BT-001 is a prescription-only digital therapeutic device intended to provide cognitive behavioral therapy to patients 18 years or older with type 2 diabetes. The device targets behavior to aid in the management of type 2 diabetes in patients who are under the care of a healthcare provider. BT-001 provides cognitive behavioral therapy as a treatment that should be used adjunctively with standard of care.
Device Story
BT-001 is a prescription-only digital therapeutic software application for Android smartphones. It delivers self-directed cognitive behavioral therapy (CBT) to patients with type 2 diabetes to assist in behavioral modifications related to diabetes management (e.g., nutrition, exercise, medication adherence). The device allows users to self-report meals, exercise, medications, and biometrics, and provides goal setting and progress visualization. It is intended for use in 90-day increments as an adjunct to standard of care. The device does not provide real-time clinical intervention but may prompt users to contact their healthcare provider based on logged data. It is not a standalone therapy or a substitute for medication. Healthcare providers use the patient's progress and self-reported data to inform clinical decision-making. The device aims to provide secondary benefit by supporting adherence to diabetes management behaviors.
Clinical Evidence
Prospective, multicenter, randomized, open-label trial (n=726; 668 analyzed). Intervention (n=325) used BT-001 + standard of care (SOC); control (n=343) used SOC. Primary endpoint: mean change in HbA1c from baseline at Day 90. Results: ITT population Day 90 HbA1c change was -0.27% (BT-001) vs 0.14% (control); Day 180 change was -0.37% (BT-001) vs -0.07% (control). Differences were statistically significant (p<0.0001). Adverse events were similar between groups; four events (tachycardia, thirst, pollakiuria, increased HbA1c) were possibly/probably device-related.
Technological Characteristics
Software-only device (mobile application) running on Android smartphones. Architecture includes smartphone App, Web Server, and Admin web application. Connectivity via APIs. No physical materials or energy sources. Cybersecurity controls include authentication and password retrieval.
Indications for Use
Indicated for patients 18 years or older with type 2 diabetes under the care of a healthcare provider. Contraindicated for individuals with unstable psychiatric disorders.
Regulatory Classification
Identification
Diabetes digital behavioral therapeutic device. A diabetes digital behavioral therapeutic device is a prescription use software device that provides digital behavioral therapy to aid in the management of diabetes. This device is intended to provide limited secondary benefit to patients with diabetes mellitus by assisting them in managing their condition. This device is not intended to replace any primary treatment, such as diet/lifestyle changes or medication.
Special Controls
In combination with the general controls of the FD&C Act, the diabetes digital behavioral therapeutic device is subject to the following special controls:
- (1) Design verification and validation must include documentation of:
- (i) Clinical data from a statistically and clinically justified sample size, fulfilling the following:
- (A) Appropriately validating the model of therapy as implemented by the device using a clinically defined endpoint, and
- (B) Demonstrating that use of the device does not adversely impact the health outcomes or health status of the intended use population. A device hazard analysis must consider all device-related adverse events observed from the clinical data collected and must demonstrate that patient risk from use of the device is minimal.
- (ii) Software verification, validation, and hazard analysis must demonstrate that the device performs as intended.
- (2) The labeling must include:
- (i) A summary of the clinical testing with the device, including a discussion of the limitations of the clinical significance of the results.
- (ii) Limiting statements that indicate:
- (A) The device is not intended for use as a standalone therapy.
- (B) The device is not a substitute for a patient's prescribed therapy or medication.
- (C) The device should not be used by people with unstable psychiatric disorders.
- (D) The device is not intended for use in the treatment of any psychiatric disorder or symptoms.
Submission Summary (Full Text)
{0}------------------------------------------------
### DE NOVO CLASSIFICATION REQUEST FOR BT-001
#### REGULATORY INFORMATION
FDA identifies this generic type of device as:
Diabetes digital behavioral therapeutic device. A diabetes digital behavioral therapeutic device is a prescription use software device that provides digital behavioral therapy to aid in the management of diabetes. This device is intended to provide limited secondary benefit to patients with diabetes mellitus by assisting them in managing their condition. This device is not intended to replace any primary treatment, such as diet/lifestyle changes or medication.
