic | 13 (4.2%) |
| Brackets - Self-ligating | 5 (1.6%) |
| Brackets - Traditional | 267 (86.2%) |
| Mixed | 1 (0.3%) |
| Mixed treatment | 23 (7.4%) |
# Final comparison between DentalMonitoring and Reference
| | | Reference Result | | |
|--------------|----------|------------------|----------|-------|
| | | Positive | Negative | Total |
| DM<br>Result | Positive | 111 | 15 | 126 |
| | Negative | 7 | 520 | 527 |
| | Total | 118 | 535 | 653 |
## Final study results for tie loss
| | Result | 95% CI Lower<br>Bound | 95% CI Upper<br>Bound |
|-------------|--------|-----------------------|-----------------------|
| Sensitivity | 93.3% | 85.7% | 97.0% |
| Specificity | 96.5% | 94.0% | 98.0% |
Scans were taken using the following phone manufacturers: Apple, Samsung, Motorola, LG, Google.
{18}------------------------------------------------
## Self-ligating clips
| Characteristic | # | (%) |
|--------------------------|-----|---------|
| Age group | | |
| ]12; 22[ years of age | 217 | (66.8%) |
| ≥ 22 years of age | 103 | (31.7%) |
| ≤ 12 years old | 5 | (1.5%) |
| Location | | |
| Central | 225 | (69.3%) |
| East Coast | 69 | (21.2%) |
| Western | 16 | (4.9%) |
| West Coast | 12 | (3.7%) |
| Out of US | 3 | (0.9%) |
| Treatment type | | |
| Brackets - Ceramic | 5 | (1.5%) |
| Brackets - Self-ligating | 262 | (80.6%) |
| Brackets - Traditional | 21 | (6.5%) |
| Mixed | 2 | (0.6%) |
| Mixed treatment | 35 | (10.8%) |
## Final comparison between DentalMonitoring and Reference
| | | Reference Result | | |
|--------------|----------|------------------|----------|-------|
| | | Positive | Negative | Total |
| DM<br>Result | Positive | 71 | 48 | 119 |
| DM<br>Result | Negative | 7 | 521 | 528 |
| DM<br>Result | Total | 78 | 569 | 647 |
## Final study results for self-ligating clip
| | Result | 95% CI Lower<br>Bound | 95% CI Upper<br>Bound |
|-------------|--------|-----------------------|-----------------------|
| Sensitivity | 91.1% | 82.5% | 95.7% |
| Specificity | 88.3% | 84.1% | 91.5% |
Scans were taken using the following phone manufacturers: Apple, Samsung, Google, LG, Motorola, Huawei, Kyocera, and OnePlus.
{19}------------------------------------------------
#### Occlusion study
DentalMonitoring evaluated the occlusion indications using 3D Models acquired with intraoral scanners as input data to the Reference Method. The study was performed using data collected prospectively. The study involved four sites located in the United States: three sites for patient enrollment and one site for generation of the Reference Method results.
The Reference Method results were generated by measuring the occlusion parameters undergoing evaluation on 3D Models of the patients enrolled in the study. Measurements were done manually and performed in three replicates using a CAD/CAM software.
Analysis was performed according to "Measurement Procedure Comparison and Bias Estimation Using Patient Samples; Clinical and Laboratory Standards Institute guideline", EP09c - Third Edition, 2018. Analysis comprised using a :Passing-Bablok regression analysis, including plots of Candidate Method verses Reference Method to calculate the slope, intercept, and correlation coefficient of the plot. Bias at levels of interest per study group were computed at the levels of interest and confidence intervals were calculated with the 95% Bootstrapping.
Data was collected in order to qualify the patient demographics.
{20}------------------------------------------------
## Results per indication
Canine class - 3D Monitoring
A total of 215 patients rendering 297 results were included in the analysis. One patient could render multiple results corresponding to left canine class and right canine class.
