EasySep™ Human Bone Marrow CD138 Positive Selection Kit (100-0748)
DEN220090 · Stemcell Technologies Canada, Inc. · QYO · Nov 6, 2023 · Immunology
Device Facts
Record ID
DEN220090
Device Name
EasySep™ Human Bone Marrow CD138 Positive Selection Kit (100-0748)
Applicant
Stemcell Technologies Canada, Inc.
Product Code
QYO · Immunology
Decision Date
Nov 6, 2023
Decision
DENG
Submission Type
Direct
Regulation
21 CFR 866.6120
Device Class
Class 2
Indications for Use
EasySep Human Bone Marrow CD138 Positive Selection Kit is an in vitro diagnostic device intended to enrich CD138+ cells from bone marrow collected from patients diagnosed with multiple myeloma. The CD138+ cells are enriched by immunomagnetic positive selection for use in validated downstream assays. The end-user is responsible for validation of this kit for use with the assay. For in vitro diagnostic use by laboratory professionals.
Device Story
Kit uses immunomagnetic positive selection to enrich CD138+ cells from bone marrow samples; intended for use by laboratory professionals in clinical settings. Input is bone marrow; output is enriched CD138+ cell population for downstream diagnostic assays. Process involves immunomagnetic bead-based selection. Device does not provide health condition information or perform cell enumeration; end-user must validate kit for specific downstream assays. Benefits include standardized pre-analytical enrichment of target cells for improved diagnostic assay performance.
Clinical Evidence
No clinical data provided. Device performance established through design verification and validation studies, including recovery, limit of detection, reproducibility, specificity, and stability testing as required by special controls.
Technological Characteristics
Immunomagnetic cell enrichment kit. Components: monoclonal antibody cocktail in PBS with Fc receptor blocking antibody and Recombinant Human Albumin; dextran-coated magnetic particles (RapidSpheres). Requires external magnet (The Big Easy). Manual operation. Storage: 2-8°C. No software or electronic components.
Indications for Use
Indicated for enrichment of CD138+ cells from bone marrow of patients diagnosed with multiple myeloma for use in validated downstream assays. For prescription use by laboratory professionals.
Regulatory Classification
Identification
A hematopoietic cell enrichment kit is an in vitro diagnostic device intended for the selection and enrichment of specific hematopoietic cells from human whole blood and/or bone marrow collected from patients with hematological malignancies using immunomagnetic bead-based selection. It is intended for use with diagnostic assays as part of the pre-analytical workflow.
Submission Summary (Full Text)
{0}------------------------------------------------
## EVALUATION OF AUTOMATIC CLASS III DESIGNATION FOR EasySep Human Bone Marrow CD138 Positive Selection Kit DECISION SUMMARY
#### I Background Information:
### A De Novo Number
DEN220090
## B Applicant
STEMCELL Technologies Canada Inc.
## C Proprietary and Established Names
EasySep™ Human Bone Marrow CD138 Positive Selection Kit
## D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
|-----------------|--------------------------------------|--------------------|------------|
| QYO | Hematopoietic cell<br>enrichment kit | 21 CFR 866.6120 | Immunology |
#### II Submission/Device Overview:
## A Purpose for Submission:
De Novo request for evaluation of automatic class III designation for EasySep Human Bone Marrow CD138 Positive Selection
## B Measurand:
CD138 (Syndecan-1)
## C Type of Test:
Tool for selection of human bone marrow CD138+ plasma cells
### III Indications for Use:
## A Intended Use(s):
See Indications for Use below.
{1}------------------------------------------------
# B Indication(s) for Use:
EasySep Human Bone Marrow CD138 Positive Selection Kit is an in vitro diagnostic device intended to enrich CD138+ cells from bone marrow collected from patients diagnosed with multiple myeloma. The CD138+ cells are enriched by immunomagnetic positive selection for use in validated downstream assays. The end-user is responsible for validation of this kit for use with the assay. For in vitro diagnostic use by laboratory professionals.
# C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only For in vitro diagnostic use
# D Special Instrument Requirements:
Not applicable: The EasySep Human Bone Marrow CD138 Positive Selection Kit was validated for manual use.
### IV Device/System Characteristics:
# A Device Description:
The EasySep CD138 Human Bone Marrow CD138 Positive Selection Kit contains reagents to enrich CD138 positive (CD138+) human bone marrow cells. The kit contains reagents for up to 60 mL of bone marrow purifications. The kit contents are listed in Table 1. Materials required but not provided are shown in Table 2.
| Component | Packaging<br>Configuration | Storage | Content |
|---------------------------------------------------------------------------------|----------------------------|----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| EasySep Human Bone<br>Marrow CD138 Positive<br>Selection Cocktail<br>(300-0394) | 3 x 1 mL vials | Store at 2 -<br>8°C. Do<br>not freeze. | A combination of monoclonal antibodies in<br>Dulbecco's Phosphate Buffered Saline. Includes an Fc<br>receptor blocking antibody and Recombinant Human<br>Albumin stabilizer. |
| EasySep Dextran<br>RapidSpheres 50105<br>(300-0400) | 3 x 1 mL vials | Store at 2 -<br>8°C. Do<br>not freeze. | A suspension of magnetic particles in water. |
## Table 1. EasySep Human Bone Marrow CD138 Positive Selection Kit Components
### Table 2. Materials required but not provided (Sold separately)
| Component | Configuration | Storage | Content |
|-----------------------------------------------------------------------|---------------------|----------------------------------------|-------------------------------------------------------------------------|
| EasySep Red Blood Cell<br>Lysis Buffer, 10x<br>Concentrate (100-0749) | 1 x 20 mL<br>Bottle | Store at<br>15 - 25°C. | A 10x concentrated red blood cell lysis<br>reagent. |
| EasySep Buffer<br>(100-0780) | 1 x 1 L | Store at<br>2 - 8°C. Do not<br>freeze. | A Dulbecco's Phosphate Buffered Saline<br>based cell separation buffer. |
| "The Big Easy" EasySep<br>Magnet (18001) | 1 | Ambient temperature | Immunomagnetic column-free magnet. |
{2}------------------------------------------------
# B Principle of Operation
The EasySep Human Bone Marrow CD138 Positive Selection Kit is designed to enrich plasma cells expressing the CD138 surface antigen from fresh human whole bone marrow aspirates. Plasma cells are targeted with antibody complexes that crosslink CD138+ cells to dextran-coated magnetic particles. When the bone marrow specimen is placed into an EasySep magnet. CD138+ cells linked to magnetic particles are retained by the magnetic field while CD138- cells remain in suspension. This allows the CD138 negative (CD138-) cell fraction to be removed by pouring off the suspension containing unlabeled, unbound cells. Enriched CD138+ cells are collected by the removal of the sample from the magnet and can then be used in downstream applications.
### V Standards/Guidance Documents Referenced:
CLSI Guideline EP05-A3 Evaluation of Precision of Quantitative Measurement Procedures
CLSI Guideline EP07-A2: Interference Testing in Clinical Chemistry
CLSI Guideline EP25-A Evaluation of Stability of In Vitro Diagnostic Reagents
### VI Performance Characteristics:
# A Analytical Performance:
- 1. Reproducibility:
## a) Reproducibility with Contrived Samples
A study was performed to demonstrate the reproducibility of the enrichment process. Due to challenges with having adequate sample for three sites, a study was performed using contrived samples generated by spiking a multiple myeloma cell line into healthy donor whole blood each day at 3 CD138+ levels based on flow cytometry events (low, medium, and high representing <3%. 3-15%. and >15%. respectively) to create 16 panel members. More panel members were studied at the low and medium levels (6 versus 4 for the high level) to demonstrate enrichment at the more challenging levels. Reproducibility was evaluated at three sites by two operators at each site for 8 non-consecutive days. Each of the three independent sites received six aliguots (from 16 panel members) from the same panel members. Each operator performed enrichments in duplicate using three different reagent lots on each panel member. After the enrichments were completed. all samples were given to a single operator to perform flow cytometry of the enriched and unenriched samples to measure CD138+ cell purity. Two different CD138 frequency contrived specimens were prepared for enrichments on each day.