NEW REGULATION NUMBER: 21 CFR 880.5735
CLASSIFICATION: Class II
PRODUCT CODE: QXC
#### BACKGROUND
DEVICE NAME: BT-001
SUBMISSION NUMBER: DEN220058
DATE DE NOVO RECEIVED: September 21, 2022
#### SPONSOR INFORMATION:
Better Therapeutics 548 Market Street, #49404 San Francisco, CA 94104
#### INDICATIONS FOR USE
The BT-001 is indicated as follows:
BT-001 is a prescription-only digital therapeutic device intended to provide cognitive behavioral therapy to patients 18 years or older with type 2 diabetes. The device targets behavior to aid in the management of type 2 diabetes in patients who are under the care of a healthcare provider. BT-001 provides cognitive behavioral therapy as a treatment that should be used adjunctively with standard of care.
{1}------------------------------------------------
# LIMITATIONS
The sale, distribution, and use of BT-001 are restricted to prescription use in accordance with 21 CFR 801.109.
The device is not intended for use as a standalone therapy.
The device is not a substitute for a patient's prescribed therapy or medication.
The device should not be used by people with unstable psychiatric disorders.
The device is not intended for use in the treatment of any psychiatric disorder or symptoms.
# PLEASE REFER TO THE LABELING FOR A COMPLETE LIST OF WARNINGS, PRECAUTIONS AND CONTRAINDICATIONS.
### DEVICE DESCRIPTION
BT-001 is a digital diabetes device that delivers cognitive behavioral therapy (CBT) to a patient with Type 2 Diabetes through an application on the patient's personal Android smartphone. BT-001 is prescription use only and is completely self-directed. The device is designed to allow patients to complete CBT without human support or intervention and is intended to be used in 90-day increments adjunctively to standard of care.
The cognitive behavioral therapy delivered by BT-001 is intended to aid users in making behavioral modifications to better adhere to behaviors associated with the management of diabetes. BT-001 focuses on users understanding barriers to their adhering to known diabetes management behaviors such as eating habits. nutrition, and exercise. The device also provides a way for users to self-report their meals, exercise, medications, and biometrics. When relevant, BT-001 will supply additional notifications for users to contact a healthcare provider, such as when a user logs a blood glucose reading.
This medical device has functions subject to FDA premarket review as well as functions that are not subject to FDA premarket review. For this De Novo request, if the product has functions that are not subject to FDA premarket review. FDA assessed those functions only to the extent that they either could adversely impact the safety and effectiveness of the functions subject to FDA premarket review or they are included as a labeled positive impact that was considered in the assessment of the functions subject to FDA premarket review.
### SUMMARY OF NONCLINICAL/BENCH STUDIES
### SOFTWARE
The device is a software only device accessible through a user's personal Android smartphone. The software of the device consists of the smartphone application (App), Web Server, and Admin. The App is the user interface for the device and features
{2}------------------------------------------------
account management, onboarding, behavioral therapy and program content, biometric entry and tracking, goal setting, and progress visualization, help and support, and user data management. The Web application provides verification of eligibility, registration of their device, and authentication and password retrieval. The Server is a web-based application that provides core functionality to the App and Web via various application programming interfaces (APIs). The Admin is a web-based application that provides administrative functionality for BT-001. Access to the Admin functionality is only available to Administrators of Better Therapeutics.
All of the elements of software and cybersecurity information as outlined in FDA's guidance documents "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" (issued May 11, 2005) and "Content of Premarket Submissions for management of Cybersecurity in Medical Devices" (issued October 2, 2014) were provided.
### PERFORMANCE TESTING - BENCH
Not applicable
#### PERFORMANCE TESTING - ANIMAL AND/OR CADAVER
Not applicable
#### SUMMARY OF CLINICAL INFORMATION
#### Pivotal Study:
The sponsor conducted a controlled, prospective, multicenter pivotal clinical trial consisting of 726 participants where 668 downloaded their assigned digital app and completed the onboarding process. The control group consisted of 343 people with type 2 diabetes who followed their prescribed standard of care therapy and the intervention group consisted of 325 people with type 2 diabetes who used the BT-001 device in addition to their prescribed standard of care therapy. Both groups were allowed to have glycemic control medications changed throughout the study by their healthcare provider.