| Characteristic | # (%) |
|-------------------------------------------|-------------|
| Gender | |
| Male | 153 (71.2%) |
| Female | 62 (28.8%) |
| Age group | |
| [6; 12] years of age | 23 (10.7%) |
| ]12; 22[ years of age | 75 (34.9%) |
| ≥ 22 years of age | 117 (54.4%) |
| Race* | |
| American Indian or Alaska Native | 13 (6.0%) |
| Asian | 10 (4.7%) |
| Black or African American | 12 (5.6%) |
| Native Hawaiian or Other Pacific Islander | 0 (0.0%) |
| White | 168 (78.1%) |
| Unknown race | 14 (6.5%) |
| Ethnicity | |
| Spanish origin | 32 (14.9%) |
| Not Hispanic or Latino | 183 (85.1%) |
| Income of the household (annual) | |
| Less than \$19,999 | 9 (4.2%) |
| \$20,000-\$37,999 | 11 (5.1%) |
| \$38,000-\$44,999 | 3 (1.4%) |
| \$45,000-\$54,999 | 7 (3.3%) |
| \$55,000-\$64,999 | 14 (6.5%) |
| \$65,000-\$74,999 | 11 (5.1%) |
| \$75,000-\$99,999 | 25 (11.6%) |
| \$100,000 or more | 69 (32.1%) |
| Unknown | 66 (30.7%) |
| Level of education | |
| < Highschool Diploma | 84 (39.1%) |
| Highschool Diploma or Equivalent | 38 (17.6%) |
| Associates Degree | 25 (11.6%) |
| Bachelors Degree | 44 (20.5%) |
| Masters Degree | 20 (9.3%) |
| Doctoral Degree | 4 (1.9%) |
{21}------------------------------------------------
| Slope | 0.95 |
|------------------|-----------------|
| | [0.92, 0.98] |
| Intercept | -0.10 mm |
| | [-0.14, -0.04]* |
| Bias at 0.00 mm | -0.10 mm |
| | [-0.18, 0.00]* |
| Bias at -2.00 mm | -0.3% |
| | [-4.9, 6.8] |
| Bias at 3.00 mm | -9.4% |
| | [-12.9, -5.3] |
| Bias at 5.00 mm | -7.9% |
| | [-11.0, -4.4] |
*Of note, the difference between the two 95% C1 is due to the fact that the bootstrap is used for bias C1 calculation, whereas for the intercept and slope it is the standard CI calculation linked to Passing-Bablok regression that is used.
{22}------------------------------------------------
Midline deviation - 2D Monitoring
A total of 277 patients rendering 277 results were included in the analysis. One patient rendered a unique result.
| Characteristic | # (%) |
|-------------------------------------------|-------------|
| Gender | |
| Male | 191 (69.0%) |
| Female | 86 (31.0%) |
| Age group | |
| [6; 12] years of age | 36 (13.0%) |
| ]12; 22[ years of age | 94 (33.9%) |
| ≥ 22 years of age | 147 (53.1%) |
| Race* | |
| American Indian or Alaska Native | 12 (4.3%) |
| Asian | 11 (4.0%) |
| Black or African American | 26 (9.4%) |
| Native Hawaiian or Other Pacific Islander | 0 (0.0%) |
| White | 215 (77.6%) |
| Unknown race | 16 (5.7%) |
| Ethnicity | |
| Spanish origin | 37 (13.4%) |
| Not Hispanic or Latino | 240 (86.6%) |
| Income of the household (annual) | |
| Less than \$19,999 | 12 (4.3%) |
| \$20,000-\$37,999 | 14 (5.1%) |
| \$38,000-\$44,999 | 6 (2.2%) |
| \$45,000-\$54,999 | 7 (2.5%) |
| \$55,000-\$64,999 | 14 (5.1%) |
| \$65,000-\$74,999 | 18 (6.5%) |
| \$75,000-\$99,999 | 29 (10.5%) |
| \$100,000 or more | 100 (36.1%) |
| Unknown | 77 (27.7%) |
| Level of education | |
| < Highschool Diploma | 100 (36.1%) |
| Highschool Diploma or Equivalent | 54 (19.5%) |
| Associates Degree | 31 (11.2%) |
| Bachelors Degree | 58 (20.9%) |
| Masters Degree | 29 (10.5%) |
| Doctoral Degree | 5 (1.8%) |
{23}------------------------------------------------
| Slope | 0.93 |
|-----------------|---------------------------|
| | [0.89, 0.97] |
| Intercept | 0.0 mm (-0.03) |
| | [0.0, 0.0] [-0.05, 0.01]* |
| Bias at 0.0 mm | 0.0 mm (0.03) |
| | [0.0, 0.0] [-0.07, 0.03]* |
| Bias at -2.0 mm | 5.3% |
| | [-0.8, 10.7] |
| Bias at 2.0 mm | -8.9% |
| | [-13.6, -3.2] |
*Of note, the difference between the two 95% CI is due to the fact that the bootstrap is used for bias CI calculation, whereas for the intercept and slope it is the standard C1 calculation linked to Passing-Bablok regression that is used.