Table 3 shows a statistical summary of the overall, site-to-site, lot-to-lot variability, and fold-enrichment from the high, medium, and low CD138+ panel members. Enrichment increased a mean of 85.7-fold in the low CD138+ initial frequency samples, 9.3-fold in the CD138+ medium frequency samples, and 3.8-fold in the high starting samples.
{3}------------------------------------------------
| CD138+ Initial<br>Frequency | N | Repeatability<br>CV (%) | Between Site<br>CV (%) | Between Lot CV<br>(%) | Reproducibility<br>CV (%) | Fold Enrichment |
|-----------------------------|---|-------------------------|------------------------|-----------------------|---------------------------|---------------------|
| Mean (Min. - Max) | | | | | | |
| High (> 15%) | 4 | 0.7 (0.5 - 0.9) | 4.0 (2.8 - 5.9) | 0.9 (0.5 - 1.6) | 4.0 (2.8 - 6.0) | 3.8 (2.6 - 5.5) |
| Med (3 - 15%) | 6 | 1.6 (0.8 - 3.0) | 4.2 (2.1 - 6.8) | 2.3 (1.0 - 3.9) | 4.5 (2.2 - 7.0) | 9.3 (5.0 - 15.4) |
| Low (< 3%) | 6 | 5.4 (3.5 - 8.9) | 9.7 (5.5 - 13.8) | 8.6 (4.5 - 14.2) | 11.9 (8.2 - 18.3) | 85.7 (20.2 - 206.0) |
Table 3. EasySep CD138 Reproducibility Study Variability and Fold Enrichment Summary
Table 4 shows the mean plus minimum and maximum CD138+ purity for unenriched and enriched samples from all panel members across the three study sites. Low CD138+ panel members had mean CD138+ cell purities ranging from 0.7% - 1.5% before enrichment and increased to mean CD138+ cell purities of 56.2% - 79.6% after EasySep enrichment. Medium CD138+ panel members had mean CD138+ cell purities ranging from 7.0% - 12.3% before enrichment and increased to mean CD138+ cell purities of 85.5% - 92.6% after EasySep enrichment. High CD138+ panel members had mean CD138+ cell purities ranging from 17.9% - 22.8% before enrichment and increased to mean CD138+ cell purities of 88.5% - 94.0% after EasySep enrichment.
| Target Freq | Panel Member | Unenriched Purity | | | Enriched Purity | | |
|-------------|--------------|-------------------|------|------|-----------------|------|------|
| | | Mean | Min | Max | Mean | Min | Max |
| Low | 1 | 1.5 | 0.8 | 2 | 56.6 | 43.1 | 74.9 |
| | 5 | 1.3 | 0.7 | 2 | 73 | 62.3 | 86.7 |
| | 8 | 0.7 | 0.6 | 0.8 | 56.2 | 43.8 | 66.5 |
| | 12 | 0.7 | 0.4 | 0.9 | 76.9 | 66.4 | 89.3 |
| | 14 | 1.2 | 0.6 | 1.8 | 79.6 | 71.4 | 91.2 |
| | 15 | 0.8 | 0.4 | 1.2 | 74.9 | 57.9 | 82.8 |
| Med | 2 | 9.4 | 7.2 | 11.3 | 89.8 | 79.6 | 95.3 |
| | 3 | 9.2 | 7 | 12.5 | 88.9 | 84.3 | 93 |
| | 7 | 7.3 | 5.3 | 9.9 | 86.3 | 81.5 | 90.4 |
| | 10 | 7 | 5.5 | 8.3 | 86.4 | 75.3 | 93.2 |
| | 13 | 12.3 | 11.1 | 14.3 | 92.6 | 89.7 | 95.4 |
| | 16 | 10.4 | 6.2 | 14.6 | 85.5 | 78.7 | 90.4 |
| High | 4 | 19 | 16.4 | 21.3 | 94 | 89.9 | 96.8 |
| | 6 | 22.8 | 19.3 | 26.6 | 92.3 | 87.8 | 96.4 |
| | 9 | 18 | 17 | 18.6 | 88.5 | 82.9 | 95.2 |
| | 11 | 17.9 | 14.3 | 20 | 93.2 | 90.3 | 96.3 |
Table 4. Unenriched and Enriched CD138+ Cell Purity for Low, Medium, and High CD138+ Panel Members as Determined by FACS Analysis by All Three Study Sites
Separate analyses of variance were performed within each of the three sites and the three kit lots. Table 5 shows analysis results by kit lot and panel member, with precision estimates for repeatability and within lot/between sites. Similarly, Table 6 shows analysis results by site and panel member, with precision estimates for repeatability and within site/between lots. In both sets of analyses, all medium and high CD138+ panel members within site and lot had CV < 15%. Similarly, all low CD138+ panel members within site and lot had CV < 35%, which met the acceptance criteria due to the expected higher variability for low CD138+ starting samples. Thus, the variability is well-controlled within lot (across site) and within site (across lot).
{4}------------------------------------------------
| Table 5. Precision Estimates by Kit Lot | | | | | | | |
|-----------------------------------------|-----------------|-----------------------|---------------------------|---------------|--------|------------|--------|
| Kit Lot | Panel<br>Member | Mean Start<br>Purity* | Mean Enriched<br>Purity** | Repeatability | | Within Lot | |
| 1 | 8 | 0.66 | 60.17 | 2.13 | 3.5 | 5.01 | 8.3 |
| | 12 | 0.71 | 72.48 | 1.21 | 1.7 | 8.39 | 11.6 |
| | 15 | 0.82 | 68.15 | 4.93 | 7.2 | 6.23 | 9.1 |
| | 14 | 1.19 | 77.56 | 1.71 | 2.2 | 4.56 | 5.9 |
| | 5 | 1.26 | 75.47 | 0.38 | 0.5 | 9.86 | 13.1 |
| | 1 | 1.47 | 55.66 | 7.13 | 12.8 | 7.18 | 12.9 |
| | 10 | 6.96 | 87.05 | 3.4 | 3.9 | 5.22 | 6 |
| | 7 | 7.27 | 87.26 | 0.33 | 0.4 | 1.17 | 1.3 |
| | 3 | 9.19 | 88.02 | 1.86 | 2.1 | 2.42 | 2.8 |
| | 2 | 9.44 | 88.91 | 0.65 | 0.7 | 7.47 | 8.4 |
| | 16 | 10.44 | 85.49 | 1.38 | 1.6 | 4.93 | 5.8 |
| | 13 | 12.34 | 92.23 | 0.71 | 0.8 | 2.1 | 2.3 |
| | 11 | 17.9 | 93.32 | 0.14 | 0.2 | 2.49 | 2.7 |
| | 9 | 18.01 | 87.11 | 1.13 | 1.3 | 5.02 | 5.8 |
| | 4 | 18.99 | 93.66 | 0.35 | 0.4 | 2.94 | 3.1 |
| | 6 | 22.84 | 92 | 0.45 | 0.5 | 3.99 | 4.3 |
| 2 | 8 | 0.66 | 55.67 | 1.68 | 3 | 9.79 | 17.6 |