| Study Feature | Description |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Title | Open-Label, Randomized, Controlled, Parallel-Group Trial of<br>a Digital Therapeutic for the Treatment of Type 2 Diabetes |
| Summary | A randomized, open label, controlled trial of 6 months at home<br>BT-001 vs. standard of care (SOC). |
| Investigational Device | BT-001 |
| Objectives | • The primary objective was to assess the safety and<br>effectiveness of BT-001 as an intervention for type 2<br>diabetes as measured by a change in HbA1c and the |
{3}------------------------------------------------
| | • occurrence, relatedness and severity of adverse events<br>after 90 days.<br>• The secondary objective was to assess the safety and<br>effectiveness of BT-001 after 180 days. |
|----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Study Design | Randomized Open-Label clinical trial with 343 control<br>subjects and 325 intervention subjects |
| Number of Sites | Fourteen US clinical sites |
| Population | Eligible subjects were patients aged 18 to 75 years with type 2<br>diabetes and HbA1c 7.0 - 10.9%. Exclusion criteria included<br>medical conditions or medications likely to affect HbA1c such<br>as active eating disorders, prior bariatric surgery, pregnancy,<br>oral corticosteroids, prandial insulin, weight loss or atypical<br>antipsychotic medications or Patient Health Questionnaire-9<br>(PHQ-9) scores ≥ 20 |
| Sample Size | The study randomized 726 participants, of whom 668<br>downloaded their assigned digital app and completed the<br>onboarding process (Control group: 343, BT-001 group: 325) |
| Treatment Group | Randomized Trial:<br>• Intervention Group: BT-001 use and daily biometric<br>tracking<br>• Control Group: Standard of Care |
| Study Duration | 6 months total |
| Protocol Overview/Synopsis | • Primary endpoint: difference of control and<br>intervention groups in mean change from baseline of<br>A1c at Day 90 between groups<br>• Secondary endpoint: difference of control and<br>intervention groups in mean change from baseline of<br>A1c at Day 180 between groups |
Baseline study demographics are summarized in the table below for the control and intervention (BT-001) groups.
| | Control<br>n=313 | BT-001<br>n=297 |
|--------------------------------------|------------------|-----------------|
| Age, years, mean (SD) | 58 (9) | 58 (9) |
| Female, n (%) | 180 (58) | 166 (56) |
| Race, n (%) | | |
| White | 192 (61) | 183 (62) |
| Black or African American | 93 (30) | 89 (30) |
| Asian | 16 (5) | 13 (4) |
| American Indian or Alaskan<br>Native | 6 (2) | 3 (1) |
{4}------------------------------------------------
| Native Hawaiian or another Pacific<br>Islander | 2 (1) | 1 (0.3) |
|----------------------------------------------------|-----------|-----------|
| Other (includes multiple races) or<br>not reported | 10 (3) | 12 (4) |
| Hispanic or Latino ethnicity, n (%) | 41 (13) | 51 (17) |
| Highest level of education, n (%) | | |
| Less than high school degree | 4 (1) | 2 (1) |
| High school degree or equivalent | 38 (12) | 30 (10) |
| Some college but no degree | 94 (30) | 80 (27) |
| Associate degree | 51 (16) | 51 (17) |
| Bachelor's degree | 74 (24) | 79 (27) |
| Graduate degree | 52 (17) | 55 (19) |
| Median income by ZIP code, \$ | 60441 | 64778 |
| Body mass index, kg/m2, mean<br>(SD) | 35 (7) | 34 (7) |
| HbA1c %, mean (SD) | 8.1 (0.9) | 8.2 (0.9) |
| Fasting glucose, mg/dL, mean<br>(SD) | 168 (55) | 170 (57) |
| Years since diagnosis of diabetes,<br>mean (SD) | 11 (8) | 11 (8) |
### Study Sample Size
The sample size was calculated assuming a two-sided, two-sample t-test would be used to compare the difference in mean change from baseline in HbA1c between the intervention and control arm assuming a common standard deviation for both groups. The assumed common standard deviation of 1.4% and a clinically significant difference in means of 0.4% were used to determine that a nominal sample size, with 90% power, was 259 subjects per treatment group. However, the Applicant also considered a 20% attrition rate and thus calculated a sample size of 324 subjects per group.
### Pivotal Study Safety Results:
Better Therapeutics defined adverse events as follows:
- Any abnormal laboratory test result (e.g., hematology, clinical chemistry, or urinalysis) or . other abnormal safety assessment (e.g., electrocardiogram, radiological scan, vital sign
{5}------------------------------------------------
measurement), that worsens from baseline and is considered clinically significant in the medical and scientific judgement of the investigator
- . Exacerbation of a chronic or intermittent pre-existing condition, including either an increase in frequency and/or intensity of the condition
- New conditions that are identified or diagnosed after the study treatment was . administered, even if those conditions may have been present prior to the start of the study
- . Signs, symptoms, or the clinical sequelae of a suspected drug-drug interaction
- . Signs, symptoms, or the clinical sequelae of a suspected overdose of a concomitant medication. Overdose per se was not reported as an AE or serious adverse event (SAE) unless it was an intentional overdose taken with possible suicidal/self-harming intent. Such overdoses were to be reported regardless of sequelae.