{24}------------------------------------------------
Midline deviation - 3D Monitoring
A total of 291 patients rendering 294 results were included in the analysis. One patient rendered a unique result.
| Characteristic | # (%) |
|-------------------------------------------|-------------|
| Gender | |
| Male | 200 (68.7%) |
| Female | 91 (31.3%) |
| Age group | |
| [6; 12] years of age | 47 (16.2%) |
| ]12; 22[ years of age | 98 (33.6%) |
| ≥ 22 years of age | 146 (50.2%) |
| Race* | |
| American Indian or Alaska Native | 12 (4.1%) |
| Asian | 6 (2.1%) |
| Black or African American | 26 (8.9%) |
| Native Hawaiian or Other Pacific Islander | 0 (0.0%) |
| White | 224 (77.0%) |
| Unknown race | 30 (10.3%) |
| Ethnicity | |
| Spanish origin | 38 (13.1%) |
| Not Hispanic or Latino | 253 (86.9%) |
| Income of the household (annual) | |
| Less than \$19,999 | 12 (4.1%) |
| \$20,000-\$37,999 | 16 (5.5%) |
| \$38,000-\$44,999 | 6 (2.1%) |
| \$45,000-\$54,999 | 7 (2.4%) |
| \$55,000-\$64,999 | 15 (5.2%) |
| \$65,000-\$74,999 | 18 (6.2%) |
| \$75,000-\$99,999 | 33 (11.3%) |
| \$100,000 or more | 101 (34.7%) |
| Unknown | 83 (28.5%) |
| Level of education | |
| < Highschool Diploma | 109 (37.5%) |
| Highschool Diploma or Equivalent | 55 (18.8%) |
| Associates Degree | 36 (12.4%) |
| Bachelors Degree | 61 (21.0%) |
| Masters Degree | 28 (9.6%) |
| Doctoral Degree | 2 (0.7%) |
{25}------------------------------------------------
| Slope | 0.98<br>[0.96, 1.00] |
|------------------|----------------------------|
| Intercept | -0.01 mm<br>[-0.02, 0.01]* |
| Bias at 0.0 mm | -0.01 mm<br>[-0.03, 0.02]* |
| Bias at -2.00 mm | 2.0 %<br>[-0.5, 4.8] |
| Bias at 2.00 mm | -2.7%<br>[-5.2, -0.4] |
*Of note, the difference between the two 95% Cl is due to the fact that the bootstrap is used for bias Cl calculation, whereas for the intercept and slope it is the standard C1 calculation linked to Passing-Bablok regression that is used. Interpretation of bias signs
{26}------------------------------------------------
## Overbite / open bite - 2D Monitoring
A total of 285 patients rendering 301 results were included in the analysis. One patient could render multiple results corresponding to both central incisor couples.
| Characteristic | # (%) |
|-------------------------------------------|-------------|
| Gender | |
| Male | 194 (68.1%) |
| Female | 91 (31.9%) |
| Age group | |
| [6; 12] years of age | 36 (12.7%) |
| ]12; 22[ years of age | 87 (30.5%) |
| ≥ 22 years of age | 162 (56.8%) |
| Race* | |
| American Indian or Alaska Native | 10 (3.5%) |
| Asian | 11 (3.9%) |
| Black or African American | 32 (11.2%) |
| Native Hawaiian or Other Pacific Islander | 0 (0.0%) |
| White | 216 (75.8%) |
| Unknown race | 21 (7.4%) |
| Ethnicity | |
| Spanish origin | 36 (12.6%) |
| Not Hispanic or Latino | 249 (87.4%) |
| Income of the household (annual) | |
| Less than \$19,999 | 11 (3.9%) |
| \$20,000-\$37,999 | 16 (5.5%) |
| \$38,000-\$44,999 | 8 (2.8%) |
| \$45,000-\$54,999 | 7 (2.5%) |
| \$55,000-\$64,999 | 12 (4.2%) |
| \$65,000-\$74,999 | 19 (6.7%) |
| \$75,000-\$99,999 | 33 (11.6%) |
| \$100,000 or more | 103 (36.1%) |
| Unknown | 76 (26.7%) |
| Level of education | |
| < Highschool Diploma | 102 (35.8%) |
| Highschool Diploma or Equivalent | 51 (17.9%) |
| Associates Degree | 31 (10.9%) |
| Bachelors Degree | 65 (22.8%) |
| Masters Degree | 32 (11.2%) |
| Doctoral Degree | 4 (1.4%) |
{27}------------------------------------------------
| Slope | 0.95<br>95% CI = [0.91, 0.99] |
|----------------|--------------------------------|
| Intercept | 0.2 mm<br>95% CI = [0.1, 0.3]* |
| Bias at 0.0 mm | 0.2 mm<br>95% CI = [0.0, 0.3]* |
| Bias at 3.0 mm | 0.9%<br>95% CI = [-2.4, 3.7] |
| Bias at 5.0 mm | -1.3 %<br>95% CI = [-4.1, 1.2] |
| Bias at 7.0 mm | -2.3 %<br>95% CI = [-5.5, 0.5] |
*Of note, the difference between the two 95% CI is due to the fact that the bootstrap is used for bias CI calculation, whereas for the intercept and slope it is the standard CI calculation linked to Passing-Bablok regression that is used.