| | 12 | 0.71 | 80.84 | 1.96 | 2.4 | 5.2 | 6.4 |
| | 15 | 0.82 | 79.59 | 1.14 | 1.4 | 1.96 | 2.5 |
| | 14 | 1.19 | 84.84 | 2.08 | 2.4 | 6.17 | 7.3 |
| | 5 | 1.26 | 72.02 | 2.72 | 3.8 | 6.6 | 9.2 |
| | 1 | 1.47 | 64.12 | 2.93 | 4.6 | 9.24 | 14.4 |
| | 10 | 6.96 | 88.4 | 2.54 | 2.9 | 4.74 | 5.4 |
| | 7 | 7.27 | 87.78 | 0.56 | 0.6 | 1.93 | 2.2 |
| | 3 | 9.19 | 90.19 | 0.85 | 0.9 | 2.19 | 2.4 |
| | 2 | 9.44 | 91.62 | 0.84 | 0.9 | 4.65 | 5.1 |
| | 16 | 10.44 | 84.9 | 0.96 | 1.1 | 3.91 | 4.6 |
| | 13 | 12.34 | 93.26 | 1.01 | 1.1 | 2.1 | 2.3 |
| | 11 | 17.9 | 93.36 | 0.78 | 0.8 | 2.54 | 2.7 |
| | 9 | 18.01 | 88.89 | 0.32 | 0.4 | 5.65 | 6.4 |
| | 4 | 18.99 | 94.27 | 0.72 | 0.8 | 3.12 | 3.3 |
| | 6 | 22.84 | 92.08 | 0.38 | 0.4 | 3.96 | 4.3 |
| 3 | 8 | 0.66 | 52.7 | 3.68 | 7 | 6.54 | 12.4 |
| | 12 | 0.71 | 77.45 | 3.58 | 4.6 | 7.67 | 9.9 |
| | 15 | 0.82 | 77.05 | 1.55 | 2 | 3.22 | 4.2 |
| | 14 | 1.19 | 76.55 | 2.44 | 3.2 | 4.08 | 5.3 |
| | 5 | 1.26 | 71.49 | 1.95 | 2.7 | 6.85 | 9.6 |
| | 1 | 1.47 | 50.17 | 1.73 | 3.5 | 6.66 | 13.3 |
| | 10 | 6.96 | 83.71 | 1.16 | 1.4 | 6.84 | 8.2 |
| | 7 | 7.27 | 83.86 | 0.93 | 1.1 | 2.56 | 3.1 |
| | 3 | 9.19 | 88.37 | 0.38 | 0.4 | 3.2 | 3.6 |
| | 2 | 9.44 | 88.72 | 0.69 | 0.8 | 5.9 | 6.7 |
| | 16 | 10.44 | 86.19 | 1.18 | 1.4 | 2.4 | 2.8 |
| | 13 | 12.34 | 92.18 | 0.81 | 0.9 | 1.68 | 1.8 |
| | 11 | 17.9 | 92.97 | 0.26 | 0.3 | 2.7 | 2.9 |
| Kit Lot | Panel<br>Member | Mean Start<br>Purity* | Mean Enriched<br>Purity** | Repeatability | | Within Lot | |
| | | | | SD | CV (%) | SD | CV (%) |
| | 4 | 18.99 | 94.11 | 0.39 | 0.4 | 2.84 | 3 |
| | 6 | 22.84 | 92.88 | 0.86 | 0.9 | 2.81 | 3 |
Table 5. Precision Estimates by Kit Lot
{5}------------------------------------------------
Start purity = unenriched CD138+ cell purity
*Starting purity for the same sample is averaged across differing measurements from the three sites.
**Mean enriched purity is the average purity of enriched samples across three study sites.
| | Panel | Start | Mean Enriched | Repeatability | | Within Site | |
|------|--------|---------|---------------|---------------|--------|-------------|--------|
| Site | Member | Purity* | Purity** | SD | CV (%) | SD | CV (%) |
| 1 | 8 | 0.57 | 60.75 | 2.69 | 4.4 | 3.81 | 6.3 |
| 1 | 12 | 0.79 | 84.73 | 2.98 | 3.5 | 3.13 | 3.7 |
| 1 | 15 | 0.92 | 76.93 | 0.72 | 0.9 | 2.73 | 3.5 |
| 1 | 14 | 1.19 | 76.97 | 2.55 | 3.3 | 3.5 | 4.5 |
| 1 | 5 | 0.65 | 81.18 | 1.08 | 1.3 | 4.6 | 5.7 |
| 1 | 1 | 1.58 | 64.33 | 6.66 | 10.4 | 10.24 | 15.9 |
| 1 | 10 | 7.11 | 87.67 | 0.43 | 0.5 | 2.2 | 2.5 |
| 1 | 7 | 6.59 | 85.89 | 0.71 | 0.8 | 2.2 | 2.6 |
| 1 | 3 | 8.06 | 89.44 | 0.94 | 1.1 | 1.6 | 1.8 |
| 1 | 2 | 9.83 | 93.35 | 0.49 | 0.5 | 0.86 | 0.9 |
| 1 | 16 | 10.52 | 85.13 | 0.59 | 0.7 | 1.12 | 1.3 |
| 1 | 13 | 11.13 | 92.7 | 0.47 | 0.5 | 0.62 | 0.7 |
| 1 | 11 | 19.39 | 92.4 | 0.14 | 0.1 | 0.35 | 0.4 |
| 1 | 9 | 17.03 | 86.78 | 0.9 | 1 | 1.22 | 1.4 |
| 1 | 4 | 19.26 | 96.03 | 0.34 | 0.4 | 0.48 | 0.5 |
| 1 | 6 | 22.61 | 92.49 | 0.35 | 0.4 | 0.73 | 0.8 |
| 2 | 8 | 0.6 | 50.2 | 2.97 | 5.9 | 5.51 | 11 |
| 2 | 12 | 0.4 | 73.25 | 2.41 | 3.3 | 6.6 | 9 |
| 2 | 15 | 0.4 | 75.6 | 1.2 | 1.6 | 7.93 | 10.5 |
| 2 | 14 | 0.6 | 77.93 | 0.88 | 1.1 | 6.25 | 8 |
| 2 | 5 | 1.1 | 71.23 | 0.85 | 1.2 | 2.31 | 3.2 |
| 2 | 1 | 0.8 | 54.1 | 2.59 | 4.8 | 5.16 | 9.5 |
| 2 | 10 | 5.5 | 80.72 | 3.9 | 4.8 | 4.91 | 6.1 |
| 2 | 7 | 5.3 | 84.77 | 0.26 | 0.3 | 2.77 | 3.3 |
| 2 | 3 | 7 | 86.23 | 0.52 | 0.6 | 1.65 | 1.9 |
| 2 | 2 | 7.2 | 82.85 | 0.78 | 0.9 | 3.15 | 3.8 |
| 2 | 16 | 6.2 | 82.32 | 1.85 | 2.3 | 3.08 | 3.7 |
| 2 | 13 | 11.6 | 90.63 | 1.23 | 1.4 | 1.23 | 1.4 |
| 2 | 11 | 14.3 | 91.18 | 0.8 | 0.9 | 0.8 | 0.9 |
| 2 | 9 | 18.6 | 84.37 | 0.31 | 0.4 | 1.37 | 1.6 |
| 2 | 4 | 16.4 | 90.63 | 0.82 | 0.9 | 0.82 | 0.9 |
| 2 | 6 | 19.3 | 88.68 | 0.89 | 1 | 1.27 | 1.4 |
| 3 | 8 | 0.82 | 57.58 | 2.19 | 3.8 | 8.56 | 14.9 |
| 3 | 12 | 0.94 | 72.78 | 1.86 | 2.6 | 4.31 | 5.9 |
| 3 | 15 | 1.15 | 72.25 | 5.1 | 7.1 | 8.58 | 11.9 |
| 3 | 14 | 1.79 | 84.05 | 2.43 | 2.9 | 6.43 | 7.7 |
| 3 | 5 | 2.04 | 66.57 | 3.08 | 4.6 | 3.08 | 4.6 |
| 3 | 1 | 2.03 | 51.52 | 3.38 | 6.6 | 7.76 | 15.1 |
| 3 | 10 | 8.28 | 90.77 | 2 | 2.2 | 2.36 | 2.6 |
| 3 | 7 | 9.93 | 88.23 | 0.85 | 1 | 1.83 | 2.1 |
# Table 6. Precision Estimates by Site
{6}------------------------------------------------
| Site | Panel<br>Member | Start<br>Purity* | Mean Enriched<br>Purity** | Repeatability | | Within Site | |
|------|-----------------|------------------|---------------------------|---------------|-----|-------------|-----|
| | 3 | 12.5 | 90.9 | 1.78 | 2 | 1.78 | 2 |
| | 2 | 11.3 | 93.05 | 0.87 | 0.9 | 1.53 | 1.6 |
| | 16 | 14.6 | 89.13 | 0.67 | 0.7 | 0.78 | 0.9 |
| | 13 | 14.3 | 94.33 | 0.67 | 0.7 | 0.93 | 1 |
| | 11 | 20 | 96.08 | 0.2 | 0.2 | 0.2 | 0.2 |
| | 9 | 18.4 | 94.28 | 1.03 | 1.1 | 1.59 | 1.7 |
| | 4 | 21.3 | 95.37 | 0.1 | 0.1 | 0.37 | 0.4 |
| | 6 | 26.6 | 95.78 | 0.41 | 0.4 | 0.44 | 0.5 |
Start purity = unenriched CD138+ cell purity
*Starting purity for the same sample is averaged across all three sites.