- . "Lack of efficacy" or "failure of expected physiological action" per se were not to be reported as an AE or SAE. Such instances were captured in the efficacy assessments. However, the signs, symptoms, and/or clinical sequelae, resulting from lack of efficacy, were reported as an AE or SAE, if they fulfilled the definition of an AE or SAE
Better Therapeutics defined adverse events that occurred after the onboarding of the study as treatment-emergent adverse events (TEAE).
A summary of the number of subjects with reportable TEAEs observed during the study (postrandomization) is provided in the following table:
| | Post - Randomization, Days 1-180 | | | |
|--------------------------|----------------------------------|-------------|-------------------|-------------|
| | Control<br>n = 343 | | BT-001<br>n = 325 | |
| | Subjects<br>n (%) | Events<br>n | Subjects<br>n (%) | Events<br>n |
| TEAE | 188 (55) | 324 | 135 (42) | 265 |
| Maximum Severity of TEAE | | | | |
| Mild | 117 (34) | 219 | 100 (31) | 215 |
| Moderate | 55 (16) | 87 | 30 (9) | 45 |
| Severe | 16 (5) | 18 | 5 (2) | 5 |
| Serious TEAE | 24 (7) | 26 | 9 (3) | 9 |
Table 1. Adverse Events by Study Treatment Group
The sponsor also provided their assessment of TEAEs that were adjudicated as being possibly or probably associated with intervention. The sponsor reported that four (4) TEAEs in three subjects were adjudicated to be possibly or probably related to use of BT-001 during the study. These events included tachycardia, thirst, pollakiuria, and increasing HbA Ic.
PHQ-9 scores were tracked throughout the study to evaluate the risks of worsening depression and suicidal ideation. Participants with scores ≥20 were excluded or discontinued from the study. Table 2 shows PHQ-9 scores that resulted in study discontinuation.
{6}------------------------------------------------
| | Post - Randomization; Days 1-<br>90 | | Post - Randomization; Days 90-180 | |
|---------------------------------|-------------------------------------|-------------------|-----------------------------------|-------------------|
| | Control<br>n = 343 | BT-001<br>n = 325 | Control<br>n = 343 | BT-001<br>n = 325 |
| PHQ-9 ≥20 | 3 | 1 | 2 | 1 |
| PHQ-9<br>Positive<br>Question 9 | 3 | 1 | 6 | 2 |
Table 2. PHO-9 scores ≥20 considered severe, and positive to Question 9 indicates suicidal ideation
# Pivotal Study Observed Results:
Overall reduction in % A1c (change from baseline) for the control and intervention groups as measured at Day 90 and Day 180 is shown in Table 3 along with standard deviation.
Table 3. Mean (standard deviation) and range (minimum, maximum) %A1c change from baseline for the intention to treat population and the per protocol population in the control and intervention arms.
| Analysis Group and Timepoint<br>(Initial sample size) | Control<br>(n=343) | BT-001<br>(n=325) |
|-------------------------------------------------------|--------------------|-------------------|
| Intention To Treat - Day 90 | | |
| Mean (SD) | 0.14 (1.23) | -0.27 (1.11) |
| (min, max) | (-4.00, 5.95) | (-3.55, 5.50) |
| Number of subjects | n=313 | n=297 |
| Intention To Treat Day - 180 | | |
| Mean (SD) | -0.07 (1.19) | -0.37 (1.36) |
| (min, max) | (-4.20, 3.80) | (-4.40, 5.95) |
| Number of subjects | n=271 | n=244 |
| Per Protocol - Day 90 | | |
| Mean (SD) | 0.09 (1.16) | -0.38 (1.02) |
| (min, max) | (-4.00, 4.90) | (-3.55, 3.70) |
| Number of subjects | n=245 | n=159 |
| Per Protocol - Day 180 | | |
| Mean (SD) | -0.03 (1.11) | -0.41 (1.10) |
| (min, max) | (-4.20, 3.80) | (-3.65, 3.15) |
| Number of subjects | n=217 | n=160 |
For the intervention group the magnitude of A1c reduction was not uniform across the participants. Participants that completed more lessons through device use observed greater reductions in % A 1 c from baseline at both day 90 and day 180. A post-hoc analysis of the data is shown in Table 4 where a range of completed lessons has a corresponding % A1c reduction.