{28}------------------------------------------------
## Overbite / open bite - 3D Monitoring
A total of 287 patients rendering 298 results were included in the analysis. One patient could render multiple results corresponding to both central incisor couples.
| Characteristic | # (%) |
|-------------------------------------------|-------------|
| Gender | |
| Male | 90 (31.4%) |
| Female | 197 (68.6%) |
| Age group | |
| [6; 12] years of age | 42 (14.6%) |
| ]12; 22[ years of age | 89 (31%) |
| ≥ 22 years of age | 156 (54.4%) |
| Race* | |
| American Indian or Alaska Native | 10 (3.5%) |
| Asian | 10 (3.5%) |
| Black or African American | 31 (10.8%) |
| Native Hawaiian or Other Pacific Islander | 0 (0.0%) |
| White | 217 (75.6%) |
| Unknown race | 26 (9.9%) |
| Ethnicity | |
| Spanish origin | 37 (12.9%) |
| Not Hispanic or Latino | 250 (87.1%) |
| Income of the household (annual) | |
| Less than \$19,999 | 11 (3.8%) |
| \$20,000-\$37,999 | 17 (5.9%) |
| \$38,000-\$44,999 | 8 (2.8%) |
| \$45,000-\$54,999 | 7 (2.4%) |
| \$55,000-\$64,999 | 12 (4.2%) |
| \$65,000-\$74,999 | 20 (7.0%) |
| \$75,000-\$99,999 | 32 (11.2%) |
| \$100,000 or more | 102 (35.5%) |
| Unknown | 78 (27.2%) |
| Level of education | |
| < High School Diploma | 101 (35.2%) |
| High School Diploma or Equivalent | 51 (17.8%) |
| Associates Degree | 33 (11.5%) |
| Bachelors Degree | 71 (24.7%) |
| Masters Degree | 29 (10.1%) |
| Doctoral Degree | 2 (0.7%) |
{29}------------------------------------------------
| Slope | 0.97<br>95% CI = [0.96, 0.99] |
|-----------------|------------------------------------|
| Intercept | 0.01 mm<br>95% CI= [-0.02, 0.05]* |
| Bias at 0.00 mm | 0.01 mm<br>95% CI = [-0.03, 0.05]* |
| Bias at 3.00 mm | -2.2%<br>95% CI = [-3.1, -1.5] |
| Bias at 5.00 mm | -2.4%<br>95% CI = [-3.3, -1.4] |
| Bias at 7.00 mm | -2.4%<br>95% CI = [-3.5, -1.3] |
*Of note, the difference between the two 95% CI is due to the fact that the bootstrap is used for bias CI calculation, whereas for the intercept and slope it is the standard CI calculation linked to Passing-Bablok regression that is used.
{30}------------------------------------------------
## Overjet - 2D Monitoring
A total of 208 patients rendering 245 results were included in the analysis. One patient could render multiple results corresponding to both central incisor couples.