** Mean enriched purity is the average purity of enriched samples across three kit lots.
The CD138 EasySep Reproducibility study showed the EasySep Human Bone Marrow CD138 Positive Selection Kit produced consistent enrichment when starting with a specimen of given volume and initial CD138+ cell frequency across three independent laboratories and three separately manufactured kit lots. The precision statistical analysis of all 16 panel members showed that the CD138 cell enrichment performance across three independent laboratories using three independent kit lots gave fold enrichment and overall variability passing the acceptance criteria, demonstrating reproducibility across sites and kit lots.
## b) Reproducibility with Clinical Specimens
A study was performed to demonstrate the reproducibility of the enrichment process using a validated FISH assay. Three (3) lots of the EasySep Human Bone Marrow CD138 Positive Selection Kit were tested on multiple myeloma (MM) patient bone marrow aspirates (BMA) in duplicate for performance in downstream FISH assays (Table 7). A total of 9 clinical MM BMA were tested, where each specimen was split in half to test enrichment with two different EasySep Human Bone Marrow CD138 Positive Selection Kit lots. The percentage of abnormal nuclei were analyzed using 5 probes which detect common MM chromosomal abnormalities, including the t(11:14) translocation (CCND1/IGH XT), chromosomes 5, 9, and 15 aneusomies (D5S23, D5S72, CEP 9, CEP 15) and 13g14 deletion.
| Specimen | Lot | Specimen | Lot |
|---------------|-------|---------------|-------|
| MM Specimen 1 | Lot 1 | MM Specimen 6 | Lot 2 |
| | Lot 2 | | Lot 3 |
| MM Specimen 2 | Lot 1 | MM Specimen 7 | Lot 1 |
| | Lot 3 | | Lot 2 |
| MM Specimen 3 | Lot 2 | MM Specimen 8 | Lot 1 |
| | Lot 3 | | Lot 3 |
| MM Specimen 4 | Lot 1 | MM Specimen 9 | Lot 2 |
| | Lot 2 | | Lot 3 |
| MM Specimen 5 | Lot 1 | | |
| | Lot 3 | | |
### Table 7. Lot-to-Lot Combinations
{7}------------------------------------------------
The percentage of cells with an abnormal FISH signal pattern(s) for each probe are listed in Table 8 are shown in bold. Three (3) BMA gave normal FISH signal patterns for all five probes tested. One BMA was abnormal for the t(11:14) translocation and 4 BMA had abnormal signal patterns for the 13q deletion. In all nine BMA tested, the percentage of abnormal cells for each probe were very similar between the plasma cells enriched with the two EasySep Human Bone Marrow CD138 Positive Selection Kit lots. The same normal/abnormal disposition for a given probe was concordant between the pairs of EasySep Human Bone Marrow CD138 Positive Selection Kit lots. Cells which are bold in Table 8 were dispositioned as abnormal with abnormal FISH signal patterns. Cells which are not bold were dispositioned as normal with abnormal FISH signal patterns that did not exceed the cutoffs for percentage of abnormal nuclei. In all cases, normal/abnormal disposition agreed, indicating that the EasySep Human Bone Marrow CD138 Positive Selection Kit is able to produce reproducible FISH results when used on the intended use specimen.
| Table 8. FISH Enumeration Results Using Five Probes on MM BMAs Following Enrichment |
|-------------------------------------------------------------------------------------|
| Using the EasySep Human Bone Marrow CD138 Positive Selection Kit |
| Sample<br>No. | Kit Lot | t(11;14)<br>% Abnormal | Trisomy | | | Tetrasomy | | | Monosomy | | | 13q<br>Deletion |
|---------------|---------|------------------------|---------|-------|--------|-----------|-------|--------|----------|-------|--------|-----------------|
| | | | Chr.5 | Chr.9 | Chr.15 | Chr.5 | Chr.9 | Chr.15 | Chr.5 | Chr.9 | Chr.15 | |
| 1 | 1 | 0 | 25.5 | 14 | 4.5 | 6.5 | 1 | 1 | 0.5 | 2 | 3.5 | 4 |
| 1 | 2 | 0 | 27 | 18 | 6.5 | 1.5 | 2 | 0.5 | 1.5 | 1 | 1 | 5 |
| 2 | 1 | 0 | 0 | 88 | 0 | 0 | 1.5 | 0 | 0 | 0 | 10 | 96 |
| 2 | 3 | 0 | 1 | 87.5 | 0 | 0 | 0.5 | 0.5 | 1.5 | 0 | 7.5 | 95.5 |
| 3 | 2 | 0 | 74.5 | 55.5 | 49 | 0.5 | 0 | 0 | 0 | 0 | 0 | 84.5 |
| 3 | 3 | 0 | 72 | 59 | 46.5 | 0 | 2.5 | 1 | 0 | 0 | 0 | 80 |
| 4 | 1 | 0 | 2 | 1 | 1.5 | 0 | 0 | 0 | 2.5 | 2.5 | 1 | 2 |
| 4 | 2 | 0 | 3 | 3 | 1 | 0 | 0 | 0 | 1.5 | 4 | 5.5 | 2.5 |
| 5 | 1 | 0 | 9.5 | 5.5 | 6.5 | 0.5 | 1 | 1 | 6 | 2.5 | 5 | 2.5 |
| 5 | 3 | 0 | 3 | 4.5 | 2 | 0.5 | 0.5 | 0.5 | 2 | 2 | 4 | 2.5 |
| 6 | 2 | 0 | 2 | 0 | 53 | 0 | 0 | 0 | 0.5 | 0 | 1 | 96.5 |
| 6 | 3 | 0 | 0.5 | 1.5 | 38.5 | 0 | 1 | 0 | 0.5 | 2 | 3 | 95.5 |
| 7 | 1 | 89 | 0 | 2 | 2 | 0 | 0 | 0 | 0 | 5.5 | 4 | 90.5 |
| 7 | 2 | 90 | 2.5 | 7 | 2 | 0 | 0 | 0.5 | 2 | 1 | 4 | 91 |
| 8 | 1 | 0 | 0.5 | 1.5 | 1 | 0 | 1 | 0 | 2 | 1.5 | 4 | 1 |
| 8 | 3 | 0 | 1.5 | 0.5 | 1 | 2.5 | 0 | 0 | 1.5 | 1 | 3 | 0.5 |
| 9 | 2 | 0 | 63 | 53 | 35 | 5 | 1.5 | 4.5 | 1.5 | 1 | 0 | 2 |
| 9 | 3 | 0 | 60 | 48.5 | 38.5 | 1 | 0 | 0 | 1 | 1.5 | 3.5 | 1 |
## 2. Repeatability
In this precision study, one operator enriched a low or medium initial CD138+ frequency panel member using one EasySep Human Bone Marrow CD138 Positive Selection Kit lot in quadruplicate over three runs using contrived samples. On three non-consecutive days, a bone marrow specimen from a healthy donor was spiked with the CD138+ cell line SK-MM-2 at a low (< 3%) and a medium (3 - 15%) frequency to generate two separate contrived sample panel members. Two operators performed EasySep enrichments on each of the study days, with each operator enriching from one of the two panel members. Each operator performed three runs of enrichments in quadruplicate on one panel member using one GMP kit lot.