Table 4. Change in %A1c at day 90 and 180 from baseline with more completed lessons from the device
| # Lessons Complete | %A1c Change | Standard Deviation |
|--------------------|-------------|--------------------|
|--------------------|-------------|--------------------|
{7}------------------------------------------------
| 0 at day 90; n=313 | 0.14 | 1.23 |
|------------------------|-------|------|
| <5 at day 90; n=93 | -0.11 | 1.31 |
| 6-11 at day 90 n=100 | -0.24 | 1.15 |
| >11 at day 90; n = 104 | -0.45 | 0.86 |
| 0 at day 180; n=313 | -0.07 | 1.19 |
| <10 at day 180; n=80 | -0.06 | 1.66 |
| 10-20 at day 180; n=77 | -0.42 | 1.20 |
| >20 at day 180; n=87 | -0.61 | 1.14 |
The second variable that correlated with differences in the magnitude of A 1c reduction was whether subjects had additional medical visits outside of the pre-planned visits for the study. The participants in the intervention group who had additional medical visits with a healthcare provider observed greater reductions in Alc compared to intervention group participants who did not have an additional medical visit with a healthcare provider. The data is shown in Table 5 with standard deviations.
| | Control | Intervention |
|------------------------------------------------|-----------------------|-----------------------|
| Additional Medical Visits at day 90<br>ITT | 0.062 (1.16)<br>n=198 | -0.40 (0.95)<br>n=189 |
| Additional Medical Visits at day 180<br>ITT | -0.07 (1.15)<br>n=253 | -0.38 (1.31)<br>n=229 |
| No Additional Medical Visits at day 90<br>ITT | 0.27 (1.32)<br>n=115 | -0.05 (1.29)<br>n=108 |
| No Additional Medical Visits at day 180<br>ITT | 0.17 (1.60)<br>n=60 | 0.01 (1.42)<br>n=68 |
| Additional Medical Visits at day 90<br>PP | 0.05 (1.15)<br>n=159 | -0.51 (0.93)<br>n=109 |
| Additional Medical Visits at day 180<br>PP | -0.08 (1.08)<br>n=182 | -0.43 (1.13)<br>n=132 |
| No Additional Medical Visits at day 90<br>PP | 0.19 (1.15)<br>n=86 | -0.09 (1.15)<br>n=50 |
| No Additional Medical Visits at day 180<br>PP | 0.19 (1.27)<br>n=35 | -0.31 (0.97)<br>n=28 |
Table 5. Change in % A1c (standard deviation) from baseline at days 90 and 180 for the Intention To Treat (ITT) and Per Protocol (PP) populations
# Pediatric Extrapolation:
In this De Novo request, existing clinical data were not leveraged to support the use of the device in a pediatric patient population.
# LABELING
{8}------------------------------------------------
Device labeling includes the device indications for use, a description of the device, warnings and precautions, a summary of the clinical testing with the device, and instructions for use of the device. The labeling is sufficient and satisfies the requirements of 21 CFR 801.109 Prescription Devices.
Per the special controls for this generic type of device, labeling includes a summary of the clinical testing with the device and certain limiting statements. The labeling includes the following statements to meet the special controls:
- . The device is not intended for use as a standalone therapy.
- . The device is not a substitute for a patient's prescribed therapy or medication.
- . The device should not be used by people with unstable psychiatric disorders.
- The device is not intended for use in the treatment of any psychiatric disorder or . symptoms.
- . Among those assigned to the device, the average change in %HbA Ic observed in the clinical study was -0.27 compared to baseline after 90 days, and -0.37% compared to baseline after 180 days.
- After 90 days of device use, while 179 patients (60%) saw numerically improved ー :: glycemic control (%HbA1c decreased compared to baseline), there were 118 patients (40%) who saw no change or numerically worsened glycemic control (%HbA1c increased compared to baseline). In comparison, 136 patients (43%) in the control group saw HbA1c improvement and 177 (57%) saw no change or worsened glycemic control. These differences between the device and control group were statistically significant (p<0.0001).