| Characteristic | # (%) |
|-------------------------------------------|-------------|
| Gender | |
| Male | 69 (33.2%) |
| Female | 139 (66.8%) |
| Age group | |
| [6; 12] years of age | 34 (16.3%) |
| ]12; 22[ years of age | 70 (33.7%) |
| ≥22 years of age | 104 (50.0%) |
| Race* | |
| American Indian or Alaska Native | 9 (4.3%) |
| Asian | 8 (3.8%) |
| Black or African American | 23 (11.1%) |
| Native Hawaiian or Other Pacific Islander | 0 (0.0%) |
| White | 159 (76.5%) |
| Unknown race | 16 (7.7%) |
| Ethnicity | |
| Spanish origin | 32 (15.4%) |
| Not Hispanic or Latino | 176 (84.6%) |
| Income of the household (annual) | |
| Less than \$19,999 | 10 (4.8%) |
| \$20,000-\$37,999 | 12 (5.8%) |
| \$38,000-\$44,999 | 6 (2.9%) |
| \$45,000-\$54,999 | 7 (3.4%) |
| \$55,000-\$64,999 | 7 (3.4%) |
| \$65,000-\$74,999 | 11 (5.2%) |
| \$75,000-\$99,999 | 21 (10.1%) |
| \$100,000 or more | 77 (37.0%) |
| Unknown | 57 (27.4%) |
| Level of education | |
| < Highschool Diploma | 87 (41.8%) |
| Highschool Diploma or Equivalent | 35 (16.8%) |
| Associates Degree | 20 (9.6%) |
| Bachelors Degree | 41 (19.8%) |
| Masters Degree | 21 (10.1%) |
| Doctoral Degree | 4 (1.9%) |
{31}------------------------------------------------
| Slope | 0.84<br>95% CI = [0.78, 0.89] |
|----------------|----------------------------------|
| Intercept | -0.3 mm<br>95% CI = [0.1, 0.4]* |
| Bias at 0.0 mm | -0.3 mm<br>95% CI = [0.1, 0.5]* |
| Bias at 3.0 mm | -7.0%<br>95% CI = [-11.5 -3.5] |
| Bias at 5.0 mm | -11.2%<br>95% CI = [-15.6, -7.1] |
| Bias at 7.0 mm | -13.0%<br>95% CI = [-18.2, -8.3] |
| Bias at 9.0 mm | -14.1%<br>95% CI = [-19.8, -8.8] |
*Of note, the difference between the two 95% CI is due to the fact that the bootstrap is used for bias CI calculation, whereas for the intercept and slope it is the standard CI calculation linked to Passing-Bablok regression that is used.
{32}------------------------------------------------
## Overjet - 3D Monitoring
A total of 263 patients rendering 292 results were included in the analysis. One patient could render multiple results corresponding to both central incisor couples.
| Characteristic | # (%) |
|-------------------------------------------|-------------|
| Gender | |
| Male | 83 (31.6%) |
| Female | 180 (68.4%) |
| Age group | |
| [6; 12] years of age | 48 (18.3%) |
| ]12; 22[ years of age | 80 (30.4%) |
| ≥ 22 years of age | 135 (51.3%) |
| Race* | |
| American Indian or Alaska Native | 10 (3.8%) |
| Asian | 9 (3.4%) |
| Black or African American | 29 (11.0%) |
| Native Hawaiian or Other Pacific Islander | 0 (0.0%) |
| White | 202 (77.2%) |
| Unknown race | 17 (6.5%) |
| Ethnicity | |
| Spanish origin | 36 (13.7%) |
| Not Hispanic or Latino | 227 (86.3%) |
| Income of the household (annual) | |
| Less than \$19,999 | 11 (4.2%) |
| \$20,000-\$37,999 | 15 (5.7%) |
| \$38,000-\$44,999 | 7 (2.7%) |
| \$45,000-\$54,999 | 7 (2.7%) |
| \$55,000-\$64,999 | 12 (4.5%) |
| \$65,000-\$74,999 | 17 (6.5%) |
| \$75,000-\$99,999 | 25 (9.5%) |
| \$100,000 or more | 90 (34.2%) |
| Unknown | 79 (30.0%) |
| Level of education | |
| < High School Diploma | 100 (38.0%) |
| Highschool Diploma or Equivalent | 45 (17.1%) |
| Associates Degree | 28 (10.6%) |
| Bachelors Degree | 61 (23.2%) |
| Masters Degree | 24 (9.2%) |
| Doctoral Degree | 5 (1.9%) |
{33}------------------------------------------------
| Slope | 1.03<br>[1.01, 1.05] |
|-----------------|--------------------------|
| Intercept | 0.14 mm<br>[0.09, 0.19]* |
| Bias at 0.00 mm | 0.14 mm<br>[0.07, 0.23]* |
| Bias at 3.00 mm | 7.2%<br>[6.0, 8.9] |
| Bias at 5.00 mm | 5.4%<br>[4.5, 6.7] |
| Bias at 7.00 mm | 4.7%<br>[3.7, 6.0] |
| Bias at 9.00 mm | 4.3%<br>[3.2, 5.6] |
*Of note, the difference between the two 95% CI is due to the fact that the bootstrap is used for bias CI calculation, whereas for the intercept and slope it is the standard CI calculation linked to Passing-Bablok regression that is used.