{8}------------------------------------------------
Table 9 shows the unenriched and enriched CD138+ cell purity from all panel members and enriched CD138+ cell purity from all panel members across all three runs. Low initial CD138+ frequency panel members had mean CD138+ cell purities ranging from 1.32 - 1.51% before enrichment and increased to CD138+ cell purities of 80.99% -87.93% after EasySep enrichment. Medium initial CD138+ frequency panel members had CD138+ cell purities ranging from 9.84% - 10.84% before enrichment and increased to mean CD138+ cell purities of 91.10% - 96.65% after EasySep enrichment.
| Table 9. Unenriched and Enriched CD138+ Cell Purity for Low and Medium Initial CD138+ |
|---------------------------------------------------------------------------------------|
| Frequency Panel Members as Determined by Flow Cvtometry Analysis |
| Initial CD138+<br>Frequency Category | Panel<br>Member | Unenriched Purity*<br>(% CD138+) | Enriched Purity (%CD138+) | | | | | |
|--------------------------------------|-----------------|----------------------------------|---------------------------|-------|-------|------|------|----|
| Low | B1 | 1.51 | 84.91 | 83.05 | 87.48 | 1.51 | 1.77 | 12 |
| Low | B4 | 1.32 | 86.77 | 85.65 | 87.93 | 0.73 | 0.84 | 12 |
| Low | B5 | 1.37 | 82.41 | 80.99 | 83.77 | 0.93 | 1.13 | 12 |
| Medium | B2 | 10.84 | 93.82 | 92.75 | 95.62 | 0.90 | 0.96 | 12 |
| Medium | B3 | 9.84 | 95.70 | 94.81 | 96.65 | 0.62 | 0.65 | 12 |
| Medium | B6 | 10.17 | 93.32 | 91.10 | 94.21 | 0.77 | 0.83 | 12 |
Table 10 shows the CD138+ cell recovery for low and medium initial CD138+ frequency panel members and the corresponding statistical analyses. Low initial CD138+ frequency panel members had CD138+ cell recoveries ranging from 34.98% - 76.69% (Mean = 53.42%). Medium initial CD138+ frequency panel members had CD138+ cell recoveries ranging from 36.89% - 65.16% (Mean = 50.62%). The overall CD138+ cell recovery within-laboratory precision estimates (%CV across three runs with four replicates. n = 12 total) ranged from 5.52% - 27.4% and 10.53% - 18.32% for low and medium initial CD138+ frequency panel members respectively.
Table 10. CD138+ Cell Recovery for Low and Medium Initial CD138 Frequency Panel Members
| Initial CD138<br>Frequency Category | Panel<br>Member | Unenriched Purity*<br>(% CD138+) | CD138+ Cell Recovery (%) | | | | | |
|-------------------------------------|-----------------|----------------------------------|--------------------------|-------|-------|-------|-------|----|
| Low | B1 | 1.51 | 55.30 | 34.98 | 76.69 | 15.15 | 27.4 | 12 |
| Low | B4 | 1.32 | 48.88 | 36.63 | 64.37 | 7.92 | 16.2 | 12 |
| Low | B5 | 1.37 | 56.09 | 50.56 | 63.00 | 3.09 | 5.52 | 12 |
| Medium | B2 | 10.84 | 50.78 | 37.58 | 58.57 | 6.64 | 13.07 | 12 |
| Medium | B3 | 9.84 | 49.04 | 36.89 | 57.99 | 5.16 | 10.53 | 12 |
| Medium | B6 | 10.17 | 52.04 | 37.88 | 65.16 | 9.53 | 18.32 | 12 |
* Mean unenriched purity measured across three runs
Table 11 shows the FISH analysis results for the selected enriched samples from each panel member and run. For low initial CD138+ frequency panel members, the number of abnormal nuclei per slide ranged from 85.5% - 94.0% and replicates from all panel members and runs resulted in abnormal FISH signal patterns. For medium initial CD138+ frequency members, the number of abnormal nuclei per slide ranged from 93% -99.5%, and replicates from all panel members and runs resulted in abnormal FISH signal patterns. Therefore, the acceptance criteria for within laboratory precision by FISH analysis were met.
{9}------------------------------------------------
Table 11. FISH Analysis Results for Enriched Samples bv Run for Low and Medium Initial CD138 Frequency Panel Members
| Initial CD138<br>Frequency<br>Category | Panel<br>Member | Run | Replicate<br>(a/b/c/d) | CD 138 Purity (% CD138+) | | FISH Results | |
|----------------------------------------|-----------------|-----|------------------------|--------------------------|----------|----------------------|---------------------|
| | | | | Unenriched | Enriched | # Abnormal<br>Nuclei | Normal/<br>Abnormal |
| Low | B1 | 1 | b | 1.51 | 87.48 | 92.5 | Abnormal |
| | | 2 | d | | 83.42 | 94 | Abnormal |
| | | 3 | a | | 84.41 | 93 | Abnormal |
| | B4 | 1 | d | 1.32 | 85.65 | 90 | Abnormal |
| | | 2 | c | | 87.93 | 89.5 | Abnormal |
| | | 3 | a | | 87.44 | 94 | Abnormal |
| | B5 | 1 | b | 1.37 | 80.99 | 85.5 | Abnormal |
| | | 2 | c | | 83.77 | 86 | Abnormal |
| | | 3 | d | | 82.12 | 89.5 | Abnormal |
| Medium | B2 | 1 | a | 10.84 | 95.62 | 99.5 | Abnormal |
| | | | 2 | b | | 93.57 | 96.5 |
| | | | 3 | d | | 92.75 | 97 |
| | B3 | 1 | b | 9.84 | 95.43 | 94.5 | Abnormal |
| | | 2 | d | | 94.81 | 95.5 | Abnormal |
| | | 3 | c | | 96.14 | 94 | Abnormal |
| | B6 | 1 | a | 10.17 | 91.1 | 93 | Abnormal |
| | | 2 | a | | 93.69 | 93.5 | Abnormal |
| | | 3 | b | | 94.21 | 97.5 | Abnormal |
The CD138 EasySep Precision study showed that the EasySep Human Bone Marrow CD138 Positive Selection Kit produced consistent enrichment within a run and across three runs when starting with a specimen of a given volume and initial CD138+ cell frequency. For low initial CD138+ frequency panel members (n=3) flow cytometry analysis, the repeatability precision estimate (n=4 per run) for enriched CD138+ purity CV ranged from 0.81% to 1.56% and the within-laboratory precision estimate (n=12 across 3 runs) for enriched CD138+ purity CV ranged from 0.84% to 1.77%. For medium initial CD138+ frequency panel members (n=3) flow cytometry analysis, the repeatability precision estimate (n= 4 per run) for enriched CD138+ purity CV ranged from 0.17% to 1.41% and the within-laboratory precision estimate (n=12 across 3 runs) for enriched CD138+ purity CV ranged from 0.65% to 0.96%. All enriched samples sent for FISH analysis (n=3 per panel member, one replicate per run) gave abnormal FISH patterns, demonstrating within-laboratory precision in downstream FISH assays. Overall, the precision of the EasySep Human Bone Marrow CD138 Positive Selection Kit met the acceptance criteria set for the precision study.
- 3. Analytical Specificity/Interference:
Three (3) fresh clinical multiple myeloma bone marrow specimens were tested to determine if excess heparin affected device performance. One specimen was processed at ~20 hours post collection, one specimen was processed at ~24 hours post collection, and the third specimen was processed ~22 hours post collection. Each specimen was split into two fractions. One fraction was treated with phosphate buffered saline (PBS, the control), while the other was spiked with 3x excess of sodium heparin. Single replicates of each fraction were then washed and enriched using the EasySep Human Bone Marrow CD138 Positive Selection Kit. CD138 purity was determined by flow cytometry and then
{10}------------------------------------------------
compared to determine whether excess heparin caused interference with CD138 enrichment and purity. The enriched CD138+ purity from the test samples with excess sodium heparin should be > 90% of the enriched CD138+ purity from the control samples without excess heparin.