- -After 180 days of device use, while 159 patients (65%) saw numerically improved glycemic control after 180 days of device use (%HbAlc decreased compared to baseline), there were 85 patients (35%) who saw no change or numerically worsened glycemic control (%HbA lc increased compared to baseline). In comparison, 122 patients (45%) in the control group saw HbAlc improvement and 149 (55%) saw no change or worsened glycemic control. These differences between the device and control group were statistically significant (p<0.0001).
- . In the clinical study with the device, observed reduction in HbA1c was correlated with increased usage of the BT-001 app. In addition, for patients that used the BT-001 app. those with additional medical visits to a health care provider had average HbA1c reductions of -0.40% at day 90 and -0.38% at day 180. In comparison, patients in the control group with additional medical visits had an average HbA 1c increase of +0.06% at day 90 and average reduction of -0.07% at day 180. The difference between groups was -0.46% at day 90 and -0.31% at day 180. Patients without additional medical visits who used the BT-001 app had average HbA1c reductions of -0.05% at day 90, and an average increase of +0.01% at day 180. In comparison, patients in the control group without additional medical visits had an average HbA1c increase of +0.27% at day 90 and +0.17% at day 180. The difference between groups was -0.32% at day 90 and -0.16% at day 180.
{9}------------------------------------------------
- . Patients who used the BT-001 app were financially compensated for taking frequent blood glucose measurements, whereas patients who used a control app were treated according to standard of care. Patients who used a control app did not take study-directed blood glucose measurements, thus their compensation was not associated with blood glucose measurements. Total compensation was equal for subjects who used the BT-001 app and those who did not use the app.
### RISKS TO HEALTH
The table below identifies the risks to health that may be associated with use of a diabetes digital behavioral therapeutic device and the measures necessary to mitigate these risks.
| Risks to Health | Mitigation Measures |
|----------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Worsening of condition due to device<br>providing ineffective treatment | Certain design verification and validation<br>activities, including clinical data<br><br>Certain labeling information, including<br>certain limiting statements |
| Treatment results in anxiety, depressed<br>mood, depression, mental disorder<br>(unspecified), stress or suicidal ideation | Certain design verification and validation<br>activities, including clinical data<br><br>Certain labeling information, including<br>certain limiting statements |
| Ineffective treatment due to use error /<br>improper use of device / device software<br>failure | Certain labeling information, including<br>certain limiting statements |
### SPECIAL CONTROLS
In combination with the general controls of the FD&C Act, the diabetes digital behavioral therapeutic device is subject to the following special controls:
- (1) Design verification and validation must include documentation of:
- (i) Clinical data from a statistically and clinically justified sample size, fulfilling the following:
- (A) Appropriately validating the model of therapy as implemented by the device using a clinically defined endpoint, and
- (B) Demonstrating that use of the device does not adversely impact the health outcomes or health status of the intended use population. A device hazard analysis must consider all device-related adverse events observed from the clinical data collected and must demonstrate that patient risk from use of the device is minimal.
{10}------------------------------------------------
- (ii) Software verification, validation, and hazard analysis must demonstrate that the device performs as intended.
- (2) The labeling must include:
- (i) A summary of the clinical testing with the device, including a discussion of the limitations of the clinical significance of the results.
- (ii) Limiting statements that indicate:
- (A) The device is not intended for use as a standalone therapy.
- (B) The device is not a substitute for a patient's prescribed therapy or medication.
- (C) The device should not be used by people with unstable psychiatric disorders.
- (D) The device is not intended for use in the treatment of any psychiatric disorder or symptoms.
# BENEFIT-RISK DETERMINATION
### Summary of Benefits:
The probable benefits of the device are related to supporting user behaviors to better adhere to known diabetes management skills. The benefits of the device are based on data collected in a clinical study described above. On average, study subjects who used BT-001 experienced a reduction in %A1c of -0.37 compared to an average reduction of -0.07 in the control group at day 180 (ITT population).
Factors that increase uncertainty in determining probable benefits for BT-001 include:
- . Improvement in %A1c was not uniform across study subjects who used BT-001. While an average improvement of -0.37 from baseline was observed at day 180 for BT-001 users, the range of %A1c change from baseline was (-4.40, 5.95) with a standard deviation of 1.36. While 159 patients (65%) saw numerically improved glycemic control after 180 days of device use (%HbA 1c decreased compared to baseline), there were 85 patients (35%) who saw no change or numerically worsened glycemic control (%HbAlc increased compared to baseline). In comparison, 122 patients (45%) in the control group saw HbA1c improvement and 149 (55%) saw no change or worsened glycemic control. As a result, there is moderate residual uncertainty regarding whether a particular user of BT-001 will experience improvement of glycemic control following device use.