{34}------------------------------------------------
#### Archwire and auxiliaries study
DentalMonitoring evaluated the passive archwire and auxiliaries indication using 3D Models acquired with intraoral scanners as input data to the Reference Method. The study was performed using data collected prospectively. The study involved six sites located in the United States: five sites for patient enrollment and one site for generation of the Reference Method results.
The Reference Method results were generated by performing a best fit on the acquired 3D Models. The best fit was performed per arch and per patient between the 3D Models acquired at two different timepoints. The best fits were done using a CAD/CAM software.
Data was collected in order to qualify the patient demographics.
{35}------------------------------------------------
| Characteristic | # (%) |
|-------------------------------------------|-----------|
| Gender | |
| Female | 172 (63%) |
| Male | 101 (37%) |
| Age group | |
| [6; 12] years of age | 24 (9%) |
| ]12; 22[ years of age | 197 (72%) |
| ≥22 years of age | 52 (19%) |
| Race* | |
| American Indian or Alaska Native | 8 (3%) |
| Asian | 11 (4%) |
| Black or African American | 48 (18%) |
| Native Hawaiian or Other Pacific Islander | 1 (0.3%) |
| White | 190 (70%) |
| Refused to disclose | 31 (11%) |
| Ethnicity | |
| Spanish origin | 64 (23%) |
| Not Hispanic or Latino | 209 (77%) |
| Income of the household (annual) | |
| Less than \$19,999 | 16 (6%) |
| \$20,000-\$37,999 | 17 (6%) |
| \$38,000-\$44,999 | 16 (6%) |
| \$45,000-\$54,999 | 16 (6%) |
| \$55,000-\$64,999 | 13 (5%) |
| \$65,000-\$74,999 | 10 (4%) |
| \$75,000-\$99,999 | 31 (11%) |
| \$100,000 or more | 76 (28%) |
| Refused to disclose | 78 (28%) |
| Level of education | |
| < Highschool Diploma | 175 (64%) |
| Highschool Diploma or Equivalent | 31 (11%) |
| Associates Degree | 18 (7%) |
| Bachelors Degree | 32 (12%) |
| Masters Degree | 11 (4%) |
| Doctoral Degree | 5 (2%) |
Passive archwire and auxiliaries - 2D Monitoring
{36}------------------------------------------------
| | | Reference Method | | |
|------------------|-------------------------|------------------|--------|-------|
| | | Passive | Active | TOTAL |
| Candidate Method | Positive class: Passive | 357 | 63 | 420 |
| | Negative class: Active | 43 | 267 | 310 |
| | TOTAL | 400 | 330 | 730 |
| | Result | 95% CI Lower<br>Bound | 95% CI Upper<br>Bound |
|-------------|--------|-----------------------|-----------------------|
| Sensitivity | 89.0% | 84.9% | 92.1% |
| Specificity | 80.4% | 75.1% | 84.8% |
{37}------------------------------------------------
| Characteristic | # (%) |
|-------------------------------------------|-----------|
| Gender | |
| Female | 166 (62%) |
| Male | 103 (38%) |
| Age group | |
| [6; 12] years of age | 27 (10%) |
| ]12; 22[ years of age | 196 (73%) |
| ≥ 22 years of age | 46 (17%) |
| Race* | |
| American Indian or Alaska Native | 9 (3%) |
| Asian | 11 (4%) |
| Black or African American | 43 (16%) |
| Native Hawaiian or Other Pacific Islander | 1 (0.4%) |
| White | 189 (70%) |
| Refused to disclose | 31 (12%) |
| Ethnicity | |
| Spanish origin | 64 (24%) |
| Not Hispanic or Latino | 209 (76%) |
| Income of the household (annual) | |
| Less than \$19,999 | 13 (5%) |
| \$20,000-\$37,999 | 14 (5%) |
| \$38,000-\$44,999 | 12 (4%) |
| \$45,000-\$54,999 | 13 (5%) |
| \$55,000-\$64,999 | 15 (6%) |
| \$65,000-\$74,999 | 9 (3%) |
| \$75,000-\$99,999 | 34 (13%) |
| \$100,000 or more | 81 (30%) |
| Refused to disclose | 78 (29%) |
| Level of education | |
| < Highschool Diploma | 174 (65%) |
| Highschool Diploma or Equivalent | 28 (10%) |
| Associates Degree | 18 (7%) |
| Bachelors Degree | 30 (12%) |
| Masters Degree | 12 (4%) |
| Doctoral Degree | 4 (1%) |
Passive archwire and auxiliaries - 3D Monitoring
{38}------------------------------------------------
| | | Reference Method | | |
|------------------|-------------------------|------------------|--------|-------|
| | | Passive | Active | TOTAL |
| Candidate Method | Positive class: Passive | 361 | 48 | 409 |
| | Negative class: Active | 39 | 282 | 321 |
| | TOTAL | 400 | 330 | 730 |
| | Result | 95% CI Lower<br>Bound | 95% CI Upper<br>Bound |
|-------------|--------|-----------------------|-----------------------|
| Sensitivity | 90.4% | 86.7% | 93.2% |
| Specificity | 85.5% | 80.8% | 89.2% |
#### Updated 3D Model study
DentalMonitoring evaluated the Updated 3D Model indication using 3D Models acquired with intraoral scanners as the Reference Method. For each case, they compared the 3D Model generated by Dental Monitoring to the baseline 3D Model using a best fit between the two 3D objects. Each best fit renders a deviation value and this value is then used to quantify the absolute difference between the two 3D objects, which was used to calculate the mean absolute error (MAE). The study was performed using data collected prospectively. The study involved seven sites located in the United States: six sites for patient enrollment and one site for generation of the study results.