Table 12 shows CD138 purity before plasma cell enrichment and after enrichment using the EasySep Human Bone Marrow CD138 Positive Selection Kit for each specimen. For specimen #1, the PBS control sample was enriched to 64.7% and the excess sodium heparin-containing sample was enriched to 58.7%. which is 91.0% of the control. For specimen #2, the PBS control sample was enriched to 32.4% and the excess sodium heparin containing sample was enriched to 33.4%, which is 103% of the PBS control. For specimen #3, the PBS control sample was enriched to 81.7% and the excess sodium heparin-containing sample was enriched to 80.1%. which is 98.0% of the control. All three specimens passed the set acceptance criteria for the study, which was that enriched CD138+ purity from the test samples with excess sodium heparin should be > 90% of the enriched CD138+ purity from the control samples without excess heparin.
| Specimen<br># | Initial CD138<br>Frequency<br>(%) | After Enrichment | | CD138 Purity:<br>Heparin/ PBS<br>Control | Meets Acceptance<br>Criteria?<br>(Pass/Fail) |
|---------------|-----------------------------------|--------------------|------------------|------------------------------------------|----------------------------------------------|
| | | PBS Control<br>(%) | + Heparin<br>(%) | | |
| 1 | 5.0 | 64.7 | 58.7 | 91.0 | PASS |
| 2 | 0.4 | 32.4 | 33.4 | 103.0 | PASS |
| 3 | 31.3 | 81.7 | 80.1 | 98.0 | PASS |
Table 12. Effect of Excess Heparin on CD138 Purity
A second study evaluated a total of 3 clinical specimens prepared by adding MM bone marrow mononuclear cells into healthy donor bone marrow at <0.2% initial CD138 frequency. Three (3) replicates were tested per specimen per condition (i.e., with/without excess sodium heparin). The enriched purities of heparin spiked specimens were not statistically different from that of the control specimens (one tailed t-test, p=0.2740).
Table 13 shows CD138 purity before plasma cell enrichment and after enrichment using the EasySep Human Bone Marrow CD138 Positive Selection Kit for each specimen. The mean enriched purity for Clinical Specimen 1 was 84.45% for the Control Specimen and 83.97% for the Heparin-spiked Specimen, which was 99% of Control. The mean enriched purity for Clinical Specimen 2 was 96.37% for the Control Specimen and 96.25% for the Heparin-spiked Specimen, which was 100% of Control. The mean enriched purity for Clinical Specimen 3 was 88.90% for the Control Specimen and 89.08% for the Heparinspiked Specimen, which was 100% of Control. The enriched purities of heparin spiked specimens were not statistically different from the enriched purities of control specimens (one tailed t-test, p=0.2740, N=3). All three clinical specimens passed the acceptance criteria that enriched CD138+ purity from the test samples with excess sodium heparin should be > 90% of the enriched CD138+ purity from the control samples without excess heparin.
{11}------------------------------------------------
| Clinical<br>Specimen | Target<br>Initial %<br>CD138 | Replicate | Enriched Purity<br>(% CD138+) | | Mean Enriched<br>Purity (% CD138+) | | Heparin/<br>Control (%) | Acceptance<br>Criteria<br>(Pass/Fail) |
|----------------------|------------------------------|-----------|-------------------------------|-------|------------------------------------|-------|-------------------------|---------------------------------------|
| 1 | 0.14 | 1 | 83.52 | 81.69 | 84.45 | 83.97 | 99 | Pass |
| 1 | 0.14 | 2 | 85.26 | 83.23 | | | | |
| 1 | 0.14 | 3 | 84.57 | 86.99 | | | | |
| 2 | 0.14 | 1 | 96.45 | 96.26 | 96.37 | 96.25 | 100 | Pass |
| 2 | 0.14 | 2 | 96.74 | 96.21 | | | | |
| 2 | 0.14 | 3 | 95.92 | 96.27 | | | | |
| 3 | 0.19 | 1 | 89 | 90.05 | 88.90 | 89.08 | 100 | Pass |
| 3 | 0.19 | 2 | 88.79 | 89.24 | | | | |
| 3 | 0.19 | 3 | NA | 87.96 | | | | |
Table 13. Summary of CD138 Purity from Low Frequency Clinical Specimens
Results show the enriched purities from the excess anticoagulant containing samples were not statistically different from those of the control samples. Therefore, the study results demonstrate that there is no interference identified from the excess anticoagulant used when enriching samples with the EasySep Human Bone Marrow CD138 Positive Selection Kit.
## 4. Stability
## Specimen Stability
The CD138 Expression Stability Study aimed to process ~six MM patient BMA at baseline (<24 hours) and at either 44 - 54 hours or ~73 hours post-draw. One kit lot of the EasySep Human Bone Marrow CD138 Positive Selection kit was used for each time point. At each time point, flow cytometry was performed for both unenriched and enriched portions of each sample, and FISH was only performed on the enriched portion of each sample. For FISH analysis, all enriched samples should have conclusive FISH results. Each of the five FISH probes should result in the same FISH disposition (i.e., normal or abnormal).
CD138 purity before and after enrichment and FISH results using five common FISH probes for each time point are shown in Table 14. Specimen #1 had an initial CD138 frequency of 10.6%, which dropped to 3.1% at 73 hours post-collection. Following enrichment, plasma cells were enriched to 89.9% at baseline, and decreased to 62.4% at 73 hours. FISH results for the three probes analyzed did not change across the time points. Specimen #2 had a very low initial CD138 frequency of 0.5%, which decreased below the limit of quantitation (LLoQ) for the flow assay at 44 - 54 hours and further decreased to below the limit of detection (LLoD) for the flow assay at 73 hours postcollection. As such, data are reported as <LLoQ and <LLoD. FISH results were consistent with baseline and 44 - 54 hours, but were more variable at ~73 hours, as there were insufficient cells to obtain conclusive FISH results for two of the probes. The initial CD138 frequency could not be determined at this time point and could therefore be below the LoD of 0.05% that has been determined for the EasySep Human Bone Marrow CD138 Positive Selection Kit. Specimen #3 also had a very low initial CD138 frequency of 0.4%. At ~73 hours post-collection, purity was again below the LLoD for the flow assay and could not be determined. FISH disposition for the three probes analyzed was consistent across the time points.
{12}------------------------------------------------
| Bone Marrow<br>Specimen | FISH Probe | Baseline (<24 h) | 44 - 54 h | ~73 h |
|-------------------------|-------------------------|------------------|-------------------------------|---------------------------|
| 1 | Initial CD138 Frequency | 10.6% | 9.1% | 3.1% |
| | Enriched CD138 Purity | 89.9 | 85.1 | 62.4 |
| | Chr. 13 (D13S319) | Abnormal | Abnormal | Abnormal |
| | CCND1/IGH XT | Normal | Normal | Normal |
| | TP53 | Normal | Normal | Normal |
| 2 | Initial CD138 Frequency | 0.5% | <LLOQ**<br>(not quantifiable) | <LLOD**<br>(not detected) |
| | Enriched CD138 Purity | 36.7 | 30.8 | 28.9 |
| | Chr. 13 (D13S319) | Abnormal | Abnormal | QNS |
| | CCND1/IGH XT | Normal | Normal | Normal |
| | TP53 | Normal | Normal | QNS |
| 3 | Initial CD138 Frequency | 0.4% | Not tested | <LLOD**<br>(not detected) |
| | Enriched CD138 Purity | 32.4 | Not tested | 23.2& |
| | Chr. 13 (D13S319) | Normal | - | Normal |
| | CCND1/IGH XT | Normal | - | Normal |
| | TP53 | Normal | - | Normal |
Table 14. Stability of CD138 Expressing Multiple Myeloma Bone Marrow Aspirates
*QNS - Quantity of cells not sufficient for analysis
** CD138 purity values are below the putative LLoO of 0.14% or LLoD of 0.14% for the flow cytometry assay performed and could not be determined.
& Note: This sample was processed at 69 hours post-collection instead of 73 hours.
Overall, FISH disposition was consistent between baseline and ~73 hours for all three specimens when initial CD138 frequencies were above the putative LLoD for the flow assay (i.e., 0.14%). Although CD138 specimens could be enriched at ~73 hours, FISH signals were not reliable when specimens with low initial CD138 frequencies were too low to be determined to be determined by flow cytometry.
### Kit Stability
### Shelf-Life and In-Use
This study was conducted to establish that the EasySep Human Bone Marrow CD138 Positive Selection Kit is able to maintain its performance characteristics over its defined shelf-life when stored at 5℃ ± 3℃ using three kit lots. The test samples were stored unopened at 5°C ± 3°C. Contrived samples were tested at or to be test at different time points (0, 3, 7, 10, 13, 19 and 25 months). Three kit lots were tested for shelf-life and 1 lot was tested for in-use stability. The in-use testing involved the mixing and removal of 25 uL from the cocktail and particle vials for 20 times over a 7-month duration before the samples were tested. Kits had to demonstrate ability to enrich CD138+ cells to a purity of >95% and >70% recovery.