- . The magnitude of the Alc reduction may be dependent on various factors, including adjustment and adherence to glycemic control medications, adherence to device use, frequency of medical visits, and adherence to other standard of care treatments such as frequent blood glucose monitoring.
{11}------------------------------------------------
- . The clinical study was limited to 6 months (180 days) in duration. As a result, the durability of the treatment effect is unclear. Diabetes mellitus is a chronic condition and clinical benefit is only expected from sustained improvement in %A1c over the long term.
- . There were patients that missed the day 90 and day 180 study visits where measurements of %A1c were made for evaluation of device efficacy. The number of BT-001 users with missing %A1c measurements was 28/325 (8.6%) at day 90 and 81/325 (25%) at day 180 in the treatment group (ITT population).
- . Patients who used the BT-001 app were financially compensated for taking frequent blood glucose measurements, whereas patients who used a control app were treated according to standard of care. Patients who used a control app did not take study-directed blood glucose measurements, thus their compensation was not associated with blood glucose measurements. Total compensation was equal for subjects who used the BT-001 app and those who did not use the app.
# Summary of Risks:
The probable risks of the device are based on data collected in a clinical study described above. Specifically:
- . Types of adverse events observed during the study were similar between groups, with BT-001 users experiencing numerically fewer events overall compared to the control group.
- . There were four (4) adverse events in three (3) subjects reported during the study that were judged to be possibly device-related, one of which was for worsening of glycemic control. In the observed case of worsening glycemic control, the subject only opened 1 lesson in BT-001 prior to day 90. and self-reported non-compliance with prescribed medication. Tachycardia reported by one subject was associated with increased physical activity as part of the BT-001 program. Pollakiuria and thirst were reported by one subject who had completed 26 lessons in BT-001.
- . Overall, device-related adverse events were not serious, and the rate and deeree of adverse events was not worse than those experienced by subjects in the control group.
- Cognitive behavioral therapy is associated with a class-effect risk of worsening . depression and suicidality. BT-001 users experienced numerically fewer of such events overall compared to the control group.
As described above, there were several factors that increased the uncertainty in determining the probable benefits that a patient is expected to experience from using the BT-001 device. Notably, 35% of patients using BT-001 saw either no change or worsening glycemic control after 180
{12}------------------------------------------------
days of use. However, given that the rate of no change or worsening glycemic control was observed to be higher in the control group (where 55% of patients saw no change or worsening control), and there were numerically fewer adverse events in the BT-001 group, it appears that the device may offer limited secondary benefit compared to standard of care alone.
# PATIENT PERSPECTIVES
The clinical study described above included use of the Patient Health Questionnaire-9 (PHQ-9) as a patient-reported outcome to evaluate the presence and severity of depression. PHQ-9 scores were evaluated at baseline and throughout the study. PHQ-9 scores of ≥20 or reported suicidal ideation were exclusion criteria for the study and subjects reporting a score of ≥20 or suicidal ideation at days 90 or 180 were removed from the study.
At baseline, PHQ-9 scores were similar between the two treatment arms (2.5 ± 2.9 in the Control group versus 2.6 ± 3.0 in the BT-001 group). At Day 90, 3 subjects in the Control group and 1 subject in the BT-001 group had scores ≥20 and were withdrawn from the study. At Day 180, 2 subjects in the Control group and 1 subject in the BT-001 group had scores >20.
### BENEFIT/RISK CONCLUSION
In conclusion, given the available information above, for the following indication statement:
BT-001 is a prescription-only digital therapeutic device intended to provide cognitive behavioral therapy to patients 18 years or older with type 2 diabetes. The device targets behavior to aid in the management of type 2 diabetes in patients who are under the care of a healthcare provider. BT-001 provides cognitive behavioral therapy as a treatment that should be used adjunctively with standard of care.
The probable benefits outweigh the probable risks for the BT-001 device. The device provides benefits and the risks can be mitigated by the use of general controls and the identified special controls.
### CONCLUSION
The De Novo request for the BT-001 is granted and the device is classified as follows:
Product Code: QXC Device Type: Diabetes digital behavioral therapeutic device Regulation Number: 21 CFR 880.5735 Class: II
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.