The study results were generated by performing a best fit between the acquired 3D Models and the 3D Model generated by DentalMonitoring, i.e. Updated 3D Model. The best fit was performed per arch and per patient. The best fits were done using a CAD/CAM software.
Data was collected in order to qualify the patient demographics.
| Characteristic | # (%) |
|-------------------------------------------|-------------|
| Gender | |
| Female | 80 (32.0%) |
| Male | 170 (68.0%) |
| Age group | |
| ]12; 22[ years of age | 114 (45.6%) |
| ≥ 22 years of age | 136 (54.4%) |
| Race* | |
| American Indian or Alaska Native | 14 (5.6%) |
| Asian | 8 (3.2%) |
| Black or African American | 12 (4.8%) |
| Native Hawaiian or Other Pacific Islander | 0 (0.0%) |
| White | 221 (88.4%) |
| Refused to disclose | 4 (1.6%) |
| Ethnicity | |
| Spanish origin | 22 (8.8%) |
| Not Hispanic or Latino | 226 (90.4%) |
| Unknown | 2 (0.8%) |
| Income of the household (annual) | |
| Less than \$19.999 | 9 (3.6%) |
{39}------------------------------------------------
| \$20,000-\$37,999 | 18 (7.2%) |
|----------------------------------|-------------|
| \$38,000-\$44,999 | 7 (2.8%) |
| \$45,000-\$54,999 | 12 (4.8%) |
| \$55,000-\$64,999 | 12 (4.8%) |
| \$65,000-\$74,999 | 13 (5.2%) |
| \$75,000-\$99,999 | 32 (12.8%) |
| \$100,000 or more | 81 (32.4%) |
| Refused to disclose | 66 (26.4%) |
| Level of education | |
| < Highschool Diploma | 106 (42.1%) |
| Highschool Diploma or Equivalent | 34 (13.5%) |
| Associates Degree | 29 (11.6%) |
| Bachelors Degree | 53 (21.0%) |
| Masters Degree | 24 (9.5%) |
| Doctoral Degree | 5 (2.0%) |
| Unknown | 1 (0.4%) |
*Of note, the race does not correspond to the total number of patients as some patients with mixed races were included in the analysis.
| # of<br>patients | # of<br>results | Mean<br>Absolute<br>Error<br>(MAE) | 95% CI |
|------------------|-----------------|------------------------------------|----------------|
| 250 | 536 | 0.10 | [0.093; 0.103] |
#### Human Factors Testing
A Human factors validation testing was performed to demonstrate the usability of the device. Considering DentalMonitoring has two external user-facing interfaces, i.e. the Dashboard and the DM App, one study was performed for each of these interfaces.
#### Dashboard study
The Dashboard study enrolled two user groups:
- . doctors; and
- . practice employees.
Doctors represented 17 participants while practice employees represented 16 participants.
Testing took place in the United States in three cities, each located in different states.
The testing shows that the Dashboard can be used safely and effectively by the user population. Most of the users used the device properly and there is no significative difference between the different groups of population. Even if some use errors and use difficulties were observed, they cannot cause any harm to the users based on the performed analysis of the root cause. Furthermore, in the Dashboard, no tasks are considered as critical.