Stability testing of three lots of EasySep Human Bone Marrow CD138 Positive Selection Kit has met all acceptance criteria at the long-term and in-use storage condition of 5℃ ± 3°C for up to 7 months and at the accelerated storage condition of 25°C ± 2°C/ 60% ± 5% RH for up to 7 days. The data therefore support a 6-month shelf life for EasySep Human Bone Marrow CD138 Positive Selection Kit stored at 5°C ± 3°C. The data also
{13}------------------------------------------------
supports the stability of the products for up to 7 days at 25℃ ± 2℃/ 60% ± 5% relative humidity.
## Shipping
This study summarizes data collected for up to 7 months at the long-term storage condition of 5°C ± 3°C from the shipping stability study conducted on one batch of EasySep Human Bone Marrow CD138 Positive Selection Kit. Shipping stability kits packaged in their final packaging configurations were subjected to summer and winter temperature profiles for 48 hours and tested for enrichment performance and FISH and monitored with data loggers. For the 48-hour Summer Profile Testing, the test article was exposed to temperature test cycles with a recorded maximum temperature of 30.5 °C and minimum of 23.8 ℃. For the 48-hour winter profile testing, the maximum recorded temperature to which the test article was exposed to is 15.1℃ while the minimum recorded temperature is 3.6 ℃. Kits had to demonstrate that CD138+ cell purity had to be >95% with at least 70% of cells recovered. All acceptance criteria were met at the longterm storage condition of 5°C ± 3°C for up to 7 months.
The stability data demonstrated that transport of the EasySep Human Bone Marrow CD138 Positive Selection Kit using the qualified 48-hour shipping configuration does not affect product performance characteristics. The data support a 6-month shelf-life following transport stress. The shipping stability study will continue for up to 25 months.
## 5. Detection Limit:
## a) Limit of Detection with Contrived Samples
The study aimed to determine the lowest initial CD138+ frequency that gives consistent abnormal FISH results in a contrived specimen after CD138+ cell enrichment using the EasySep Human Bone Marrow CD138 Positive Selection Kit. On six non-consecutive days, a whole blood (WB) or BM sample from a healthy donor was spiked with the CD138+ cell line SK-MM-2 at 20% (acceptable range 26% -14% by flow cytometry) CD138+ initial frequency and then serially diluted by a factor of two with WB or BM to achieve ~20%, 10%, 5%, 2.5%, 1.25%, 0.63%, 0.31%, 0.08%, and 0% (i.e., WB or BM not spiked with SK-MM-2 cells). Each sample was enriched in duplicate using the EasySep Human Bone Marrow CD138 Positive Selection Kit and all unenriched and enriched specimens were assessed by flow cytometry.
Table 15 shows the results from the EasySep CD138 Limit of Detection Study including the unenriched and enriched CD138+ cell purity measured by flow cytometry and the t(11:14) disposition analyzed by FISH. All contrived specimens spiked with SK-MM-2 cells, targeting as low as 0.08% initial CD138+ frequency, resulted in abnormal t(11;14) disposition in FISH analysis.
The enriched CD138+ purity measured by flow cytometry positively correlated with the percentage of abnormal nuclei observed in downstream FISH analysis. The average enriched CD138+ purity that resulted in abnormal t(11;14) disposition in FISH was 33% CD138+, enriching from specimens with an average unenriched purity of 0.07% CD138+. For 0.00% initial target CD138+ frequency specimens, four out of the six donors resulted in normal t(11:14) disposition in FISH analysis, however one BM and one WB donor only had one replicate deemed normal and the other replicate inconclusive
{14}------------------------------------------------
due to insufficient number of cells recovered. Since the lowest targeted initial CD138+ frequency specimen achieved abnormal t(11;14) disposition by FISH, the limit of detection (i.e., the lowest initial CD138+ frequency in a contrived specimen that gave a consistent abnormal t(11;14) disposition in FISH after CD138+ enrichment) was determined to be in the range of 0.05 - 0.15% CD138+ measured by flow cytometry. The lower end of the range was below the limit of detection of the flow cytometer. Therefore, the limit of detection was set at the limit of detection of the flow cytometer, which was 0.05% CD138+.
| Target<br>CD138+<br>Initial<br>Frequency | Flow Cytometry Analysis | | | | FISH Analysis | | | | | | | | | |
|------------------------------------------|------------------------------|-------|-------|------------------|----------------------------|-------|-------|----|----------------------------------------------|------|-------|------------------------|-------------|--|
| | Unenriched CD138+ Purity (%) | | | | Enriched CD138+ Purity (%) | | | | %Abnormal Nuclei and t(11;14)<br>Disposition | | | | | |
| | Mean | Min | Max | N | Mean | Min | Max | N | Mean | Min | Max | Normal/<br>Abnormal | N | |
| 0.00% | NA | ND | <LoQ | ND: 5<br><LoQ: 1 | 13.33 | <LoQ | 32.48 | 12 | 0.0 | 0.0 | 0.0 | Normal<br>Inconclusive | N=10<br>N=2 | |
| 0.08% | 0.09* | 0.05* | 0.15* | 6 | 33.03 | 6.31 | 68.07 | 12 | 51.7 | 30.5 | 73.0 | Abnormal | 12 | |
| 0.16% | 0.18* | 0.10* | 0.30* | 6 | 42.57 | 11.52 | 76.2 | 12 | 68.0 | 40.0 | 82.0 | Abnormal | 12 | |
| 0.31% | 0.32* | 0.21* | 0.40* | 6 | 55.42 | 20.73 | 83.82 | 12 | 83.4 | 73.5 | 90.5 | Abnormal | 12 | |
| 0.63% | 0.69 | 0.41 | 0.94 | 6 | 67.2 | 30.97 | 90.12 | 12 | 87.6 | 80.0 | 92.5 | Abnormal | 12 | |
| 1.25% | 1.38 | 0.87 | 1.86 | 6 | 76.55 | 51.55 | 90.43 | 12 | 93.1 | 85.5 | 97.0 | Abnormal | 12 | |
| 2.50% | 2.17 | 1.41 | 2.84 | 6 | 87.07 | 76.24 | 93.86 | 12 | 95.8 | 89.5 | 99.0 | Abnormal | 12 | |
| 5% | 5.28 | 3.37 | 7.25 | 6 | 91.71 | 88.26 | 95.01 | 12 | 97.0 | 93.5 | 100.0 | Abnormal | 12 | |
| 10% | 10.71 | 8.16 | 13.21 | 6 | 94.39 | 91.2 | 96.91 | 12 | 98.0 | 96.5 | 99.5 | Abnormal | 12 | |
| 20% | 20.01 | 14.73 | 25.06 | 6 | 95.01 | 92.93 | 98.2 | 12 | 98.1 | 96.5 | 100.0 | Abnormal | 12 | |
Table 15. EasySep CD138 Limit of Detection Flow Cytometry and FISH Analysis Result Summary
Table 16 shows the unenriched and enriched purities for BM and WB contrived specimens for each target initial frequency. The healthy donor BM only specimens (0% initial CD138+ frequency) had enriched CD138+ purities in the range of 16.7% - 32.5%.