#### DM App study
The DM App study was divided into two parts: validation of the healthcare professional-facing interface
De Novo Summary (DEN230035)
{40}------------------------------------------------
which offers some of the features of the Dashboard, and validation of the patient-facing interface.
The DM App healthcare professional-facing interface study enrolled one user group constituted of doctors and practice employees. A total of 17 doctors along with 16 practice employees composed a testing group of 33 participants.
Testing took place in the United States in three cities, each located in different states.
The testing shows that the DM App healthcare professional-facing interface can be used safely and effectively by the user population. Most of the users used the device properly and there is no significative difference between the different groups of population. Furthermore, even if some use errors and use difficulties were observed, they cannot cause any harm to the users based on the performed analysis of the root cause.
The DM App patient-facing interface study enrolled five user groups:
- . adults 22 years of age or older:
- . adolescents between 13 and 21 years old;
- children: pairs of pediatric patients (6 to 12 years of age) and lay caregivers (22 years of age or . older):
- adults with disabilities: pairs of adult patients with disabilities who cannot independently use . the product (22 years of age or older) and lay caregivers (22 years of age or older); and
- healthcare professionals (includes doctors and practice employees). .
| Population group | adults | adolescents | children | adults with<br>disabilities | healthcare<br>professionals |
|-------------------|--------|-------------|----------|-----------------------------|-----------------------------|
| # of participants | 16 | 15 | 15 | 17 | 17 |
Testing took place in the United States in five cities, each located in different states. This study included use of the DM Check Retractor and DM ScanBox which are mandatory hardware products to be used with the DM App in order to appropriately acquire Scans.
The testing shows that the DM ScanBox, the DM Cheek Retractor and the DM App, can be used safely and effectively by the user population. Most of the users used the device properly and there is no significative difference between the different groups of population. Furthermore, some use errors and use difficulties were observed; however, they are not expected to cause any harm to the users based on an analysis of the root cause that was conducted. Users were able to execute the relevant tasks directly or in workarounds within the patient profiles. The task errors identified were found to be non-critical and the testing was able to demonstrate that the device can be adequately used by the health care professional and patient according to the instructions for use.
#### LABELING
Labeling has been included. It comprises a user manual dedicated to healthcare professionals on one hand and a user manual dedicated to patients on the other hand.
Labeling also includes the two distinct labels embedded in the Dashboard, and in the DM App.
#### RISKS TO HEALTH
{41}------------------------------------------------
The table below identifies the risks to health that may be associated with use of and the measures necessary to mitigate these risks.
| Identified Risk | Potential Mitigation Measure |
|-----------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------|
| Adverse tissue reaction | Biocompatibility evaluation |
| Use error/ improper device use leading to<br>incorrect oral health information conveyed to<br>healthcare professional | Labeling<br>Usability testing |
| Software malfunction leading to inaccurate patient<br>monitoring and diagnosis | Software verification, validation, and hazard<br>analysis. |
| Software failure to identify or monitor the correct<br>oral health condition leading to delayed treatment | Clinical validation<br>Software verification, validation, and hazard<br>analysis |
| Device failure to adequately capture image<br>leading to incorrect identification or monitoring of<br>oral health condition | Non-clinical performance testing<br>Technological specifications |
#### SPECIAL CONTROLS
In combination with the general controls of the FD&C Act. the Dental Image Analyzer is subject to the following special controls:
(1) The patient-contacting components of the device must be demonstrated to be biocompatible.
(2) Non-clinical performance testing data must demonstrate that the device performs as intended under anticipated conditions of use and must include full specifications of the hardware requirements and testing to demonstrate the specified hardware ensures adequate data for validated and accurate measurements; (3) Clinical performance data must demonstrate the accuracy and precision of the diagnostic and monitoring device algorithms to perform as intended under anticipated conditions of use.
(4) Usability testing must demonstrate that the user can correctly use the device, based solely on reading the directions for use.
(5) Software verification, validation, and hazard analysis must be performed.
(6) Labeling must include:
(i) Information on the patient population for which the device has been demonstrated to be effective.
(ii) Information on how the device operates during the typical course of treatment
(iii) A detailed summary of the device technical parameters, including compatible operating systems, hardware, and/or accessories.
(iv) Instructions for cleaning of any reusable components.
(v) Warning that the device is not intended to be used independently for diagnosis and that the device does not replace clinical decision making.
- (7) Patient labeling must be provided and must include:
(i) Relevant contraindications, warnings, precautions, adverse effects/complications, and when to seek help from a dental healthcare professional.
(ii) Information on how the device operates during the typical course of treatment.
(iii) A de…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.