| Table 16. Unenriched and Enriched Purities for BM and WB Contrived Specimens for Each | | | |
|---------------------------------------------------------------------------------------|--|--|--|
| Target Initial Frequency | | | |
| Target<br>Initial<br>CD138+<br>Frequency | Sample Type | | | | | | | |
|------------------------------------------|-----------------------------------------|---------------------------------------|--------------------------------------|--------------------------------------|-----------------------------------------|---------------------------------------|--------------------------------------|--------------------------------------|
| | BM | | WB | | | | | |
| | Unenriched Purity<br>(% CD138+)<br>Mean | Enriched Purity<br>(% CD138+)<br>Mean | Enriched Purity<br>(% CD138+)<br>Min | Enriched Purity<br>(% CD138+)<br>Max | Unenriched Purity<br>(% CD138+)<br>Mean | Enriched Purity<br>(% CD138+)<br>Mean | Enriched Purity<br>(% CD138+)<br>Min | Enriched Purity<br>(% CD138+)<br>Max |
| 0 | 0.04 | 26.21 | 16.66 | 32.48 | 0.00 | 0.46 | 0.10 | 0.98 |
| 0.08 | 0.07 | 54.67 | 39.73 | 68.07 | 0.07 | 11.40 | 6.31 | 16.10 |
| 0.16 | 0.18 | 65.4 | 54.28 | 76.2 | 0.12 | 19.75 | 11.52 | 33.01 |
| 0.31 | 0.31 | 75.61 | 69.86 | 83.82 | 0.21 | 35.23 | 20.73 | 55.57 |
| 0.63 | 0.71 | 82.35 | 74.86 | 90.12 | 0.67 | 52.05 | 30.97 | 72.51 |
| 1.25 | 1.29 | 85.35 | 78.84 | 90.43 | 1.47 | 67.74 | 51.55 | 81.51 |
| 2.5 | 2.33 | 89.78 | 83.92 | 93.86 | 2.01 | 84.37 | 76.24 | 91.95 |
| 5 | 5.82 | 91.87 | 89.21 | 95.01 | 4.74 | 91.56 | 88.26 | 93.93 |
| 10 | 10.41 | 93.75 | 91.20 | 95.94 | 11.01 | 95.03 | 92.83 | 96.91 |
| 20 | 19.84 | 94.34 | 92.93 | 96.87 | 20.18 | 95.68 | 94.06 | 98.2 |
{15}------------------------------------------------
### b) Limit of Detection with Clinical Multiple Myeloma Bone Marrow Specimens
Five (5) low CD138 frequency specimens with pre-enrichment frequencies of 0.2%, 0.5%, 0.7%, 1.5%, and 1.7% CD138+ (Table 17). All five specimens were successfully enriched to purity ≥ 18.5% CD138+ with fold enrichment > 29.6-fold. All five specimens had an increase in the percentage of abnormal nuclei in FISH analysis for at least one FISH probe. Three of the five specimens only had percentages of abnormal nuclei above FISH cutoff after plasma cell enrichments (highlighted in bold). The data confirm that clinical MM BM specimens with initial CD138+ frequency close to the limit of detection set in this study were able to have plasma cells successfully enriched and gave interpretable FISH results.
## Table 17. CD138 Purity and Percentage of Abnormal Nuclei Detected by FISH Probes Pre- and Post-EasySep CD138 Enrichment for Low CD138 Initial Frequency Multiple Myeloma Bone Marrow Aspirates
| MM<br>Patient<br>Sample | CD138<br>Freq* | Pre- Enr<br>CD138<br>Purity | Enr<br>CD138<br>Purity | Fold<br>Enr& | % Abnormal Nuclei | | | | | | | | | | | |
|-------------------------|----------------|-----------------------------|------------------------|--------------|------------------------------------|-----|------------------------------------|-----|----------------------------------|-----|-----------------------------------------------|-----|-----------------------------------------|------|---------------------------------------|-----|
| | | | | | Probe<br>(D5S23)<br>Cutoff =<br>1% | | Probe<br>(D15Z4)<br>Cutoff =<br>1% | | Probe<br>(MYC)<br>Cutoff =<br>3% | | Probe<br>(D13S319,<br>13qter) Cutoff<br>= 14% | | Probe<br>(IGH/CCND1 XT)<br>Cutoff=3-15% | | Probe (TP53,<br>D17Z1) Cutoff<br>= 9% | |
| | | | | | Pre-<br>Enr | Enr | Pre-<br>Enr | Enr | Pre-<br>Enr | Enr | Pre-<br>Enr | Enr | Pre- Enr | Enr | Pre-<br>Enr | Enr |
| BM 118 | low | 0.7 | 64.5 | 91.1 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 4 | 85.5 | 0 | 0 |
| BM 121 | low | 0.2 | 18.5 | 91.5 | 1.5 | 48 | 0 | 0 | 0 | 0 | 0 | 50 | 0 | 0 | 0 | 0 |
| BM 122 | low | 1.7 | 52.0 | 29.6 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 80 | 0 | 0 | 0 | 0 |
| 5 | low | 1.5 | 80.6 | 52.7 | 0 | 0 | 0 | 4 | 0 | 0 | 0 | 49 | 0 | 77 | 0 | 0 |
| 8 | low | 0.5 | 32.8 | 64.6 | 5 | 53 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 |
Enr = enrichment
*low = < 3% CD138 start frequency
& Fold enrichment is calculated using the following formula: [Enriched %CD138 Purity / Pre-enriched %CD138 Purity] - 1
The limit of detection study showed that the EasySep Human Bone Marrow CD138 Positive Selection Kit was able to consistently enrich from contrived specimens with initial CD138+ frequency of 0.02% - 25.06% and achieved consistent abnormal FISH patterns even in the lowest initial target CD138+ frequency specimens spiked with the SK-MM-2 cell line. Therefore, the limit of detection for the EasySep Human Bone Marrow CD138 Positive Selection Kit was set to 0.05% of initial CD138+ frequency. which is the limit of detection of the flow cytometer used.
## c) Verification of Limit of Detection with Clinical Multiple Myeloma Bone Marrow Specimens
The goal of this study was to verify the LoD at 0.05% initial CD138 frequency, determined in a previous LoD study on contrived specimens in one clinical specimen using two kit lots, having 3 replicates per kit lot. CD138+ cell enrichment was performed on a clinical specimen produced by spiking a multiple myeloma patient BM aspirate into a healthy donor BM aspirate at 0.05% initial CD138 frequency. The clinical specimen was washed and split into 6 fractions for EasySep CD138 cell enrichments, using two kit lots with three replicates per kit lot. Enriched samples were split into two fractions: one fraction was used for plasma cell purity assessment by flow cytometry and the second fraction was used for FISH.
{16}------------------------------------------------
Table 18 shows plasma cell purity before and after enrichment using two kit lots of the EasySep Human Bone Marrow CD138 Positive Selection Kit with three replicates per kit lot. The specimen was enriched from 0.05% to a range of 4.1% to 7.4%. Overall enriched purity CV was 19.7%, passing the acceptance criteria of overall CV ≤ 35%. All enriched samples had normal FISH signal patterns for Chromosome 5 and 15 polysomy (D5S23. D15Z4), MYC breakapart, and the CCND1/IGH XT probes; and were abnormal for Chromosome 13q deletion (D13S319) and TP53. The enriched samples' FISH disposition was the same as the undiluted clinical specimen's FISH disposition, with 100% of the replicates being abnormal for Chromosome 13g deletion (D13S319) and TP53.
| Sample | Unenriched<br>Plasma Cell<br>Purity (%<br>per 105<br>events) | Enriched Plasma Cell Purity by<br>Flow Cytometry | | | | Genomic Abnormality by FISH | | | | | |
|---------------------|--------------------------------------------------------------|--------------------------------------------------|------|-----|------|-----------------------------|--------|--------|--------------------|----------------|------------------------|
| | | % Plasma<br>Cell per<br>105<br>events | Mean | SD | %CV | D5S23 | D15Z4 | MYC | D13S319,<br>13qter | TP53,<br>D17Z1 | IGH/CC<br>ND1<br>(LDT) |
| Undiluted<br>MM BMA | 12.9 | - | - | - | - | Normal | Normal | Normal | Abnormal | Abnormal | Normal |
| Lot 1 - A | 0.05 | 4.1 | 6.2 | 1.2 | 19.7 | Normal | Normal | Normal | Abnormal | Abnormal | Normal |
| Lot 1 - B | | 7.4 | | | | Normal | Normal | Normal | Abnormal | Abnormal | Normal |
| Lot 1 - C | | 7.0 | | | | Normal | Normal | Normal | Abnormal | Abnormal | Normal |
| Lot 2 - A | | 5.8 | | | | Normal | Normal | Normal | Abnormal | Abnormal | Normal |
| Lot 2 - B | | 7.0 | | | |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.